terahertz spectroscopy for off-gas detection and … also goes to other labmates, arathi...
TRANSCRIPT
Terahertz Spectroscopy for off-gas detection and analysis in the steel-making industry
by
Yuhui Song
A thesis submitted in conformity with the requirements for the degree of Master of Applied Science
Mechanical and Industrial Engineering University of Toronto
© Copyright by Yuhui Song 2014
ii
Terahertz Spectroscopy for off-gas detection and analysis in
steel-making industry
Yuhui Song
Master of Applied Science
Mechanical and Industrial Engineering University of Toronto
2014
Abstract
Terahertz (THz) radiation has the advantage of transmitting through gases with high particle
loading. In this thesis, we introduce the use of THz spectroscopy to the measurement of water
vapor at high temperatures in the combustion off-gas. We establish a high temperature water
vapor measurement system where a monochromatic continuous wave THz source is used and the
heating system enables the temperature of water vapor in the gas cell to go up to 500 ℃. At high
temperature, peaks at 0.67 THz, 0.88 THz emerge and grow with increasing the water
concentration. Whereas major absorption peaks (0.557 THz, 0.753 THz) shrink with the
increasing temperature. Plotting out the area calculated underneath the major absorbance peak
against the water concentration, we observed a linear correlation. The same trend is found for the
small new peak with large uncertainty however due to instrumental challenges.
iii
Acknowledgments First I would like to thank my supervisor Professor Murray Thomson for his constant guidance
and support throughout my endeavors in the graduates study. His insights and encouragement are
indispensable for me to face all the challenges and overcome difficulties I encountered in during
the research.
My labmate, Zhenyou Wang is owed a great deal of thanks for his dedication in mentoring me
and helping me in carrying out the research. From the building of the experimental set-up, to
trouble-shooting, experiments and paper writing, he is providing me with continuous support and
enlightenment.
Credit also goes to other labmates, Arathi Padmanabhan, Tommy Tzanetakis,
Bobby Borshanpour, Jamie Loh and Hansen Wang. They played important roles in introducing
me to the project, helping me understanding of spectroscopy theory, the design and building
experimental set-up and carrying out tests.
I am grateful to the staff in machine shop of the Mechanical Engineering department. Their work
in the fabrication of the experimental facilities and their guidance regarding mechanical
problems enables me to carry out complex experiments with ease.
Appreciation is also expressed to the industrial collaborator Tenova Goodfellow Inc. and the
instrument supplier TeraView. They provided me with first-hand information in industrial
practices and helped me identify and resolve the issues encountered during the operation of the
terahertz system.
Last but not least, I would pay my sincere gratitude to my parents, my family and friends. They
are always a constant source of love and happiness in my life.
iv
Table of Contents Acknowledgments .......................................................................................................................... iii
Table of Contents ........................................................................................................................... iv
List of Tables ................................................................................................................................ vii
List of Figures .............................................................................................................................. viii
List of Appendices ......................................................................................................................... ix
Chapter 1 INTRODUCTION .......................................................................................................... 1
1 Motivation and objectives .......................................................................................................... 1
1.1 In-situ Quantification of Off-gas in Industry ...................................................................... 1
1.2 High Temperature Measurement of Water Vapor .............................................................. 3
2 Research Background ................................................................................................................. 4
2.1 Fundamentals of Terahertz Radiation ................................................................................. 4
2.2 State of the Art of Terahertz Technology ........................................................................... 5
2.3 Terahertz Spectroscopy and Application ............................................................................ 7
3 Organization of Thesis ............................................................................................................... 9
Chapter 2 EXPERIMENTAL INSTRUMENTATION ................................................................ 11
1 Introduction .............................................................................................................................. 11
2 The light Source ....................................................................................................................... 12
2.1 Continuous Wave THz Sources ........................................................................................ 12
2.2 Photomixing Technology .................................................................................................. 13
2.3 Operating Principles of TeraView CW Spetra 400 ........................................................... 13
3 Gas Cell .................................................................................................................................... 16
3.1 Gas Cell Design ................................................................................................................ 16
3.2 Piping and Liquid Injection .............................................................................................. 17
3.3 Heating and Thermal Insulation ........................................................................................ 18
v
4 Purging Chamber and Faraday Cage........................................................................................ 18
4.1 Purging System ................................................................................................................. 18
4.2 Compressed Air Cooling ................................................................................................... 19
4.3 Faraday Cage .................................................................................................................... 20
5 Summary .................................................................................................................................. 20
Chapter 3 METHOD DEVELOPMENT ...................................................................................... 21
1 Introduction .............................................................................................................................. 21
2 Development of Testing Procedure .......................................................................................... 21
2.1 System Warm-up and Stabilization .................................................................................. 21
2.2 Background Measurement ................................................................................................ 23
2.3 Baseline Drift and Reference Measurement ..................................................................... 25
2.4 Water Signal Measurement ............................................................................................... 27
3 Effect of Ambient E/M interference ........................................................................................ 27
3.1 Day-night Fluctuation in Signal Quality ........................................................................... 27
3.2 Introduction of Faraday Cage ........................................................................................... 29
4 Data Processing Techniques .................................................................................................... 33
4.1 Multiple scans, Averaging and Subtraction ...................................................................... 33
4.2 Techniques for Removing Etalon Fringes ........................................................................ 34
4.3 Curve Fitting and Peak Area Calculation ......................................................................... 35
5 Summary .................................................................................................................................. 37
Chapter 4 RESULTS AND DISCUSSION .................................................................................. 38
1 Introduction .............................................................................................................................. 38
2 HITRAN Spectral Simulation .................................................................................................. 38
2.1 Modelling of Water Spectrum in Gas Cell ....................................................................... 38
2.2 Variation of Water Absorbance Spectra with Temperature .............................................. 39
2.3 Variation of Water Absorbance Spectra with Humidity ................................................... 40
vi
2.4 Correlation between Absorbance Peak Area and Water Concentration ........................... 41
2.5 Two-cell simulation and peak area-concentration correlation .......................................... 43
3 Experimental Study .................................................................................................................. 45
3.1 Beam scattering test with different light sources .............................................................. 45
3.2 Variation of Water Absorbance Spectra with Water Concentration ................................. 47
3.3 Experimental Correlation Between Absorbance Peak Area and Water Concentration .... 49
4 Summary .................................................................................................................................. 52
Chapter 5 CONCLUSION ............................................................................................................ 54
1 Summary .................................................................................................................................. 54
2 Industrial application ................................................................................................................ 55
3 Current Challenges ................................................................................................................... 56
4 Recommendations for Future Work ......................................................................................... 56
References or Bibliography .......................................................................................................... 58
Appendix ....................................................................................................................................... 63
viii
List of Figures Figure 1.1. Schematic diagram of an optical measurement and process control system in dusty
EAF.
Figure 1.2. LINDARC™ off-gas analysis system [52]
Figure 1.3. Variation of transmission in dusty environment with wavelength., courtesy of
colleague Amirhossein Alikhanzadeh.
Figure 2.1. Schematic diagram of the experimental setup for high-temperature water vapor
measurement with the THz spectroscopy.
Figure 2.2. The fiber optic scheme [53]
Figure 2.3. Difference frequency generation and detection: (a) Illustration of photomixing
scheme. Two laser heads working at different temperatures have different frequencies. By
controlling the temperature, the frequency difference between the two laser head can be adjusted.
(b) Illustration of the THz generation scheme. (c) Illustration of the THz detection scheme [53].
Figure 2.4. 3-D model of the gas cell assembly.
Figure 2.5. Section view of the one end of the gas cell.
Figure 2.6. Schematic diagram of piping and water interjection system.
Figure 2.7. Purging chamber and compressed air ventilation scheme.
Figure 2.8. Compressed air cooling for the optical system inside the purging chamber.
ix
List of Appendices A. Humidity and Partial Pressure Conversion Table
B. AutoCAD Engineering Drawing for Experimental Facilities
C. MatLab Codes for Data Processing
1
Chapter 1 Introduction
1.1 Motivation and objectives
1.1.1 In-situ Quantification of Off-gas in Industry
In large industrial combustion processes where timely and accurate measurement of off-gas
species is needed to optimize the process, in-situ non-destructive diagnostic technologies like
optical sensors show great advantage over extractive methods, in collecting more precise and
real-time data. Among them, infrared, visible, and ultraviolet light have been widely utilized in
industry and are able to identify a variety of gas species in a broad spectrum [1]–[12]. However,
in industrial furnaces, where the environment is harsh and the exhaust is heavily-laden with
particles and aerosols, it is hard to do measurements with traditional diagnostic methods; the
beam scattering caused by particles renders it difficult for the light to penetrate the dust and
identify gaseous species. Figure 1 shows the schematic diagram of a typical optical measurement
and process control system in an electric arc furnace (EAF) in a steel making plant. The optical
path intersecting the off-gas flow is severely contaminated with dust, making it difficult to detect
the laser beam travelling across the exhaust duct.
Figure 1.1. Schematic diagram of an optical measurement and process control system in dusty EAF.
2
In industry, different methods have been developed to tackle the beam scattering problem, one of
them being to shorten the beam path. Figure 2 illustrates the LINDARCTM off-gas analysis
system using the technique of “Tunable Diode Laser Absorption Spectroscopy” (TDLAS). With
the support of two protruding arms, the emitter and detector are placed closer in the middle of the
exhaust duct. As consequence the power loss due to beam scattering is significantly reduced. The
concentration of a selected species can be determined by the quantity of light absorbed at its
specific spectral range even with very high dust contents. However, the cost involved in the
modification of the hardware and facility is considerable and molten slag can build up and plug
the gap.
However some interesting prospects are found in another frequency domain. As shown in Figure
1.3, i.e. scattering simulation, in dusty environment, transmission increases with the wavelength.
Terahertz (THz) light which has relatively longer wavelength is able to penetrate gases with high
particle loading and allows the gas phase concentration measurement to be made in dusty media
without compromising the signal power.
Figure 1.2. LINDARC™ off-gas analysis system [52].
3
1.1.2 High Temperature Measurement of Water Vapor
Among all combustion gases, water vapor, as a major combustion product, has gained particular
amount of attention from scholars across the field. Up to now, plenty of research has been
devoted to the investigation of THz absorption spectrum of water vapor at different pressures or
water concentrations at room temperature [13]–[16]. Many distinct absorption peaks showing up
at room temperature have been observed and reported in literature [17]–[20]. However, water is
in such great abundance in the atmosphere that as the THz beam travels across an open space, the
absorption spectrum is likely to saturate due to water redundancy. It can be extremely difficult to
distinguish between ambient (room temperature) water vapor and combustion-produced (high
temperature) water vapor by focusing only on the well-identified peaks which are usually tall and
large and are prone to saturation at even room humidity. In industrial furnaces, the typical water
Figure 1.3. Variation of transmission in dusty environment with wavelength., courtesy of colleague Amirhossein Alikhanzadeh.
4
vapor concentration ranges from 16% to 20%, which would inevitably lead to saturation
problems, making it impossible to quantify the precise amount of water present.
At high temperatures, studies have shown that a number of new rotational transitions would
occur to water molecules at highly excited states [21]–[25]. This provides the opportunity of
measuring high-temperature combustion water vapor without the interference of ambient
humidity. Therefore, in the present study, we focus on the measurement of water vapor at high
temperatures. The characteristics of the large absorbance peaks as well as the new emerging
peaks at high temperature would offer guidance in water vapor detection and analysis in
industrial furnaces [26]–[35].
1.1.3 Objectives of Research
The objects of the research are:
1) Ground -up design and building of the laboratory facilities for THz spectroscopy study.
2) Theoretical study of water vapor absorption in both room temperature and high
temperatures based on HITRAN spectral simulation.
3) Experimental investigation on water vapor absorption at different temperatures and
concentrations to validate simulation results.
1.2 Research Background
1.2.1 Fundamentals of Terahertz Radiation
Terahertz (THz) radiation, which fills the gap in the electromagnetic spectrum between the
microwave and the infrared light, is referred to as the frequency range from 0.1 THz to 10 THz
corresponding to a wavelength of 3 mm to 30 μm . We cannot see THz radiation but we can feel
its warmth as it shares its spectrum with far-infrared radiation. Figure 1.4 illustrates the terahertz
band in the electromagnetic spectrum. The THz band merges into neighboring spectral bands
such as the millimeter-wave band, which is the highest radio frequency band known as
Extremely High Frequency (EHF), the submillimeter-wave band, and the far-IR band.
In the THz region, innumerable spectral features show up due to fundamental processes such as
rotational transitions of molecules, large-amplitude vibrational motions of organic compounds,
5
lattice vibrations in solids, intraband transitions in semiconductors and energy gaps in super-
conductors. These unique characteristics of material responses to THz radiation lead to THz
applications in various fields.
Based on optical properties at THz frequencies, condensed matter is generally classified into
three categories: water, metal and dielectric. Water, a strongly polar liquid, is highly absorptive
in the THz region. Due to high electrical conductivity, metals are highly reflective at THz
frequencies. Nonpolar and nonmetallic materials, i.e., dielectrics such as paper, plastic, clothes,
wood and ceramics that are usually opaque at optical wavelengths, are transparent to THz
radiation [36]. The optical properties of each material type enable THz radiation to be utilized in
different applications and they will be discussed in the following section.
1.2.2 State of the Art of THz Technology
This field has remained quite unexplored until recent years when substantial advancements in
photonics have taken place, making it possible to generate and detect THz beams with much
higher efficiency. There are three major approaches for developing THz sources. The first is
optical THz generation, which has been the frontier of THz research for the past few decades.
The second is known as THz-Quantum Cascade Laser (THz-QCL) which emerged in recent
years and is still in development. The third uses solid-state electronic devices which are quite
established at low frequencies [37]. Another approach is the free electron method which requires
huge and complicated facility and is not as widely applied as the three major sources.
Figure 1.4. Terahertz band in the electromagnetic spectrum [53].
6
There are two general categories for the optical generation of THz radiation using either pulsed
or continuous wave laser. The first involves generating an ultrafast photocurrent in a
photoconductive switch or semiconductor using electric-field carrier acceleration or the photo-
Dember effect. In the second category, THz waves are generated by nonlinear optical effects
such as optical rectification, difference-frequency generation or optical parametric oscillation.
Some of the nonlinear media gaining attention are GaAs, GaSe, GaP, ZnTe and LiNbO3 and
research to find more effective materials is underway. Femtosecond lasers are used mainly for
THz-Time Domain Spectroscopy (THz-TDS) while frequency-domain spectroscopy and imaging
systems employ other lasers. Enhancing the output power, reducing the system size and
increasing the speed of frequency sweep and data acquisition are among the current objectives
for future sources.
The second THz generation technology, the THz-QCL has been developed along with the rapid
advancement in nanotechnology in recent years. The THz waves are emitted by means of
electron relaxation between subbands of quantum wells; for examples, between several few-nm-
thick GaAs layers separated by AlxGas1-xAs barriers, whose emission blocks are serially
connected to generate THz waves. At present the research mainly focuses on reduction in
threshold currents reduction and lasing frequencies, increasing the operational temperatures and
frequency range to obtain higher quality beam modes. Extensive studies on new design of
structures, gratings and waveguides are being carried out to achieve these objects.
The third approach, solid-state electronic devices mainly dominate the low frequency end of the
THz regime. As one of the most promising technology, uni-traveling-carrier photodiode (UTC-
PD) produces high-quality sub-THz waves by means of photomixing. The optical beat of the
light from two different wavelength laser diodes (LDs) are used to produce the THz waves in a
UTC-PD. The difference between the two wavelengths determines the emission frequency. So
far the technology has found application in sub-THz wireless communication and photonic local
oscillators.
On the side of detection, much attention has been paid to GaAs grown at low temperature which
is often used as photoconductive antenna. Alternatively, electro-optic sampling technologies are
available for ultrawideband time-domain detection. It is feasible to measure over 100 THz using
a 10-fs-laser and a thin nonlinear crystal such as GaSe. Other traditional THz detectors include
7
DTGS, crystals, bolometers, SBDs and SIS junctions. Also, a single-photon THz detector has
been developed using a single electron transistor. [26], [27], [36], [38]–[44]
1.2.3 Terahertz Spectroscopy and Application
For a long period of time, THz technology has interested astronomers because approximately
one-half of the total luminosity and 98% of the photons emitted since the Big Bang fall into the
submillimeter and far-infrared region and THz can be used to help study these photons. Different
types of THz facilities have been installed in observatories around the world for astronomy,
environmental monitoring and plasma diagnostics [45]. In addition to its long-standing
applications in astronomy, physics and chemistry, the THz technology is now finding its use in a
much wider range of sectors: information and communications technology (ICT); biology and
medical sciences; non-destructive evaluation; homeland security; quality control of food and
agricultural products among others.
The fact that many organic and biological molecules have spectral signatures in the THz region
makes THz spectroscopy a very effective tool in identifying unknown species. The see-through
capability of THz waves has attracted the attention of many security agencies to explore the use
of THz for explosive detection. THz-TDS has proved to be reliable in identifying and
characterizing many explosive and related compounds (ERCs). Figure 1.4 illustrates the THz
spectra of many ERCs and the distinct spectral signatures can be used to identify them.
Compared with other technologies such as Fourier Transform Infrared Spectroscopy (FTIR) and
Raman, THz has the advantage of penetrating covering materials and interrogating materials that
are not optically visible. Therefore THz technology has been employed in security applications
such as mail scanning, illicit drugs inspection and short distance standoff detection of solid or
powder ERCs. It achieved even better success in full body scanning for detection of hidden
objects and has been deployed in many airport checkpoint in the U.S. Since full body scanners
seldom perform spectroscopic analysis for material identification, continuous wave (CW)
systems shows better capability due to its narrowband nature and could be more sensitive in
detection.
Another major application for THz TDS lies in pharmaceutical characterization and pill coating.
Polymorphs which are different crystallizations of the same molecule may show different
dissolution rates and can affect the stability and efficacy of a drug. THz is used to identify
8
polymorph compounds and provide proof of authenticity of pills. Also the thickness of a coating
is very important for the proper release of the drug inside the body and it can affect the efficacy
and potential side effects of the drug. THz provides a non-destructive and non-contact method by
using the time of flight information contained in the waveform to measure the thickness of the
coatings of the pills.
In the food industry, THz spectroscopy is utilized to detect contaminants, pesticides, foreign
objects and the presence of antibiotics. The see-through capability of THz waves enables the
inspection of packaged food. As the moisture level is an indicator of the conditions of bacteria
growth, THz offers the potential to increase food safety by measuring the moisture level in the
food. Figure 1.5 shows the capability of THz in detection of foreign objects such as glass, plastic
and ceramic as small as a few mm in a chocolate bar.
Figure 1.5. Spectral characteristics of different ERCs in THz range, Figure adapted from Ref. [42]
9
Figure 1.6. Foreign objects detection in chocolate bars Figure adapted from Ref. [46]
Furthermore, with the combined advantage of see-through capabilities and excellent spatial
resolution, THz can also be utilized in NDE (Non-Destructive Evaluation) imaging. Specific
applications include detection of defects in insulation, plastic and ceramic materials, inspection
of corrosion in both metal and non-metallic substrates under insulation or coatings, fire damage
and other structure defects (cracks and delaminations) in composite materials, turbine blades and
art pieces inspection and restoration. THz-time domain systems are always employed in such
occasions based on the same principles of ultrasound in studying the structure of layers of
samples. CW systems are equally attractive since they can provide faster data rates.
In recent years, researchers are getting keen on exploring THz applications in medical area such
as skin, breast and liver cancer diagnosis and skin burn evaluation with promising results. These
applications are essentially based on the high sensitivity of THz waves to the presence of water.
In general, cancerous tissue tends to accumulate more water and display stronger THz absorption
in the spectrum. One of the big challenges currently is that the high water content in biological
samples and its high variability make measurement difficult. Also proper samples preparation
and measurement protocols need to be developed for THz experiments. Once applications in the
research phase are successful, THz would have a significant impact on early cancer diagnosis
and prevention.
1.3 Organization of Thesis
The thesis comprises five major sections:
10
Chapter 1 Introduction
Chapter 2 Experimental Instrumentation
Chapter 3 Methodology Development
Chapter 4 Results and Discussion
Chapter 5 Conclusion
In particular, we would like to elaborate on the experimental instrumentation and methodology
development in more detail since they have been the most challenging work in the entire project
and a comprehensive description will lead to a better understanding of the entire system.
In Chapter 2 we will present the experimental set-up at length including every subsystem that
performs heating, purging, cooling and ventilation for the system respectively. In the ensuing
Chapter 3 we will talk about how the system functions, what problems came up and how to
resolve them or optimize the situation. Next we will move on the discuss results obtained from
both spectral simulation and experiments, analyze the leads and accuracy and draw conclusions
accordingly.
Finally we will come to a summary of the whole research, evaluate the pros and cons of the THz
spectroscopy technology and its feasibility in industry, propose plans for improvements and cast
our vision into future prospects.
11
Chapter 2 Experimental Instrumentation
2.1 Introduction
In this chapter, we introduce the experimental set-up we designed for the high temperature
measurement of water vapor with THz spectroscopy. Figure 2.1 shows a schematic diagram of
the entire measurement system.
Figure 2.1. Schematic diagram of the experimental setup for high-temperature water vapor measurement with the THz spectroscopy.
The whole system comprises a couple of sub-systems, namely the light source, the gas cell, the
heating system, the purging system, and the ventilation and cooling system.
The THz beam leaving the emitter travels in an open path before reaching the window of the gas
cell. To eliminate the influence of room humidity, the entire optical path is enclosed in an airtight
box, purged constantly with dry air. The gas cell is a stainless steel cylinder 1 meter in length,
with 2 inch inner diameter, equipped with an Omega digital pressure gauge and several K-type
thermocouples. The two end windows are 5 mm-thick z-cut quartz disks and are tilted at an angle
of 45º. The gas cell body is wrapped with Omega Ultra-High Temperature Heating Tapes
(STH051-060) on top of which sits a thick blanket of fibre-glass as thermal insulation. A brief
description of the experimental procedure is as follows. Firstly we block the THz beam with a
piece of aluminum foil to collect the background signal. The gas cell is then filled with pure
nitrogen for reference measurement. To investigate the water spectrum at a certain water
12
concentration, we inject a calculated amount of distilled water though a septum fitting using a
gastight syringe. At the end of each water absorption measurement, we evacuate and flush the
gas cell with N2 gas 10 times to remove any residual water vapor. The cell is then backfilled with
the reference N2 to start the next test.
In the following sections, we will describe the design and function of each facility. Specific
experimental techniques and procedure will be explained along with the set-up description. A
more systematic testing procedure development will be elaborated in the next chapter.
2.2 The light Source
2.2.1 Continuous Wave THz Sources
As has been mentioned in the previous section, the recent advancement in optical technology
lead to the development of THz sources of all kinds. Among them, THz pulses based on femto-
second pulse lasers have been applied to objects imaging, and spectroscopy of gas, liquid and
solid materials. In particular the time-domain spectroscopy (TDS) system based on THz pulses
has been well-established as a laboratory standard for the THz spectroscopy and is widely
commercialized. The Fourier transformation is applied to the time-domain data in a THz-TDS
system to obtain frequency characteristics [47].
Over the past years enormous attention has been paid to tunable continuous wave (CW)
spectrometer which uses monochromatic sources. Compared with pulsed systems, CW systems
are more appropriate for imaging and spectroscopy because they provide a higher spectral
resolution (up to 100 MHz).
There are several requirements for the CW THz sources: they should provide high output power
to enable a wide dynamic range and the radiation should be monochromatic and tunable in
frequency. Moreover it is desirable to have a high degree of polarization for the characterization
of anisotropic samples. For microspectroscopy they also should serve as diffraction-limited THz
point sources, a property that can be best represented by M2 value [41].
13
2.2.2 Photomixing Technology
Photomixing denotes heterodyne difference frequency generation in high-bandwidth
photoconductors. The output of two continuous-wave lasers converts into terahertz radiation [26],
[37], [48]–[51].
S. Martens and scholars in Germany have described the principle and mechanism involved in the
photomixing technology [41]. A photomixer usually utilizes a tunable two-wavelength laser as a
source, which is stabilized by external cavities and includes an antireflection-coated commercial
laser diode, a collimator, a diffraction grating cylinder lens and a V-shaped mirror. In a tapered
optical amplifier the output beam is amplified to about 20-50 mW and is then focused on the
photomixer. The emitted THz radiation is then collimated by a hyperhemispherical Si lens that is
installed at the back of the LT-GaAs chip, which is used to collimate the beam. The optical
elements are made of Polyethylene terephthalate and Teflon lenses. Polarizers consist of wire
grids with thick wires and other wire grids serve as partially reflecting mirrors in the Fabry-Perot
setup [41].
Key advantages of photomixing systems include a high frequency resolution, spectral selectivity
and large signal-to-noise ratios. Typical applications utilizing these properties are high-resolution
gas spectroscopy, solid-state spectroscopy with the benefit of determining a sample’s complex
dielectric constant and spectrally sensitive imaging.
2.2.3 Operating Principles of TeraView CW Spectra 400
The basic principle of operation of the THz spectrometer is as follows. Two near-infrared diode
lasers are precisely tuned to offset their relative wavelengths, producing a beat signal at the
difference frequency when coupled into the same fiber. The output fiber is connected to a THz
photomixer emitter, converting this beat signal into coherent THz radiation. Similarly, the THz
beat signal is delivered as a reference to the THz photomixer receiver, which allows ultra-
sensitive coherent detection of the incident THz radiation, even at sub-nanowatt power levels.
This scheme is phase sensitive, compact, and requires no cryogenic cooling. The fiber-optic
scheme is shown in Figure 2.2.
14
Figure 2.2. The fiber optic scheme [53].
The THz spectrum is then achieved by incrementally varying the difference frequency via a
smooth, mode-hop free tuning of the near-infrared diode lasers, according to the temperature-
tuning technique (Figure 2.3a). The signal amplitude and phase are measured at each discrete
frequency point, from which the power can be derived. The spectrometer system comprises: two
distributed feedback (DFB) diode lasers, with electric temperature and current controllers; fiber
connections to all components; electronic source for terahertz emitter drive circuit; detection
amplifier system; one terahertz emitter (Figure 2.3b); one terahertz receiver (Figure 2.3c).
15
Figure 2.3. Difference frequency generation and detection: (a) Illustration of photomixing scheme. Two laser heads working at different temperatures have different frequencies. By controlling the temperature, the frequency difference between the two laser head can be adjusted. (b) Illustration of the THz generation scheme. (c) Illustration of the THz
detection scheme [53].
At each frequency point in the spectrum, a sinusoidal waveform is acquired using a time domain
sweep, executed using fiber stretching technology. This allows each frequency measurement to
be acquired in approximately 30 ms. When executing a frequency sweep, the scan time hold-off
is user variable, according to the desired signal-to noise ratio.
16
Terahertz pulsed instruments require little maintenance and are relatively compact and mobile as
there is no need for sophisticated cooling solutions. The power of the terahertz radiation used for
the measurements is below 1 μW.
2.3 Gas Cell
2.3.1 Gas Cell Design
The gas cell is made of a one meter long, two-inch-diameter stainless steel tube. The two end
windows are 5 mm-thick disks made of z-cut quartz. The windows are tilted at an angle of 45º as
calculations indicate that the power loss is less than 5% at this incident angle. The 45º is chosen
because it is close to Brewster’s angle which is calculated as 66º for the current case and it also
easy to be manufactured. Figure 2.4 shows the 3-D model of the gas cell assembly.
Figure 2.4. 3-D model of the gas cell assembly.
Flanges tilted at the same angle are used at both ends of the gas cell to mount the windows. To
seal the windows, we chose graphite gasket that is compatible with the window material in its
hardness and is resistant to high temperatures. Due to manufacturing difficulties, the surface of
the flange cannot be made perfect flat. This requires the bolting to be well-performed so that the
force on the flange disk can be evenly distributed. Figure 2.5 displays the section view of the
flange, gasket and window at one end of the gas cell. After the manufacturing and assembling,
we carried out air-tightness tests under both pressurized and vacuum conditions. It is confirmed
that the pressure inside the gas cell can stay stable without any significant drop for a considerable
period of time.
17
Figure 2.5. Section view of the one end of the gas cell.
2.3.2 Piping and Liquid Injection
Figure 2.6 shows the schematic diagram of the piping and water injection system. There are two
openings along the cylindrical body of the gas cell. One is used for gas ventilation and the other
water injection.
Figure 2.6. Schematic diagram of piping and water interjection system.
The water injection port is a septum fitting that provides good sealing during the water injection
process. A 50/250 μl gastight syringe is used to inject a known amount of water into the gas cell.
The other port is designed for gas ventilation. The T-junction connects the gas piping branch and
the evacuation piping branch. The gas cylinder is currently only nitrogen as pure N2 is selected
as a non-polar gas to flush the gas cell. For future experiments that involve multiple gases, a
manifold that connects the gas cell to multiple gas cylinders can be introduced. The vacuum
18
pump serves for dual purposes in the current set-up: evacuation of the gas cell and the purging
and cooling of the optical path. By control of a three-way ball valve which connects the pump to
both the gas cell and the purging box, we are able to use the pump for different purpose under
difference circumstances.
2.3.3 Heating and Thermal Insulation
One of the most important features of the system is its achievement of the high temperature
measurement. The gas cell body is wrapped with Omega Ultra-High Temperature Heating Tapes
(STH051-060) on top of which sits a thick blanket of fibre-glass as thermal insulation. Two
heating tapes have been used, each of them controlled by a Variac. The heating system enables
the system to be warmed up to 500 ℃ and stay relatively stable at any temperature below Tmax.
Two K-type thermocouples are installed at the openings of the gas cell body to monitor the
temperature inside and provide real-time feedback for the temperature control.
2.4 Purging Chamber and Faraday Cage
2.4.1 Purging System
To investigate the water absorption spectra at different water concentrations inside the gas cell at
high temperatures, we have to keep the optical path dry and clean to remove the interference of
room humidity. To do this, we introduce the compressed air provided by the building where the
laboratory is located. Shown below is a demonstration of the purging scheme.
Figure 2.7. Purging chamber and compressed air ventilation scheme.
19
A clear Acrylic chamber is built to enclose the whole optical system. A rotameter is used to
control the volume flow rate of the compressed dry air for purging. The air flushes through the
chamber from both ends of the box and exits from the middle of the lid. Four humidity indicators
are placed at different positions in the chamber to ensure that the humidity is controlled well
below the atmospheric level.
2.4.2 Compressed Air Cooling
The compressed air that is used for purging also serves as a cooling medium to keep the
temperature of the chamber within acceptable limits. The optical sensor, especially the emitter
and receiver which are placed inside the purging box, are extremely delicate and high-
maintenance. The ambient temperature they are exposed to has to be kept at around 20-40℃. As
the radiation of the heated gas cell is constantly increasing the air temperature inside the purging
chamber, the flush air is not only removing the humidity but also the excessive heat. Figure 2.8
illustrates how the cooling scheme works.
Figure 2.8. Compressed air cooling for the optical system inside the purging chamber.
To mitigate the radiation heat transfer from the blazing body of the gas cell, we place two
thermal baffles between the gas cell and the optical heads. The material of the baffle plates is
calcium sulfide and they are 10 mm in thickness. A one-inch hole is chiseled on each plate to
allow the propagation of the THz beam. The compressed air enters the purging box from the left
and right ends, flows though the holes, travels along the gas cell and exits from the multiple tiny
20
openings on the purging box. The warm air is thus constantly exhausted from the chamber with
fresh cool air coming as replacement.
2.4.3 Faraday Cage
Tests have shown that the system is susceptible to ambient electromagnetic interference, which
will be discussed in detail in section 3.3. Therefore we built a Faraday cage that protects each
optical head, especially the receiver, from the E/M influence. The schematic diagram of the
Faraday cage is shown in Figure 2.8. It is achieved by wrapping each optical head with
aluminum foil and grounding it with a thin wire.
Note that the current Faraday cage is a modified version of a previous larger cage. Instead of
using two small cages to enclose the optical heads separately, we originally wrapped the whole
purging box with a seamless layer of Al foil. The large cage worked well in reducing the E/M
interference and improving the quality of the signal. However the heat dissipation was
significantly affected by the cage that it is hard to maintain the temperature close to the optical
heads to be below 40℃. Furthermore the plastic lid warped and melted due to the overheating
and repulsive odor was emitted from the melting material that consequently affected the water
spectrum. Therefore we removed the large cage and designed the current small separate cages
that perform equally well in signal improvement without causing any overheating.
2.5 Summary
In this chapter, we described the design and function of the experimental set-up. The whole
measurement system consists of the light source, which is a continuous wave THz spectrometer,
the 1-m long stainless steel gas cell, the heating and thermal insulation, the piping and water
injection, the compressed air purging and cooling system and the Faraday cage. The system
enables the measurement of the THz absorption spectra water vapor of different concentrations
at temperatures from room temperature to 500℃.
21
Chapter 3 Method Development
3.1 Introduction
In this chapter, we will focus on the method development for the THz spectroscopic
investigation on water vapor absorption. The THz spectrometer we purchased from TeraView
was originally fabricated as a standard product that aims to measure the transmission within the
path-length of 20 cm. In the present study, we extended the path-length to 120 cm which requires
a certain extension of the optical fiber connecting the receiver/emitter to the laser. However, this
modification to the system doesn’t come without a cost. The power amplitude dropped to 0.001
of its original value due to the diffraction of THz beam. Normally as the frequency goes higher,
the power drops accordingly. Fortunately the frequency range (below 1 THz) we are focusing for
the current study has relatively high power signal.
Since the technology is still in its infancy, the modification of the system further adds to its
unpredictability. The current research is largely devoted to method development; considerable
efforts have been spent on identifying the issue, trouble-shooting and finding approaches to
mitigate or resolve the problem before commencing the real investigation.
3.2 Development of Testing Procedure
3.2.1 System Warm-up and Stabilization
One important character about a THz system is its repeatability. The THz spectrometer we
acquired proves to have good repeatability as has been demonstrated in its Site Acceptance Test.
However, in our application where multiple scans are required for the averaging and analysis,
highly repeatable spectra are needed that could suppress the noise. But for the sake of saving
time, we would use single scan as it also provides a good enough SNR. In the previous tests, we
found it is hard for the system to produce repeatable data right after it is turned on. It needs a
considerable time to warm up and stabilize. This is due to the fact that the physical state of the
fiber stretcher that is crucial to accurate time delay control is dependent on the time of its
operation. Usually it takes a couple of hours or even a whole night to reach a steady state. Figure
3.1 shows the repeatability of the signal before and after the system reaches its stable state.
22
Figure 3.1. (a) Screenshots of the power output of the system before warm-up. (b) Power output of the system after warm-up.
The screenshots on the top display the difference in the repeatability of the power output
between the system in its pre-warmup and post-warmup states. As can be observed, the unstable
state on the top presents very fluctuating patterns in the spectra; the etalon fringes are highly
non-repeatable in the sequential tests. As a consequence, the averaged spectrum can be very
noisy after obtaining multiple scans of the non-repeatable data. By contrast, the screenshot at
Figure 3.1 (b) shows the spectra after the system is stabilized. The etalon fringes in adjacent tests
coincide with one another well with no apparent discrepancy, leading to minimal noise in the
averaged data.
Therefore, before collecting data for analysis in each test, we turn on the system a day
beforehand and run the warm-up scans for a whole night to make sure it becomes fully stabilized
23
once we kick off the test. To prevent the system from being worn down due to over-operation,
we set the emitter bias at 0.0 V during the system warmup and change it back to 0.55 V when
starting the data collection.
3.2.2 Background Measurement
The raw data we obtained from the spectrometer is THz power. To observe the absorbance
spectrum we have to carry out some subtractive calculations and data post-processing. To do this,
we employ the following equation,
0
ln IAI
= −
(1)
Where A is the absorbance,
I and I0 mean the signal and reference intensity respectively.
We employ the equation in the experiments and divide the signal with reference for each test.
The following figure shows the absorption spectrum we obtained from high temperature (500 ℃)
water vapor measurement at different water concentrations following the data subtraction
mentioned above.
Figure 3.2. High temperature water absorption spectrum without background measurement.
24
As can be observed in Figure 3.2, the spectrum turned out to be extremely noisy; the spectral
baselines which should be coinciding or at least close to each other are scattered in a random
manner on the plot. It is hard to carry out further calculation and analysis based on the current
spectra. Therefore, we introduce the background measurement. Instead of going straight into
reference and signal collection, we set off each test with the measurement of background noise.
This is achieved by blocking the receiver with a thin layer of aluminum foil so that the THz
beam couldn't be detected and only the ambient noise signal is collected. Thus rather than
directly dividing the signal with reference, we subtract each value with background noise, as
given by the equation below,
noise
0 noise
ln I IAI I
−= − −
(2)
where Inoise stands for the power intensity of background noise.
Figure 3.3. High temperature water absorption spectrum with background measurement.
In Figure 3.3 we perceive the difference the measurement of background noise had on the high
temperature water. Not only do the baselines coincide with one another, but they also appear
much cleaner in the whole absorption spectrum. Therefore, we apply the baseline measurement
in the tests and carry out the spectral calculations and analysis accordingly.
25
3.2.3 Baseline Drift and Reference Measurement
The water absorption signal in each test is referred to a reference signal which is obtained by
measuring the absorption spectrum of the gas cell filled with nitrogen. The gas cell is evacuated
and flushed with dry N2 for 10 times to remove any residual water from the previous test.
Previously, we started off each set of tests by measuring the background noise and then the N2
reference. Each of the following water tests was referred to the reference test conducted at the
beginning. The results we obtained from this procedure can be found in figure 3.3. The baselines
of the spectra are slant and keep drifting upwards as we increase the water concentration of the
gas cell.
To investigate this phenomenon, we performed repetitive reference tests and refer each reference
to the first reference test to observe the change in the baseline over a period of time. The results
are shown in the figure below.
Figure 3.4. Spectral baseline drift with the time.
In figure 3.4, the bold lines are filtered spectral plots as opposed to the raw data in the back. As
all the reference data are referred to reference 1, we could see as time elapses the baseline tends
to climb upwards. Such phenomenon might be caused by the intrinsic power drift with time in
the system. To mitigate the influence of such systematic defect, we reduce the time interval
26
between the reference and signal tests by conducting a reference test for each humidity test.
Figure 3.5 shows the difference the improvement in the testing procedure had on the quality of
the spectrum.
Figure 3.5. (a) Slanted baseline due to power drift over-time. (b) Rectified baseline after designating reference for each signal test.
The change in the testing procedure makes a significant difference in the baseline of the spectra;
after conducting a reference test for each humidity test, we are able to rectify the tilted baseline
and move on with curve-fitting and analysis.
27
3.2.4 Water Signal Measurement
The water signal measurement is carried out after each reference test. The experimental
procedure is, firstly close the pump valve, leave the gas cell in vacuum condition and record the
pressure value. Then open the valve connecting the septum fitting and inject a known amount of
water (range: 30 μl-450 μl) with a gastight syringe. Wait at least half a minute for the water to
fully evaporate and the pressure to stabilize. Record the pressure again and the increase in the
pressure corresponds to the partial pressure of water vapor or the water volume fraction inside
the gas cell which should be consistent with the amount of water injected.
After the water injection, we fill the gas cell back with nitrogen until it reaches atmospheric
pressure. Then we start the scans. Note that in the initial tests when we raised the temperature,
the water condensation on the gas cell windows was so severe that it was extremely difficult for
the THz beam to pass though the gas cell and get detected at the other end. This is due to the
large temperature difference between the two sides of the gas cell windows. However as the
temperature gets high enough, usually over 120 ℃ , the condensed water would vaporize
completely and the test could be conducted smoothly. Therefore it is highly important that we
inject the water after the gas cell is fully warmed up and the temperature inside is stabilized.
3.3 Effect of Ambient E/M interference
3.3.1 Day-night Fluctuation in Signal Quality
As we ran tests throughout the day, we found an intriguing phenomenon that the scans done
during daytime and evening produce completely different results; the data acquired during the
day turns out to be very noisy while evening tests give us a much cleaner spectrum. Figure 3.6
illustrates how the signal appears totally different in the morning (Figure Figure 3.6 (a)) and the
evening tests (Figure 3.6 (b)). The experimental conditions are kept exactly the same for both
tests, the gas cell temperature being 500 ℃ and the water volume fraction 69.5%.
28
Figure 3.6. Water absorption spectrum obtained in the (a) morning test and (b) evening test.
Same trends occurred when we repeated the tests on different days and it is confirmed that the
scans taking place every day after 5 pm, regardless of workday or weekend, prove to be
significantly lower in noise and show a much cleaner baseline. To look into this phenomenon,
we extracted the raw data of each scan and the following figures show how the discordance in
etalon fringes leads to the untidy spectrum in the morning tests.
In figure 3.7 we can see the power amplitudes for the reference, signal and noise in the morning
and evening tests respectively. The difference in the absorption spectra mainly consists in the
discrepancy of the etalon fringes of the reference and signal tests. The larger the mismatch in the
etalon fringes, the noisier the spectrum gets. Therefore we suspect the interference of some
external electromagnetic source during the daytime is root cause of such patterns in the signal.
29
Figure 3.7. Power amplitude of one sample tests and the resulting absorption spectrum: (a) Power amplitude data for morning test on 43%/m water at 500 ℃. (b) Absorption spectrum for the morning test. (c) Power amplitude data for
evening test on 43%/m water at 500 ℃; (d) Absorption spectrum for the evening test.
3.3.2 Introduction of Faraday Cage
In order to reduce the impact of ambient E/M interference, we built a Faraday cage around the
measurement system. The cage is made of a layer of aluminum foil which is wrapped seamlessly
around the purge box, shielding the system inside against any external E/M interference. The
Faraday cage is grounded with a thin copper wire. Figure 3.8 and 3.9 show the experimental set-
up before and after adding the Faraday cage.
30
Figure 3.8. High temperature water measurement system without Faraday Cage.
Figure 3.9. High temperature water measurement system with Al foil Faraday cage.
To investigate the effect of the Faraday cage, we set the system running day and night
continuously for a whole week and use the root mean square (RMS) method to analyze the
fluctuation of the data. The Faraday cage was used at the beginning and removed for a few days
31
and put back on for the rest of the test. Figure 3.10 illustrates how the RMS of data varies with
time.
Figure 3.10. RMS performance of the system during the whole-week operation.
In figure 3.10, the blue dots represent the tests done with the Faraday cage while the green ones
are obtained after removing the Faraday cage. The test started from 1st May and ended on 7th
May. The day and night time is denoted by the light and dark background. We can distinguish
the day-night cyclic variations in the RMS value and the quality of signal without the cage
appears much inferior to that with the cage; the fluctuation in RMS without the cage is much
higher than that with the cage. Figure 3.11 further demonstrates the difference the Faraday cage
has on the signal quality.
32
Figure 3.11. (a) RMS performance of the system with Faraday cage (b) RMS performance of the system without Faraday cage.
As we can observe from the figures above, in both cases the signal during the day is much
noisier than it is at night. Applying the Faraday cage which serves to block the external E/M
interference, the noise during both day and night is significantly lower than when the cage is
removed. Therefore we conclude that the Faraday cage is necessary in improving the signal
quality and lowering the noise.
33
One problem imposed by the Faraday cage is the heat dissipation of the system. The reflective
metal surface provides a layer of thermal insulation for the system that severely affects the
original cooling and ventilation. The temperature inside the purging chamber immediately goes
up, making it hard to control the temperature surrounding the optical heads below 35 ℃. Thus
instead of wrapping the whole purging box with aluminum foil, we made two small Faraday
cages covering only the emitter and receiver without affecting the heat transfer of the system
(See figure 3.12 for illustration). It has been tested and confirmed that the small separate cages
have the same effect on improving the signal quality.
Figure 3.12. Mini Faraday cage on the receiver.
3.4 Data Processing Techniques
3.4.1 Multiple scans, Averaging and Subtraction
After running the tests and exporting the raw THz power amplitude data, we move on to process
and analyze the data. According to the principles of measurement, the noise of the data is
inversely proportional to the number of repetitive measurement by a factor of n , where n is the
number of scanning. Therefore as a balance between the reduction of noise and the time spent on
each test, we choose to run 10 repetitive scans for each experimental condition and average the
data.
34
The next step is to divide the signal from the reference data. Equation (2) in section 3.2.3
indicates how the absorbance is calculated from the power signal.
3.4.2 Techniques for Removing Etalon Fringes
When we first plot out the absorbance spectrum, it appears extremely noisy most of the time. It is
hard to carry on with further analysis with such data. Therefore smoothing or denoising is needed
to process the data and make it analyzable. One of the most efficient techniques is Fast Fourier
Transform (FFT). Figure 3.11 compares the absorption spectrum before and after the FFT
filtering.
Figure 3.13. Water absorption spectra for different humidity levels (a) before and (b) after FFT filtering.
The reason why we choose the FFT method among all other smoothing techniques is because the
oscillation of the etalon signal in the water spectrum has a distinct frequency. If we do the FFT
analysis on a given spectrum, we would find, as is shown on the left of figure 3.12, the etalon
signal has a dominant frequency of 3.02. Therefore, we apply the low pass filter and set the
cutoff frequency at different values ranging from 1.0-3.5. The plot on the right of figure 3.12
displays how different cutoff frequency affects the filtering results. The optimal cutoff frequency
is found to be approximately 3.
35
Figure 3.14. (a) FFT analysis in water spectrum signal processing. (b) FFT filtering results with different cutoff frequency.
Since applying the filtering of any kind to the data inevitably leads to a compromise in its
accuracy, we are more intent on improving the quality of raw signal than doing signal processing.
As has been mentioned in section 3.3, doing tests in the evening usually produces good results in
terms of signal quality. We thus are inclined to use the evening data if it is good enough without
applying any signal processing.
3.4.3 Curve Fitting and Peak Area Calculation
We carry out our investigation on the spectrum signature of water vapor and its significance in
water quantification based on the Beer Lambert Law which is given by [49],
P0 tot( / ) exp[ ( ) ( )]I I XP S T Lυ υ φ υ= − (3)
or,
P0 totln( / ) ( ) ( )I I XP S T Lυ υ φ υ− = (4)
where 0( / )I I υ is the measured signal intensity ratio; X is the mole fraction of the absorbing
species; Ptot is the total absolute system pressure in atm; P ( )S Tυ is the line strength of the
transition in cm-2atm-1; L is the absorption path length in cm; ( )φ υ is the line shape function of
the transition in cm-1.
36
The formula on the left denotes the absorbance whose integration is given by,
P P0 tot totln( / ) ( ) ( ) ( ) ( )I I XP S T L XP S Tυ υ υφ υ φ υ− = =∫ ∫ ∫ (5)
where
( ) 1φ υ =∫ (6)
Thus we obtain,
P0 totln( / ) ( )PeakArea I I XP S Tυ υ= − =∫ (7)
where 0ln( / )I I υ−∫ stands for the absorbance peak area and totXP combined stands for the
volumetric concentration of species. As P ( )S Tυ is fixed at a particular wavenumber, the peak area
is in linear correlation with the gas concentration.
To verify the theory, we need to analyze the water spectrum from both simulation and
experiments based on curve-fitting and area integration. According to the theory in spectroscopy,
we apply Voigt-fitting to the absorbance peaks. OriginPro 8.0 is used as the data analysis tool for
curve-fitting and area calculations. Figure 3.13 show the curve-fitting results for simulation and
experimental data. In Chapter 4 results and discussion, we will compare the simulation line
strength with the experimental results in detail.
Figure 3.15. Voigt fitting results for (a) simulation and (b) experimental water absorption spectrum at 500 ℃ and at 0%/m water volume fraction.
37
3.5 Summary
In this chapter, we presented the method development of current research, which comprises a
major part of the whole project. As the application of THz technology in gas spectroscopy is still
in its infancy and the system we purchased has been modified for the current study, we have
been encountering many an issue along the way.
On the hardware side, we have dealt with the system stabilization, the background noise effect,
the power drift and the E/M interference to the system. On the software side, we addressed the
signal processing and data analysis. On the whole, we are able to identify various problems
coming up during the tests, carry out trouble-shooting and finding the best possible way to
optimize the situation.
38
Chapter 4 Results and Discussion
4.1 Introduction
In this chapter, we will present and discuss the results obtained from the experiments following
the experimental procedure and data analysis method developed in the previous chapters. The
results consist of two categories, the spectral simulation and the experimental study. Theoretical
simulations were carried out first by employing the commercial spectral simulator SpectraCalc
which draws on data from the 2008 edition of HITRAN database. The simulation results
essentially serve as a guidance for us to locate and pinpoint the new peaks which emerge at high
temperatures and also to justify the experimental outcome. By comparing the experimental
results with simulation, we will be able to find the correlation between the water concentration
and peak absorption features. Beam scattering experiments were conducted as an auxiliary test to
justify the motivation mentioned in the first chapter. The characteristics of the emerging new
peaks would also be discussed. These results will help us in the use of THz spectroscopy to
quantify water vapor produced from combustion in industrial processes.
4.2 HITRAN Spectral Simulation
4.2.1 Modelling of Water Spectrum in Gas Cell
The simulation conditions are set exactly the same as the experimental study. Details of the
simulation settings are displayed in Table 4.1.
Although we control the crucial factors exactly the same as the experimental condition, the
current simulation, as of the experimental study, is in fact a simplified version of the real
industrial processes. For one thing, the open optical path is constantly purged with dry air in the
experiment and is eliminated in the spectral simulation. Whereas in industry, the path might be
exposed to room humidity and the spectrum would change to some extent accordingly. This will
be further addressed in the coming sections and both simulation of single gas cell and 2-cells are
carried out and compared to investigate the possible challenges in industry.
39
Table 4.1 Simulation conditions for the water absorption experiment
Simulation Conditions
Waveband 15-27 cm-1
Pressure 1013.25 mbar
Temperature 773K
Gas H2O+N2
Path Length 100 cm
Volume Mixing Rate 0.791
4.2.2 Variation of Water Absorption Spectra with Temperature
Figure 4.1 shows the simulation results for water absorbance at temperatures ranging from 25 ℃
to 500 ℃. The water volume mixing ratio is set as 10%.
Figure 4.1. (a) Variation of water absorbance spectra with the increasing temperature; (b) Zoom-in spectra at the small new peak.
40
In the figure we can observe the peaks due to rotational transmission of water molecules and that
as the temperature goes up the large peaks at 18.57 cm-1 (0.573 THz) and 24.53cm-1 (0.88 THz)
shrink with the temperature in both their heights and widths. The small peak at 21.94cm-1 (0.667
THz) emerges at high temperature and grows as the temperature increases.
4.2.3 Variation of Water Absorption Spectra with Humidity
According to the Beer-Lambert law, the water absorbance peak area is in proportion to the water
concentration. We thus performed a series of simulation at the same temperature 500 ℃ with
water concentration varying from 5%-m to 80%-m which corresponds to 1.25%-m to 20%-m in
a typical 4-meter-long off-gas duct in the industrial furnace. Figure 4.2 shows the results of
variation of water absorbance with humidity.
Figure 4.2. Effect of water volume fraction on the water absorbance spectrum
In the figure, the water absorbance peak is growing monotonically with the increasing humidity.
In the absorption spectrum the large peaks are prone to saturation at higher concentrations;
instead of growing higher, the absorption peaks stop at 1 and start extending horizontally and
become fatter. By contrast, in the absorbance spectrum, the peak keeps growing upwards with
the humidity instead of stretching too much horizontally. Same trend applies to both the large
41
and small peaks. Another remarkable feature is the rise of baseline. As the humidity increases,
the water spectrum goes up and encloses the curves underneath which is due to the tail of the
major peaks.
4.2.4 Correlation between Absorbance Peak Area and Water Concentration
Following the methodology described in Chapter 3, we perform curve fitting and analysis for the
water absorbance spectrum. Figure 4.3 demonstrates the sample calculation for the large peak at
18.57 cm-1 (0.557THz).
Figure 4.3. Sample calculation for the large peak at 18.57 cm-1(0.573 THz).
The transmittance data obtained from the spectral simulator is processed based on equation (2) to
get the absorbance spectrum. Then we use two different means to perform curve-fitting for the
absorbance peaks. The curve-fitting shown on the top is achieved by the multiple-peaks Voigt
fitting algorithm in OriginLab. With the fitted curve we do the integration and get the peak area.
Plotting out the peak area against the water concentration we perceive a very near linear
correlation (details shown in Figure 4.5). The slope is calculated to be 0.04458 and the adjusted
R square value is approximating 1, indicating the goodness of fitting is very satisfactory.
The fitting at the bottom is achieved by a self-developed subroutine written in Matlab. The Voigt
profile is generated by an approximation of the Voigt function. Instead of curve-fitting all four
42
peaks in the whole spectrum, the fitting is applied to individual peaks and integration obtained
accordingly. Similar results of the peak area-water concentration correlation is observed
following this curve-fitting method. As the accuracy of the fitting is better with the OriginLab
method, we choose to use the OriginLab software to perform the curve-fitting and further
analysis. The same sample calculations for the small new peak at 0.667 THz is shown in figure
4.4. The linear correlation of the water concentration and peak area for both the large and small
peaks are displayed in figure 4.6 (a) and (b) respectively.
Figure 4.4. Sample calculation for the small new peak at 21.94 cm-1(0.667 THz).
43
Figure 4.5. Linear fitting of the water concentration-peak area correlation for (a) Large absorbance peak at 0.557 THz.
(b) Small peak at 0.667THz.
4.2.5 Two-cell simulation and peak area-concentration correlation
To explore the possibility of industrial application, we carry out the 2-cell simulation as a
simplified version of the real industry. Shown below is the schematic diagram of the 2-cell
simulation.
Figure 4.6. Schematic diagram of the two-cell water absorption simulation.
The 0.2 m room temperature cell is recognized as the open path the laser travels outside the
furnace. The 1 m high temperature cell is seen as the harsh environment in the furnace. The
humidity for the room cell is selected as the atmospheric humidity on a typical day (23 ℃, 1.3 %
water) and the humidity inside the high temperature cell is varied from 5%-90%.
44
Figure 4.7. Variation of water absorbance spectrum with humidity in two-cell simulation.
The same trend of variation of water absorbance spectrum with humidity occurs in the two-cell
simulation; all absorbance peaks grow monotonically with the increasing humidity. The room
temperature has a slight effect on the profile by raising the baseline and consequently the entire
spectra higher. The absorbance peaks appear wider and higher in the 2-cell simulation spectra.
After applying the curve-fitting and integration we obtain the correlation between the water
concentration and the peak area for both large (a) and small peaks (b), as shown in figure 4.8.
45
Figure 4.8. Correlation between water concentration and peak based in the two-cell simulation (a) Large absorbance
peak at 0.557 THz; (b) Small peak at 0.667THz.
From figure 4.8 we see that the peak area in the two-cell simulation also displays a linear
correlation with the water volume fraction for both the large and small peaks. The slopes of the
fitted lines match well with those in the single cell simulation. A slight offset of the line can be
observed on the plot for the large peak while for the new peak, the fitted line goes across point
zero as has been found in the single-cell simulation. Therefore we conclude that the room
humidity exerts insignificant influence on the water concentration-peak area correlation
especially for the new peak that shows up only at high temperature.
4.3 Experimental Study
4.3.1 Beam scattering test with different light sources
Before carrying out experiments to verify the simulation results, we first conduct the beam
scattering test as a recapitulation of what we stressed as the advantage of THz in the motivation
of the research. Three light sources are employed, namely the THz source, the visible and near-
infrared lasers. The following figure is a photo showing the experimental set-up of the beam
scattering test.
46
Figure 4.9. Experimental set-up for the beam scattering test with different light sources.
In the beam scattering testing system, the emitters and receivers of the three light sources are
placed at the right and left side of the purging box respectively. In the middle there put a plastic
tube with a vibrating sieve dropping fine Talcum powder with an estimated particle size of 0.01
mm.
As the purpose of the beam scattering test is to verify THz capability of penetrating dusty
environment, we select four typical frequencies, namely 0.2 THz, 0.6 THz, 1.0 THz and 1.25THz
and carry out four tests accordingly to compare the transmittance in the THz domain with that in
the visible and near-infrared. Results are shown in figure 4.10 where the green dots denote the
THz beam, the red and blue lines stand for the visible and near-infrared respectively. The steep
dip of signal marks the beginning of the particle drop. As we can observe from all four plots, the
visible and NIR transmittance dropped to 50% of its original value due to beam scattering loss,
or particularly MIE scattering. Whereas for THz, especially in lower frequency, there is no or
only a slight dip of signal. At higher frequency some scattering occurred, but transmittance is
always at or above that of the Visible and NIR lasers. Note that due to limitations of the THz
instrument, at higher frequency the power drops significantly, see Figure 3.1, which leads to the
uncertainty of data at 1.0 THz and 1.25 THz.
47
Figure 4.10. Transmittance of the difference light sources in the beam scattering test.
4.3.2 Variation of Water Absorbance Spectra with Humidity
Here we start the investigation on the relationship between the water absorption spectra with the
increasing humidity. Measurements are carried out when the gas cell is heated up to 500 ℃ at the
water concentrations ranging from 8%-90%. Results are displayed in figure 4.11 and 4.12. In the
aspects of peak location and height, the experimental spectra show good agreement with the
simulation results.
48
Figure 4.11. Experimental water spectra varying with increasing humidity at 500 ℃.
Figure 4.12. Variation of water absorbance peak at 18.57 cm-1(0.573 THz) with humidity.
49
4.3.3 Experimental Correlation Between Absorbance Peak Area and Water Concentration
After obtaining high temperature water absorbance spectra at different humidity we move on to
analyze the correlation between the peak area and water volume fraction. The figure below
demonstrates the calculation procedure for the experimental data.
Figure 4.13. Sample calculation of experimental data for the large peak at 18.57 cm-1(0.573 THz): (a) Power amplitude data obtained from one scan. (b) Absorbance Spectrum. (c) Multiple peak Voigt Curve-fitting. (d)
Correlation between water volume fraction and peak area.
In figure 4.13 we first export the power signal data from the multiple scans. Averaging and
subtraction are done afterwards to obtain the absorbance spectrum. Then we come to the
multiple-peak curve-fitting and peak area integration. Finally the peak area is plotted against the
water concentration and as we expected, there exists a linear correlation between the two
according to the theoretical prediction. More repetitive tests are performed to confirm the results.
Figure 4.14 show the experimental results obtained from tests on different days and compare
them with the HITRAN prediction.
50
Figure 4.14. Experimental results versus HITRAN prediction of tests: (a) Correlation between injected water volume and peak area for peak at 0.557THz. (b) Correlation between injected water volume and peak area for peak at
0.753THz.
There is a good agreement between the experimental and theoretical results for the two large
peaks at 0.557 THz and 0.753 THz. If we do a linear fitting for the peak of interest (0.557 THz)
and compare the results with the predicted linear correlation, we found the slopes of the
experimental (0.04413) and simulation (0.04458) linear correlations match with each other
within 1%.
Due to low output power at the new peak position (0.67 THz), we used two parabolic mirrors
(one with 4’ focal length, the other 1’ focal length, with protected gold coating) to collimate the
THz beam and reduce the diffraction. Figure 4.15 illustrates the upgraded measurement system
layout. It is found that with the beam collimation the THz signal strength increase by 100 times.
As a consequence, the water absorbance spectra appear much cleaner and the small new peak
clearly visible, as can be observed in Figure 4.16. After the data processing we obtain the linear
correlation between the water concentration and the new peak area shown in figure 4.17. The
slope is calculated as 0.00307 which is about 8% different from the simulated value 0.00284.
This is a promising lead that sheds light on the possibility of quantifying high temperature water
vapor in industrial processes by small peak measurement.
51
Figure 4.15. THz high temperature water measurement system parabolic mirror upgrade.
Figure 4.16. High temperature water absorbance spectra measured after the system upgrade.
52
Figure 4.17. Experimental data for small peak at 21.94cm-1 (0.667 THz) and linear fitting results.
4.4 Summary
In this chapter, we present the results obtained from the spectral simulation and experimental
study. In both simulation and experiments, it is found that the strong absorption peaks at 0.557
THz (18.57 cm-1) and 0.753 THz (24.85 cm-1) shrink with the increasing temperature, and new
peaks at 0.667 THz (21.94 cm-1), 0.88 THz (29.04 cm-1) emerge at high temperature. This
phenomenon can be explained by the change in the rotational rates and state populations
triggered by the change in temperature. The absorbance spectra have been collected within the
frequency range of 0.5-0.8 THz at a series of humidity levels ranging from 8% to 70%. We
observed that absorbance peaks grow with the humidity. The areas underneath the peaks of
interest are calculated by integration and hence we are able to plot out the peak area against a
series of water concentrations.
For the gas cell humidity ranging from 5% to 78%, the absorbance peak area at 18.57 cm-1
(0.557 THz) and 24.85 cm-1(0.753THz) displays good linear correlation with the water
concentration, which can be justified by the Beer-Lambert Law. The simulation results are then
verified by experimental study; a clean linear correlation between the water concentration and
the absorbance peak area was found and displayed in figure 3b. The slopes of the experimental
(0.04413) and simulation (0.04458) linear correlations match with each other within 1%.
53
To increase the signal strength of the THz system, we added two parabolic mirrors for beam
collimation. The power is improved significantly and the experimental results for the small new
peak at 0.667 THz proved to be in good agreement with the theoretical prediction. Such leads
give us more confidence in the future industrial application.
54
Chapter 5 Conclusion
5.1 Summary
Terahertz (THz) radiation, which fills the gap in the electromagnetic spectrum between the
microwave and the infrared light, has the advantage of transmitting through gases with high
particle loading with reduced light scattering. THz has remained quite unexplored until most
recent years when fast advancements in photonics took place, making it possible to generate and
detect THz beams with much more efficiency.
In the current research, we focus on the measurement of water vapor at high temperatures using
THz spectroscopy. A monochromatic continuous wave THz source with wide tunable frequency
range and high resolution is employed. The heating system we established allows us to carry out
experiments at temperatures from room temperature to as high as 500 ℃. The gas cell is 1 meter
long and made of a stainless steel tube at the end of which mounted two z-cut quartz windows.
The optical path is constantly purged with dry air to remove the room humidity and also to
increase heat dissipation and keep the temperature inside the purging chamber stable. Two
separate Faraday cages are applied on the optical heads to block the E/M interference. The
system enables the measurement of water vapor of different concentrations at temperatures from
room temperature to 500℃.
As the application of THz technology in gas spectroscopy is still in its infancy, we encountered
many issues when carrying out our research. On the hardware side, we dealt with the system
stabilization, the background noise effect, the power drift and the E/M interference to the system.
On the software side, we addressed the signal processing and data analysis. Up to now, we are
able to identify various problems coming up during the tests, carry out trouble-shooting and
finding the best possible way to optimize the situation.
HITRAN simulation for frequency range 0.5-0.8THz is first carried out by employing the
commercial spectral simulator SpectraCalc which draws on data from the 2008 edition of
HITRAN database. Results indicate that the strong absorption peaks at 0.557 THz (18.57 cm-1)
and 0.753 THz ( 24.85 cm-1) shrink with the increasing temperature, whereas new peaks at 0.667
55
THz (21.94 cm-1), 0.88 THz (29.04 cm-1) emerge at high temperatures and grow as the
temperature increases.
For the gas cell humidity ranging from 5% to 78%, the absorbance peak area at 18.57 cm-1
(0.557 THz) and 24.85 cm-1(0.753THz) displays good linear correlation with the water
concentration, which is in consistence with the Beer-Lambert Law. The simulation results are
then verified by experimental study; a clean linear correlation between the water concentration
and the absorbance peak area was observed. The slopes of the experimental (0.04413) and
simulation (0.04458) linear correlations match with each other within 1%.
Two parabolic mirrors for beam collimation were added to increase the signal strength of the
THz system. The power is improved significantly and the experimental results for the small new
peak at 0.667 THz proved to be in good agreement with the theoretical prediction. Such leads
give us more confidence in the future industrial application.
5.2 Industrial application
In a typical industrial furnace, the optical path length is usually 4 meters. The water
concentration in a steel making plate ranges from 15%-25% which corresponds to 60%-100% in
a 1-m measurement system. The temperature is around 1200-1500 ℃ which is much higher than
what can be achieved in the current laboratory facility and beyond the calculation scope of
HITRAN.
As has been mentioned in the motivation, THz provides a non-destructive, in-situ off-gas
detection and analysis method. It also has the advantage over other light sources for being
capable of penetrating environment heavily-laden with particles. Many gaseous species have
spectral signatures in the THz range and among them, water vapor shows distinct absorption
lines that can be utilized for diagnostic purpose. Moreover, at high temperature, new absorption
peaks emerge in the water spectrum and grow with the increasing humidity. This feature makes it
possible to quantify water vapor in the furnace without the influence of the atmospheric humidity
in the optical path.
However, since the continuous wave THz technology is still immature, there are considerable
challenges that need to be addressed before its industrial application. For one thing, in the
56
industrial configuration, the optical path is too long for the THz beam to travel without suffering
significant diffraction. Parabolic mirrors could be used to improve collimation purpose which
will increase cost and add to the complexity. Another challenge is that the THz instrument is
extremely delicate and susceptible to the harsh environment and exterior E/M interference. The
lasers and optical heads must be placed in air-conditioned environment to ensure its good
performance. Also the strike of electric arc in a steel-making furnace poses substantial hazard for
the THz instrument and good E/M protection must be considered. Last but not least, the current
system is very bulky and heavy and can be extremely vulnerable to vibration and damage in the
industrial application. It is highly desirable to have a more compact and rugged system.
5.3 Current Challenges
The E/M interference on the system is one of the most indecipherable challenges under
investigation. Although the application of Faraday cage makes a difference in improving the
signal quality, the quality of the data is quite unpredictable and the day-night fluctuation in signal
still remains despite the presence of the Al shield. In order to obtain good data for analysis, we
are obliged to carry out evening tests as in general the evening data appear cleaner than day-time
data.
Although the power strength is improved substantially within the frequency range of interest, the
signal drops drastically beyond 0.8 THz, posing huge challenges for investigation at higher
frequency range.
5.4 Recommendations for Future Work
In the future study we would like to continue focusing on the explorative study in the laboratory
and also plan out field trials when opportunity comes.
The laboratory work includes exploring the temperature dependency of line strength and line-
width of water absorption peaks. Also, as the strengths of peaks like 0.557 THz and 0.753 THz
decrease with the increasing temperature and the strength of peaks at 0.667 THz and 0.88 THz
increase with temperature, it could be used to measure the temperature. Other combustion gases
such as CO will be investigated after some modification to the current facility.
57
For the filed trial, we would like to collaborate with the THz instrument supplier for the
development of a more compact and robust system and the improvement of its overall
performance. Once the technology is ready, we would install the system in the off-gas duct of a
furnace and test its performance in the real steel-making process.
58
References [1] T. Aizawa, T. Kamimoto, and T. Tamaru, “Measurements of OH radical concentration in combustion
environments by wavelength-modulation spectroscopy with a 1.55-micron distributed-feedback diode
laser.,” Appl. Opt., vol. 38, no. 9, pp. 1733–41, Mar. 1999.
[2] T. K. Balasubramanian, A. P. Mishra, and R. J. Kshirsagar, “Absorption intensities of rovibronic transitions
in the A-band of 16O2: Analysis and calculation of magnetic dipole and electric quadrupole contributions,”
J. Quant. Spectrosc. Radiat. Transf., vol. 133, pp. 91–98, Jan. 2014.
[3] S. I. Chou, D. S. Baer, and R. K. Hanson, “Diode laser absorption measurements of CH(3)Cl and CH(4)
near 1.65 micron.,” Appl. Opt., vol. 36, no. 15, pp. 3288–93, May 1997.
[4] J. Graul and T. Lilly, “Coherent Rayleigh-Brillouin scattering measurement of atmospheric atomic and
molecular gas temperature,” Opt. Express, vol. 22, no. 17, p. 20117, Aug. 2014.
[5] a. Grosch, H. Wackerbarth, O. Thiele, T. Berg, and L. Beckmann, “Infrared spectroscopic concentration
measurements of carbon dioxide and gaseous water in harsh environments with a fiber optical sensor by
using the HITEMP database,” J. Quant. Spectrosc. Radiat. Transf., vol. 133, pp. 106–116, Jan. 2014.
[6] R. M. Mihalcea, D. S. Baer, and R. K. Hanson, “Tunable diode-laser absorption measurements of NO(2)
near 670 and 395 nm.,” Appl. Opt., vol. 35, no. 21, pp. 4059–64, Jul. 1996.
[7] R. M. Mihalcea, D. S. Baer, and R. K. Hanson, “Diode laser sensor for measurements of CO, CO(2), and
CH(4) in combustion flows.,” Appl. Opt., vol. 36, no. 33, pp. 8745–52, Nov. 1997.
[8] V. Nagali, S. Chou, and D. Baer, “Tunable diode-laser absorption measurements of methane at elevated
temperatures,” Appl. Opt., vol. 35, no. 21, 1996.
[9] a. Padmanabhan, T. Tzanetakis, a. Chanda, and M. J. Thomson, “Study of pressure broadening effects of H2
on CO2 and CO in the near infrared region between 6317 and 6335cm−1 at room temperature,” J. Quant.
Spectrosc. Radiat. Transf., vol. 133, pp. 81–90, Jan. 2014.
[10] A. Pogány, O. Ott, O. Werhahn, and V. Ebert, “Towards traceability in CO2 line strength measurements by
TDLAS at 2.7µm,” J. Quant. Spectrosc. Radiat. Transf., vol. 130, pp. 147–157, Nov. 2013.
[11] D. M. Slocum, E. J. Slingerland, R. H. Giles, and T. M. Goyette, “Atmospheric absorption of terahertz
radiation and water vapor continuum effects,” J. Quant. Spectrosc. Radiat. Transf., vol. 127, pp. 49–63, Sep.
2013.
59
[12] D. M. Sonnenfroh and M. G. Allen, “Absorption measurements of the second overtone band of NO in
ambient and combustion gases with a 1.8-mum room-temperature diode laser.,” Appl. Opt., vol. 36, no. 30,
pp. 7970–7, Oct. 1997.
[13] J. Ge, R. Wang, S. Hu, N. Dai, D. Li, H. Ma, and G. Ma, “Investigation of the strengths of vapor rotational
absorption lines with different humidity by far infrared absorption spectra,” 2009, vol. 7385, p. 73851L–
73851L–11.
[14] H. Hoshina, T. Seta, T. Iwamoto, I. Hosako, C. Otani, and Y. Kasai, “Precise measurement of pressure
broadening parameters for water vapor with a terahertz time-domain spectrometer,” J. Quant. Spectrosc.
Radiat. Transf., vol. 109, no. 12–13, pp. 2303–2314, Aug. 2008.
[15] V. B. Podobedov, D. F. Plusquellic, and G. T. Fraser, “Investigation of the water-vapor continuum in the
THz region using a multipass cell,” J. Quant. Spectrosc. Radiat. Transf., vol. 91, no. 3, pp. 287–295, Mar.
2005.
[16] K. Su, L. Moeller, R. B. Barat, and J. F. Federici, “Experimental comparison of terahertz and infrared data
signal attenuation in dust clouds.,” J. Opt. Soc. Am. A. Opt. Image Sci. Vis., vol. 29, no. 11, pp. 2360–6, Nov.
2012.
[17] V. B. Podobedov, D. F. Plusquellic, K. E. Siegrist, G. T. Fraser, Q. Ma, and R. H. Tipping, “New
measurements of the water vapor continuum in the region from 0.3 to 2.7THz,” J. Quant. Spectrosc. Radiat.
Transf., vol. 109, no. 3, pp. 458–467, Feb. 2008.
[18] V. . Podobedov, D. . Plusquellic, and G. . Fraser, “THz laser study of self-pressure and temperature
broadening and shifts of water vapor lines for pressures up to,” J. Quant. Spectrosc. Radiat. Transf., vol. 87,
no. 3–4, pp. 377–385, Sep. 2004.
[19] Y. Yang, M. Mandehgar, and D. Grischkowsky, “Time domain measurement of the THz refractivity of
water vapor.,” Opt. Express, vol. 20, no. 24, pp. 26208–18, Nov. 2012.
[20] Y. Yang, A. Shutler, and D. Grischkowsky, “Measurement of the transmission of the atmosphere from 0.2 to
2 THz.,” Opt. Express, vol. 19, no. 9, pp. 8830–8, Apr. 2011.
[21] B. J. Drouin, S. Yu, J. C. Pearson, and H. Gupta, “Terahertz spectroscopy for space applications: 2.5–
2.7THz spectra of HD, H2O and NH3,” J. Mol. Struct., vol. 1006, no. 1–3, pp. 2–12, Dec. 2011.
[22] S. Eliet, A. Cuisset, M. Guinet, F. Hindle, G. Mouret, R. Bocquet, and J. Demaison, “Rotational spectrum of
formaldehyde reinvestigated using a photomixing THz synthesizer,” J. Mol. Spectrosc., vol. 279, pp. 12–15,
Sep. 2012.
60
[23] X. Xin, H. Altan, a. Saint, D. Matten, and R. R. Alfano, “Terahertz absorption spectrum of para and ortho
water vapors at different humidities at room temperature,” J. Appl. Phys., vol. 100, no. 9, p. 094905, 2006.
[24] S. Yu, J. C. Pearson, and B. J. Drouin, “Terahertz spectroscopy of water in its second triad,” J. Mol.
Spectrosc., vol. 288, pp. 7–10, Jun. 2013.
[25] S. Yu, J. C. Pearson, B. J. Drouin, M.-A. Martin-Drumel, O. Pirali, M. Vervloet, L. H. Coudert, H. S. P.
Müller, and S. Brünken, “Measurement and analysis of new terahertz and far-infrared spectra of high
temperature water,” J. Mol. Spectrosc., vol. 279, pp. 16–25, Sep. 2012.
[26] R. Engelbrecht, “A compact NIR fiber-optic diode laser spectrometer for CO and CO(2): analysis of
observed 2f wavelength modulation spectroscopy line shapes.,” Spectrochim. Acta. A. Mol. Biomol.
Spectrosc., vol. 60, no. 14, pp. 3291–8, Dec. 2004.
[27] a. Farooq, J. B. Jeffries, and R. K. Hanson, “CO2 concentration and temperature sensor for combustion
gases using diode-laser absorption near 2.7 μm,” Appl. Phys. B, vol. 90, no. 3–4, pp. 619–628, Jan. 2008.
[28] K. Kohse-Höinghaus, R. S. Barlow, M. Aldén, and J. Wolfrum, “Combustion at the focus: laser diagnostics
and control,” Proc. Combust. Inst., vol. 30, no. 1, pp. 89–123, Jan. 2005.
[29] X. Zhou, X. Liu, J. B. Jeffries, and R. K. Hanson, “Development of a sensor for temperature and water
concentration in combustion gases using a single tunable diode laser,” Meas. Sci. Technol., vol. 14, no. 8, pp.
1459–1468, Aug. 2003.
[30] B. L. Upschulte, D. M. Sonnenfroh, and M. G. Allen, “Measurements of CO, CO2, OH, and H2O in room-
temperature and combustion gases by use of a broadly current-tuned multisection InGaAsP diode laser.,”
Appl. Opt., vol. 38, no. 9, pp. 1506–12, Mar. 1999.
[31] T. Uno and H. Tabata, “In situ Measurement of Combustion Gas Using Terahertz Time Domain
Spectroscopy Setup for Gas Phase Spectroscopy and Measurement of Solid Sample,” Jpn. J. Appl. Phys.,
vol. 49, no. 4, p. 04DL17, Apr. 2010.
[32] K. Schäfer, J. Heland, D. H. Lister, C. W. Wilson, R. J. Howes, R. S. Falk, E. Lindermeir, M. Birk, G.
Wagner, P. Haschberger, M. Bernard, O. Legras, P. Wiesen, R. Kurtenbach, K. J. Brockmann, V. Kriesche,
M. Hilton, G. Bishop, R. Clarke, J. Workman, M. Caola, R. Geatches, R. Burrows, J. D. Black, P. Hervé,
and J. Vally, “Nonintrusive optical measurements of aircraft engine exhaust emissions and comparison with
standard intrusive techniques.,” Appl. Opt., vol. 39, no. 3, pp. 441–55, Jan. 2000.
[33] G. B. Rieker, H. Li, X. Liu, J. T. C. Liu, J. B. Jeffries, R. K. Hanson, M. G. Allen, S. D. Wehe, P. a. Mulhall,
H. S. Kindle, a. Kakuho, K. R. Sholes, T. Matsuura, and S. Takatani, “Rapid measurements of temperature
61
and H2O concentration in IC engines with a spark plug-mounted diode laser sensor,” Proc. Combust. Inst.,
vol. 31, no. 2, pp. 3041–3049, Jan. 2007.
[34] C. McDonagh, C. S. Burke, and B. D. MacCraith, “Optical chemical sensors.,” Chem. Rev., vol. 108, no. 2,
pp. 400–22, Feb. 2008.
[35] X. Liu and Y. Xu, “New rovibrational bands of the Ar–H2O complex at the ν2 bend region of H2O,” J. Mol.
Spectrosc., vol. 301, pp. 1–8, Jul. 2014.
[36] Y.-S. Lee, Principles of Terahertz Science and Technology: Proceedings of the International Conference,
Held in Mainz, Germany, June 5-9, 1979, vol. 170. Springer, 2009.
[37] A. Deninger, “State-of-the-Art in cw Photomixing Systems,” in International Conference on Fibre Optics
and Photonics, 2012, p. T3A.2.
[38] J. T. C. Liu, J. B. Jeffries, and R. K. Hanson, “Wavelength modulation absorption spectroscopy with 2 f
detection using multiplexed diode lasers for rapid temperature measurements in gaseous flows,” Appl. Phys.
B Lasers Opt., vol. 78, no. 3–4, pp. 503–511, Feb. 2004.
[39] J. T. C. Liu, J. B. Jeffries, and R. K. Hanson, “Large-modulation-depth 2f spectroscopy with diode lasers for
rapid temperature and species measurements in gases with blended and broadened spectra.,” Appl. Opt., vol.
43, no. 35, pp. 6500–9, Dec. 2004.
[40] X. Liu, J. B. Jeffries, R. K. Hanson, K. M. Hinckley, and M. a. Woodmansee, “Development of a tunable
diode laser sensor for measurements of gas turbine exhaust temperature,” Appl. Phys. B, vol. 82, no. 3, pp.
469–478, Dec. 2005.
[41] S. Martens, B. Gompf, and M. Dressel, “Characterization of continuous-wave terahertz sources: laser
mixing versus backward-wave oscillators.,” Appl. Opt., vol. 48, no. 29, pp. 5490–6, Oct. 2009.
[42] A. Redo-Sanchez, N. Laman, B. Schulkin, and T. Tongue, “Review of Terahertz Technology Readiness
Assessment and Applications,” J. Infrared, Millimeter, Terahertz Waves, vol. 34, no. 9, pp. 500–518, Jul.
2013.
[43] G. Whitenett, G. Stewart, H. Yu, and B. Culshaw, “Investigation of a Tuneable Mode-Locked Fiber Laser
for Application to Multipoint Gas Spectroscopy,” J. Light. Technol., vol. 22, no. 3, pp. 813–819, Mar. 2004.
[44] X. C. Zhang, “Terahertz wave imaging: horizons and hurdles.,” Phys. Med. Biol., vol. 47, no. 21, pp. 3667–
77, Nov. 2002.
62
[45] T. Kampfrath, M. Battiato, P. Maldonado, G. Eilers, J. Nötzold, S. Mährlein, V. Zbarsky, F. Freimuth, Y.
Mokrousov, S. Blügel, M. Wolf, I. Radu, P. M. Oppeneer, and M. Münzenberg, “Terahertz spin current
pulses controlled by magnetic heterostructures.,” Nat. Nanotechnol., vol. 8, no. 4, pp. 256–60, Apr. 2013.
[46] M. Tonouchi, “Cutting-edge terahertz technology,” Nat. Photonics, vol. 1, no. 2, pp. 97–105, Feb. 2007.
[47] T. Nagatsuma, A. Kaino, S. Hisatake, K. Ajito, and H. Song, “Continuous-wave Terahertz Spectroscopy
System Based on Photodiodes,” pp. 390–394, 2010.
[48] A. Dobroiu, M. Yamashita, Y. N. Ohshima, Y. Morita, C. Otani, and K. Kawase, “Terahertz imaging system
based on a backward-wave oscillator.,” Appl. Opt., vol. 43, no. 30, pp. 5637–46, Oct. 2004.
[49] M. G. Allen, “Diode laser absorption sensors for gas-dynamic and combustion flows.,” Meas. Sci. Technol.,
vol. 9, no. 4, pp. 545–62, Apr. 1998.
[50] N. Docquier and S. Candel, “Combustion control and sensors: a review,” Prog. Energy Combust. Sci., vol.
28, no. 2, pp. 107–150, Jan. 2002.
[51] V. Ebert, T. Fernholz, C. Giesemann, H. Pitz, H. Teichert, J. Wolfrum, and H. Jaritz, “SIMULTANEOUS
DIODE-LASER-BASED IN SITU DETECTION OF MULTIPLE SPECIES AND TEMPERATURE IN A
GAS-FIRED POWER PLANT,” vol. 28, pp. 423–430, 2000.
[52] “Lindarc | More,” 25-Aug-2014. [Online]. Available: http://www.more-oxy.com/en/products/lindarc.html.
[53] TeraView Documentation, received 7-August-2014 .
63
Appendices Appendix A. Humidity and Partial Pressure Conversion Table
Temperature/ K Water Volume/ μl Partial Pressure/ %
773.00 50.00 8.692984107
773.00 100 17.38596821
773.00 150 26.07895232
773.00 200 34.77193643
773.00 250 43.46492053
773.00 300 52.15790464
773.00 350 60.85088875
773.00 400 69.54387285
773.00 450 78.23685696
773.00 500 86.92984107
773.00 550 95.62282517
67
Appendix C. MatLab Codes for Data Processing
Main Program: clc close all clear all %% For March 13th 14th tests different humidities with different references name0={'bg2', 'ref50_2', 'sig50_2','ref80_2','sig80_2','ref100_2','sig100_2','ref150_2','sig150_2','ref180_2','sig180_2','ref200_2', 'sig200_2','ref250_2','sig250_2','ref280_2','sig280_2','ref300_2', 'sig300_2',... 'ref350_2','sig350_2','ref380_2','sig380_2','ref400_2', 'sig400_2','ref450_2'};%,'sig450_2','ref480_2','sig480_2','ref550_2','sig550_ for k=1:length(name0) name{k}=name0{k}; end co='cbrkbgkkmgrgkmcbkrcybbyrkbyrk'; % Reading data and Averaging for k=1:length(name0) [Xave{k},Yave{k}]=xiaohuifun3(name{k},co(k)); end figure for k=[1 10 11] semilogy(Xave{k},Yave{k},co(k)); hold on end for k=3:2:length(name0) y{k-2}=-log((Yave{k}-Yave{1})./(Yave{k-1}-Yave{1})); % Subtraction] yy{k-2} = sgolayfilt(y{k-2},3,17); % SGolay Smooth rms_y(k-2)=sqrt(var(cell2mat(y(k-2)))); end rms_y=rms_y'; figure for k=[19 21 23 ] plot(Xave{1}./0.0303,y{k},co(k)); hold on end figure for k=[19 21 23 ] plot(Xave{1}./0.0303,yy{k},co(k)); hold on
68
end yy=cell2mat(yy); C=cell2mat(Yave); C=[Xave{1}./0.0303 C]; xlabel('wavenumber(cm-1)'); ylabel('Absorption'); A=[Xave{1}./0.0303 yy]; B=[Xave{1}./0.0303 y]; % for TTF filtering B=cell2mat(B); Y=cell2mat(y);
Functions: function [Xave, Yave]=xiaohuifun3(name,color); pq='July_4_diff_hum'; % pq='Apr_8_diff_hum'; if pq(end)~='\' pq=[pq '\']; end % f=dir([pq '*' name '*.csv']); f=dir([pq name '\*.csv']); n=length(f); X=[]; Y=[]; for k=1:n k; fname=[pq name '\' f(k).name] fdata.data=csvread(fname,1,0); z=size(fdata.data,1) z=304; % for 0.5-0.8THz scans X=[X,fdata.data(1:z,1)]; Y=[Y,fdata.data(1:z,4)]; % end Xave=mean(X,2); Yave=mean(Y,2); figure plot(Xave,Yave,color,'LineWidth',1.5); end