Definition

Thermal Ablation (TA) is the use of extreme hiperthermic temperatures to destroy diseased tissues.

T.A. is performed in Interventional Radiology as an alternative to the surgical treatment of tumors in:

2

Page 3: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Different TA technologies

• Radiofrequency

• Microwaves

• Laser

• Cryoablation

33 3

Page 4: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

RF Current Flow

• RF Currents that goes through the tissue from the tip of the applicator to the ground pad.

Side effect: skin burning. 4 4

Page 5: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

MW Energy

Microwaves radiate only near the tip of the applicator.

An applied MW field causes water molecules to rotate bilions of time per second , which leads to

heat generation.

5

Page 6: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Comparison between RF and MW thermal ablation

6

• Heat sink effect

• Ground pad required

• Difficult to control the increase

of ablation

• Less influenced by veins

• Ground pad is not required

• Predictable and controllable ablation

• Compatible with pacemaker

Page 7: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

7

TATO :

Multiprobe microwave

thermal ablation System

Rev. 09 del 27.10.2016

Page 8: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

8

Main features of the system

1 • No cooling system

2

• Ergonomic and easily transportable

3

• Multi-probes with detachable cable

Page 9: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

1. No cooling system

The NIC (Not Internally Cooled) coax needles have a high energy efficiency that allows operating with less MW power for treatment up to 15 minutes.

Cooling system not needed

Less complexity

Higher reliability

Greater safety

Better applicator handling

Reducing procedure dead time

Easy control of the procedure

9

Page 10: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

10

Switch Unit Roll Stand

2.Ergonomic and easily transportable

Main Unit- Microwave generator

Articulated Arm extendable up to 1,5 m

Tato cart version Tato travel version

Switch Unit Roll Stand

Page 11: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

11

Main unit Handle

On/off button

Touchscreen 7,1 inch

Yellow Led: lighting indicates that MWs are ON

Red Led: when lighting up, it indicates a fault

2,45 GHz

120 W total power

IEC standards

compliant

USB port

Page 12: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Extendable arm and switch cover

12

Extendable up to 1.5 m

Sterile cover for arm

• Total power 120 W • 1 to 4 power outputs

Page 13: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

TATO1 vs TATO2

13

FEATURES TATO1 TATO2

TOTAL POWER 100W 120W

DATA LOGGING

DATA

DOWNLOAD

TRACK ABLATION

PROTOCOL

REFLECTED

POWER

MONITORING

SELECTION

NEEDLE GAUGE

Page 14: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

14

11 G

14 G

17 G

18 G

Multi-Gauge

Page 15: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

15

DIAMETER

G mm

3 mm

2,1 mm

1,5 mm

1,2 mm

11 G

14 G

17 G

18 G

The thinnest applicator currently

available in the world market

Page 16: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Version with infusion tip

Holes 0,3 mm Ø

16

• 14 G and 17 G version • Applicator with 4 micro-holes in the tip for fluid infusion • A syringe can be used to inject fluids during the treatment

Page 17: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

17

• Simple radiating monopole

• High efficiency

• Lightweight

(~35g w/o cable, 50g w/cable)

• Detachable feeding cable:

advantage under CT scan

Main structural features

Page 18: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Massa - Patient with HCC

Usability of TATO under CT-scan

Extendable arm

18

Page 19: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

19

Echogenic needle

Page 20: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Video of needle insertion

20

Page 21: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

• All steel tips are tested with an applied force of 5kg.

• No fragile ceramic parts are used

• All probes are Teflon covered, non-sticky.

• Available in different lengths from 10 up to 50 cm.

• Centimeter marked cannula

21

Main structural features

Page 22: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

How to insert the needle

22

E

Lesion

Ablation

E

Insert the tip of the needle into the lesion and use length E parameter (indicated in the chart) as reference.

Page 23: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

23

Control of increasing ablation volume : By employing lower MW power, the procedure’ s safety is increased, thus eliminating a lot of bubble gas and ensuring a better control under ultrasound.

Gauge Max power W

11 60

14 40

17 30

18 15

Main functional features

In order to operate safely, antennas must be used respecting the assigned power limits:

Page 24: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Increasing ablation

24

Page 25: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

TATO 2 HMI

25

Page 26: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

26

TATO2 HMI Home Screen

Page 27: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Setting

27

Language

Date and time

LOG Download

Services

Output mating cycles :

1

4

3

2

Page 28: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

28

TATO2 Home Screen

Page 29: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Possibility to insert Patient ID, Physician ID and note

29

Page 30: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Needle choice

30

Page 31: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Therapy Setting

31

Page 32: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Therapy summary

32

Page 33: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Therapy running

33

REFLECTED POWER COEFFICIENT OK : NEEDLE IS WORKING WELL AIR : NEEDLE IS IN THE AIR MISMATCH : NON-OPTIMAL FUNCTIONING OF NEEDLE NOT CONNECTED : NEDDLE IS NOT CONNECTED

Page 34: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Therapy ended - Summary

34

TRACK

ABLATION

Page 35: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Track Ablation

35

Page 36: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Video for Track ablation

• 14 G, 60 W, 25 sec

36

Page 37: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Ex – vivo and in – vivo charts of Tato System

37

Page 38: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

CHARTS TATO2

38

ABLATION ZONE EXAMPLES* – 1 APPLICATOR Ex – vivo Bovine Liver

11 G 60 W

Duration (min) 10:00 15:00

L (cm) 6,0 7,0

D( cm) 4,3 5,0

E (cm) 1,3 1,5

14 G 40 W

Duration (min) 5:00 10:00 15:00

L (cm) 5,0 5,4 6,2

D( cm) 2,6 4,0 4,5

E (cm) 0,9 1,0 1,1

17 G 30 W

Duration (min) 5:00 10:00 15:00

L (cm) 4,6 5,0 5,6

D( cm) 2,5 3,4 3,8

E (cm) 0,8 1,0 1,2

18 G 10 W 15 W

Duration (min) 5:00 10:00 5:00 10:00

L (cm) 2,0 2,6 2,3 3,3

D( cm) 1,4 1,7 1,5 2,3

E (cm) 0,4 0,6 0,6 0,6

Page 39: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

CHARTS TATO2

39

Page 40: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

CHARTS TATO2

40

Page 41: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

CHARTS TATO2

41

Page 42: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

E.g. HCC or liver metastasis

• Examples of MW applications in different liver tumors with same values of watt and time.

• HCC : «oven effect» due to the tissue’s consistency that is rich in water and to the encapsulated tumor’s morphology . The result is a bigger ablation.

• Metastasis: no «oven effect» because of the tissue has less water content and is not encapsulated. The result is a smaller ablation.

42

Page 43: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Frequently used parameters In vivo most frequently used values of power (W) and time (min) • D = diameter

43

Diameter

of tumor

0,5 <D<1 1<D<2 2<D<2,8 2,8 <D<4 D>4

Gauge

11

60 W

Open

surgery

1 needle

10 min

2 needles

15 min

14

40 W

1 needle

10 min

1 needle

15 min

2 needles

15 min

3 needles

15min

4 needles

10 min

17

30 W

1 needle

10 min

2 needles

15 min

18

15W

1 needle

10 min

Page 44: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Microwave ablation system Cross- reference

44

Page 45: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Padova – TATO use in laparoscopy

• Lesion: 7 cm

• 2 x 14 G, 35 W, 15 min

• And Repositioning 1 x 14G, 35 W, 4.5 min

45

Page 46: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

46

Treatment on HCC (percutaneous)

Dimension of the tumor 5 X 3 cm Used 2 x 17G

Page 47: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Padova – Open surgery

47

9 tumors in total RESECTION AND ABLATION Data of the first 4 ablations: 18 mm, 35W, 10min 19 mm, 35W, 2.3min and 40W, 1.30min 20 mm, 40W, 6min 18 mm, 40W, 6.2min

Page 48: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Osteoid osteoma

48

Osteoid osteoma in vertebra 18G, 10W, 6min

Page 49: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

Aarau - Patient with kidney tumor

Multi-probe case Tumor size: 2.4 x 2.2 cm (+2 cm margin) Number of antennas: 4 x 14G Power: 100W total Time: 15min

49

Page 50: THERMAL ABLATION TATO SYSTEM - AMT Medi · Padova – TATO use in laparoscopy •Lesion: 7 cm •2 x 14 G, 35 W, 15 min •And Repositioning 1 x 14G, 35 W, 4.5 min 45 . 46 Treatment

50