Thermal Testing of TOB/TEC Hybrids at UCSB. Why Thermal Test?. The CMS detector will run at –20 C in order to lessen radiation damage to the silicon. Why Thermal Test Hybrids?. Some Hybrids fail at –20 C APVs do not output Hybrid fails to initialize - PowerPoint PPT Presentation

TRANSCRIPT

  • Thermal Testing of TOB/TEC Hybrids at UCSB

  • The CMS detector will run at 20 C in order to lessen radiation damage to the siliconWhy Thermal Test?

  • Why Thermal Test Hybrids?Some Hybrids fail at 20 CAPVs do not outputHybrid fails to initializeNoise increases in cold (peltier current?) APV: Analog Pipeline Voltage Mode Chip

  • The 4 Hybrid Cold BoxCold box design and photo courtesy of:Bruno Wittmer and Giacomo Squazzoni @ CERNHow to thermally test hybrids?4 at a time.

  • The Set-up at CERNPicture courtesy of CERN

  • The Set-up at CERN

  • A question at Tracker week

  • Conclusion

    Hey folks, is there someone willing to test hybrids on a large scale?!?!

    G.Sguazzoni

  • Building our own Construction

    &

    Repair

  • Slow ControlFirst order of business: hooking things up 5 temperature sensors 1 humidity sensor

    1 Water Flow Sensor 1 Air Flow Sensor

  • Interfacing to a PC: ACDC

  • Cooling DownCooling Mechanism: Peltier UnitSupercool Peltier:Voltage = 28 V Current = 13 A Power = 350 W

  • Cooling DownHow long should/does it take to get to 20?Cold chamberPeltier + PlateChillerAir Air Flow(0.05 L/min)Water Flow (14 L/min)

  • Simplifying thingsS= surface area of enclosure (m )k=thermal conductivity of walls (W/m C)h=convection factor (W/m C) x=thickness of walls (m)

  • Simplifying thingsAssumptions:

    Most of the heat transferred to the system goes into cooling the aluminum peltier plate (specific heat of aluminum is much greater than that of air)m= 2.0 kgc= 910 J/kg C

  • Simplifying thingsAssumptions2) The convection factor can be calculated by waiting until the system comes to equilibrium

    Temp at equilibrium

  • Simplifying thingsWaiting (I was too impatient to wait for t )h= 7.55 (W/m C)

  • Calculating time to -20First order linear differential equation:Solution: (T(t=0)=295K)

  • Not quiteCalculation says that plate should reach 20 in ~14 minutes.

    Actual time: ~21 minutes

    Possible problems: Unaccounted heat influx through leaks or grounding strips connecting plate to outside assumptions made are flawed Missing term in differential equation

  • -20 Reached: On to Testing Set-up in Clean RoomChillerCold BoxPC for Monitoring+ ControllingElectronicsGas/Water Flow Control

  • Testing Software: ARCSAPV Readout Control SoftwareCommunicatesWith ACDC via TCP/UDP

  • G.Sguazzoni

  • What are we testing for?Shorted channels: On each clock, charge injected into every 8th channel. Shorted channels share, ADC amplitude is diminished.128 Channels per APV-6 APV hybrids: 768-4 APV hybrids: 512

  • What are we testing for?Open Channels:Inject charge throughCapacitive coupling of pitch adapter to antenna carrying TTL PulseAntennaPitch Adapter

  • Easier said than doneSlide courtesy of G.Squazzoni

  • ARC

    NIMpulser

    trigger

    NIMtoTTL

    start

    out

    50W

    antenna very close (~1mm) to the bond pads

    Peak - InvOFF

    Pulser ONPeak ModeOptimal latency

    OPENS

    G.Sguazzoni

  • We thought it was the pulserARC BoardLVDSNIM LogicNIMTTL

  • LVDS Signal?ARC Boards with functioning External BusSer 29Ser 31

    ARC Boards with non-functioning External BusSer 56Ser 57Ser 59

    LVDS SignalNoise

  • ARC Controller Problem

    Bad Batch of ARC BOARDS: LVDS Driver and LVDS Receiver U32 and U33 were switched during board assembly. Problem was never discovered.

    External Buss connector and associated ICs on lower left of photo. U32

  • With Pulser OperationalOpens detectable in Scope Mode Intentionally Plucked Bonds Show up in Pedestal Profile

  • With Pulser Operational.Opens are also detectable in Pedestal & Noise Test Plucked Bonds

  • Grounding IssuesSame Problem CERN was having.Huge spikes in noise at APV edges

  • GroundingFerrite Ring around Coax G10/copper strips connect peltier plate to front end ground

  • Much BetterOld GroundingNew Grounding

  • Thermal Cycle TestOne Button test

    Three Sets of Tests:Pedestal & NoisePulserShort Detection

  • Automated Test ResultsFinal Report After Thermal Cycle Test:Lists -Shorts -Opens -Noisy Channels

  • The testing goes onSo far, about 100 hybrids have been thermal cycled.

    At maximum capacity: ~28 per day

    And the cycling will continue