— LCD
—
—
VFD V & Hz readings.
Record MAX/MIN regular readings.
Display Hold Function.
Backlight Display
GENERAL SPECIFICATIONS :
í
í
í
í
í
í
í
í
í
í
í
í
í
í
í
í
í
í
í
í
Sensing : AC, True RMS
Display : 3-5/6 digits 6000 Counts Backlight LCD Display.
61 Segments Bar Graph : 40 per second max.
I
Auto Power Off.
Polarity : Automatic
Update Rate : 5 per second nominal
Operating Temperature : -10ºC ~ 50ºC
Relative Humidity : Maximum relative humidity 90% for temperature
up to 28ºC decreasing linearly to 50% relative humidity at 50ºC.
Pollution degree : 2
P Rating : IP40
Storage Temperature : -20ºC ~ 60ºC, < 80% R.H. (with battery removed)
Altitude : Operating below 2000m
Temperature Coefficient : Nominal 0.1 x (specified accuracy)/ °C @(-10ºC~18ºC or 28ºC~50ºC),
or otherwise specified.
Low Battery: approx. 4.6V
APO Timing: Idle for 20 minutes
APO Consumption: 20mA typical
Power Supply: Four Alkaline AA batteries +Hz Power Consumption: 6.5mA typical except the followings : VFD ACV : 8mA
Insulation Resistance @ 1mA Test Current :
50V output Voltage : 25mA, 100V output Voltage : 45mA
250V output Voltage : 85mA, 500V output Voltage : 170mA
1000V output Voltage : 440mA
Tester can perform at least 950 Insulation Tests with new alkaline batteries at room temperature.
These are standard tests of 1000V into 1MW with a duty cycle of 5 seconds on and 25 seconds off.
Dimension: 208(L) X 103(W) X 64.5(H) mm with holster
Weight: 635 gm with holster.
TRUE RMS DIGITAL INSULATION MULTIMETER
MODEL KM 887
Preliminary Data
D:/Chhaya/My Documents/Chhaya/backup/catlogs/NEW CAT/2015 new arrival catalogs/KM887.cdr
An ISO 9001:2008 Company
®®
SPECIAL FEATURES :
All Specifications are subject to change without prior notice
NEW— Dual
—
—
—
—
Display +Hz readings.
Relative zero mode.
LOCK-Test mode for Insulation Resistance
BeepJack audible & visible input warning.
High resolution 60.00mV & 60.00W ranges.
TM
SAFETY :
í
í
í
í
í
Safety : Double insulation per IEC / UL / EN61010-1 Ed. 3.0, IEC / UL / EN61010-2-030 Ed. 1.0, IEC / UL / EN61010-2-033 Ed. 1.0,
IEC / UL / EN61010-031 Ed. 1.1 & the corresponding CAN / CSA-C22.2 regulations to Measurement CAT III 1000V AC & DC & CAT IV 600V AC & DC.
Compliance to IEC / EN61557:2007 : (per CE requirements, not certified by UL or ETL) : IEC / EN61557-1 & IEC / EN61557-2 .
Overload Protections :
Insulation Resistance, mA & mA : 0.4A/1KV, IR 30kA, F Fuse, or better
A : 11A/1KV, IR 20kA, F Fuse or better
V : 1100Vrms
mV, W & Others : 1000 Vrms
E.M.C. : Meets EN61326-1:2013
In an RF field of 3V/m:
Total Accuracy = Specified Accuracy + 25 digits.
Performance above 3V/m is not specified.
Transient Protection : 8KV(1.2/50mS Surge)
ACCESSORIES :
ACCESSORIES : BKB32 banana plug to type-K socket plug adaptor,
Test probe pair, BRP21S2-C Remote probe, Alligator clip pair, Holster, Bkp60 banana plug type-K thermocouple & User manual.
OPTIONAL Magnetic hanger, & BMP-86x Carrying Case.
D:/Chhaya/My Documents/Chhaya/backup/catlogs/NEW CAT/2015 new arrival catalogs/KM 887.cdr
ELECTRICAL SPECIFICATIONS : KM 887
o oAccuracy is ± (% of reading digits + number of digits) or otherwise specified, at 23 C ± 5 C & less than 80%relative humidity.
True RMS voltage & current accuracies are specified from 1 % to 100 % of range or otherwise specified.
Maximum Crest Factor < 1.8:1 at full scale & < 3.6:1 at half scale, and with frequency components fall within the specified frequency bandwidth
for non-sinusoidal waveforms.
Range
VFD AC VOLTAGE
1) Accuracy Resolution
10Hz ~ 45Hz
±(4%rdg + 5dgts)100 mV600.0 V
45Hz ~ 200Hz
±(2.5%rdg + 5dgts)100 mV600.0 V
Input impedance : 10M , 110pF nominal.1) Unspecified for fundamental frequency > 440Hz2) Accuracy linearly decreases from 2.5% + 5d @ 200Hz to 9.0% + 5d @ 440Hz
W
200Hz ~ 440Hz
±(9.0%rdg + 5dgts)2)100 mV600.0 V
All Specifications are subject to change without prior notice
RESISTANCE
60.00W2)
AccuracyResolutionRange1)
0.01 mW ±(0.5%rdg + 5dgts)
6.000kW±(0.2%rdg + 2dgts)
1 W
60.00kW 10 W
1)
2)
3)
4)
5)
6)
Open Circuit Voltage : 1.7VDC typical
Specified assumes input lead resistance been offset by RELD or Shrt (short)
feature.
Constant Test Current : 0.2mA typical
Constant Test Current : 0.2mA typical
Add 1% @ >20MW
Add 2% @ operation temperature >35°C
600.0kW 100 W
6.000M 3)W 1 kW ±(1%rdg + 3dgts)
60.00M 4)W 10 kW ±(1.5%rdg + 6dgts)5) 6)
600.0W 100 mW ±(0.2rdg + 3dgts)
±(0.3%rdg + 2dgts)
AUDIBLE CONTINUITY TESTER
Audible Threshold Between 20 and 350W W
Response Time < 30ms approx.
Input Impedance : 10MW, 110pF nominal1)
Add 20d @ >80% of range2)
Unspecified @ <5% of range
Range
AC VOLTAGE
Accuracy Resolution
50Hz ~ 60Hz
±(0.7%rdg + 4dgts)1 mV6.000 V
10 mV60.00 V
100 mV600.0 V
1 V1000 V
0.01 mV60.00 mV
0.1 mV600.0 mV
40Hz ~ 1kHz
±(1.3%rdg + 4dgts)1 mV6.000 V
10 mV60.00 V
100 mV600.0 V
1 V1000 V
0.01 mV60.00 mV
0.1 mV600.0 mV
±(2%rdg + 4dgts)
1kHz ~ 5kHz
1)±(2%rdg + 4dgts) 1 mV6.000 V
10 mV60.00 V
100 mV600.0 V
1 V1000 V
0.01 mV60.00 mV
0.1 mV600.0 mV
Unspecified
2)5kHz ~ 20kHz
±(2%rdg + 20dgts)1 mV6.000 V
10 mV60.00 V
100 mV600.0 V
1 V1000 V
0.01 mV60.00 mV
0.1 mV600.0 mV
Unspecified
Unspecified
Unspecified
60.00 mV
600.0 mV
±(0.1%rdg + 2dgts)
DC VOLTAGE
AccuracyResolutionRange
0.01 mV
0.1 mV
Input Impedance : 10MW, 110pF nominal
6.000 V
60.00 V
1 mV
10 mV
600.0 V 100 mV
1000 V±(0.2%rdg + 3dgts)
1 V
±(0.2%rdg + 3dgts)
RECORD MODE
This mode records standard measurement Max & Min
readings on most functions, Manual or Auto-ranging
where available.
Nominal response & accuracy : Same as standard
measurement.
2.700 V ±(1.5%rdg + 4dgts)
DIODE TESTER
AccuracyResolutionRange
1 mV
Test Current : 0.4mA typically
Open Circuit Voltage : < 2.8VDC typically
G-17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai - 400 015. INDIA.Sales Direct.: 022-24156638, Tel. : 022-24124540, 24181649, Fax : 022-24149659Email : [email protected], Website : www.kusamelectrical.comAn ISO 9001:2008 Company
®®
D:/Chhaya/My Documents/Chhaya/backup/catlogs/NEW CAT/2015 new arrival catalogs/KM 887.cdr
ELECTRICAL SPECIFICATIONS : KM 887
All Specifications are subject to change without prior notice
1) mA / mA DC accuracies will be affected by extreme interior temperatures of the meter.
for rated accuracies, allow 6 to 20 minutes cool down interval after measuring A-currents of
3 to 10A continuously.2) £ 400mA continuous : >400mA for < 1.1 hours on per > 20 minutes off3) 10A continuous upto ambient 35°C; < 15 mins on per >5 mins off @ 35°C~50°C.
>10A to 20A for < 30 second on per >5 mins off.
DC CURRENT
Range Resolution Accuracy Burden Voltage
1)600.0 mA
60.00 mA1)
600.0 mA1) 2)
±(0.7%rdg + 2dgts)
0.2mV / mA
3mV / mA
30mV / A
6000 mA1)
6.000 A
10.00 A3)
0.1 mA
0.01 mA
0.1 mA
1 mA
0.001 A
0.01 A
±(0.2%rdg + 4dgts)
±(0.2%rdg + 4dgts)
±(0.2%rdg + 2dgts)
±(0.3%rdg + 3dgts)
±(0.5%rdg + 4dgts)
INSULATION RESISTANCE
3.000M , 30.00M , 55.0MW W W
3.000M , 30.00M , 110.0MW W W
3.000M , 30.00M , 275.0MW W W
1) Actual output voltage : 100% ~ 120% of Test Voltage
Live Circuit Detector : Inhibit test and display voltage reading instead if terminal voltage > 30V
prior to initialization of test.
Display Voltage Accuracy : DCV : 1.5% + 5d , ACV : 3.0% + 5d @50Hz ~ 60Hz
Specified measuring range is 0.020MW....25.0GW for percentage operating uncertainly B(%)
£ ± 30% per IEC/EN61557-2 requirements.
3.000M , 30.00M , 300.0MW W W,
50 V
100 V
1)Test Voltage
250 V
500 V
Range Test Current
1mA @50kW
1mA @100kW
1mA @250kW
1mA @500kW
Accuracy
±(1.5%rdg + 5dgts)
3.000M , 30.00M , 300.0MW W W
1000 V 3000MW
25.0GW
1mA @1MW
±(10%rdg + 5dgts)
±(1.5%rdg + 5dgts)
±(2.0%rdg + 5dgts)
550.0MW
AC CURRENT
Range Resolution Accuracy Burden Voltage
1)
2) 10A continuous upto ambient 35°C; < 15 mins on per >5 mins off
@ 35°C~50°C. >10A to 20A for < 30 second on per >5 mins off.
£ 400mA continuous : >400mA for < 1.1 hours on per > 20 minutes off
600.0 mA
60.00 mA
600.0 mA1)
0.2mV / mA
3mV / mA
30mV / A
6000 mA
6.000 A
10.00 A2)
0.1 mA
0.01 mA
0.1 mA
1 mA
0.001 A
0.01 A
±(1%rdg + 3dgts)
50Hz ~ 60Hz
600.0 mA
60.00 mA
600.0 mA1)
0.2mV / mA
3mV / mA
30mV / A
6000 mA
6.000 A
10.00 A2)
0.1 mA
0.01 mA
0.1 mA
1 mA
0.001 A
0.01 A
±(2%rdg + 3dgts)
40Hz ~ 3kHz
600.0 mA
60.00 mA
600.0 mA1)
0.2mV / mA
3mV / mA
30mV / A
6000 mA
6.000 A
10.00 A2)
0.1 mA
0.01 mA
0.1 mA
1 mA
0.001 A
0.01 A
3kHz ~ 5kHz
Unspecified
±(2%rdg + 5dgts)
~ Hz Line Level Frequency
Accuracy : ±(0.02%rdg + 4dgts)
Sensitivity(Sine RMS)
Range
4 mV 6Hz ~ 50kHz
40 mV 10Hz ~ 100kHz
0.4 V10Hz ~ 50kHz
4 V
40 V
10Hz ~ 5kHz400 V
40 V 10Hz ~ 440Hz
40 mA
Function Range
60 mV
600 mV
6 V
60 V
600 V
1000 V
VFD 600 V
600 mA
6000 mA 400 mA10Hz ~ 5kHz
600 mA 40 mA
0.6 A10Hz ~ 3kHz
6 A
10 A 6 A
60 mA 4 mA
10Hz ~ 30kHz
CAPACITANCE
Range Resolution Accuracy1)
2.000 mF 2)±(1.5%rdg + 5dgts)
20.00 mF
200.0 mF
±(5%rdg + 5dgts)
2000 mF
20.00 mF
1) Accuracies with film capacitor or better2) Specified from 0.200mF
0.001 mF
0.01 mF
0.1 mF
1 mF
0.01 mF
1)
the ambient (isothermal stage) for a correct junction voltage compensation. Allow enough time to reach the isothermal stage for a significant change of ambient temperature. It can take up to an hour for changes > 5°C.2) Type-K thermocouple range & accuracy not included.
Accuracies assume meter interior has the same temperature of
o o-40.0 C to 0.0 C 1% + 2 Co
TEMPERATURE
Range Accuracy 1) 2)
o o0.0 C to 50.0 C
o o50.0 C to 537.0 C
o o-40.0 F to 32.0 F
o o32.0 F to 122.0 F
o o122.0 F to 999.0 F
2.2 Co
1% + 2 Co
1% + 3.6 Fo
4 Fo
1% + 3.6 Fo
®®
AN ISO 9001:2008 COMPANY
MODEL - KM 887
TRMS DIGITAL
INSULATION MULTIMETER
OPERATION MANUAL
G 17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai - 400 015. INDIA.
Sales Direct : (022) 24156638Tel. : (022) 24124540, 24181649. Fax : (022) 24149659
Email : [email protected] Website : www.kusamelectrical.com
www.kusam-meco.co.in
®
LIST OF PRODUCTS
¬ Digital Multimeter ¬ Digital AC & AC/DC Clampmeter
¬ AC Clamp Adaptor ¬ AC/DC Current Adaptor
¬ Thermo Anemometer ¬ Thermo Hygrometer
¬ Distance Meter ¬ Digital Lux Meter
¬ Network Cable Tester ¬ Power Factor Regulator
¬ Earth Resistance Tester ¬ Digital Panel Meters
¬ DC Power Supplies ¬ High Voltage Detector
¬ Calibrators ¬ Gas Analysers
¬ Frequency Counter ¬ Function Generator
¬ Phasing Sticks ¬ Battery Tester
¬ Waterproof Pen Testers ¬ Solar Power Meter
¬ EMF Detector
¬ Wood, Paper & Grain Moisture Meter
¬ Transistorised Electronic Analog & Digital Insulation
Resistance Testers(upto 10 KV)
¬ Digital Sound Level Meter & Sound Level Calibrator
¬ Digital contact & Non-contact Type Tachometer
¬ Digital Non-contact (infrared) Thermometer
¬ Maximum Demand Controller/Digital Power Meter
¬ Digital Hand Held Temperature Indicators
®
TRMS DIGITAL
INSULATION MULTIMETER
MODEL - KM 887
®
TABLE OF CONTENTS
TITLE PAGE NO.
I. .............................
II. Cenelec Directives......................................
Safety..............................
IV. General Specifications...............................
III. Special Features.........................................
V. Product Description
. Test Certificate...........................................
IX. Warranty......................................................
...................................
VI. Operation....................................................
VII. Maintenance................................................
VIII
01
03
04
05
14
15
25
27
28
01
®
I. SAFETY :
Terms in this manual :
WARNING : Identifies conditions and actions that could result in serious injury or even death to the user.
This manual contains information and warnings that must be followed for operating the instrument safely and maintaining the instrument in a safe operating condition. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired.
CAUTION : Identifies conditions and actions that could cause damage or malfunction in the instrument.
WARNING :To reduce the risk of fire or electric shock, do not expose thisproduct to rain or moisture. The Meter is intended only for indooruse, Observe proper safety precautions when working with voltagesabove 33 Vrms, 46.7 Vpeak or 70 VDC. These voltage levels posea potential shock hazard to the user. Before and after hazardousvoltage measurements, check the voltage function on a knownsource such as line voltage to determine proper meter functioning.The meter meets UL/IEC/EN61010-1 Ed. 3.0, CAN/CSA C22.2No. 61010-1 Ed. 3.0, UL/IEC/EN61010-2-033 Ed. 1.0 toMeasurement CAT-III 1kV and CAT-IV 600V, AC & DC. The meteralso meets the relevant Parts of EN61557. In particular, Part 1Ed. 2.0, Part 2 Ed. 2.0 and Part 10 Ed. 2.0. These relevant parts ofEN61557 are not covered by the ETL certification. The accompaniedtest probe assembly meets UL/IEC/EN61010-031 Ed. 1.1 to thesame meter ratings or better. IEC 61010-031 requires exposed
conductive test probe tips to be ≤ 4mm for CAT III & CAT IV ratings.
Refer to the category markings on your probe assemblies as well ason the add-on accessories (like detachable Caps or Alligator Clips),if any, for applicable rating changes.
®
02
Keep your hands/fingers behind the hand/finger barriers (of the meter and the test probe assembly, where applicable) that indicate the limits of safe access of the hand-held parts during measurements. Inspect lead wires, connectors, and probes for damaged insulation or exposed metal before using the meter. If any defects are found, replace them immediately. Only use the probe assembly provided with the meter or a UL Listed Probe Assembly to the same meter ratings or better.
CAUTION :Disconnect the test leads from the test points before changingfunctions.
INTERNATIONAL ELECTRICAL SYMBOLS :
Marking of Electrical and Electronic Equipment (EEE).
Do not dispose of this product as unsorted municipal
waste. Contact a qualified recycler
Caution ! Refer to the explanation in this Manual
Caution ! Risk of electric shock
Earth (Ground)
Double Insulation or Reinforced insulation
Fuse
AC--Alternating Current
DC--Direct Current
Three-phase Alternating Current3
03
®
II. CENELEC DIRECTIVES :
The instruments conform to CENELEC Low-voltage directive
2014/35/EC, Electromagnetic compatibility directive 2014/30/EU &
RoHS directive 2011/65/EU.
Measurement Category
Measurement Category IV is applicable to test & measuring circuits
connected at the source of the building’s low-voltage MAINS
installation. Examples are measurements on devices installed before
the main fuse or circuit breaker in the building installation.
Measurement Category III is applicable to test & measuring circuits
connected to the distribution part of the building’s low-voltage MAINS
installation. Examples are measurements on distribution boards
(including secondary meters), circuit-breakers, wiring, including
cables, bus-bars, junction boxes, switches, socket-outlets in the fixed
installation, & equipment for industrial use & some other equipment
such as stationary motors with permanent connection to the
fixed installation
Measurement Category II is applicable to test & measuring circuits
connected directly to utilization points (socket outlets & similar points)
of the low-voltage MAINS installation. Examples are measurements
on MAINS CIRCUITS of household appliances, portable tools &
similar equipment.
04
®
IV. GENERAL SPECIFICATIONS :Display : 3-5/6 digits 6,000 countsPolarity : AutomaticUpdate Rate : 5 per second nominal61 Segments Bar graph : 40 per second maxOperating Temperature : -10°C to 50°CRelative Humidity : Maximum relative humidity 90% for temperat-ure up to 28°C decreasing linearly to 50% relative humidity at 50°CPollution Degree : 2IP Rating : IP40Storage Temperature : -20°C to 60°C, < 80% R.H. (with batteryremoved)Altitude : Operating below 2000mTemperature Coefficient : nominal 0.1 x (specified accuracy) /°C @(-10°C ~ 18°C or 28°C ~ 50C), or otherwise specifiedSensing : AC, True RMSSafety : ETL certified per IEC/UL/EN61010-1 Ed. 3.0,IEC/UL/EN61010-2-030 Ed. 1.0, IEC/UL/EN61010-2-033 Ed. 1.0,IEC/UL/EN61010-031 Ed. 1.1 and the corresponding CAN/CSA-C22.2 regulations to Measurement Categories:CAT III 1000 V AC & DC and Category IV 600V AC & DCCompliance to IEC/EN61557:IEC/EN61557-1 Ed. 2.0, IEC/EN61557-2 Ed. 2.0 & IEC/EN61557-10 Ed. 2.0(Tested per CE requirements, not covered by ETL certification)Overload Protections :Insulation Resistance, μA & mA: 0.4A/1KV, IR 30kA,F Fuse; or betterA: 11A/1KV, IR 20kA, F Fuse; or betterV: 1100VrmsmV, & Others: 1000 VrmsTransient Protection : 8kV (1.2/50s surge)E.M.C.: Meets EN61326-1:2013In an RF field of 3V/m:Total Accuracy = Specified Accuracy + 25 digitsPerformance above 3V/m is not specified
®
05
Power Supply : Four Alkaline AA batteries (IEC Lr6)Power Consumption : 6.5mA typical except the followings:VFD ACV +Hz: 8mAInsulation Resistance @1mA test current:50V output voltage: 25mA100V output voltage: 45mA250V output voltage: 85mA500V output voltage: 170mA1000V output voltage: 440mATester can perform at least 950 insulation tests with new alkaline batteries at room temperature. These are standard tests of 1000 V into 1 M with a duty cycle of 5 seconds on and 25 seconds off.Low Battery : approx. 4.6VAPO Timing : Idle for 20 minutesAPO Consumption : 20A typicalDimension : 208(L) X 103(W) X 64.5(H) mm with holsterWeight : 635 gm with holsterAccessories : Test probe pair, Alligator clip pair, BRP21S2-C Remote probe, Holster, User's manual, Bkp60 banana plug type-K thermocoupleOptional Accessories : BKB32 banana plug to type-K socket plug adaptor, BMH-01 magnetic hanger; BMP-86x soft carrying pouch
Special Features :
• Record MAX/MIN regular readings
• Relative Zero
• Display Hold
• LCD Backlight
• VFD V & Hz readings
• Dual display +Hz Readings
• High resolution 60.00mV & 60.00 ranges
• Lock-Test mode for Insulation resistanceTM• BeepJack audible & visible input warning
®
ELECTRICAL SPECIFICATIONS :
Accuracy is ± (% reading digits + number of digits) or otherwisespecified, at 23°C ± 5°C & less than 80% relative humidity. TrueRMS voltage & current accuracies are specified from 1 % to 100 %of range or otherwise specified. Maximum Crest Factor < 1.8:1 atfull scale & < 3.6:1 at half scale, and with frequency componentswithin the specified frequency bandwidth for non-sinusoidalwaveforms.
AC Voltage :
Range Resolution Accuracy
50Hz ~ 60Hz3)60.00mV4)600.0mV
6.000V
60.00V
600.0V
1000V
0.01
0.1
0.001
0.01
0.1
0.1
±(0.7%rdg + 4dgts)
40Hz ~ 1kHz
±(2%rdg + 4dgts)
3)60.00mV4)600.0mV
6.000V
60.00V
600.0V5)1000V
0.01
0.1
0.001
0.01
0.1
0.1
±(1.3%rdg + 4dgts)
06
®
07
1kHz ~ 5kHz
Unspecified
3)60.00mV4)600.0mV
6.000V
60.00V
600.0V
1000V
0.01
0.1
0.001
0.01
0.1
0.1
1)±(2%rdg + 4dgts)
2)5kHz ~ 20kHz
Unspecified
60.00mV4)600.0mV
6.000V
60.00V
600.0V
1000V
0.01
0.1
0.001
0.01
0.1
0.1
1)±(2%rdg + 20dgts)
Unspecified1)±(2.5%rdg + 20dgts)
Input impedance: 10MW, 110pF nominal
1) Add 20d @ >80% of range
2) Unspecified @ <5% of range
3) Signal peak absolute values, including DC bias,
less than 110mVpeak
4) Signal peak absolute values, including DC bias,
less than 1100mVpeak
VFD AC Voltage :
Range Resolution1)Accuracy
10Hz ~ 45Hz
45Hz ~ 200Hz
0.1
0.1
600.0V
600.0V
±(4.0%rdg + 5dgts)
±(2.5%rdg + 5dgts)
08
®
200Hz ~ 440Hz
0.1600.0V2)±(9.0%rdg + 5dgts)
Input impedance: 10MW, 110pF nominal
1) Unspecified for fundamental frequency > 440Hz
2) Accuracy linearly decreases from 2.5% + 5d @200Hz to
9.0% + 5d @440Hz
Range Resolution Accuracy
DC Voltage :
0.0160.00mV ±(0.2%rdg + 3dgts)
Input impedance: 10MW, 110pF nominal
600.0mV
6 000V.
60.00V
60.00mV
600.0V
1000V
0.1
0.001
0.01
0.01
0.1
1
±(0.1%rdg + 2dgts)
±(0.2%rdg + 3dgts)
Ohms :1)Range Resolution Accuracy
0.0160.00W2) ±(0.5%rdg + 5dgts)
600.0W
6 000k. W
60.00kW
600.0kW
6.000 3)MW
60.00M 4)W
0.1
0.001
0.01
0.1
0.001
0.01
±(0.2%rdg + 2dgts)
±(1%rdg + 3dgts)
±(0.2%rdg + 3dgts)
±(0.3%rdg + 2dgts)
5)6)±(1.5%rdg + 6dgts)
®
09
1) Open Circuit Voltage: 1.7VDC typical
2) Specified assumes input lead resistance been offset by
REL or Shrt (short) feature
3) Constant Test Current: 0.2µA Typical
4) Constant Test Current: 0.02µA Typical
5) Add 1% @ >20MW
6) Add 2% @ operation temperature >35°C
Audible Continuity Tester
Audible threshold: between 20W and 350W
Response time: < 30ms
Diode Tester
Test Current
Open CircuitVoltage
AccuracyResol.Range
2.700V 0.4mA < 2.8 VDC±(1.5%rdg + 4dgts)0.001
Capacitance :
Range Resolution1)Accuracy
2)2.000µF
20.00µF
200.0µF
2000µF
20.00mF
±(1.5%rdg + 5dgts)
±(5%rdg + 5dgts)
0.001
0.01
0.1
1
0.01
1) Accuracies with film capacitor or better
2) Specified from 0.200µF
10
®
Temperature :
Range 1)Accuracy
1% + 2°C-40.0°C to 0.0°C
0.0 C to 50.0 C° °
50.0 C to 537.0 C° °
-40.0 F to 32.0 F° °
32.0 F to 122.0 F° °
122.0 F to 999.0 F° °
2.2 C°
1% + 2 C°
1% + 3.6 F°
4 F°
1% + 3.6 F°
1) Accuracies assume meter interior has the same temperature
of the ambient (isothermal stage) for a correct junction voltage
compensation. Allow enough time to reach the isothermal
stage for a significant change of ambient temperature.
It can take up to an hour for changes > 5°C.
2) Type-K thermocouple range & accuracy not included
DC current :
Range Resolution Accuracy
1)2)600.0mA
±(0.2%rdg + 4dgts)0.1
1
0.01
0.1
0.001
Burden Voltage
6.000A3)10.00A
1)60.00mA
1)6000 A µ
1)600.0 A µ
0.01
±(0.2%rdg + 2dgts)
±(0.2%rdg + 4dgts)
±(0.3%rdg + 3dgts)
±(0.5%rdg + 4dgts)
±(0.7%rdg + 2dgts)
0.2mV / Aµ
3mV / Am
30mV / A
1) A/mA DC accuracies will be affected by extreme interior
temperatures of the meter. For rated accuracies, allow
6 to 20 minutes cool down interval after measuring A-currents
of 3 to 10A continuously.
µ
®
11
2) 400mA continuous : >400mA for <1.1 hours on per >20
minutes off
3) 10A continuous up to ambient 35°C; <15 mins on per >5 mins
off @ 35°C ~ 50°C. >10A to 20A for <30 seconds on per >5
mins off
_<
AC current :
Range Resolution Accuracy Burden Voltage
±(1%rdg + 3dgts)
50Hz ~ 60Hz
600.0 Aµ
6000 Aµ
60.00 Am
600.0 A 1)m
6.000A
10.00A 2)
0.2mV / µA
3mV / mA
30mV / A
0.1
1
0.01
0.1
0.001
0.01
±(2%rdg + 3dgts)
40Hz ~ 3kHz
600.0 Aµ
6000 Aµ
60.00 Am
600.0 A 1)m
6.000A
10.00A 2)
0.2mV / µA
3mV / mA
30mV / A
0.1
1
0.01
0.1
0.001
0.01
±(2%rdg + 5dgts)
3kHz ~ 5kHz
600.0 Aµ
6000 Aµ
60.00 Am
600.0 A 1)m
6.000A
10.00A 2)
0.2mV / µA
3mV / mA
30mV / A
0.1
1
0.01
0.1
0.001
0.01Unspecified
12
®
1) 400mA continuous : >400mA for <1.1 hours on per >20
minutes off
2) 10A continuous up to ambient 35°C; <15 mins on per >5 mins
off @ 35°C ~ 50°C. >10A to 20A for <30 seconds on per >5
mins off
_<
~ Hz Line Level Frequency :
Function Range RangeSensitivity (Sine RMS)
10Hz ~ 100kHz
60mV
600mV
6V
60V
600V
1000V
VFD 600V
600 Aµ
6000 Aµ
60mA
600mA
6A
10A
4mV
40mV
0.4V
4V
40V
4V
400V
40 Aµ
400 Aµ
4mA
40mA
0.6A
6A
6Hz ~ 50kHz
10Hz ~ 50kHz
10Hz ~ 30kHz
10Hz ~ 5kHz
10Hz ~ 440Hz
10Hz ~ 5kHz
10Hz ~ 3kHz
Accuracy : ±(0.02%rdg + 4dgts)
Record mode
This mode records standard measurement Max and Min readings
on most functions, Manual or Auto-ranging where available.
Nominal response and accuracy : Same as standard
measurements
®
13
Insulation Resistance : 1)Test Voltage Range Test Current Accuracy
50V
100V
250V
500V
1000V
1mA @50kW
1mA @100kW
1mA @250kW
1mA @500kW
1mA @1MW
±(1.5%rdg + 5dgts)
±(1.5%rdg + 5dgts)
±(2.0%rdg + 5dgts)
±(10%rdg + 5dgts)
3.000MW
30.00MW
55.0MW
3.000MW
30.00MW
110.0MW
3.000MW
30.00MW
3.000MW
30.00MW
3.000MW
30.00MW
275.0MW
300.0MW
550.0MW
300.0MW
3000MW
25.0GW
1) Actual output voltage : 100% ~ 120% of Test Voltage
Live Circuit Detector : Inhibit test and display voltage reading
instead if terminal voltage > 30V prior to initialization of test.
Display voltage accuracies : DCV : 1.5% + 5d,
ACV : 3.0% + 5d @50Hz ~ 60Hz
Specified measuring range is 0.020M … 25.0G for percentage
operating uncertainty B[%] ±30% per IEC/EN61557-2
requirements.
®
14
III. PRODUCT DESCRIPTION :Note : Top of the line model is used as representative for illustrationpurposes. Please refer to your particular model for functionavailability.
1. 3-5/6 digits 6000 counts dual displays.
2. Push-buttons for special functions & features.
3. Selector to turn the PowerOn or Off and Select a function.
4.
5. C o m m o n ( G r o u n dreference) Input Jack forall functions EXCEPTInsulation Resistance function.
6. Input Jack for all functionsEXCEPT A, mA and µAfunctions.
7. Input Jack for A functionpositive input.
Input Jack for insulationfunction Ground reference INS. or mA/µA function positive input.
–
1
2
3
4
5
6
7
. . .
0 6. . .
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
Analog bar-graph :
The analog bar graph provides a visual indication of measurement like a traditional analog meter needle. It is excellent in detecting faulty contacts, identifying potentiometer clicks, and indicating signal spikes during adjustments.
®
16
W Resistance, Continuity
Press the SELECT button momentarily to toggle the functions. Lastselection will be saved as power up default for repeat measurementconvenience. Continuity function is convenient for checking wiring connections & operation of switches. A continuous beep tone indicates a complete wire.
W
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
0 6.
– +
1SELECT
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
0 6.
W
+–
Using resistance or continuity function in a live circuit will produce false results and may damage the instrument. In many cases the suspected component must be disconnected from the circuit to obtain an accurate reading.
CAUTION :
Auto Leads Resistance Calibration
When entering the 60 range manually by RANGE button for highprecision low resistance measurement, this feature will prompt youto short the inputs for temporary test lead resistance calibration onthis range.
W
15
®
IV. OPERATION
CAUTION :
Before and after hazardous voltage measurements, test the voltage function on a known source such as line voltage to determine proper meter functioning.
+Hz +HzACV & VFD ACV Functions :Press the SELECT button momentarily to toggle the subject functions. Last selection will be saved as power up default for repeat
+Hz measurement convenience. For ACV function, press the RANGE button momentarily to select other ranges when needed.
+HzFor VFD ACV function, only 600V range is available to best cope with the range of most Variable Frequency Drives (VFD) measurements.
DCV Function :Turn Rotary Knob to DCV position for measurement.
V+Hz V+Hz
VFD
V
.
0 6.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
0 6.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
0 6.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidthV
– +VFD
1SELECT
VFD
HzVVFD
17
®
The fastest way is to short the leads to auto-range to the 60range, then press the RANGE button momentarily. The displayshows “Shrt”. Keep shorting the leads for further 3 seconds untilthe display shows zero. The lead resistance is then temporarilycompensated.
W
The compensation value stays until the next meter power reset,
and can be as high as 5W. If you need a compensation value thatis higher than that, Relative Zero mode is recommended.
Diode Test, Capacitance Function
Press the SELECT button momentarily to toggle the functions. Lastselection will be saved as power up default for repeat measurementconvenience.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
– +
– +
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
FORWARD BIAS
1SELECT
– +
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
REVERSE BIAS
– +
Discharge capacitors before making any measurement. Large valuecapacitors should be discharged through an appropriate resistanceload.
CAUTION :
®
18
Normal forward voltage drop (forward biased) for a good silicondiode is between 0.400V to 0.900V. A reading higher than thatindicates a leaky diode (defective). A zero reading indicates a shorted diode (defective). An OL indicates an open diode (defective).Reverse the test leads connections (reverse biased) across the diode. The digital display shows OL if the diode is good. Any otherreadings indicate the diode is resistive or shorted (defective).
+HzDCmV, ACmV functions :
Press the SELECT button momentarily to toggle the subjectfunctions. Last selection will be saved as power up default forrepeat measurement convenience.
mVHz
mV
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
0 6.
1SELECT
– +
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
0 6.
mVHz
– +
19
®
Temperature functions :
Press SELECT button momentarily to toggle ºC and ºF readings.Last selection will be saved as power up default for repeat measurement convenience.
ºC
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
Temp
+
–
+
–
– +
ºF
Note : Be sure to insert the banana plug type-K temperature beadprobe Bkp60 with correct + – polarities. You can also use a plugadapter Bkb32 (Optional purchase) with banana pins to type-Ksocket to adapt other standard type-K mini plug temperatureprobes.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
A, mA, µA Current functions :+HzPress the SELECT button momentarily to toggle the DC or AC of
the subject functions. Last selection will be saved as power up default for repeat measurement convenience.
1SELECT
+
–
+
–
– +
®
20
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
0 6.
A AHz
mAHz
µAHz
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
0 6.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
0 6.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
0 6
Insulation Resistance function :
The LCD icons used together throughout in this manual is
referred as active measurements of Insulation Resistance
Function through the activation of the TEST button on the meter or
on the Remote Probe. The sources a user selectable test
voltage of 50V, 100V, 250V, 500V or 1000V to measure Insulation
Resistance values. The in a flashing manner warns against
test voltage is being output. Use extreme caution when operating
the to avoid electric shock.
TEST
TEST
TEST
is inhibited when the meter sounds 3-beeps and displays
the detected voltage value plus warning against energized circuit
of more than 30V is being connected, before the is active.
Measurements shall only be carried out on parts of an installation
or equipment that are de-energized.
WARNING
TEST
TEST
–+ –+ –+–+
21
®
Connecting to energized circuits when the is active will
produce false results and may damage the instrument. Always check
with voltage functions and remove power from the circuits before
carrying out the .
TEST
TEST
Select an intended test voltage of 50V, 100V. 250V. 500V or 1000V.
press the RANGE (Test Voltage) button momentarily
to select the voltage in sequence. Last selection will be saved as
power up default for repeat measurement convenience.
The secondary display shows the selected voltage for 1 second right
after the selection, and then displays the actual detected voltage
readings. The voltage annunciator by the secondary display remains
indicating the voltage selected.
Setup as illustrated below.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
100V
50V...1000VINSULATION
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
250V
TestVoltage
1
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
®
22
Default startup primary display reading is "-.---". Allow enough
time for a good measuring result. After the ink, is
released, the measuring loop starts to discharge the testing voltage.
The last measured resistance reading stays on the primary display
until the next or a function change. The secondary display
keeps showing the actual detected voltage readings.
TEST TEST
TEST
The is active as long as the TEST button is pressed and
hold. The TEST buttons on the meter and on the Remote Probe
work alike. The Insulation Resistance readings are shown on
the primary display.
TEST
NOTE. Maximum display reading of each Insulation Resistance
range is subjected to the test voltage selected. They are 55.0MW,
110.0MW, 275MW, 550MW & 25.0GW for 50V, 100V, 250V, 500V &
1000V respectively. Over-range is indicated as > maximum
display-reading.
Lock-Test mode is recommended for continuous measurements.
To apply, press the LOCK button momentarily to display the
annunciator LOCK before pressing the TEST button
momentarily. The LCD will show both LOCK & to indicate
continuous measurement is active. Press again either button
momentarily to release the Lock-Test mode.
TEST
®
24
Backlighted display :
Press button momentarily to toggle the LCD backlight. The backlight will also be turned off automatically after approximate 10 minutes to extend battery life.
Auto- or Manual-ranging (Volts, Current & Ω functions only) :
Press the RANGE button momentarily to select manual-ranging,and the meter will remain in the range it was in, the LCD turnsoff. Press the button momentarily again to select an adjacent range.Press and hold the button for 1 second or more to resume auto-ranging.
AUTO
Hold: The hold feature freezes the display for later view. turns on. Press the HOLD button momentarily to toggle the hold feature. This feature does not apply to Insulation Resistance Functions.
LCD “ ”
MAX/MIN Recording Mode: Press REC button momentarily to activate MAX/MIN recording mode. The LCD “R” & “MAX MIN” turn on. The meter beeps when new MAX(maximum) or MIN(minimum) reading is updated. Press the button momentarily to read the Real-time, MAX & MIN readings in sequence. Press the button for 1 second or more to exit MAX/MIN recording mode. When activated,Auto-Power-Off is disabled automatically. This feature does not apply to Insulation Resistance functions.
Smooth mode (Insulation Resistance function only) :
Smooth mode displays the running average of the last eight
measured readings having changes within 300 counts in sequence.
On the contrary, it displays directly, without smoothing, the measured
reading that is beyond 300 counts in changes comparing to its former
one. Press the button momentarily to enable with LCD annuciator
“ ” turned on. Press momentarily again to disable.
23
®
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
Test Voltage
SELECT RANGE
INSULATION TEST REC R
LOCK
REL
OFF
Hz
VFD V
V
W
Hz
mVmAA Hz
Hz
µAINSULATION
50V...1000V
INSULATION
1000V CAT IIIMAX
600V CAT IVHBC FUSED
MAX 0.6A HBC FUSEDMAX 10A
COMAJACK
JACK
INS.µA mA
INS.
W V
HOLD
20kHz ACVBandwidth
.
+
1
TEST
LOCK
TESTflash
1TEST
250V
LOCK
LOCK1
LOCK
TEST flash
+
1
TEST1TEST
LOCK
TEST flash flashTEST
TESTPress& Hold
+
TEST
Press& Hold
25
®
Relative-Zero ( ) mode :
Relative-Zero allows the user to offset the meter consecutive measur-ements with the main display displaying reading as the reference value. LCD “ ” turns on. Press the REL button momentarily to toggle Relative-Zero mode.
Beep-Jack™ Input Warning :The meter beeps as well as displays“InEr” to warn the user against possible damage to the meter dueto improper connections to the “ INS. µA mA” or “A” input jackwhen other functions, especially voltage function, is selected.
Set Beeper Off : Press the RANGE button while turning themeter on to temporarily disable the Beeper feature. Turn the rotaryswitch OFF and then back on to resume.
Auto-Power-off (APO) : The Auto-Power-off (APO) mode turns the meter off automatically to extend battery life after approximately 20 minutes of no rotary switch or push button operations. To wake up the meter from APO, press the SELECT or button momentarily or turn the rotary switch OFF and then back on. Always turn the rotary switch to the OFF position when the meter is not in use.
Disabling Auto-Power-off :
Press the SELECT button while turning the meter on to temporarilydisable the Auto-Power-Off feature. Turn the rotary switch OFF andthen back on to resume.
MAINTENANCE :WARNING :To avoid electrical shock, disconnect the meter from any circuit,remove the test leads from the input jacks and turn OFF the meterbefore opening the case. Do not operate with open case. Installonly the same type of fuse or equivalent
Calibration :
Accuracy is specified for a period of one year after calibrationPeriodic calibration at intervals of one year is recommended to mai-ntain meter accuracy.
®
26
Cleaning and Storage :
Periodically wipe the meter and the test probe assembly with a damp cloth and mild detergent. Do not use abrasives or solvents. Allow to dry completely before operating. If the meter is not to be used for periods of longer than 60 days, remove the battery and store it separately.
Trouble Shooting :
If the instrument fails to operate, check battery, fuses, leads, etc.,and replace as necessary. Double check operating procedure asdescribed in this manual.
Battery and Fuse replacement :Loosen the screws from the access cover of the case bottom. Lift theaccess cover. Replace the batteries or fuse. Re-fasten the screws.
Battery and Fuse replacement :Battery use : Four 1.5V AA battery (IEC LR6)Fuses use :Fuse (F2) for INS. µA mA input : 0.4A/1000Vac & Vdc, IR 30kA or better, F fuse;Dimension : 6 x 32 mmFuse (F3) for A input : 11A/1000Vac & Vdc, IR 20kA or better, F fuse;Dimension : 10 x 38 mm
AA
+
AA
+
AA
+
F2
F3
27
®
MUMBAI
TEST CERTIFICATE
TRMS DIGITAL INSULATION MULTIMETER
This Test Certificate warrantees that the product has been
inspected and tested in accordance with the published
specifications.
The instrument has been calibrated by using equipment
which has already been calibrated to standards traceable
to national standards.
MODEL NO. ________________KM 887
SERIAL NO. ___________
DATE: ___________
ISO 9001REGISTERED
QC
PASS
KUSAM-MECO
®
28
WARRANTYEach “KUSAM-MECO” product is warranted to be free from defects in
material and workmanship under normal use & service. The warranty
period is one year (12 months) and begins from the date of despatch of
goods. In case any defect occurs in functioning of the instrument,
under proper use, within the warranty period, the same will be rectified
by us free of charges, provided the to and fro freight charges are borne
by you.
This warranty extends only to the original buyer or end-user customer
of a “KUSAM-MECO” authorized dealer.
This warranty does not apply for damaged Ic’s, fuses, burnt PCB's,
disposable batteries, carrying case, test leads, or to any product which
in “KUSAM-MECO’s” opinion, has been misused, altered, neglected,
contaminated or damaged by accident or abnormal conditions of
operation or handling.
“KUSAM-MECO” authorized dealer shall extend this warranty on new
and unused products to end-user customers only but have no authority
to extend a greater or different warranty on behalf of “KUSAM-MECO”.
“KUSAM-MECO’s” warranty obligation is limited, at option, free of
charge repair, or replacement of a defective product which is returned
to a “KUSAM-MECO” authorized service center within the warranty
period.
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY
AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR
IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. “KUSAM-MECO” SHALL NOT BE LIABLE
F O R A N Y S P E C I A L , I N D I R E C T, I N C I D E N TA L O R
CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF
DATA, ARISING FROM ANY CAUSE WHATSOEVER.
All transaction are subject to Mumbai Jurisdiction.