UC Santa Cruz @ VOCALS?. PI: Patrick Chuang Instrument: Phase-Doppler interferometer Measures cloud drop size velocity inter-drop spacing on a single drop basis. Size range: 3 to 150 um. PDI mounted on the CIRPAS TwinOtter. Pre-flight testing of the PDI. - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: UC Santa Cruz @ VOCALS?](https://reader036.vdocuments.net/reader036/viewer/2022062322/5681502b550346895dbe1e55/html5/thumbnails/1.jpg)
UC Santa Cruz @ VOCALS?
• PI: Patrick Chuang• Instrument: Phase-Doppler
interferometer• Measures cloud drop
– size– velocity – inter-drop spacing
on a single drop basis.• Size range: 3 to 150 um
PDI mounted on the CIRPAS TwinOtter
![Page 2: UC Santa Cruz @ VOCALS?](https://reader036.vdocuments.net/reader036/viewer/2022062322/5681502b550346895dbe1e55/html5/thumbnails/2.jpg)
Pre-flight testing of the PDI
![Page 3: UC Santa Cruz @ VOCALS?](https://reader036.vdocuments.net/reader036/viewer/2022062322/5681502b550346895dbe1e55/html5/thumbnails/3.jpg)
Sample data from the Marine Stratocumulus Experiment (MASE). Note the occurence of bi-modal distributions most likely caused by collision-coalescence, a key process in drizzle formation.
![Page 4: UC Santa Cruz @ VOCALS?](https://reader036.vdocuments.net/reader036/viewer/2022062322/5681502b550346895dbe1e55/html5/thumbnails/4.jpg)
Example of bimodal distribution in stratocumulus clouds.
![Page 5: UC Santa Cruz @ VOCALS?](https://reader036.vdocuments.net/reader036/viewer/2022062322/5681502b550346895dbe1e55/html5/thumbnails/5.jpg)
Comparison of view volume diameter derived from in situ measurements and theoretically-derived curve.
The theoretical curve uses two parameters:
1. The laser beam waist diameter
2. The minimum scattered light intensity required for detection by photodetectors.
Fitting data to theoretical curve retrieves (1), leaving (2) as the only free parameter.
![Page 6: UC Santa Cruz @ VOCALS?](https://reader036.vdocuments.net/reader036/viewer/2022062322/5681502b550346895dbe1e55/html5/thumbnails/6.jpg)
Comparisons of PDI and Gerber derived LWC.
(A) No roll-off correction applied to Gerber data.
(B) Normal roll-off correction.
(C) Roll-off shifted by 7 um (towards smaller sizes)
(D) PDF of 1 Hz LWC data during MASE.