Bio|MAX UCLA abutment with Integrated Impression Transfer

Remove healing abutmentfrom implant

Assemble screw driver with Thumb Knob

Bio|MAX UCLA abutment with Integrated Impression Transfer

Insert the UCLA abutmentinto the implant ensuringthe appropriate orientation

Tighten the screw with the driver assembly with light finger force

Bio|MAX UCLA abutment with Integrated Impression Transfer

Remove screwdriver assembly

Apply wax to the top of UCLA abutment

Bio|MAX UCLA abutment with Integrated Impression Transfer

Inject impression material around the abutment

Bio|MAX UCLA abutment with Integrated Impression Transfer

Insert tray with impression materialSeat impression to capture dental arch

Bio|MAX UCLA abutment with Integrated Impression Transfer

Remove the impression from the mouth when material has set

Bio|MAX UCLA abutment with Integrated Impression Transfer

Remove UCLA abutmentfrom implant

Use a sharp instrument to remove the wax from the UCLA abutment

Bio|MAX UCLA abutment with Integrated Impression Transfer

Healing abutment in place

Insert healing abutment into implant

Bio|MAX UCLA abutment with Integrated Impression Transfer

Insert driver into UCLA abutment Insert abutment/analog assembly into

impression

Assembled abutment with implant level analog

Bio|MAX UCLA abutment with Integrated Impression Transfer

Align the flats of UCLA Abutment with the flats within the impression and insert the abutment/analog assembly into the impression

UCLA Abutment/analogassembly completely seated inside impression

Bio|MAX UCLA abutment with Integrated Impression Transfer

Apply soft tissue replica material around implant level analog

Pour dental stone into impression

Bio|MAX UCLA abutment with Integrated Impression Transfer

Bio|MAX UCLA abutment with Integrated Impression Transfer

UCLA abutment in place

Bio|MAX UCLA abutment with Integrated Impression Transfer

Cast and finish custom UCLA abutment

Create a wax up of the desired shape of the abutment by subtracting or adding to the resin of the UCLA abutment

Bio|MAX UCLA abutment with Integrated Impression Transfer

Seal the screw opening of the UCLA Abutment

Apply die spacer and wax up undercasting on the UCLA Abutment

Bio|MAX UCLA abutment with Integrated Impression Transfer

Porcelain crown on model

Cast wax up using a conventional technique and insert it in the casting on the stone model

Stack and fire the Ceramic in the usual manner

Bio|MAX UCLA abutment with Integrated Impression Transfer

Insert the abutment into implant orienting it according to the instructions from the laboratory and tighten with driver

Bio|MAX UCLA abutment with Integrated Impression Transfer

Insert driver into the abutment screw tighten the screw until torque of 30 Ncm is reached and the head of the ratchet releasesInsert the hex driver into the

30Ncm torque ratchet

Bio|MAX UCLA abutment with Integrated Impression Transfer

Cement restoration on abutment