uk journal of pharmaceutical and biosciences vol....
TRANSCRIPT
UK Journal of Pharmaceutical and Biosciences Vol. 2(6), 09-21, 2014 RESEARCH ARTICLE
New Two Sensors PVC- Membrane and Chemically Carbon Paste for Determination of Antidepressant Drug Escitalopram Oxalate in Bulk, Cipralex and Human Fluids Amal F. Khorshid*
Department of Pharmaceutical Analytical Chemistry, Faculty of Pharmacy, Nahda University, NUB, Beni-Sueff, 082-Egypt
Article Information
Received 22 September 2014
Received in revised form 25 Dec 2014
Accepted 27 Dec 2014
Abstract
Preparation and construction of a new polyvinyl chloride PVC-membrane and chemically
modified carbon paste (CMCP) sensors based on ion-pair exchanger escitalopram-silicotungstic
(Es-ST) (Sensor 1) and escitalopram-silicomolybdic (Es-SM) (Sensor 2). The effect of several
plasticizers and composition of ion-exchangers on the performance characteristics of the sensors
was studied. Sensors(1, 2) with 1% composition and dioctyl phthalate (DOP) as plasticizer in
PVC-membrane showed Nernstian slopes ranged from 57.5- 59.5 ±0.1 mV/decade over the
concentration ranged from 5.0 x 10-7-1.0 x 10
-2 M with the life span not less than two months and
pH 2.5-7.5 with a detection limit 0.1 nM. While in (CMCP) sensors (1, 2) exhibited with 3%
composition and tricresylphthalate (TCP) as binder an excellent Nernstian slopes 59.5, 60.5±0.5
mV/decade for 1 and 2 respectively and a wide concentration range from 1.0 x 10-7-1.0 x 10
-2 M
and pH 2.5-7.5 with a detection limit 0.5 nM. The sensors reflected high selectivity towards
different anions, cations, sugars and amino acids and was recommended by IUPAC. The
standard addition, the calibration curve and potentiometric titration methods were used for
determination escitalopram in its bulk powder, pharmaceutical tablets Cipralex, human
plasma/urine and monitoring profile for the tablet in vitro-dissolution rates. The recoveries were
excellent and with good agreement compared with British Pharmacopoeia.
Keywords:
Escitalopram Oxalate (Cipralex),
PVC-membrane,
Carbon-paste Sensor,
Potentiometry,
Vitro-dissolution rates,
British Pharmacopoeia
*Corresponding Author:
E-mail: [email protected]
Tel.: 01280875558
1 Introduction
Escitalopram Oxalate is an antidepressant drugwith a potent
selective serotonin reuptake inhibitor (SSRI) which is used for the
treatment of depression, generalized anxiety, social anxiety and
panic disorders. Escitalopram is a single isomer of the pure S-
enantiomer for the racemic phthalane bicyclic derivative of
citalopram. Escitalopram oxalate has a chemical structure S-(+)-1-
[3-(amino-dimethyl-)propyl]-1-(p-fluorophenyl)-5- oxalate of phthalan
carbonitrile as shown in Fig 1.
The empirical formula is C20H21FN2O.C2H2O4 with its weight 414.40.
Escitalopram oxalate is a white to slightly yellowish fine powder and
is freely soluble in methanol and (DMSO) dimethyl sulfoxide , soluble
in isotonic of saline solution, sparingly soluble in water and ethanol ,
insoluble in heptane with slightly soluble in ethyl acetate1,2
.
Literature survey showed several methods for the determination of
escitalopram including chromatographic HPLC3-6
, HPTLC7-10
, TLC11
,
spectrophotometric12-18
, fluorimetry19-20
, LC-MS21-24
, LC-MS/MS25
,
enantiomeric separation26
, CE27
and only one ISE method28
. A major
advantage of ISE is that it can be used very rapidly without having to
change range to measure samples with large batches covering wide
concentrations range. Achieving sufficient selectivity to measure, a
specific ion in the presence of others for the research29
. In addition, a
chemically modified sensor (CMEs) is one of the most important
electrochemical methods, which have been widely used as sensitive
and selectiveanalytical methods for the detection of the trace
amounts of biologically important compounds30,31
. CMEs have
important properties of their ability to enhance the selectivity in the
electroanalytical methods with respect for the potential and the
relatively selective interaction of the sensor mediator with the target
analyte in a coordination medium.
This work explains the formation and construction with potentiometric
characterization and analytical application of two types of ISSs for
the determination of escitalopram Oxalate. The first type based on
the formation of PVC- plastic membrane with ion-pair exchangers
escitalopram-silicotungstic (Es-ST) (Sensor 1) and escitalopram-
UK Journal of Pharmaceutical and Biosciences
Available at www.ukjpb.com ISSN: 2347-9442
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 10
silicomolybdic (Es-SM) (Sensor 2) and dioctyl phthalate (DOP) as
the best plasticizer, while the second type depends on the
construction of CMCPSs with the same ion-pair exchangers(Sensor
1 , 2) and tricresyl phthalate (TCP) as the best binder. The studying
in-vitro dissolution rate of bioavailability of poor water-soluble drug
escitalopram-oxalate that showed it is important in biopharmaceutical
quality of the product in its lifecycle. In addition, the analytical
determination of the drug in its pure state, tablets (Cipralex® 10
mg/tablet), biological fluids and the statistical treatments of the
results with respect to the official method and compared with sensors
previously reported28
.
NC
O
F
NCH3
CH3
C2H2O4
Fig 1 The chemical structure of Escitalopram Oxalate
2 Materials and Methods
2.1The Electrochemical system
The potentiometric measurements mode was carried out with a
Jenway 3510 digital pH/mV meter at 25±1 oC and Jenway 3505
digital pH/mV meter for measuring the ruggedness and pH. A
saturated calomel electrode (SCE) was used as an external
reference electrode. A techno circulator thermostat was used to
control the temperature of the test solution and for shaken the
samples. USP XXXII32
method was used to study the dissolution with
apparatus II33
, which was supplied in vitro dissolution testing for
controlled/modified-release preparations, and more uniform flow
profile. The equipment used is "SR8Plus", model CA USA Hanson
Research; with number "73-100-116" and the spectrophotometer
instrument UV-1800 Shimadzu (Japan).
2.2 Reagents and materials
All reagents used were chemically pure grade. Doubly distilled water
was used throughout all the experiments. Escitalopram Oxalate (Es-
Ox,M.wt = 414.40), and its pharmaceutical preparations (Cipralex®
10mg/tablet) were provided by H. Lundbeck Company, Cairo-
A.R.E.under License of Denmark. Silicotungstic acid (STA)
H4[SiW12O40], silicomolybdic acid (SMA) H4[SiMo12O40],
Dibutylphthalate (DBP), dioctylphthalate (DOP), tricresylphosphate
(TCP), tributylphosphate (TBP), graphite powder, poly (vinyl chloride)
(PVC) of high relative molecular weight, acetone, and
tetrahydrofuran (THF) were obtained from Aldrich.
0.5 M chloride solution for each of the following: Na+, K
+, NH4
+, Ca
2+,
Mg2+
, Ba2+
, Mn2+
, Zn2+
, Co2+
, Ni2+
, Cu2+
, Cd2+
, Pb2+
, Sr2+
, Cr3+
, Al3+
and Fe2+
solutions (1000 µg ml-1) , glucose anhydrous, lactose
monohydrate, maltose, urea, ascorbic acid, citric acid, L-threonine,
L-lysine, L-cystine, L-glycine, arginine, and L-alanine were prepared
and supplied from Aldrich. Silver nitrate and hydrochloric acid are
from local stores of NODCAR. Plasma was provided by VACSERA
(Giza, Egypt) and was used within 24 h while urine samples were
obtained from healthy volunteers.
2.3 Preparation and construction of plastic membrane sensors
Different compositions of membranes sensors were prepared. The
ion-exchangers were changed to cover the ranges of percentages
0.5-5% of each Es-ST, Es-SM. The membranes were prepared by
dissolving and mixing amounts of PVC and DOP in 5 ml THF as
solvent in Petri-dish (7.0 cm diameter) with the ion-exchangers Es-
ST or Es-SM. The total weight of constituents is fixed at 0.350 g. To
obtain uniform thickness, and homogenous the membranes were left
to dry freely in air (not less than 24 h).
For each composition, a membrane disk with 12 mm diameter was
punched from the large membrane and glued by a mixture of
(PVC+THF) to the polished end of a 2 cm long PVC plastic cap
attached by the same mixture to one end of a 10 cm glass tube
homemade sensor body. The sensors were filled with a mixture
solution of (10-1 M KCl and 10
-3M Es-Ox drug) as internal solution
and soaking in 10-3M Es-Ox drug solution. The electrochemical
system is represented as follows: Ag/AgCl/filling solution/PVC
membrane/test solution//SCE.
2.4 Preparation of carbon paste sensors
Fresh surfaces of the sensors wereprepared directly for
potentiometric measurements by wiped the surplus paste out and
squeezing more out of the paste. The exposed surface was polished
on a paper until the freshly surface showed shiny appearance.
Chemically modified carbon paste sensors were as previously
described30,31,34
.
2.5 Construction of the sensor calibration
A standard drug solutions were added to 50 ml doubly distilled water
to cover the concentration range 1.0x10-7
-1.0x10-2 M. Lower
Concentrations were prepared by appropriate dilutions. The sensor
and the reference sensor were immersed in the solution, and the emf
value was recorded at 25±l˚C, after each addition, the values were
plotted versus the negative logarithmic value of the drug
concentration (pDrug) and the resulting graph was used for
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 11
subsequent determination of unknown drug concentration from the
liner part of the curve (calibration curve method).
2.6 Effect of soaking on life span of the sensors and regeneration
The investigated conventional sensor (PVC membrane) was soaked
in 10-3 M Es-Ox drug solution at 25˚C. A calibration graph was
constructed for each sensor after different time intervals covering the
range from ½ h up to more than two months for each sensor. The
measurements were stopped when the slope of the calibration graph
deviated largely from Nernstian value and the sensor recovery
becomes out of range.
The sensor(s) regeneration was examined by freshly the reformation
of the ion-exchangers on the external gel layer of membrane. The
regeneration of the Es PVC membrane was achieved successfully by
soaking the exhausted sensor(s) for 24 h in a 1.0 x 10-2 M STA or
SMA solution, followed by soaking for 3 h in 1.0 x 10-2M Es-Oxdrug
solution.
2.7 Effect of pH on the sensor potential
The effect of pH on the test solution at different concentrations of the
Es-Ox (1.0x10-3, 1.0x10
-4 and 1.0x10
-5 M) on the potential values of
the sensor system was studied. For each concentration, 50 ml of the
drug solution were transferred to 100 ml measured cell, and the IS
sensor in conjunction with the calomel reference sensor, combined
with glass sensor were immersed in the same solution. The pH and
mV readings were recorded simultaneously. The pH for each
concentration was varied from 1.0-10.0 by adding very small
amounts of 2 M HCl and/or (0.1-1.0 M) NaOH solution. The pH-
values for each concentration were plotted against the mV-readings.
2.8 The selectivity of sensors
The matched potential method35,36
was recommended by IUPAC as a
method that gives analytically relevant practical selectivity coefficient
values. To study the selectivity coefficients of different interfering ion
for the IS sensors, a reference solution (aA) is added to the drug so
give a final concentration of (a-A), the change in potential (ΔΕ) is thus
measured. In addition, a reference solution containing the same
concentration (aA), the additional amount of the interference ion with
concentration not less than tenth times the concentration of the drug
that cause or reached the same change (ΔΕ) value so determined
(aj). The following equation is used to calculate the selectivity values
ofpot
JEs, zKlog :
Where: aA- is the initial concentration of drug, adrug is the activity of the
added drug, and aj is the activity of the added interfering ion
producing the same increase in potential.
2.9.1 Potentiometric determination of Es-Ox
The standard additions method was achieved by adding certain
volumes of standard drug solution to 50 ml water containing different
volumes of the Es-Ox drug in its pure state, pharmaceutical
preparation (tablets), and in human biological fluids plasma/urine
samples spiked with known volumes of the Es-Ox drug. The jump in
mV reading was recording for each increment and used to calculate
the concentration of the drug in sample solution using the following
equation [37]:
Vs Vx Cx = Cs (ـــــــــــــــ) (10
n (ΔΕ//S)(ــــــــــــــ -
-1
Vx+VS VS+Vx
Where Cx is the concentration to be calculated, Vx is the volume of
the original sample solution, Vs and Cs are the volume and
concentration of the standard solution added to the sample to be
analyzed, respectively, ΔΕ is the change in potential after addition of
certain volume of standard solution, and S is the slope of the
calibration graph.
2.9.2 Analysis of pharmaceutical preparation
20 tablets of (Cipralex® 10 mg/tablet) were powdered and weighed
(200-250 mg) portion from each was mixed with 50 ml doubly
distilled water, using the mechanical shaker for shaken in an about
30 min, filtered into a 100 ml volumetric flask and the solution was
filled to the mark with doubly distilled water and shaken. Different
known amounts of the solution (1.0-10 ml) were putted and
subjected to the potentiometric determination.
2.9.3 In spiked human plasma
Half ml drug of Es-Ox solution was taken from the concentration
1x10-3, 1x10
-4 and 1x10
-5 M in centrifugation shaking tubes 20-ml
stoppered and spiked with 4.5 ml plasma separately in each tube.
Adjusted the pH with phosphate buffer 6 then immersed the modified
sensor in conjunction with the calomel electrode taken the reading of
mV for each tube and washed with water between measurements.
From emf produced for each solution, the concentration of Es-Ox
drug was determined from the corresponding calibration and
standard addition methods.
2.9.4 In spiked human urine
Different quantities of the concentrations from 1.0x10-6 to 5.0x10
-4 M
Es-Ox drugs were putted in 100 ml volumetric flask spiked with five
ml urine and shaking for 5 min, then completed to the mark with
doubly redistilled water. A small addition of 0.01 M HCl (0.1– 2.0 ml)
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 12
was putted to adjust the pH from 4 to 5. The spiked urine was
determinedby the standard addition method for drug determination.
2.10 Content uniformity of Es tablets
Ten tablets of Cipralex® 10 mg/tablet was putted in 100 ml
measuring flask dissolved with doubly bidistilled water until 50 mL
mark then each sensor was immediately immersed in the sample
solution three times, washed between each individual measurement
with redistilled water to reach steady potential. Using the mean
potential from the calibration graph the content uniformity was
determined while for the spectrophotometric measurements by using
the standard solution employing UV absorbance λ max 240 nm.
2.11 Dissolution
In vitro release study, using USP XXXII32
apparatus II(paddle
method), one tablet of Cipralex® 10 mg/tablet was placed in gastric
media so the dissolution medium pH 1.2 (900 ml of 0.01 M HCl) was
maintained at 37±0.5˚C for 1 h.The revolution per minute dissolution
speed of the escitalopram tablet is measured at (90 rpm), close to
coated tablet with its physiological conditions. At intervals of time, the
potential values were recorded from the cell containing the
investigated sensor in conjunction with (SCE) reference sensor and
the amount of releasing was calculated from the calibration graph. In
the spectrophotometric measurements, five ml of the dissolution
solution were withdrawn then filtered and diluted with 0.01 M HCl
sothe concentration of samples was used in analyzed UV
spectrophotometer (1800, Shimadzu, Japan) and the absorbencies
were measured at λ max 240 nm. For drug release calculation the
calibration graph was used.
3 Results and Discussions
3.1 The Optimum composition of PVC membrane/ CMCPS
In plastic membrane sensors, the amount of the plasticizer should be
to the extent that produces a membrane of good physical properties
and at the same time plays efficiently its role as a solvent mediator
for the ion-exchanger(s) (lipophilic salts) [38].In this work, the ratio of
the plasticizer to polymer (PVC) was always 1:1 w/wand containing
one of the lipophilic salt, escitalopram-silicotungstic (Es-ST) (Sensor
1) and/or escitalopram-silicomolybdic (Es-SM) (Sensor 2).The
composition of each of these membranes or pastes was varied to
reach the optimum composition exhibiting the best performance
characteristics, (slope of calibration graph, concentration range and
reproducibility of the results) table 1.
Five plasticizers with differentpolarities, including DBP, TBP, TCP,
DOP and 2-NPPE was used to study the influence of the plasticizer
in PVC membrane or binder in CMCPs.The sensors in PVC
membrane containing DOP generally showed better potentiometric
responses, while in CMCPs TCP is the best, i.e. the sensitivity and
the linearity of the calibration plots. This is attributed to DOP in PVC
or TCP in CMCPs, as a low polarity mediator, for incorporation of the
highly lipophilesEs+ ion into the membrane prior or the paste binder
to its exchange with the soft ion-exchanger(s).
In addition among the different compositions studied1% composition
and dioctyl phthalate (DOP) as plasticizer in PVC-membrane showed
Nernstian slopes ranged from 57.5-59.5±0.1 mV/decade over the
concentration ranged from 5.0 x 10-7-1.0 x 10
-2 M with the life span
not less than two months and pH 2.5-7.5 with a detection limit 0.1
nM. While in (CMCP) sensors (1, 2) exhibited with 3%
compositionand tricresyl phthalate (TCP) as binder an excellent
Nernstian slopes 59.5, 60.5±0.5 mV/decade for 1 and 2 respectively
and a wide concentration range from 1.0 x 10-7
-1.0 x 10-2
M and pH
2.5-7.5 with a detection limit 0.5 nM.
3.2 Effect of soaking on life span and regeneration
In the PVC membrane, the effect of soaking on the performance
characteristics of the Es-sensors was studied for variable intervals of
time. Continuous soaking of the sensors for prelongated intervals of
time affected negatively their response to the drug cation. The
soaking has negative value effect may be due to the ion-
exchanger(s) leaching with plasticizer in the bathing solution that is
related to the diffusion rates and the distribution equilibria. The other
explanation can be attributed to the solvation by the water molecules
from the bulk solution into the surface of the membrane where the
lipophilic salts increase slow solvation, which leads to slowly leached
out with limiting the sensor life39
. As shown in table 2 and fig. 2.
After regeneration of the exhausted sensor characteristics (linear
concentration range, limit of detection, and slope 1x10-5-1x10
-2,
7.6x10-6 M and 38, 36 mV/decade for sensors1,2 respectively, were
changed to 1x10-5-1x10
-2, 3.5x10
-6 M and 55, 54 mV/decade for
sensors 1, 2 respectively. This can be attributed to the formation of
Es-ST or Es-SM after transferring the sensor from the drug to STA
and SMA solutions.
The effect of time on the performance of the CMCPS was studied by
measuring the slope at variable intervals of time starting from 30 min
reaching to 3 months. The results indicated that the life span (t), of
the sensors, in general, is more than those of the similar
conventional plastic membrane sensors. It is obvious that after
cutting and polishing the sensors surface, the slopes of the sensors
increase again as new sensor. The life span of the surface of the Es-
CMCPE sensor is at least or more than 90 days. As shown in table 2
3.3 Reproducibility of the Sensor
The repeatability examination for the potential reading of the Es-ST
sensor (1) and Es-SM sensor (2)/ PVC membrane was studied by
the subsequent measurements in 1.0 x 10-3 M Es-Ox solution then
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 13
followed by measuring the first set of at 1.0 x 10-4 M solution Es-Ox.
The standard deviation values for sensors are given in table 3 by
measuring emf for six replicate measurements. The obtained values
indicate the repeatability of the potential response of the sensors is
shown excellent results. In addition, the slope of the calibration graph
forEs-ST or Es-SM /CMCPS was nearly constant due to polishing
for any time taken days but starts to decrease gradually without
polishing so it consider as a new sensor with cutting and polishing to
the sensor. Table 3 exhibited the standard deviation values of
measuring emf for six replicate measurements obtained for each
type.
The dynamic response time40
was taken for any sensor by recording
the time required to achieve a steady-state potential (within ± 1 mV)
after subsequence immersions of the sensor in a series of Es- drug
solutions, each having a 10-fold increase in concentration from 1.0 x
10-7 to 1.0 x 10
-2 M. The practical response time was followed inEs-
Ox concentration up to be 10-fold.The sensors reached steady
potential differ in PVC than in CMCP where in CMCPS is less than
PVC. At the solution–paste interface, the kinetics of association–
dissociation of the ionophores with escitalopram ion is probably fast
exchange. The potential–time plot for the response of the sensor Es-
/PVC orCMCPs is shown in Fig. 3.
3.5 Effect of pH
The potential of the pH profile is obtained in the studying Es-
PVC/CMCPs and the responses of the sensors are constant over the
pH range 2.5–7.5. Inthat range the sensor can be safely used for the
determination of the Es-drug and there is no need to adjust the pH or
use buffer solution, as the drug solution is in the allowable or working
range of the pH of the sensors. It can be seen from Fig. 4 that at pH
values lower than the 2.5 pH ranges, the decrease of potential
readings can be related to interference of hydronium ion while at pH
values higher than pH 7.5 the formation of free base of escitalopram
at pH values higher than 7.5 decreases of the protonated species in
the test solutions leads to gradually decrease of the potential
readings.
3.6 Selectivity of the sensors
The sensors response towards different substances and ionic
species such as inorganic / organic cations, sugars, and amino acids
that may be present in the pharmaceutical preparations was
examined by MPM conditions for sensors 1, 2 in PVC and/ or CMCP
so the values of selectivity coefficients were used to evaluate of their
interference. The inorganic cations do not interfere due to their
mobility and permeability is difference both in PVC and/ or CMCP as
compared to escitalopram cation. For sugars and amino acids,the
high selectivity is related to the difference in polarity and lipophilic
nature of their molecules relative to escitalopram cation as shown in
table 4.
3.7 Validation of the proposed method
3.7.1 Linearity and detection limit (LOD)
The value of LOD for the proposed method is indicting the sensitivity
for detection the very small concentrations of Es reach to 0.1, 0.5nM
for PVC and/or CMCPS respectively. The correlation coefficient (r)
and other statistical parameters were listed in table 1.
3.7.2 Accuracy
The accuracy of the proposed PVC/CMCPS methods were
investigated by the determination of Es-Ox in its pharmaceutical
preparations without interfering from the co formulated adjuvant as
indicated by the mean recovery value of 99.75±0.08, 97.65±0.05
mV/decad for sensor 1, 2 in PVC and 99.87±0.04, 99.39±0.06
mV/decad for the investigated sensors 1, 2 in CMCP.
3.7.3 Precision
The relative standard deviation (% RDS) of the PVC/CMCPS
methods was tested by repeating the proposed method for analysis
of the investigated Es-Ox in intra-day (within the day) and inter-day
(consecutive days) to six replicates. The precision measured as
percentage of % RSD values obtained which are less than 2%,
indicating good precision.
3.8 Analytical applications
3.8.1 In PVC
For PVC sensors the analytical determination of the drug in its pure
state, tablets, and plasma/urine by the standard addition method,
calibration curve37
and potentiometric titration which are frequently
the most, applied in analytical application. The methodssuccessfully
were proved for the determination of escitalopram ions. The data
reflect the high validation for linearity, the investigated sensors for
determination of escitalopram ion and its statistical data treatments
in comparison with official methods1,2
. As shown in tables 5 and table
6.
3.8.2 In CMCP
The determination of Es shows that a wide concentration range of
the drug can be determined by the investigated sensors 1, 2 by
CMCP with high precision and accuracy. In addition, plasma/urine
samples the standard addition technique was applied to overcome
the matrix effects in these samples. Also, the response times of the
CMCP sensors are instant (within10 s), so the sensors are very
rapidly transferred back and forth between the biological human
samples and washed by the bi-distilled water between
measurements to protect the sensing component from adhering to
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 14
the surface of some matrix components. It is concluded that the
CMCP sensors can be applied successfully in vitro studies or for
clinical use. This confirms that limit of quantification (LOQ) and the
sensitivity are adequate for determination of escitalopram oxalate in
pharmacokinetic studies and the data are summarized in table 5.
Table 1 Composition and slope of calibration curves for Es- PVC membrane and CMCP sensors at 25.0±0.1˚C
Composition % (w/w) Slope
Es-ST PVC DOP C.R. (M) LOD (M) (mV/decade) R(s)
0.5 49.75 49.75 5.55 x 10-7- 1.0 x 10
-2 7.90 x 10
-7 58.5±0.1 ≤14
1.0* 49.50 49.50 5.00 x 10
-7- 1.0 x 10
-2 1.00 x 10
-8 59.5±0.1
* ≤12
3.0 48.50 48.50 6.35 x 10-7- 1.0 x 10
-2 7.90 x 10
-7 57.5±0.1 ≤14
5.0 47.50 47.50 7.45 x 10-7- 1.0 x 10
-2 7.90 x 10
-7 56.5±0.1 ≤15
7.0 46.50 46.50 7.79 x 10-7- 1.0 x 10
-2 7.90 x 10
-7 55.5±0.1 ≤15
Es-SM
0.5 49.75 49.75 6.25 x 10-7- 1.0 x 10
-2 7.90 x 10
-7 56.5±0.1 ≤14
1.0* 49.50 49.50 5.00 x 10
-7- 1.0 x 10
-2 1.25 x 10
-8 57.5±0.1
* ≤12
3.0 48.50 48.50 7.50 x 10-7- 1.0 x 10
-2 7.99 x 10
-7 56.5±0.1 ≤15
5.0 47.50 47.50 7.94 x 10-7- 1.0 x 10
-2 8.65 x 10
-7 55.5±0.1 ≤15
7.0 46.50 46.50 8.35 x 10-7- 1.0 x 10
-2 8.90 x 10
-7 54.5±0.1 ≤15
CMCPS
Composition % (w/w)
Es-ST graphite TCP C.R. (M) LOD (M) Slope R(s)
(mV/decade)
0.5 55.0 44.5 2.50 x 10-7-5.00 x 10
-3 6.50 x 10
-7 57.5±0.5 ≤8
1.0 55.0 44.0 5.00 x 10-7-1.00 x 10
-3 1.00 x 10
-7 58.5±0.5 ≤8
3.0* 55.0 42.0 1.00 x 10
-7-1.00 x10
-2 5.00 x 10
-8 60.5±0.5
* ≤5
5.0 55.0 40.0 5.00 x 10-7-1.00 x 10
-3 1.25 x 10
-7 57.5±0.5 ≤7
Es-SM TCP
0.5 55.0 44.5 7.99 x 10-7-5.00 x 10
-3 7.90 x 10
-7 56.8±0.5 ≤10
1.0 55.0 44.0 6.55 x 10-7-8.99 x 10
-3 5.50 x 10
-7 57.6±0.5 ≤7
3.0* 55.0 42.0 3.50 x 10
-7-1.00 x 10
-2 4.00 x 10
-8 59.5±0.5⃰ ≤5
5.0 55.0 40.0 7.50 x 10-7-7.31 x 10
-3 5.00 x 10
-7 57.0±0.5 ≤8
Continued Table 1: Response characteristics of the Es-ion exchangers at 95% confidence intervals, average of six replicates at
25.0±0.1 °C
Parameters
Sensors
PVC membrane CMCPS
Sensor 1
Es-ST
Sensor 2
Es-SM
Sensor 1
Es-ST
Sensor 2
Es-SM
Composition
(W/W %)
1 % Es-ST
+49.50 % PVC
+49.50 % DOP
1 % Es-SM +49.50 PVC
+49.50 DOP
3 % Es-ST
+ 55.0 % G
+ 42.0 % TCP
3 % Es-SM
+ 55.0 % G
+ 42.0 % TCP
Slope (mV/decade) 59.5±0.1 57.5±0.1 60.5±0.5 59.5±0.5
Correlation coefficient
(r) 0.998 0.997 0.999 0.998
LOD (M) 1.00 x 10-8 1.25 x 10
-8 5.00 x 10
-8 4.00 x 10
-8
Linear range (M) 5.0x10-7
-1.0x10-2
5.0x10-7-1.0x10
-2 1.0x10
-7-1.0x10
-2 3.5x10
-7-1.0x10
-2
Working pH range 2.5-7.5 2.5-7.5
Response time (s) ≤12-15s
≤5-10 s
Life span (days) 72 ds 66 ds 90 ds 90 ds
Recovery (%) ± S.D*
99.75±0.08 97.65±0.05 99.87±0.04 99.39±0.06
Robustness 99.88±0.02 99.27±0.05 99.87±0.08 99.70±0.05
Ruggedness 99.65±0.06 99.45±0.07 99.95±0.05 99.55±0.03
* Six replicates
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 15
Table 2 Effect of soaking on Es-PVC membrane sensor using 1.00x10-3
M Es-Ox at 25.0±1.0˚C
Soaking time Slope (mV/decade) Linear-range (M) Response time
Es-ST/ PVC (tresp). (s)
1/2 hr 59.5±0.1 5.00x10-7
-1.00x10-2
≤12
1 59.5±0.1 5.00x10-7
-1.00x10-2
≤12
7 59.5±0.1 1.00x10-7-1.00x10
-2 ≤20
14 59.5±0.1 5.00x10-6
-5.00x10-3
≤20
21 52.0±0.1 1.00x10-6
-5.00x10-3
≤20
28 48.5±0.1 5.00x10-5
-3.50x10-3
≤25
45 45.0±0.1 1.00x10-5
-2.50x10-3
≤25
60 42.5±0.1 5.00x10-4
-2.50x10-3
≤30
70 38.0±0.1 1.00x10-4
-1.25x10-3
≤30
Soaking time Slope (mV/decade) Linear-range (M) Response time
Es-SM/ PVC (tresp). (s)
1/2 hr 57.5±0.1 5.00x10-7
-1.00x10-2
≤12
1 57.5±0.1 5.00x10-7
-1.00x10-2
≤12
7 55.5±0.1 1.00x10-7-1.00x10
-2 ≤20
14 55.5±0.1 5.00x10-6-5.00x10
-3 ≤20
21 52.0±0.1 1.00x10-6
-5.00x10-3
≤20
28 48.5±0.1 5.00x10-5
-3.50x10-3
≤25
45 45.0±0.1 1.00x10-5
-2.50x10-3
≤30
60 42.5±0.1 5.00x10-4
-2.50x10-3
≤30
66 38.0±0.1 1.00x10-4
-1.25x10-3
≤30
Effect of soaking on Es-CMCPE using 1.00x10-3
M Es-Ox at 25.0±1.0˚C
Soaking Time Slope (mV/decade) Linear range (M) Response time
(tresp), (s)
CMCP Es-ST Es-SM
1/2 hr 59.5±0.1 57.5±0.1 5.00x10-7
-1.00x10-2
≤ 6
3 days 59.5±0.1 57.5±0.1 5.00x10-7
-1.00x10-2
≤ 6
10 days 59.5±0.1 57.5±0.1 1.00x10-7-1.00x10
-2 ≤ 7
15 days 58.5±0.1 57.5±0.1 5.00x10-7
-1.00x10-2
≤ 7
20 days 58.5±0.1 56.5±0.1 5.00x10-7
-1.00x10-2
≤ 8
35 days 57.5±0.1 56.5±0.1 1.00x10-6-1.00x10
-3 ≤ 8
50 days 57.5±0.1 56.5±0.1 5.00x10-6
-1.00x10-3
≤ 9
75 days 57.5±0.1 55.5±0.1 5.00x10-6
-1.00x10-3
≤ 9
85 days 57.5±0.1 55.5±0.1 1.00x10-6-1.00x10
-3 ≤ 10
90 days 57.5±0.1 55.5±0.1 5.00x10-6
-1.00x10-3
≤ 10
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 16
Table 3 Standard deviation values of measuring emf for six replicate measurements obtained for each type
Sensors Standard deviation S.D
1.0 x 10-4
1.0 x 10-3
Es- Sensors
(1) Es-ST/ PVC 0.23 0.42
(2) Es-SM/ PVC 0.54 0.76
(3) Es-ST/ CMCP 0.29 0.47
(4) Es-SM/ CMCP 0.33 0.55
Table 4 Selectivity coefficient values for pot
zJDrug,K log - Es-PVC/CMCPS
Sensor 2 Sensor 1 Sensor 2 Sensor 1
CMCP PVC CMCP PVC Interferent CMCP PVC CMCP PVC Interferent
4.22 4.10 4.28 4.12 Glucose
3.73 3.42 3.84 3.55 Na+
4.17 4.00 4.21 4.07 Lactose 3.30 3.15 3.49 3.27 K+
4.15 4.06 4.20 4.10 Maltose 3.80 3.82 3.92 3.64 NH4+
4.02 4.01 4.06 4.05 L-Lysine 3.79 3.87 3.83 3.90 Cu2+
4.10 4.06 4.15 4.08 L-cystine 3.71 3.50 3.87 3.67 Zn2+
4.03 4.01 4.07 4.01 L-Glycine 3.63 3.80 3.98 3.95 Co2+
4.14 4.03 4.16 4.06 L-Theronine 2.40 2.13 2.42 2.14 Fe2+
4.27 4.27 4.31 4.35 Urea 3.60 3.73 3.88 3.86 Fe3+
4.13 4.21 4.18 4.24 Ascorbic acid 3.59 3.68 3.69 3.79 Ni2+
4.15 4.01 4.17 4.05 Asparagine 3.66 3.69 3.68 3.75 Mn2+
4.05 4.21 4.36 4.18 L-arginine 3.83 3.77 3.89 3.84 Mg2+
4.32 4.26 4.42 4.53 L-proline 3.84 3.90 4.07 3.97 Cr3+
4.22 4.27 4.52 4.31 L-valine 3.94 3.89 4.25 4.12 Ba2+
Each value is the average of three determinations
Table 5 Determination of escitalopramoxalate in bulk solutions, tablet, plasma and urine applying the standard addition method and
potentiometric titrations in PVC membrane and CMCPS
Standard addition method
Bulk solutions
Potentiometric titration
Bulk solutions
Potentiometric titration
Cipralex® 10 mg/tablet
PVC CMCP PVC CMCP PVC CMCP
Taken
(mg)
R (%) R.S.D
%
R (%) R.S.D
%
Taken
(mg)
R (%) R.S.D
%
R (%) R.S.D
%
R (%) R.S.D.
%
R (%) R.S.D
%
Sensor 1 Es-ST Sensor 1 STA as titrant
2.07 98.7 0.66 99.7 0.74 12.43 99.8 0.62 99.9 0.48 101.7 0.44 101.9 0.48
4.14 98.4 0.58 99.8 0.62 20.72 99.6 0.47 99.7 0.52 101.8 0.52 101.7 0.59
6.21 98.8 0.49 99.4 0.60 29.01 99.7 0.79 99.8 0.70 101.5 0.60 101.4 0.67
8.28 98.1 0.82 98.9 0.45 41.44 98.4 0.93 99.6 0.43 101.9 0.48 101.7 0.71
Sensor 2 Es-SM Sensor 2 STA as titrant
2.07 98.7 0.46 99.4 0.26 6.21 99.1 0.54 99.7 0.52 101.6 0.57 101.4 0.43
4.14 98.6 0.59 99.6 0.65 8.28 99.3 0.33 99.0 0.47 102.0 0.48 101.3 0.69
6.21 98.5 0.66 99.3 0.39 12.43 99.0 0.43 99.6 0.79 101.8 0.53 101.0 0.47
Cipralex® 10 mg/tablet
Sensor 1
20.72
29.01
100.9
101.3
0.50
0.33
99.3
98.6
0.65
0.53
101.4
101.7
0.38
0.22
101.2
101.5
0.41
0.52
2.07 99.1 0.71 99.8 0.44
4.14 99.4 0.58 99.5 0.76 Sensor 1 SMA as titrant
6.21 98.8 0.56 99.0 0.93 6.21 100.1 0.32 100.5 0.83 100.7 0.59 101.4 0.39
Sensor 2 8.28 100.5 0.50 100.6 0.78 101.5 0.49 101.7 0.66
2.07 99.0 0.38 99.7 0.59 12.43 100.2 0.57 100.3 0.40 101.3 0.29 102.0 0.61
4.14 98.7 0.84 99.4 0.73 20.72 100.7 0.46 100.8 0.69 101.8 0.77 101.6 0.53
6.21 99.1 0.56 99.0 0.51 29.01 100.5 0.62 101.3 0.48 101.5 0.66 101.7 0.48
Spiked human plasma Sensor 1 41.44 101.4 0.74 101.8 0.67 101.9 0.80 101.2 0.42
2.07 98.1 0.88 98.9 0.42 Sensor 2 SMA as titrant
4.14 98.8 0.57 99.0 0.59 2.07 100.5 0.38 100.4 0.46 101.3 0.43 101.6 0.78
6.21 98.5 0.70 98.7 0.53 4.14 100.7 0.44 100.7 0.64 101.5 0.61 101.7 0.61
Spiked urine Sensor 1 8.28 100.4 0.37 101.0 0.36 101.7 0.71 101.9 0.63
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 17
Table 6 Statistical treatment of data obtained for the determination of escitalopram oxalate applying the standard addition method and
potentiometric titration by comparison with official methods and using PVC / CMCPs
X±S.E.: Recovery± standard error, F-tabulated is 6.39 at 95.0% confidence limit, t-tabulated is 3.143 at 99.0% confidence limit and 6 degrees of freedom
pEs
2468
E, m
V
50
100
150
200
250
300
350
400
450
2468
2468
2468
2468
2468
2468
2468
2468
Fig. 2 Calibration graphs obtained at 25 o C after soaking [1] Es-ST and [2] Es-SM PVC- membrane sensors for 1/2 hr (a), 1 d (b), 7 d (c), 14 d (d), 21 d (e),
28 d (f), 45 d (g), 60 d (h) and 70 days (i)
pEs
2468
E, m
V
50
100
150
200
250
300
350
400
2468
2468
2468
2468
2468
2468
2468
2468
(a)(b)
(c) (d)
(e)
(f)
(g)
(h)
(i)
(a) (b)
(c)
(d)
(e)(f)
(g)
(h) (i)
[1] [2] Es-SMEs-ST
Time, sec0 200 400 600
E, m
V
100
200
300
400
500
600
700
Es-ST /PVC
Es-ST /CMCPS
Es-SM /PVC
Es-SM /CMCP
Fig. 3 The response time-potential plot for Es-Ox
10-7
10-6
10-5
10-4
10-3
10-2
M
M
M
M
M
M
2.07 98.0 0.66 99.4 0.51 12.43 100.8 0.58 101.5 0.63 101.5 0.85 101.6 0.48
4.14 98.7 0.54 99.6 0.48 20.72 100.9 0.17 101.3 0.49 101.3 0.55 101.5 0.71
6.21 98.4 0.49 99.2 0.63 29.01 101.4 0.47 101.7 0.77 102.0 0.44 101.9 0.55
Standard addition method
Sample Official (1)
method
PVC CMCP
Sensor 1 Sensor 2 Sensor 1 Sensor 2
Bulk solutions
X±S.E. 98.5±0.2 98.7±0.3 98.4±0.1 99.2±0.6 98.5±0.1
F value 4.55 3.93 5.33 5.25
t value 2.78 2.57 4.61 3.44
Cipralex® 10 mg/tablet
X±S.E. 101.7±0.2 101.1±0.7 101.4±0.1 101.3±0.5 101.7±0.5
F value 4.98 5.63 4.70 6.35
t value 3.61 4.98 3.32 4.15
Potentiometric titrations
Bulk solutions
X±S.E. 98.5±0.2 99.4±0.2 99.5±0.1 99.7±1.0 99.8±0.4
F value 2.30 2.44 3.38 3.11
t value 2.45 2.22 2.78 2.50
Cipralex® 10 mg/tablet
X±S.E. 101.7±0.2 101.8±0.7 100.7±0.5 100.8±0.3 100.9±0.5
F value 3.13 2.35 3.45 4.66
t value 2.45 1.77 2.72 4.65
Spiked human plasma
X±S.E. 101.7±0.2
101.5±0.4 101.2±0.5 100.7±0.2 100.9±0.5
F value
3.66 3.55 3.05 3.68
t value
2.83 2.45 2.44 2.76
Spiked human urine
X±S.E. 101.7±0.2
101.5±0.8 101.7±0.5 101.0±0.6 101.4±0.5
F value
2.83 2.75 3.65 4.55
t value
2.66 2.67 2.12 2.94
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 18
pH0 2 4 6 8 10
E,
mV
50
100
150
200
250
300
350
400
Es-ST / (PVC)
0 2 4 6 8 10
0
100
200
300
400
500
Es-SM / (PVC)
E,
mV
pH
0 2 4 6 8 10
0
100
200
300
400
500
600
E,
mV
pH
Es-ST /CMCPs
0 2 4 6 8 10
0
100
200
300
400
500
E,
mV
pH
Es-SM /CMCPs
Fig. 4 Effect of pH on the potential response of Es-Ox for PVC and CMCP sensors
10-5
10-4
10-3
10-5
10-4
10-3
M
M
M
M
M
M
10-5
M
10-4
M
10-3
M
10-5
M
10-4
10-3
M
M
0 10 20 30 40 50 60 70
0
20
40
60
80
100
120
CMCP
Fig. 5 Dissolution profiles of 10 mg escitalopram tablets obtained by (A) potentiometric:
sensor Es-PVC/CMCPE, and (B) spectrophotometric measurements at 240 nm.
Time in min
Rel
ease
%
PVC
spectro
Statistical treatment of results
The recoveries of the results for Es applying the calibration curve,
standard additions method and the potentiometric titration were
evaluated statistically and were compared with the values obtained
with the (0fficial) pharmacopeia method by applying the F-tests41,42
.
The values obtained table 6 show that the determination methods
have a precision comparable to that of the pharmacopeia method.
However, the PVC/ CMCP methods are more practical regarding
time of analysis, consumption of solvents and sample pretreatment
requirements for spectrophotometric or chromatographic analysis of
escitalopram oxalate.
3.9 Robustness and ruggedness
The robustness method was examined by the replacement of
aqueous solution with phosphate buffer pH 6±0.5 for the Es-
PVC/CMCPSs. The results are in a good agreement with those
obtained from standard drug solution as shown in table (1). In
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 19
addition, the ruggedness was checked by using another type of pH-
meter model (Jenway, 3505) for each sensor (1, 2) in table 1.
3.10 Content uniformity of Cipralex® tablets
The content uniformity assay for Es-PVC/CMCPS methods
described good accuracy, precision and reproducible resultsfor the
quality control tests so the sensors can be working for the
quantification determination of escitalopram ions and the percentage
of the recovery of Es-Ox is mainly in acceptance quantitatively.
3.11 Potentiometric monitoring of Cipralex®dissolution
The changes of the dissolution medium caused by the dissolving
drug can be detecting by the sensors PVC/ CMCP monitors to the
dissolution of the tablet. The changes are converted into % of the
concentration in dissolved drug via dedicated (“Potential measured in
mV, to Concentration”). The results are compared to analysis using
UV spectrophotometric. It shows that escitalopram releases
immediately after capsule was ruptured and the releasing was
achieved during 15 min nearly 50% was released, then than 74%
drug was released within 20 min and complete dissolution in the third
stage within 35 min according to USP1.
For the UV spectrophotometric assay, fixed volumes of the
dissolution medium were withdrawn, diluted with 0.01 M HCl,
measured at λ max 240nm and compared with a calibration graph.
Fig. 5 shows the dissolution profiles of escitalopram tablet using both
measurement techniques. The results obtained by PVC/CMCP ISS
and spectrophotometric are almost identical but the use of the ISS
methods sensors have the advantage of more sensitive due to
overcome the matrix effect.
5 Acknowledgements
I gratefully acknowledge the support of National Organization for
drug control and research (NODCAR) in Giza-Egypt, by the
chemicals and dissolution equipment acknowledge the supporting of
Central Research Lab at Nahda University with the other equipments
at the practical work.
6 Competing interests
The sensors developed (PVC/CMCP) are superior as compared
with the escitalopram ISS (PVC) described in the literature28
.
7 Author’s contributions
MA is participated in collection of escitalopram oxalate as raw
material from (NODCAR) and in literature survey.
8 References
1. The United States Pharmacopeia. May 1, 2010 Authorized
USP Pending MonographVersion 2.
2. Sweetman SC. Martindale -The Complete Drug
Reference,PharmaceuticalPressLondon2009; 391:996.
3. Gandhi SV, Dhavale ND, Jadhav VY,Sabnis
SS.Spectrophotometric and Reversed‐phase
High‐performance Liquid Chromatographic Methods For
Simultaneous Determination of Escitalopram Oxalate And
Clonazepam In Combined Tablet Dosage Form.Journal Of
Association of Official Analytical Chemists
International.2008; 91(1): 33-38.
4. Syama B,Suneetha A. Development And Validation of
Liquid Chromatographic Method For Estimation of
Escitalopram Oxalate In Tablet Dosage
Forms.International Journal of Pharma And Bio
Sciences.2011; 2(1): 140-146.
5. Tapobana S, Suddhasattya D, Bhusan HS, Bharat Kumar
D, Mohanty DL,Bhar K. RPHPLC Method For The
Estimation Of Escitalopram In Bulk And In Dosage
Forms.International Journal of Chemistry Research.2011;
2(2): 11-15.
6. Chakole RD, Charde MS, Bhavsar N,Marathe RP.
Simultaneous Estimation of Escitalopram and Clonazepam
By RP-HPLC In Pharmaceutical Formulation.International
Journal of Phytopharmacy. 2012; 2 (1): 25-29.
7. Mahadik MV, Dhaneshwar SR,Kulkarni MJ. Application of
Stability Indicating HPTLC Method For Quantitative
Determination of Escitalopram Oxalate In Pharmaceutical
Dosage Form.Eurasian Journal of Analytical Chemistry.
2007; 2(2): 101 –117.
8. Dhavale N, Gandhi S, Sabnis S,Bothara B. Simultaneous
HPTLC Determination Of Escitalopram Oxalate And
Clonazepam In Combined Tablets.Chromatographia,
2008; 67(5‐6): 487‐ 490.
9. Kakde R, Satone D, Bawane N. HPTLC method for
simultaneous analysis of escitalopram oxalate and
clonazepam in pharmaceutical preparations. JPC- J Planar
Chromat. 2009; 2(6):417-20.
10. Kakde RB, Satone D,Bawane N. HPTLC Method for
Simultaneous Analysis of Escitalopram Oxalate and
Clonazepam In Pharmaceutical Preparations.Journal of
Planar Chromatography.2009; 22(6): 417–420.
11. Soliman SM,Youssef NF. Enantiomeric thin-layer
chromatographic assay of escitalopram in presence of “in-
process impurities”.Journal of Planar
Chromatography.2011; 24(6): 475-481.
12. Shete Y, Pimpodkara N, Mahajana NS, Pore YV, Jadhava
RL,Kuchekar BS. Spectrophotometric Estimation of
Escitalopram Oxalate In Pharmaceutical Dosage
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 20
Forms.International Journal of Chemical Sciences.2009;
7(1): 235-237.
13. Kakde RB,Satone DD. Spectrophotometric Method for
Simultaneous Estimation of Escitalopram Oxalate and
Clonazepam In Tablet Dosage Form.Indian Journal of
Pharmaceutical Sciences.2009; 71(6): 702–705.
14. Chaudhari BG,Parmar HR. Spectrophotometric Method
For Determination Of Escitalopram Oxalate From Tablet
Formulations.International Journal of Pharmaceutical
Quality Assurance.2010; 2(1): 9‐12.
15. Vetrichelvan T, Arul K, Sumithra M,Umadevi B.
Colorimetric Method For The Estimation Of Escitalopram
Oxalate In Tablet Dosage Form. Indian Journal Of
Pharmaceutical Sciences.2010;72(2): 269‐271.
16. Sharma S, Rajpurohit H, Sonwal C, Bhandari A,
Choudhary VR,Jain T. Zero Order Spectrophotometric
Method For Estimation Of Escitalopram Oxalate In Tablet
Formulations.Journal of Young Pharmcist.2010; 2(4): 420–
423.
17. Fatema K, Rahman MZ, Biswas S,Akter S. Development
of UV Spectroscopic Method For Nefopam And
Escitalopram As INN Drugs In Tablet Dosage
Form.Stanford Journal Of Pharmaceutical Science.2010;
3(1): 4-10.
18. Sharma S, Rajpurohit H, Sonwal C, Sharma P,Bhandari A.
Simultaneous Spectrophotometric Determination of
Escitalopram Oxalate and Clonazepam Using Multi-
component Mode of Analysis.Journal Of Pharmacy
Research.2010; 3(9): 2303-2305.
19. Serebruany V, Malinin A, Dragan V, Atar D,Zyl L.
Fluorimetric Quantitation of Citalopram And Escitalopram
In Plasma: Developing An Express Method To Monitor
Compliance In Clinical Trials.Clinical Chemical Laboratory
Medicine.2007; 45(4): 513–520.
20. Elham AT, Nahla NS,Shudong W. Micelle Enhanced
FluorimetricAnd Thin Layer Chromatography
Densitometric Methods For The Determination Of (±)
Citalopram And Its S – Enantiomer Escitalopram.Analytical
Chemistry Insight.2009;4:1–9.
21. Singh SS, Shah H, Gupta S, Jain M, Sharma K, Thakkar
P,Shah R. Liquid Chromatography–electrospray Ionisation
Mass Spectrometry Method For The Determination Of
Escitalopram In Human Plasma And Its Application In
Bioequivalence Study.Journal Of Chromatography B.2004;
811(2): 209-215.
22. Johannesson N,Bergquist J. Rapid on-line extraction and
quantification of escitalopram from urine using sol–gel
columns and mass spectrometric detection.Journal of
Pharmaceutical and Biomedical Analysis.2007;43: 1045–
1048.
23. Zheng H, Yang L,Tang WY. Determination of Plasma
Escitalopram With Liquid Chromatography-mass
Spectrometry.Nan Fang Yi Ke Da XueXueBao.2008;
28(11): 2044-2046.
24. Dhaneshwar SR, Mahadik MV,Kulkarni MJ. Column Liquid
Chromatography-ultraviolet And Column Liquid
Chromatography/Mass Spectrometry Evaluation of Stress
Degradation Behavior of Escitalopram Oxalate Journal of
Association of Official Analytical Chemists International
2008; 92(1): 138-147.
25. Suresh PS, Giri S, Husain R,Mullangi R. A Highly
Sensitive LC-MS/MS Method for The Determination of S-
citalopram In Rat Plasma: Application to A
Pharmacokinetic Study in Rats.Biomedical
Chromatography.2010; 24(10): 1052–1058.
26. Greiner C, Hiemke C, Bader W,Haen E. Determination Of
Citalopram And Escitalopram Together With Their Active
Main Metabolites Desmethyl(es-)citalopram In Human
Serum By Column-switching High Performance Liquid
Chromatography (HPLC) And Spectrophotometric
Detection.Journal Of Chromatography B.2007; 848: 391–
394.
27. Sungthong B, Jac P,Scriba G. Development and validation
of a capillary electrophoresis method for the simultaneous
determination of impurities of escitalopram including the R-
enantiomer.Journal of Pharmaceutical and Biomedical
Analysis.2008; 46: 959–965.
28. Al-Amri FMG, Alarfaj NA,AlyAF.Development of New
Sensors for Determination of Escitalopram Oxalate in
Dosage Forms and Biological Fluids.Int.J.Electrochem.Sci.
2013; 8: 10044 – 10058.
29. Issa YM, Khorshid AF. Using PVC ion-selective electrodes
for the potentiometric flow injection analysis of distigmine
in its pharmaceuticalformulation and biological fluids.J. of
Advanced Research.2011; 2:25–34.
30. Ibrahim H, Khorshid A.Modified Carbon Paste Sensor for
Cetyltrimethylammonium Ion Based on Its Ion-associate
with Tetrachloropalladate(II).Analytical Sciences. 2007;
23:573-579.
31. Khorshid A, Amin RR,Issa YM. Fabrication of a Novel
Highly Selective and Sensitive Nano-Molar Cu(I), Cu(II)
Membrane Sensors Based on Thiosemicarbazide and
Acetaldehydethiosemicarbazone Copper Complexes.J.of
ChemicaActa. 2012; 152-58
32. United States Pharmacopeia.National Formulary USP 32–
NF 27. Convention, Inc.: Rockville, MD. 2009.
Khorshid New two sensors pvc- membrane and chemically carbon paste
UK J Pharm & Biosci, 2014: 2(6); 21
33. Liddell M, Deng G, Hauck W, Brown W, Wahab S,
Manning R, (2007) Dissolution Technologies, 14(1):28.
34. Khorshid A.F, Issa YM.Modified carbon paste sensor for
the potentiometric determination of neostigmine bromide in
pharmaceutical formulations, human plasma and urine.
Biosensors and Bioelectronics. 2014; 51: 143–149
35. Umezawa Y, Buhlmann P, Umezawa K, Tohda K,
Amemiya S.Potentiometric selectivity coefficients of ion-
selective sensors. Part I. Inorganic cations (technical
report).Pure Appl Chem. 2000; 72: 1851-55.
36. Umezawa Y, Umezawa K, Sato H.Selectivity coefficients
for ion-selective sensors: recommended methods for
reporting KA,B pot values (technical report).Pure Appl
Chem. 1995; 67: 507-511.
37. Guilbault G, Drust R.A, Frant MS, Freiser H, Hansen, et
al.Recommendations for nomenclature of ion-selective
sensors. Pure Appl Chem. 1976; 48: 127-129.
38. Ammann D, Pretsch E, Simon W, Lindner E, Bezegh A,
Pungor E. Lipophilic salts as membrane additives and their
influence on the properties of macro- and micro-electrodes
based on neutral carriers. Anal ChimActa.
1985;171(C):119–129.
39. OeshU, SimonW..Effect of lipophilic salts increase slow
solvation. Anal. Chem.1980; 52:692.
40. Linder E, Umezawa Y.Performance evaluation criteria for
preparation and measurement of macro- and micro
fabricated ion-selective sensors.Pure Appl Chem. 2008;
80(1): 85–104.
41. Skoog DA, Holler FJ, Neiman TA.Principles of
Instrumental Analysis, 5thed. Harcourt Brace College
Publishers. London. 1997.
42. Miller JC, Miller JN.Statistics for Analytical Chemistry.Ellis
Horwood.Chichester.England. 1994.