UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System Training Modules Name_________________________ This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in any other manual.

Upload: others

Post on 16-Mar-2020

13 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

UniCel DxH Slidemaker Stainer

Coulter Cellular Analysis System Training Modules

Name_________________________

This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in any

other manual.

Page 2: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DXH SLIDEMAKER STAINER TRAINING MODULES

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

ii of iv

WARNINGS AND PRECAUTIONS

READ ALL PRODUCT MANUALS AND CONSULT WITH BECKMAN COULTER-TRAINED PERSONNEL BEFORE ATTEMPTING TO OPERATE INSTRUMENT. BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY EQUIPMENT.

INTENTION FOR USE

This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in any other manual.

TRADEMARKS

THE BECKMAN COULTER logo, COULTER, DxH, and UniCel are trademarks of Beckman Coulter; the BECKMAN COULTER logo COULTER, and UniCel are registered in the USPTO.

All other trademarks, service marks, products or services are trademarks or registered trademarks of their respective holders.

Copyright Beckman Coulter, Inc. 2014 All Rights Reserved

Find us on the World Wide Web at: www.beckmancoulter.com

Beckman Coulter Ireland, Inc. Mervue Business Park, Mervue Galway, Ireland 353 91 774068

Made in USA

Page 3: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DXH SLIDEMAKER STAINER TRAINING MODULES

Unicel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

iii of iv

REVISION STATUS

Revision AA Initial Release (March 2012) Software release 1.0 and higher

Revision AA.1 (March 2013) Software release 1.0 and higher

Revision AA.2 (June 2014) Software release 1.0 and higher Presented June 2014, Miami, Florida

Page 4: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DXH SLIDEMAKER STAINER TRAINING MODULES

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

iv of iv

Page 5: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

1 of 6

COURSE GUIDE

This document is not intended to replace the information in your Instrument Instructions for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in any

other manual.

UniCel DxH Slidemaker Stainer

Basics

Troubleshooting

Additional Topics

System Operations

Page 6: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN A96923AA.2 (June 2014)

2 of 6

Page 7: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

3 of 6

TRAINING TOPICS

BASICS

SYSTEM OPERATIONS

TROUBLESHOOTING

ADDITIONAL TOPICS

Page 8: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN A96923AA.2 (June 2014)

4 of 6

Page 9: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

5 of 6

GROUP DISCUSSIONS The following discussions take place in the classroom/laboratory. Ending times noted are approximate.

Tuesday

General Information

Learn about each other

Safety: Personal Protective Equipment and Fire Evacuation

Introduction to the course and materials

Location of additional reference materials

Registration Questionnaire

Instrument Demo

Wednesday

SMS Presentation

Thursday

Additional Topics

Friday

Additional Topics

Evaluations

Graduation

Page 10: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN A96923AA.2 (June 2014)

6 of 6

You will see the following symbols used throughout the learning modules.

Important information you should know

Techie-type information you may be interested in knowing

Do these steps now!

Timely Tip

You reached the goal!

I do not understand…Questions

Resources, Help

Page 11: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-1 of 26

COURSE MAP – BASICS

BEGIN HERE

Basics

Daily Checks &

Shutdown

Page 12: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-2 of 26

PERSONAL PROGRESS SUMMARY

Basics Modules Date Initial Facilitator

Basics

Daily Checks & Shutdown

Page 13: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-3 of 26

BASICS

OBJECTIVES Given an operating DxH Slidemaker Stainer:

Log On and Log Off the System Manager

Identify the main hardware of the system by labeling a diagram

Identify the areas of the optional floor stand by labeling a diagram

Locate and identify the areas and icons used on the screen by labeling a diagram

State the meaning of icon colors and how to respond to them

Access System HELP using the Help icon

Find information using the Contents, Index and Search features

Use hypertext jumps within Help

Print Help topics

Create Favorites in Help

Why Is It Important?

This module introduces the basics of the DxH Slidemaker Stainer. A basic hardware overview allows identification of the general areas of the system.

These topics form the building blocks for the rest of your learning journey.

Resources

DxH Slidemaker Stainer

System HELP

Page 14: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-4 of 26

Page 15: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-5 of 26

INFORMATION/PRACTICE

Logoff/Logon

In order to access the System Manager, you must Log on. Both an Operator ID and a Password are required. The system administrator will assign these to you in the lab.

Step Action

1 Select the Logoff icon at the top right of the screen.

2 On the Logoff screen, select the logoff button.

3 On the Logon screen, type in the Operator ID: BCI (case sensitive), press tab.

4

Type in the Password: bci (case sensitive).

NOTE: Neither the Operator ID nor the password “bci” is present in the system. Upon installation, Operator IDs and Passwords must be set up.

5 Select the Log On button or Press Enter.

You will now be on the DxH Slidemaker Stainer “home” screen. Once you select any button or icon to move to a different screen, the home screen disappears. You will only see it after a Log On or if you press the F8 key.

System Manager Statistics

The System Manager features touch screen capability that lets you select the Status icons and Menu buttons on the screen.

Your System Manager (computer) consists of the following:

LCD flat panel monitor with touch screen capability

Computer with a CD/DVD RW drive running Microsoft Windows XP Professional operating system and the DxH Solutions software

Standard Keyboard

Optical Mouse

Hand-held barcode scanner

Small Form Factor Computer

Page 16: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-6 of 26

Access System HELP

Find the Help icon along the top right of the screen, next to the Logoff icon. Select this icon to enter the System HELP. To close Help, select “x”.

Step Action

1

Select the System Status icon to give you an overview of the instrument. On this screen, you will see the status of the SMS, LIS, and Printer. We will explain this screen further in the Daily Checks Module.

2 Select the System HELP icon .

3 Select the Index Tab. Type the words home screen in the keyword field.

4 A list displays, double click on the words home screen in the list or select the words and select the “Display” button at the bottom.

5 Take some time to study the Home Screen section to learn the names of the areas on this screen. You will label a diagram with the correct names as part of the skill check later.

6 To place this topic in Favorites as a reference for you, select the Favorites tab (to the right of the Index tab).

7 Then, select the “Add” button near the bottom.

8 You may now refer to this topic any time you select Favorites.

9 To delete a topic from Favorites, highlight it and select Remove.

10 The Help screen may be moved and resized to place it in a better position on the screen while you are working. Try it now!

11 The font icon at the top of the Help screen allows you to increase or decrease the size of the font used to make it easier for you to read. Try it now!

12

The print icon at the top of Help allows you to print any information that you want from Help. It will print in the font size chosen. Please feel free to print any topics that you want this week.

13 To close Help, select “x”.

Page 17: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-7 of 26

Software Overview

System Status, Worklist, Daily Checks, Supplies, Event Log

To review the software areas, reference the HELP, Chapter 1. System Overview>Viewing System Status/System Overview.

System Icons

Step Action

1 Click the Start icon to place the Slidemaker Stainer online for processing.

2 Click the Stop icon to place the Slidemaker offline (inactive state); stainer and STM are still active.

Note If an Audible Alarm is triggered, click the Turn Off Alarm icon to shut the audible alarm off.

Displays if unit is Online, Offline, in Diagnostics, or Stopped

Active = online and ready Inactive = offline and not ready Disabled = maker or stainer not available for use Enabled = maker and/or stainer available for use

Printer Status: Status displays Auto Print Disabled if Disable All Auto Printed Reports is selected from Print> Print Printer Setup

LIS Status: Green = LIS Interface ON Gray = LIS Interface OFF

Alert Status Icon color meaning: Red = error

Yellow = warning

Neutral = OK

Page 18: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-8 of 26

Hardware Overview

The DxH Slidemaker Stainer is a slidemaking module, a slidestaining module and a System Manager. It is precisely integrated to provide process control, slidemaking and staining, and cassette or single-tube delivery of specimens.

System HELP can provide a quick overview:

Step Action

1 Select the Help icon .

2 Select the Contents tab.

3 Find Chapter 1 System Overview.

4

Use “Next” on the top right corner of the help screen to navigate the different Figures of the Slidemaker Stainer. You will label diagrams with the correct names as part of the skill check at the end of this module.

5 Place this info into your Favorites.

Note The three Beacon colors are Red (Error), Yellow (Warning), or Green (OK to Run).

Optional Floor Stand

Step Action

1 Select the Index tab.

2 Type floor stand.

3 Double click Floor Stand Features from the list.

4 Take some time to study the Figures and their descriptions. You will label a diagram with the correct names as part of the skill check at the end of this module.

Congratulations on completing this portion of your training. Please continue to the Skill Check on the next page.

Page 19: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-9 of 26

BASICS SKILL CHECK 1. Label the sections of the Home screen:

1. ________________________ 5. _________________________

2. ________________________ 6. __________________________

3. ________________________ 7. _________________________

4. ________________________ 8. _________________________

2. Label the main areas of the SMS System

1. ______________

2. ______________

3. ______________

1

2 3 4 5

6

7

8

1

2 3

Page 20: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-10 of 26

3. Label the parts of the Slidemaker Stainer:

1. ______________________ 6. ______________________

2. ______________________ 7. ______________________

3. ______________________ 8. ______________________

4. ______________________ 9. ______________________

5. ______________________ 10. _____________________

12

3

4

5

6

78

10

9

Page 21: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-11 of 26

4. Left side of the Slidemaker Stainer:

5. Right side of the Slidemaker Stainer:

1. ______________________ 5. ______________________

2. ______________________ 6. ______________________

3. ______________________ 7. ______________________

4. ______________________

6. Show facilitator your favorites tab. Demonstrate how to add and remove a topic.

1. ______________________

2. ______________________

3. ______________________

4. ______________________

1

2

3

4

5

6

7

2

3

4

1

1

2

3

4

Page 22: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-12 of 26

7. Show facilitator how to Log off and Log on.

8. Label the parts of the Floor Stand:

1. ______________________ 7. ______________________

2. ______________________ 8. ______________________

3. ______________________ 9. ______________________

4. ______________________ 10. _____________________

5. ______________________ 11. _____________________

6. ______________________ 12. _____________________

9. What do the colors mean? ___________________________________

_____________________________________

_____________________________________

1

2

3 4

5

7

8

9

10

11

12

6

Page 23: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-13 of 26

DAILY CHECKS & SHUTDOWN

OBJECTIVES Given an operational DxH Slidemaker Stainer and using the System HELP:

Point to the last time a Daily Checks or Shutdown was run using the System Status screen

Access the Daily Checks Summary Screen

Request a Daily Checks cycle

Set up automatic Daily Checks after Shutdown

View Daily Checks Information

Add a comment to Daily Checks results

Set up automatic printing of the Daily Checks results

Perform the Shutdown procedure using DxH Cleaner

State the minimum time the instrument should be in Shutdown and how often Shutdown should be performed

Set up automatic Shutdown to occur for five days at 17:30pm

Why Is It Important?

The Daily Checks screen shows you the results of running the Daily Checks diagnostic tests. The summary screen gives a quick overview of the results and any alerts due to failed tests.

The Shutdown procedure must be performed once every 24 hours that the instrument is used to ensure optimal cleaning of the instrument.

Resources

DxH Slidemaker Stainer

System HELP

Page 24: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-14 of 26

Page 25: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-15 of 26

INFORMATION/PRACTICE

Daily Checks

The Daily Checks testing is a once-a-day procedure normally performed after running a Shutdown cycle. Both of these cycles have auto-program features. We will discuss this and set it up later in the module. Even when you have the cycle pre-programmed, you may still request to perform a Daily Checks cycle at any time (for troubleshooting, etc.). For additional Reference, refer to Chapter 3, Daily Checks.

Requesting a Daily Checks Cycle

Step Action

1

Select the System Status icon . Look at the SMS Status screen to give you an overview of basic information about the instrument.

2 Click on the Details Status button or double click on the instrument picture on the screen.

LIS STATUS

STAINER & MAKER STATUS

PRINTER STATUS

UNIT STATUS

Page 26: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-16 of 26

If the date and time are a neutral color, all is OK. If they are red, there is a problem.

3

Look at the Maker and Stainer status for Active, Inactive, or Disabled state.

This is where you view the status of the maker and stainer.

Note

Transport Status may be Online, if processing samples.

Offline, for special cycles and procedures.

Diagnostics or Standby, for certain diagnostic tests.

Stopped, when system is commanded by the operator to stop the process.

Disconnected, when the system is powered off.

4 Select the Daily Checks alert status icon or Menu>QA>Daily Checks

5 Select Daily Checks from the local navigation bar.

DATE & TIME of last DAILY CHECK and SHUTDOWN Cycles MAKER & STAINER

STATUS

TRANSPORT STATUS DEFAULT SLIDE ORDER CASSETTE PREPARATION SINGLE-TUBE PREPARATION

Page 27: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-17 of 26

6

When the system displays: You have requested to perform Daily Checks.

Select OK to run Daily Checks For classroom practice, Perform a Daily Checks now.

7 This cycle takes approx. 3 minutes.

Viewing Daily Checks Information

Step Action

1

Look at the Daily Checks alert status icon . If it is neutral, all is well. If it is red, then something failed and needs review. You can view the daily checks result information for the Summary, SMS, and Supplies.

Note

The Daily Checks must pass or be reviewed in order to run specimens. If Daily Checks passes, the Review button is disabled.

Page 28: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-18 of 26

2 Select the Review button from the local navigation bar to clear any system failures and to return to neutral.

3 Select the View Log button to review the Daily Checks tab of the History Log.

4 After resolving the cause, repeat Daily Checks.

5 For classroom practice, use View Log to find a Daily Checks with a Fail entry, select it and then select the Details button.

Note

When you click “Review” for failed results, you are acknowledging the failure and the need to repeat the Daily Checks. Be sure to document per your regulatory requirements. You could add a comment in the History Log.

If you review a failed Daily Checks result, the system will prompt the operator with the following message: Are you sure you want to acknowledge the Daily Checks failure?

Press OK to continue.

Enable Automatic Daily Checks

Step Action

1

From the Daily Checks screen, select Auto Configuration>Configure Daily Checks to display the Auto Daily Checks Configuration dialog box.

2

1

Page 29: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-19 of 26

2 Select Enable Automatic Daily Checks to enable running daily checks automatically on your system.

3

You have 2 options to choose from:

Perform Daily Checks After Shutdown Perform Daily Checks at a Specified Time

For classroom practice, select Perform Daily Checks at a Specified Time. Hour will be 6, Min will be 30, and check the Daily check box.

4

If you selected Perform daily checks after shutdown, select OK to save your selection. or If you selected Perform daily checks at a specified time, select a time and a day, and select OK to save your selection.

Viewing Past Results/Adding Comments

Step Action

1 Select the View Log button from the Local Navigation Bar.

2 You may select one date at a time to review.

3

Once highlighted in blue, select the Details button below to view all tabs of those results. Select View Log to return to the log or the large back arrow on the upper left to go back one screen.

Do this at this time!

Page 30: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-20 of 26

4

With any date highlighted, select the Comments button below to add a comment. For classroom practice, add a comment.

5 Select OK to save comment.

6

Once saved, you will see a yellow, page icon appear to let you know there is a comment there. You may go back to a comment to edit it as well.

7 Be sure to save at least one comment for any Daily Checks result as part of this exercise.

Page 31: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-21 of 26

Printing/Auto-Printing

Step Action

1 You may manually print individual results by selecting the print icon from the utility area at the upper right .

2

On the daily checks screen, you may print the Daily Checks Auto Report.

If you are on Daily Checks: View Log>Log Type>Data Summary Logs, you may print: Entire log, All filtered, or Selected entries by clicking the print icon from the utility area at the upper right.

3 You may decide to have the Daily Checks print automatically.

4 To do this, return to the Daily Checks screen by selecting the Daily Checks icon.

5

Select the Auto Report button from the Local Navigation bar. This opens the Daily Checks Auto Report window.

Page 32: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-22 of 26

6

You may select to print automatically by checking the Enable auto report generation box.

Shutdown

The DxH Slidemaker Stainer allows you to program the length of time the instrument is in Shutdown. The default time is 30 minutes. This is the minimum time required for Shutdown.

At the end of your programmed time, the system removes DxH Cleaner and replaces it with DxH Diluent. It then awaits a Daily Checks cycle, performed manually or automatically.

If a Shutdown cycle needs to be interrupted, just cancel and then request a Daily Checks cycle.

Step Action

No other cycles or System Manager actions can be performed until the Shutdown cycle has completed. Once the Shutdown cycle is complete, a countdown timer bar appears near the bottom of the Daily checks screen to let you know how much time remains for the cleaning.

1 Select or Select Menu>QA>Daily Checks.

2 On the Daily Checks Screen, select the Shutdown button from the local navigation bar.

3

For classroom practice, select OK to begin the shutdown or In the event that you choose not to do a Shutdown, select Cancel to go back to the previous screen.

Note Refer to Chapter 5 Shutdown in HELP for further information on the Shutdown Procedures.

Page 33: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-23 of 26

Enable Automatic Shutdown

Step Action

1 From the Daily Checks screen, select Auto Configuration>Configure Shutdown to display the Auto Shutdown Configuration dialog box.

2 Select Enable Automatic Shutdown to enable running Shutdown automatically on your system.

Note

In your Laboratory you may set-up shutdown as follows:

Type the time for automatic shutdown in Hour and Minutes in the text box.

Select Daily or select specific days of the week to run Shutdown automatically.

Type the time the cleaner should remain in the DxH Slidemaker Stainer during the shutdown procedure in Hours and Minutes.

Select OK to save your selections.

4 For classroom purposes, type in one hour for time in cleaner.

5 Type in 17:30 PM for the time the cycle should begin and select Monday through Friday.

6 Select OK.

Note Refer to Chapter 6 Setup in HELP for further information on Enabling Automatic Procedure.

Please continue to the Skill Check on the next page.

Page 34: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-24 of 26

Page 35: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

B-25 of 26

DAILY CHECKS & SHUTDOWN SKILL CHECK

1. Show your printouts from Daily Checks of today’s result and one from a week ago to the facilitator.

2. Go to the DxH SMS Stainer Status screen and point to the time of the last Daily Checks cycle.

3. Show facilitator that you set up automatic Daily Checks using Auto Configuration.

4. Show facilitator the comment you added to a Daily Checks results.

5. Show facilitator that you set up automatic printing of Daily Checks.

6. What is the minimum time for the DxH SMS to be in shutdown? ________________________________________________________

7. How often should a Shutdown cycle be performed? ________________________________________________________

8. What is the reagent used for Shutdown?

________________________________________________________

Page 36: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

B-26 of 26

Page 37: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-1 of 62

COURSE MAP – SYSTEM OPERATIONS

Supplies

Setting Up the

Slidemaker

BEGIN HERE

Setting Up the

Stainer

Disabling/Enabling

the Stainer & Maker

Sample

Processing

Page 38: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-2 of 62

PERSONAL PROGRESS SUMMARY

System Operations Modules Date Initial Facilitator

Supplies

Setting Up the Slidemaker

Setting Up the Stainer

Disabling/Enabling the Stainer & Maker

Sample Processing

Page 39: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-3 of 62

SUPPLIES

OBJECTIVES Using an operating DxH Slidemaker Stainer and System HELP:

Setup and replace supplies using Beckman Coulter reagents

Setup and replace other supplies manually

Loading Slides

Why Is It Important?

An up-to-date supplies screen shows that your laboratory is aware of good Quality Assurance procedures. Even if the same lot number of a reagent replaces the previous container, the supplies setup must be updated to show the correct open container expiration date.

Resources

DxH Slidemaker Stainer

System HELP

Page 40: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-4 of 62

Page 41: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-5 of 62

INFORMATION/PRACTICE Note: For any non-Beckman Coulter reagent, select Setup Other and manually enter the bar code information. You can enter up to five bar codes.

Setting Up DxH Slidemaker Reagents (Diluent & Cleaner)

Step Action

1 Select to display the Supplies screen.

2 Select the Slidemaker tab.

3

On the Supplies screen, select the Setup button from the local navigation bar.

Page 42: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-6 of 62

Configure Low-Level Notification

Step Action

1 Select to display the Supplies screen.

2 From either the Slidemaker or Stainer tab, select the Config Low Level button from the local navigation bar on the Supplies screen.

3 On the Configure Low Level Condition screen, select a low-level percentage from the drop-down box (default is 10%).

4 Select OK to save your changes.

Note In the classroom, we set the low-level warning to 10%. When this occurs, the Supplies Alert Status Icon is yellow. If a supply is out, the icon is red.

4 On the Setup Supplies dialog box, scan or manually enter the bar codes into Bar Code fields 1 and 2, and select OK.

5

If replacing Diluent, choose Diluent 1 or Diluent 2 from the Supply Type drop-down list and select OK.

If replacing Cleaner, select OK.

Note: The fields on the Setup Supplies screen are automatically populated from the bar code information. The Supply screen updates the percent (%) remaining for the supply bottle graphics.

6 Select OK to verify the reagent has been replaced.

7 Repeat this procedure for each remaining supply.

Page 43: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-7 of 62

Setting Up Beckman Coulter Stainer Reagents

Follow these steps to set up reagents for the stainer.

Note: For any non Beckman Coulter reagent, select Setup Other and manually enter the bar code information. You can enter up to five supplies.

Step Action

1 Select to display the Supplies screen

2 Select the Stainer tab.

3

On the Supplies screen, select the Setup button from the local navigation bar.

Page 44: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-8 of 62

4

Scan or manually enter the barcodes into Bar Code fields 1 and 2, and select OK.

For any non Beckman Coulter reagent, select Setup Other and manually enter the bar code information. You can enter up to five supplies.

5 On the Supply field, choose the reagent to be replaced from the drop-down list.

6

On the displayed message, select OK to verify that the reagent has been replaced.

For non Beckman Coulter reagents, after selecting OK the system will display the Setup Supplies screen. Select OK (no need to enter any information in the barcode fields).

7 You would repeat steps 3 to 5 above for each remaining supply.

8

Select Details from the local navigation bar to display information for each supply and select OK.

Note: Slidestainer Supplies Details indicates the supplies required for the active staining protocol set for the instrument.

Page 45: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-9 of 62

Supplies History Logs

Step Action

1 Select the from the local navigation bar on the Supplies screen.

There are two history logs associated with Supplies. One is located under Event Logs and the other is located under Data Summary Logs.

Supplies Event Log:

This log records any changes or alerts associated with the supplies.

Supplies Data Summary Log:

This log records all the information on each new container set up on the System.

2 Find both of these logs and examine the entries in each one.

Page 46: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-10 of 62

Loading Slides

The software supply icon turns yellow when the slide level is low.

The software supply icon will turn red when the slides are completely out. The system will go offline until more slides are loaded.

Go back online to continue processing the slides.

Risk of injury and biohazardous contamination. Handle slides with care to avoid skin puncture. Clean up any broken glass or blood spill as quickly as possible. Dispose of all contaminated disposable cleaning materials and broken slides in accordance with your local regulations and acceptable laboratory practices.

Note: There are 2 sensors on the backside of the slide chute. One of the sensors, checks the slide inventory below ½ (approx. 75 slides). When this happens, the Supplies Icon will be in Yellow. The other sensor will stop the Slidemaker Stainer. This will occur when there are 12 slides left in the chute. The Supplies Icon will be in red.

Step Action

1 Open the slide ejector door.

2

Remove slides from the box handling them by the painted end.

Page 47: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-11 of 62

3

Load the slides into the slide chute with the painted side up.

Note: You can load up to 172 slides. Be aware that slides exposed to high humidity may stick together.

4

Close the slide ejector door.

You reached the goal!

Please continue to the Skill Check on the next page.

Page 48: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-12 of 62

Page 49: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-13 of 62

SUPPLIES SKILL CHECK

1. Show facilitator the steps to setup Slidemaker reagent and Stainer reagent.

2. Show facilitator both history logs associated with Supplies.

3. The Supply icon will turn __________ when the slide level is low.

The Supple icon will turn __________ when the slides are completely out.

4. Show facilitator the slides you placed in the slide chute.

Page 50: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-14 of 62

Page 51: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-15 of 62

SETTING UP THE SLIDEMAKER

OBJECTIVES Using the System HELP: Chapter 6 Setup

Set up Default Slide Order/Set Number of Slides

Set up Basket Settings

Set Pre-Stain Drying Type

Configure the Slide Label

Set Up Basket Location Availability

Set Up Stain Only Priority

Set Up Agitation

Set Up Heater Availability

Set Up Smear Settings

Why Is It Important?

Setting up the slidemaker according to your laboratory will ensure maximum instrument performance.

Resources

DxH Slidemaker Stainer System

System HELP

Page 52: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-16 of 62

Page 53: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-17 of 62

INFORMATION/PRACTICE

Set Default Slide Order

Use Default Slide Order for samples that do not have a slide order from the LIS or samples without a slide request from the worklist. Runs with default slide order are done with batching enabled.

Step Action

1 Select to display the System Status screen.

2 Select the Details Status from the local navigation bar.

3

Select Default Slide Order and the number of slides from the drop-down list for Cassette: Preparation and Single-Tube Preparation.

For classroom practice, select Make Slides and Stain and Number to 1.

Note

When you select Cassette Preparation, the maximum number of slides obtained per sample is 12. When you select Single Tube Preparation, the maximum number of slides obtained per sample is 4.

Page 54: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-18 of 62

Select Basket Settings

Use basket settings to select the number and distribution of slides in each basket, the pre-stain slide drying time, and the time to advance baskets with different priorities through the Stainer module.

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the DxH Slidemaker Stainer Status screen.

2 Select Details Status from the local navigation bar.

3 Select SMS Configuration from the local navigation bar.

4 Select Basket Settings from the local navigation bar on the DxH Slidemaker Stainer Configuration screen to display the Basket Settings dialog box.

5

For classroom practice, select Every other position and set the Number of Slides to 10.

6

Configure the Advance Timeout options by entering the minutes for each timeout (after which a basket will advance).

For classroom practice, set the STAT Timeout to 2 minutes, ASAP Timeout to 3 minutes, and Routine Timeout to 4 minutes.

Note: Timeout is the amount of time set to move the basket after pre-drying (drying before staining). The timeout must be equal or greater than the amount of time for pre-drying.

Page 55: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-19 of 62

7

Enter the Pre-Staining Drying Time in minutes and seconds by using the arrows.

For classroom practice, set the Pre-Staining Drying Time to 1 minute.

8 Select OK to save the setting.

Set Pre-Stain Drying Type

Basket Settings allows you to select a Pre-Stain Drying Time.

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status Screen, select SMS Configuration from the local navigation bar.

4

Select Pre-Stain Drying from the local navigation bar to display the Pre-Stain Dryer Settings dialog box.

5

For classroom practice, select Heat and Fan Drying.

NOTE: Selecting Fan Drying powers off the heaters and allows the smears to dry without heat.

6 Select OK to save the settings.

Page 56: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-20 of 62

Configure the Slide Label

This feature lets you assign sample information to print on the slide for identification.

Note: The system must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status Screen, select SMS Configuration from the local navigation bar.

4

Select Slide Labeling from the local navigation bar to display the Label Configuration screen.

5

Select Insert Field from the local navigation bar to add a field to the slide label from the displayed list or Select Remove Field to remove a field from the slide label.

Note: The Specimen ID and Date/Time Made cannot be removed.

TIP To improve visibility of STAT slides in a basket, ***STAT*** will print on a slide if: the specimen has STAT priority or the Priority field is configured to print on the slide.

Page 57: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-21 of 62

6

Select the individual field to move to the desired location by using drag and drop or by using the arrows to move the field on the grid.

A red outline will be displayed around the fields when the fields overlap.

7 Use the arrow on the Font Size drop-down list to change the font size to Small, or Large.

8 From the Selected Field, select Specimen ID from the labeling configuration screen.

9

Enable the Print with Barcode checkbox to include the Specimen ID bar code on the slide label.

You can select to print the bar code label identical to the sample or you can select a Data Matrix (2D) bar code label.

For classroom practice, select the 2D bar code.

Note: Slide bar code labels will default to 2D when the number of characters with certain symbologies exceed the printable area of the slides.

10 Move the Darkness scroll bar to adjust lighter or darker print.

11 Select the Option A or Option B radio button to determine the orientation of the label.

12

Select Save Changes to save the label configuration updates or Select Discard Changes to ignore the label configuration updates or Select Restore Default to display the original instrument slide label configuration.

13

Select OK to confirm that you are discarding changes or restoring the default

Note: If you are saving changes, the system will not ask for your confirmation.

Page 58: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-22 of 62

Set Up Basket Location Availability This feature lets you enable and disable positions to place a basket in the I/O drawer and the internal storage area during operation.

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4

Select Basket Location Availability from the local navigation bar on the DxH Slidemaker Stainer Configuration screen to display the Basket Location Availability screen.

5

Click to enable or disable the positions for basket placement in the I/O drawer and the internal storage area (displayed to the right of the I/O drawer on the Basket Availability screen).

For classroom practice, select one of the positions of the I/O drawer and disable it by clicking on it.

The internal storage area holds seven additional storage positions that may be used when the I/O drawer is full or no remaining positions are available.

6 Select OK to save the settings.

Page 59: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-23 of 62

Set Up Stain Only Priority

The stain-only feature is used to prioritize the staining cycle when the information is not downloaded from the LIS.

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4 On the SMS Configuration screen, select Stain-Only Priority from the local navigation bar.

5

Select the appropriate radio button for your stain-only processing priority.

STAT – immediately ASAP – as soon as possible Routine - as configured by your lab.

For classroom practice, select Routine as the Stain-only Processing Priority.

6 Select OK to save your changes.

Page 60: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-24 of 62

Set Up Agitation

Follow these steps to facilitate the staining process:

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4 On the SMS Configuration screen, select Agitation Settings from the local navigation bar.

5

On the Agitation Settings dialog box, verify that a checkmark appears next to Stainer Agitation Enabled to enable agitation.

6 Select OK to save your changes.

Page 61: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-25 of 62

Set Up Heater Availability

Follow these steps to warm up the smears for faster drying:

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4 On the SMS Configuration screen, select Heater Availability from the local navigation bar.

5

On the Heater Availability dialog box, verify that a checkmark appears next to your selections for enabling heaters.

Both heaters enable drying by heat.

Pre-Stain Heater provides heat for drying smears that just received printing information and are waiting in the basket elevator.

Post Stain Heater provides heat for drying smears in baskets located in the stainer dryers.

Note: Slides made while heaters are not selected will generate slide exceptions. See the Slide Exceptions in CHAPTER 7. Use this configuration only when instructed by your Beckman Coulter Representative.

6 Select OK to save your changes.

Page 62: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-26 of 62

Set Up Smear Settings

Follow these steps to manipulate the smear as needed for testing and user preferences.

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4 On the SMS Configuration screen, select Smear Settings from the local navigation bar.

5

On the Smear Settings dialog box, use the arrows to make your selections for the smear's length, thickness, edge, wicking position as it relates to the amount of blood drop used, and blood drop position near or away from the sample information area.

For classroom practice, set the Smear Settings Length to Longer and Edge to the Square.

6 Select OK to save your changes.

You reached the goal!

Page 63: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-27 of 62

SETTING UP THE SLIDEMAKER SKILL CHECK

1. Show the facilitator your SMS Status screen for Default Slide Order and Number of slides.

2. What is the maximum number of slides for:

Cassette Presentation _____________________________________

Single Tube Presentation ___________________________________

3. Show the facilitator the Basket Settings

4. Show the facilitator the Basket Location Availability settings.

Page 64: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-28 of 62

Page 65: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-29 of 62

SETTING UP THE STAINER

OBJECTIVES Given an operating DxH Slidemaker Stainer System and using the System HELP

Copy a Staining Protocol

Delete a Staining Protocol

Add/Edit a Staining Protocol

Configure Baths (mapping) to the Bath Location

Set Active Staining Protocol

Set Up a Quick Rinse (read only)

Why Is It Important?

Setting up appropriate stain protocols for your laboratory ensures that slides are properly stained for review of normal and abnormal cell population results.

Changing staining protocols can accommodate various types of stains needed in a laboratory.

Resources

DxH Slidemaker Stainer System

System HELP

Page 66: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-30 of 62

Page 67: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-31 of 62

INFORMATION/PRACTICE

Setting up the stainer includes the following: Adding or copying a stain protocol Installing reagents according to a selected protocol Setting an active protocol

Any protocol can be edited except for the default protocols in bold in the Staining Protocols table.

When you change a protocol and reagents for a new active protocol, perform steps in the Flushing Reagent Lines and Stainer with Methanol procedure in Chapter 9, Cleaning Procedures to remove current reagents and clear the lines.

Copy a Staining Protocol

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4 Select Staining Protocols from the local navigation bar on the DxH Slidemaker Stainer screen.

5 Select a protocol to copy from the displayed list.

6 Select Copy Protocol from the local navigation bar. The Staining Protocol window for the protocol you selected is displayed.

7 Rename the copy.

8 Select OK to save your settings.

Note

When OK is selected to save your settings, the system w ill give you the following message:

The selected protocol will be set as the active protocol. Select OK to continue.

Page 68: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-32 of 62

If you select a different protocol with a different consumable configuration that is not consistent with the consumable configuration of the instrument, the system will prompt you with:

Do you still want to set it as the active protocol? Click YES

Delete a Staining Protocol

The instrument must be offline and no staining should be in progress in order to perform this procedure.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4 Select the Staining Protocols from the local navigation bar on the DxH Slidemaker Stainer screen.

5 Select the Protocol you want to delete from the Staining Protocols list.

6 Select Delete Protocol from the local navigation bar.

Note This option is allowed only in user-defined protocols. Default protocols (In bold text) cannot be altered.

7 Select OK to delete the protocol.

Page 69: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-33 of 62

Add/Edit a Staining Protocol

Important Create or edit a stain protocol before placing reagents on a new system or adjusting a reagent setup on an existing system.

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure. This option is allowed in user-defined protocols. Default protocols (In bold text) cannot be altered.

Step Action

1 Select to display the System Status Screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4 Select Staining Protocols from the local navigation bar.

5 Select the Protocol.

6 Select the Edit Protocol button.

Configure Baths (Mapping) to the Bath Location

IMPORTANT: Create or edit a stain protocol before placing reagents on a new system or adjusting a reagent setup on an existing system.

Step Action

1 Select the Wright Stain Protocol that you created.

2 Select Configure Baths on the Staining Protocol screen.

3 Select the Blue underlined names for the supplies and select the supply type for that location. Repeat for all supplies.

Page 70: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-34 of 62

4

Connect all the baths to their supply types:

Select the bath. The supply types available for connection are shown in bold type.

Select the supply type(s) you need. (The operator may select None, select a supply type from the list, and select Select to save the supply type). The lines connecting the bath and supply will be displayed in the same color as the supply type bottle.

Note

Configuration is required for all baths that are associated with a supply type. For baths 2, 3, and 4 you may configure up to two supplies into one bath. Baths 1 and 5 can only accommodate one type of supply.

Note The system will display *Configuration Required to the left of the Supply Type graphic above the Configure Bath button. An asterisk will also be displayed next to the bath name.

5 Select Configure Bath.

6

Configure the bath content in Bath Content and the frequency required in Auto Drain and Refill. When more than one supply type is configured for a bath, you can define the ratio of one supply to the other in the % fields. The bath content for combined supplies must equal 100%. Set the time that the basket will stay in the bath in the Duration field.

From the When drop-down list, you can set Auto Drain and Refill to occur:

Page 71: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-35 of 62

At a specified time - lets you select times when the baths will drain and refill at the same time

After an elapsed time – tells the system to perform this function upon completion of a time duration

For every specified slide carrier – counts the number of baskets carrying slides into the stainer bath

Note: After an elapsed time and For every specified slide carrier may cause the drain and refill function to occur independently from each stainer bath.

7 Select OK to confirm your selection.

8

Select the Preview Protocol checkbox to display the selected mapping for the staining protocol.

Deselect the checkbox to return to the Staining Protocol screen.

9 Select OK to go back to the Staining Protocol screen, Protocol Options.

10

Select the post-stain drying duration (amount of time the slide is allowed to dry after staining).

Select OK to return to the Staining Protocol table.

Page 72: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-36 of 62

Set Active Stain Protocol

This selection will activate the staining protocol to run the stainer.

Note: The instrument must be offline and no staining should be in progress in order to perform this procedure.

Risk of instrument errors. Remove empty baths from the tray to prevent instrument errors and facilitate cleaning in the event of an overflow into the tray. See Clean Stainer Baths in Chapter 9, Cleaning Procedures for information on cleaning.

Step Action

1 Select to display the System Status screen.

2 Select Details Status from the local navigation bar.

3 On the SMS Status screen, select SMS Configuration from the local navigation bar.

4 Select Staining Protocols from the local navigation bar.

5 Select the protocol to be used from the list displayed on the screen

6 Select Set Active Protocol from the local navigation bar.

7 Select OK to set the selected protocol.

Note

When an Active Stain Protocol is set-up and the system has another configuration of reagents installed, the system will display the following message: The selected protocol is not consistent with the consumable configuration of the instrument. Do you still want to set it as the active protocol? If YES is selected, the operator must verify that the stain supplies are appropriate for the selected protocol on the Slidestainer Supplies tab on the Supplies screen.

See Setting Up Stainer Reagents in HELP.

Page 73: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-37 of 62

Set Up a Quick Rinse (read only)

Follow the procedure, when needed, to reduce stain precipitate on the slide for the Wright Giemsa QR(quick rinse) default stain protocol.

Risk of personal injury or contamination. Failure to properly shield yourself while using or servicing the instrument may result in injury or contamination. To prevent possible injury or biological contamination, you must wear proper laboratory attire including gloves, and a laboratory coat, and eye protection.

You will need the following materials:

Methanol Deionized Water Empty reagent bottle with a 2L capacity, provided in the accessory

kit Funnel Beaker with a minimum volume capacity of 2L

Step Action

PREPARATION

1 Obtain an empty 2L reagent bottle and label it QR for Quick Rinse or Supply 3.

2 Mix 500 mL of methanol and 1500 mL of deionized water in a beaker.

3 Use the funnel to transfer the 2L mixture from the previous step into the QR quick rinse bottle.

Note If you have a floor stand, load the quick rinse bottle on the floor stand behind Supply 3.

4 Carefully insert the Supply 3 pickup tube assembly into the new quick rinse bottle and tighten the cap.

Note The pickup tube assembly has several tubes that may bend when you insert them into the new container. Use a slight twisting motion if you experience difficulty.

5 Access the Fill Bath option by selecting Menu>Diagnostics>DxTools>Slidestainer Tab>Fluidics.

6 Select Fill Bath 4 from the drop-down list and select Start from the local navigation bar.

Page 74: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-38 of 62

Note Drain and Fill Bath 4 one additional time to prime the system.

7 Select Stop to exit the diagnostic mode and place the instrument back online to resume operation.

Note See Set Up/Edit Other Reagents for information on how to enter the reagent information manually.

Step Action

PROCESSING

1 Ensure that the active protocol requires a quick rinse supply.

2 Prepare the quick rinse according to the instructions in this section.

3 Load the quick rinse reagent bottle on the instrument.

4 Place the instrument online. The instrument automatically primes and fills all the baths.

5 Load the sample(s) to run the active stain protocol with the quick rinse.

Step Action

REVIEW RESULTS

1 Review the slides for smear quality if this is your initial setup.

2 Refill the quick rinse bottle, as needed.

You reached the goal!

Please continue to the Skill Check on the next page.

Page 75: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-39 of 62

SETTING UP THE STAINER SKILL CHECK

1. Show the facilitator where the Stain Protocols are displayed and point out which Stain Protocol is active.

2. Show the facilitator how you would Copy a Staining Protocol.

3. Show the facilitator how you would Delete a Staining Protocol.

4. What would you do to reduce Stain Precipitate on your slides when using the Wright Giemsa Stain Protocol?

________________________________________________________

Page 76: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-40 of 62

Page 77: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-41 of 62

DISABLING/ENABLING THE STAINER & MAKER OBJECTIVES

Given an operating DxH Slidemaker Stainer System and System HELP

Disable the Slidestainer

Enable the Slidestainer

Disable the Slidemaker

Enable the Slidemaker

Why Is It Important?

Knowing how to follow the appropriate procedure to enable/disable the slidestainer and/or slidemaker is important when troubleshooting the system.

Resources

DxH Slidemaker Stainer

System HELP

Page 78: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-42 of 62

Page 79: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-43 of 62

INFORMATION/PRACTICE Disable the Stainer

Step Action

1 Select to display the System Status screen.

2 Select Details Status.

3 Select Module>Disable Stainer from the local navigation bar.

4 On the Disable Module dialog box, select the checkbox and select OK to disable the stainer.

5 Select YES when the system displays: You have requested to temporarily disable a module. Do you want to continue with the request?

Enable the Stainer

Step Action

1 Select to display the System Status screen.

2 Select Details Status.

3 Select Module>Disable Stainer from the local navigation bar.

4 On the Disable Module dialog box, deselect the check box and select OK to enable the stainer.

5

Select YES when the system displays: You have selected to enable a module. The System will automatically perform a Prime. Performance verification must be performed before patient samples can me analyzed. Do you want to Continue?

Page 80: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-44 of 62

Disable the Maker

Step Action

1 Select to display the System Status screen.

2 Select Details Status.

3 Select Module>Disable Maker from the local navigation bar.

4 On the Disable Module dialog box, select the check box and select OK to disable the maker.

5 Select YES when the system displays: You have requested to temporarily disable a module. Do you want to continue with the request?

Enable the Maker

Step Action

1 Select to display the System Status screen.

2 Select Details Status.

3 Select Module>Disable Maker from the local navigation bar.

4 On the Disable Module dialog box, deselect the check box and select OK to enable the maker.

5

Select YES when the system displays: You have selected to enable a module. The System will automatically perform a Prime. Performance verification must be performed before patient samples can me analyzed. Do you want to Continue?

You reached the goal!

Please continue to the Skill Check on the next page.

Page 81: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-45 of 62

DISABLE/ENABLE SLIDESTAINER & SLIDEMAKER SKILL CHECK

1. Show facilitator how to disable and enable the Slidestainer.

2. Show facilitator how to disable and enable the Slidemaker.

Page 82: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-46 of 62

Page 83: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-47 of 62

SAMPLE PROCESSING

OBJECTIVES Given an operating DxH Slidemaker Stainer, access to System HELP and blood specimens collected in EDTA anticoagulant:

State the aspiration volume for sampling, carryover, throughput, and known limitations

Identify the five types of cassettes (A, A+, B, C, D) and the tubes to use in each

Process bar coded samples using cassette presentation

Process bar coded samples using single-tube presentation

Process a STAT

Process a specimen in a microtainer using single-tube presentation (optional)

Adding a Test Order

Basket Management & Details

Advance Baskets

Stain Next & Cancel Staining

Why Is It Important?

Running samples is a daily task that occupies much of your time in the lab. Learning how to process samples efficiently on the DxH Slidemaker Stainer saves you time.

Resources

DxH Slidemaker Stainer

System HELP

Blood specimens collected in EDTA

Bar code labels

Cassettes

Microtainers

Page 84: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-48 of 62

Page 85: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-49 of 62

INFORMATION/PRACTICE

The DxH Slidemaker Stainer makes blood smears based on the criteria and rules set up for the sample results on the System Manager, on downloaded test orders from the host, or on the addition of a Test Order at the System Manager.

The sample aspiration probe aspirates 90 μL for up to four smears. A maximum of four smears can be obtained from a single-tube aspiration (single-tube or cassette presentation). A maximum of 12 smears can be obtained from three aspirations in the cassette presentation.

Tube dead volume varies depending on the tube’s physical characteristics (length and diameter), the tube bottom adjustments, and the sampling presentation.

Throughput

Under a steady state, the DxH Slidemaker Stainer achieved a throughput of 140 slides (unstained dried blood films) per hour using normal whole blood under the following conditions:

Consumables and waste containers were at acceptable levels Each specimen was presented as a single aspiration and a single

slide Cassettes were fully loaded

Refer to Chapter 1, System Overview, Chapter 2, Operation Principles, and Chapter 4, Sample Processing in HELP for further information on Sample Processing

Page 86: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-50 of 62

Running Samples in Cassette Presentation

(LIS Bidirectional Download to Pending)

Step Action

1

Obtain barcode labels from your facilitator. This section assumes bidirectional LIS download and positive identifier set to Specimen ID.

Affix the barcode labels supplied to the specimens provided.

2 Ensure the label is flattened smooth against the tube. Press the label down securely, including all the edges and the corners. Ensure that no part of the label is loose.

3 Ensure your specimens have been collected and stored properly as per CLSI guidelines.

4

Slide each sample firmly into the cassette ensuring that the bar codes are facing up within the cassette window.

5 Place the cassettes into the input buffer.

6 Place the instrument online. The instrument will automatically begin cycling the cassettes.

Page 87: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-51 of 62

Running Samples in Single-Tube Presentation

Obtain another bar-coded specimen. You will use the handheld bar-code scanner to scan the bar-code label on the sample, you may manually enter the Sample ID, or you can place the sample in the grooved area in front of the barcode reader.

Step Action

1 Double-click the instrument icon to display the default order for single-tube aspiration or select the DETAIL STATUS from the local navigation bar on the system status screen.

2

Select Single-Tube Presentation.

The system displays Preparing for Single Tube presentation, please wait and then the Single-Tube Station dialog box is displayed.

3

At the Single-Tube Station: Place the specimen in the grooved area in front of the barcode reader or Enter the Sample ID manually (You may use alpha or numeric characters) or Use the bar-code scanner to scan the label. Your handheld bar-code scanner is a camera that takes a picture of the bar-code.

Note: Move the label closer if the bar-code scanner does not read the information. Once the bar-code is read, you will hear a beep.

4 Press Enter.

5

Follow the instructions in the highlighted area of the dialog box. If the system recognizes the barcode as a previous run barcode, it will prompt the operator to either choose: Re-run or Reflex. Re-run does not work. Either type in a new ID or Reflex.

6

Place the tube in the correct cradle position. The left (lavender) position adapts to both open and closed vial tubes from 12 to 13 mm in diameter. The right (green) position holds microtainer tubes. The cradle retracts and the instrument processes the specimen. When the instrument is ready to process the next specimen, the system sends the cradle back out again.

7 Select the Exit button on the Single-Tube Station dialog box to exit from the Single-Tube Station when processing is complete.

Page 88: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-52 of 62

Adding a Test Order (Menu>Worklist)

Step Action

1 Select or Select Menu>Worklist.

2

Select the Slides tab, Slides Pending for the Filter Name, and Add Order button from the local navigation bar or Select the Slides tab and select the Add Order button from the local navigation bar.

3

On the Add Order screen, enter the specimen ID manually or by using the bar-code scanner.

4 Select the desired radio buttons for Make, or Make and Stain.

5 Select the number of slides to be made, and/or made and stain using +/- buttons or editing the default value of 1.

6 If desired, select the Patient button and Add Patient from the local navigation bar to add patient demographics.

7 Enter patient information and select OK.

8 Select the Submit button from the local navigation bar.

Page 89: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-53 of 62

View Basket Details

Step Action

1 Select to display the System Status screen.

2 Select the Details Status from the local navigation bar.

3

Select Basket Management from the local navigation bar to view the Basket Management screen. On this screen, you can manage baskets or get basket information anywhere in the Stainer module and the Basket Transport Module.

4 Click on a basket that has slides which have been made by the Slidemaker.

5

Select Basket Details from the local navigation bar to display the Basket Details dialog box containing specific patient information for the slides in the selected location.

6 Select OK when you are done.

Page 90: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-54 of 62

Advance Baskets

Follow these steps to advance a selected basket to the I/O drawer.

Step Action

1 Select to display the System Status screen.

2 Select the Details Status from the local navigation bar on the DxH Slidemaker Stainer Status screen or double click on the instrument icon.

3

Select Basket Management from the local navigation bar to view the Basket Management screen.

4

Select the basket to advance.

Note: The operator may Advance a basket from the elevator position or from the Stain Dryer positions.

5

Select Advance Basket from the local navigation bar to display the Basket Details dialog box.

Note: The baskets/slides may be wet if you select Advance Basket before the pre-stain drying cycle is complete. The baskets will be transferred to the I/O drawer.

6 On the DxH Solutions dialog box, select OK to forward the basket to the I/O drawer.

Page 91: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-55 of 62

7 Select OK to save your changes.

Select Basket to Stain Next

Follow these steps to schedule the next basket with slides for staining.

Step Action

1 Select to display the System Status screen.

2 Select the Details Status from the local navigation bar.

3

Select Basket Management from the local navigation bar to view the Basket Management screen.

4 Select the basket to stain next.

5 Select Stain Next from the local navigation bar to display the Stain Next dialog box.

Page 92: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-56 of 62

6 Select OK to save you changes.

Cancel Staining

This command cancels all staining in progress and all slides waiting to be stained. Slides that are in the stainer will advance to the dryer and then to the I/O drawer. Slides in the basket elevator will advance to the I/O drawer.

Step Action

1 Select to display the System Status screen.

2 Select the Details Status from the local navigation bar.

3

Select Basket Management from the local navigation bar to view the Basket Management screen.

4 Select Cancel Staining from the local navigation bar to display the Cancel Staining dialog box.

Page 93: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-57 of 62

5 Select OK to save your changes.

Note If cancel staining is selected, the stainer Status will become inactive. Select System Status>Details Status>Start to make the Stainer Active.

Loading or Removing Baskets at the Input/Output (I/O) Drawer

Risk of personal injury and contamination. The processing areas may contain prepared smears and/or broken glass with biohazardous material. Handle with care and avoid skin contact.

Risk of personal injury and contamination. The instrument could continue to move or could have delayed movement. Wait a few seconds to ensure that all movement has ended before pulling the I/O drawer.

Step Action

1

Press the I/O drawer button button to access the input/output (I/O) drawer.

Page 94: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-58 of 62

2

Once the drawer has stopped moving (it may take a few seconds), pull the handle towards you until it stops.

3

Remove and/or load the baskets. Ensure that the baskets are thoroughly dry prior to loading. Note: Ensure that two slots are empty or available in the I/O drawer at all times.

The indicator lights are: Blinking green – when the basket is ready for you to

remove it Steady green – when the basket is ready to be used by

the system Blinking red – when the basket is in an error state Blinking blue – when a basket is loaded with slides for

staining only

4

Push the handle toward the instrument until it stops. The drawer will automatically return to its position inside the instrument.

Page 95: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-59 of 62

Staining Manually Prepared Slides (Stain-Only Priority)

Risk of personal injury and contamination. The processing areas could contain prepared smears and/or broken glass with biohazardous material. Handle with care and avoid skin contact.

Follow these steps to manually stain a sample.

Step Action

1

Fill an empty basket with slides for a Stain Only cycle as follows:

• Load the slides with the painted end on top to prevent the staining of sample information.

• Load all smears in the same direction facing the back of the next smear.

2

Open the I/O drawer and place the basket back in the I/O drawer in any open/empty position from 1 to 6.

Page 96: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-60 of 62

3

Press the number where the basket is placed.

4

Push the drawer in for staining to begin.

Making Slides (No Staining)

Risk of injury and biohazardous contamination. Clean up any broken glass or blood spill as quickly as possible. Handle with care. Avoid skin puncture. Dispose of all contaminated disposable cleaning materials and broken slides in accordance with your local regulations and acceptable laboratory practices.

Risk of contamination. The prepared smears contain biohazardous materials. Handle with care and avoid skin contact.

Follow these steps to make slides that do not require staining:

Step Action

1 From the Status Screen, select Make Slide on the Default Test Order for the applicable cassette or single-tube presentation or Add or edit a test from the Worklist Pending tab or Load samples for the processing of downloaded Test Orders from the LIS.

Please continue to the Skill Check.

Page 97: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

S0-61 of 62

SAMPLE PROCESSING SKILL CHECK

1. What is the sample aspiration volume for the DxH Slidemaker to make up to four smears? _________ Aspiration Volume to make twelve smears? _________

2. Show stained slides of the samples run to your facilitator.

3. Navigate the Worklist screen tabs with your facilitator.

Page 98: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

SO-62 of 62

Page 99: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-1 of 48

COURSE MAP – TROUBLESHOOTING

Removing System Covers

Components and Sample Flow

Cleaning Procedures

Replacement Procedures

Optimize Stain Protocols

BEGIN HERE

Using Diagnostics

Page 100: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-2 of 48

PERSONAL PROGRESS SUMMARY

Troubleshooting Modules Date Initial Facilitator

Removing System Covers

Components and Sample Flow

Cleaning Procedures

Replacement Procedures

Using Diagnostics

Optimize Stain Protocols

Page 101: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-3 of 48

REMOVING SYSTEM COVERS

OBJECTIVES Using an operating DxH Slidemaker Stainer System and System HELP:

Locate and Lift/Lower the Front Cover

Locate and Remove/Install the Transport Shield

Locate and Remove/Install the Slide Printer Cover

Locate and Remove/Install the Upper Right-Side Cover

Locate and Remove/Install the Stainer Shield

Why Is It Important?

Knowing how to remove and install the system covers is important when you are troubleshooting. It will help you get to the source of the problem when something is not working. Using the System Help as a primary resource can help you solve problems.

Resources

DxH Slidemaker Stainer System

System HELP

Screwdriver

Page 102: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-4 of 48

Page 103: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-5 of 48

SETUP Be sure to wear full-face protection (either a full-face shield OR

protective eyewear with a face mask) along with a lab coat and gloves when running a sample with covers open.

Have the facilitator help you to remove the covers.

Note: this exercise is for training purposes only. You would not normally open the covers of the Slidemaker Stainer, and never when running samples.

INFORMATION/PRACTICE

Step Action

Removing and Installing the Transport Shield

Note

Risk of personal injury. To avoid personal injury, do not bypass any of the safety interlocks and do not open covers while the instrument is operational.

For some cleaning procedures, you need to remove the transport shield.

1

Grasp the transport shield on both sides and pull it away from the instrument.

2 Silence any alarms.

3 To reinstall transport shield, locate the pins on the shield and their corresponding rails on the front of the instrument.

Page 104: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-6 of 48

4

Insert the pins into the rails and push the transport shield into place.

Step Action

Lifting and Lowering the Front Cover

1

Grasp the transport shield on both sides and pull it away from the instrument Grasp the sides of the front cover and lift upward.

2

Grasp the sides of the front cover and push downward. Reinstall the transport shield.

Page 105: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-7 of 48

Step Action

Removing and Installing the Slide Printer Cover

Note

Risk of damage to dispense probe. Ensure that the SAM is moved all the way to the left before pulling the Slidemaker.

1 Go to Menu>Diagnostics>Dx Tools. Click on the Release SAM button from the Local Navigation bar.

2 Remove the Transport Shield.

3 Lift the Front Cover.

4 Manually move the SAM to the left.

5

Pull the Slidemaker release handle.

Note: Releasing the SAM and pulling SM out is not required to access the printer cover unless there is no access like in a connectivity or if instrument up against a corner with no access.

6

Pull the Slidemaker out to the maintenance position.

Page 106: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-8 of 48

7

Pull the release pin down to unlock the Slidemaker.

Note

Risk of hand injury. Use caution when pushing the Slidemaker back into position.

8 Push the Slidemaker back until it locks into operating position.

9

Unfasten the captive screw at the rear of the slide printer cover.

10

Pull the cover toward the front of the instrument to disengage the cover hangers.

Page 107: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-9 of 48

11

To reinstall the Slide Printer Cover, position the cover to align the cover hanger bracket with the three cover hangers.

12

Slide the cover toward the rear of the instrument to engage the cover hanger bracket with the cover hangers.

13

Fasten the captive screw at the rear of the cover.

14 Lower the Front Cover and install the Transport Shield.

Page 108: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-10 of 48

Step Action

Removing and Installing the Upper Right Side Cover

1 Remove the Transport Shield.

2 Lift the Front Cover.

3 Remove the Slide Printer Cover.

4 Grasp the front and back of the cover.

5 Slide the cover forward and then up and off the instrument.

6

To reinstall the Upper Right-Side Cover, grasp the front and back of the cover.

7 Lower the cover to engage the top slides in their mounting brackets.

IMPORTANT: To engage the round slide on the top and bottom of the cover in its cutout in the frame, hold the cover almost vertical during installation. It is critical to align all four sides at the same time.

Page 109: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-11 of 48

8

Align the two lower slides with their cutout in the frame.

9

Slide cover toward rear of instrument to engage all four slides.

10 Lower the Front Cover and install Transport Shield.

Step Action

Removing and Installing the Stainer Shield

1

Go to Menu>Diagnostics>Dx Tools. Click on the Release SAM button at the bottom of the screen.

Manually move the SAM to the left.

2 Remove the Transport Shield.

3 Lift the Front Cover.

4 Pull the stainer shield forward to disengage it from the alignment pins and magnets.

Page 110: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-12 of 48

5

Slide the shield down and out.

6 To reinstall the Stainer Shield, support the e-chain out of the way.

7

Slide the shield up and into place behind the e-chain.

Page 111: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-13 of 48

8

Attach the shield in line with the alignment pins and magnets.

9 Lower the Front Cover.

10 Install the Transport Shield.

You reached the goal!

Please continue to the Skill Check on the next page.

Page 112: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-14 of 48

Page 113: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-15 of 48

REMOVING SYSTEM COVERS SKILL CHECK

Show facilitator that all the covers have been put back and no alarms or errors are being triggered.

Page 114: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-16 of 48

Page 115: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-17 of 48

COMPONENTS AND SAMPLE FLOW

OBJECTIVES Using an operating DxH Slidemaker Stainer and System HELP:

Locate and name major modules of the DxH Slidemaker Stainer

Locate and name the components of the Specimen Transport Module

Locate and name the components of the Sample Aspiration Module

Locate and name the components of the Slidemaker Module

Locate and name the components of the Stainer Module

Locate the Bar Code Readers

List the steps the Slidemaker Stainer performs to process a stained slide

Why Is It Important?

Knowing module and component names and their locations as well as their functions is important when you are troubleshooting. Recognizing and knowing what components should be doing in a cycle helps you troubleshoot when something is not working. Using the System Help as a primary resource can help you solve problems.

Resources

DxH Slidemaker Stainer System

System HELP

Page 116: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-18 of 48

Page 117: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-19 of 48

SETUP Be sure to wear full-face protection (either a full-face shield OR

protective eyewear with a face mask) along with a lab coat and gloves when running a sample with covers open.

Have the facilitator help you to remove the transport shield and raise the front cover.

Note: This exercise is for training purposes only. You would not normally open the covers of the Slidemaker Stainer, and never when running samples.

INFORMATION/PRACTICE

Step Action

1 Open System Help .

2 Expand Chapter 1, read the information in this Chapter to familiarize yourself with the System Overview.

4 Expand Chapter 2 Operation Principles.

5 Read the information in this section to learn the sample flow from Request through Basket Transfer.

Note You will label system components at the end of this module as part of the Skill Check.

Page 118: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-20 of 48

DxH Slidemaker Stainer System

DxH Slidemaker Stainer

System Manager (SM)

Handheld Bar Code Scanner

Specimen Transport Module (STM)

Input Buffer Pushers

Input Buffer

STAT Cassette Position

Single-Tube Station

Bar Code Reading Station

Mix Stations

Output Buffer Pusher

Output Buffer

Sample Aspiration Module (SAM)

Aspiration Syringe (Cassette and Single-tube stations)

Hemasphere - examines the blood left behind in the tubing to determine characteristics for speed and acceleration used in the smear preparation.

Blood Detector - distinguishes between blood, air, and diluent.

Aspiration Probe – samples from both closed and open specimen tubes.

Dispense Probe - carries sample used for the smearing process.

Slidemaker Module

Ejector with Slide Chute

Smear Truck

Smear Shuttle

Print Shuttle

Slide Printer

Basket Elevators

Heat Dryer

Fan Dryer

Page 119: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-21 of 48

Stainer Module and Basket Transport

Input/Output (I/O) Drawer

Stain Drawer

Baths (within Stain Drawer)

Drying Stations

Robot Arm

Bar Code Scanners

The Slidemaker Stainer has 3 bar code readers. It has a bar-code reader on the single-tube station, the SAM (Specimen Aspiration Module) bar code reader, and the STM (Specimen Transport Module) bar code reader.

These barcodes have the following functions:

Provide for positive identification of all specimen tube IDs in the form of an image.

Scan the cassette and specimen ID twice prior to aspiration.

Support symbologies as small as 5 millimeters and include Code 128, Codabar, NW7, Code 39 and Interleaved 2 of 5.

Scan a maximum number of 22 characters plus checksum or whatever can be printed in the viewable height, whichever is less.

Sample Flow (Chapter 2>Operation Principles)

Cassette Presentation

Samples in cassettes are loaded in the input buffer. Each specimen is identified by a bar code scanner. Cassettes are transferred to one of the two mixing stations. The aspirated sample is transported through the hemasphere, the blood detector, and finally into the dispense probe. Twenty-five ul of the sample is discarded. The probe moves to the drop placement position. Approximately 4 uL drop of blood is placed on the slide located on the smear shuttle. A second slide is picked up by the smear truck and used to spread the blood drop on the first slide using the wedge technique. The prepared slide is transferred from the smear shuttle into the print shuttle, and then into the basket elevator for drying.

Single-Tube Presentation

The specimen is manually mixed, identified, and placed in the single-tube station. The specimen is aspirated and follows the same path through the hemasphere, and so forth as described above in the Cassette Presentation.

Page 120: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-22 of 48

Review Chapter 2 >Operation Principles:

Slide Request:

Cassette Presentation

Single-Tube Presentation

Sample Aspiration Volume

Initial Slide Transport

Smear Preparation

Slide Labeling

Smear Drying

Basket Transfer

Stain Cycle

Method History:

Slidemaking

Staining

You reached the goal!

Please continue to the Skill Check on the next page.

Page 121: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-23 of 48

.

COMPONENTS AND SAMPLE FLOW SKILL CHECK

Use Job Aid to complete the Skill Check.

Get numbered tabs from the facilitator and place on the components listed below.

1. Bar Code Reading Stations 10. Smear Truck

2. Input Buffer 11. Smear Shuttle

3. Mix Stations 12. Print Shuttle

4. Aspiration Probe 13. Slide Printer

5. Single Tube Presentation 14. Basket Elevators

6. Hemasphere 15. Stain Drawer

7. Blood Detector 16. Baths

8. Dispense Probe 17. Fill Probe

9. Slide Chute 18. I/O Drawer

SAMPLE FLOW

1. State the sample aspiration volume _______________.

2. State the maximum number of smears from each aspiration for:

Single-tube presentation _________

Cassette presentation ___________

3. How does the dispense probe prepare for a clean drop of blood?

_____________________________________________________

_____________________________________________________

Score: ____ / 18

Page 122: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-24 of 48

Page 123: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-25 of 48

CLEANING PROCEDURES

OBJECTIVES Use the System HELP

Clean the Aspiration Probe

Clean the Dispense Probe

Clean the STM

Clean the STM Optical Sensors

Clean the STM Bar Code Scanner

Clean the Vacuum Trap

Clean the Slide Chute

Empty the Broken Slide Bin

Clean the Baskets

Flushing Reagent Lines and Stainer with Methanol

Cleaning the Stainer Baths and Tray

Clean the Stainer Fill Probes, Drain Probes, and Level Sense Probes

Extensive Cleaning of Fill Probes and Drain Probes (read only)

Why Is It Important?

Knowing how to clean certain components on an as-needed basis will keep your system running smoothly.

Resources

DxH Slidemaker Stainer System

System HELP

Lint-free tissue/cloth and lint-free swabs

Screwdrivers and Pliers

Page 124: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-26 of 48

Page 125: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-27 of 48

INFORMATION/PRACTICE

Reminder: Please wear lab coats, gloves and full-face protection for these procedures.

Step Action

1 Open System Help .

2 Expand Chapter 9 Cleaning Procedures. Read the information in this Chapter to familiarize yourself with the Cleaning Procedures.

3 Follow the procedures for:

Cleaning the Aspiration Probe

Cleaning the Dispense Probe

Cleaning the STM

Cleaning the STM Optical Sensors

Clean the STM Bar Code Scanner

Cleaning the Vacuum Trap

Cleaning the Slide Chute

Emptying the Broken Slides Bin

Cleaning the Baskets

Flushing Reagent Lines and Stainer with Methanol

Clean Stain Baths and Tray

Clean Stainer Fill Probes, Drain Probes, and Level Sense Probes

Extensive Cleaning of Fill Probes and Drain Probes (read only)

Page 126: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-28 of 48

You reached the goal!

Please continue to the Skill Check on the next page.

Page 127: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-29 of 48

CLEANING PROCEDURES SKILL CHECK

Have facilitator verify all cleaning procedures done properly:

Cleaning the Aspiration Probe

Cleaning the Dispense Probe

Cleaning the STM

Cleaning the STM Optical Sensors

Cleaning the STM Bar Code Scanner

Cleaning the Vacuum Trap

Cleaning the Slide Chute

Emptying the Broken Slides Bin

Cleaning the Baskets

Flushing the Reagent Lines and Stainer with Methanol

Clean Stain Baths and Tray

Clean Stainer Fill Probes, and Level Sense Probes

Page 128: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-30 of 48

Page 129: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-31 of 48

REPLACEMENT PROCEDURES

OBJECTIVES Using the System HELP:

Replace a Bath

Replacing the Fill and Drain Probes

Replace the Ribbon

Replace the Aspiration Probe (read only)

Replace the Dispense Probe (read only)

Replacing or Refilling the Reagent Containers (read only)

Replacing the Biohazard Waste Container (read only)

Replacing the Stain Waste Container (read only)

Replacing the Stainer Reagent Line Filters (read only)

Why Is It Important?

Knowing how to replace some basic components will minimize instrument down time and require fewer service calls.

Resources

DxH Slidemaker Stainer System

System HELP

Pliers

Screwdrivers

Page 130: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-32 of 48

Page 131: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-33 of 48

INFORMATION/PRACTICE Reminder: Please wear lab coats, gloves and full-face protection for these procedures.

Step Action

1 Open System HELP .

2 Expand Chapter 10 Replacement/Adjustment Procedures. Read the information in this Chapter to familiarize yourself with the Replacement Procedures.

Note It is important to follow all instructions in Help.

Note When in Diagnostics, do not remove the shield and raise cover until directed to do so. You must select START to begin a procedure and FINISH to end.

3

Follow the replacement procedures for:

Bath

Fill and Drain Probes

Printer Ribbon

Remove and replace the same part.

4

Read only: Replace the Aspiration Probe, Replace the Dispense Probe, Replacing or Refilling the Reagent Containers, Replacing the Biohazard Waste Container and Stain Waste Container, Replacing the Stainer Reagent Line Filters.

Do not remove the shield or raise cover on the instrument until told to do so in Diagnostic Tools.

You reached the goal!

Please continue to the Skill Check on the next page.

Page 132: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-34 of 48

Page 133: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-35 of 48

REPLACEMENT PROCEDURES SKILL CHECK

Have facilitator verify all replacements:

Bath

Fill and Drain Probes

Printer Ribbon

Page 134: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-36 of 48

Page 135: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-37 of 48

USING DIAGNOSTICS

OBJECTIVES Use the Dx Tools and System HELP to perform the following troubleshooting procedures:

Access the System Monitor

Verifying the Blood Detector

Retrieving a Lost Basket (read only)

Event Messages (read only)

Why Is It Important?

Using the Diagnostics available in the System will help you in any troubleshooting efforts, especially when working with BCI Representatives.

Resources

DxH Slidemaker Stainer System

System HELP

Page 136: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-38 of 48

Page 137: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-39 of 48

INFORMATION/PRACTICE

When you troubleshoot problems that may occur, the Diagnostic features available to you through the software will help you to solve the problem and resume operation. In your lab, you may do these procedures at the direction of a BCI representative.

IMPORTANT: The DxH Slidemaker Stainer must be offline to run diagnostic procedures. If the system is online, the following message is displayed:

Putting the instrument offline will stop processing specimens and any cassette in progress will be routed to the output buffer. Do you still want to continue with the request?

Select OK to continue or select Cancel to cancel the procedure and remain online.

When you are finished running diagnostics and want to return to online processes, select Finish from the local navigation bar on the Diagnostics screen.

Step Action

1 Open System HELP .

2 Expand Chapter 8 Troubleshooting. Read the information in this Chapter to familiarize yourself with the General Troubleshooting-Diagnostic Procedures.

3 Under Individual Troubleshooting Procedures, perform the following:

System Monitor

Verifying the Blood Detector

Retrieving Lost Basket (read only)

4 Review Chapter 8>Event Messages. There are suggested actions for error events that you may be able to perform.

Page 138: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-40 of 48

You reached the goal!

Please continue to the Skill Check on the next page.

Page 139: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-41 of 48

USING DIAGNOSTICS SKILL CHECK Have facilitator verify all diagnostic procedures:

System Monitor

Verifying the Blood Detector

Page 140: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-42 of 48

Page 141: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-43 of 48

OPTIMIZE STAIN PROTOCOL

OBJECTIVES Given an operating DxH Slidemaker Stainer and access to System HELP

Perform a Stain Quality Adjustment

Why Is It Important?

Having stained slides, which meet each individual laboratory colorization is very important. Colorization is subjective and optimizing the stain protocol will satisfy each institution’s expectation.

Resources

DxH Slidemaker Stainer

System HELP

Blood specimens collected in EDTA

Bar code labels

Cassettes

Page 142: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-44 of 48

Page 143: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-45 of 48

INFORMATION/PRACTICE

Overall Stain Quality

Beckman Coulter has developed COULTER TruColor reagents and default protocols to be used on the UniCel DxH Slidemaker Stainer. The Beckman Coulter default protocols have been optimized to a predefined colorization specification.

Since colorization is subjective, optimization of the protocols and/or reagents may be necessary to meet your individual colorization criteria. Ensure that both the methanol and water you use meet the specifications detailed in your system’s online help.

Typical time adjustment increments:

Stain – 30 seconds

Buffer – 1 minute

Coloration (red/blue) is primarily controlled by the pH of the buffer.

Stain intensity and differential staining characteristics are controlled by:

The length of staining times (longer = more intense)

Decreasing the buffer strength

Concentration of stain in buffer (more stain = darker, more intense)

Step Action

1 Open APPENDIX B.

2 Read all the information in that section.

3 Select Default Wright Stain Protocol and copy it.

4 Rename it to STAINCHANGE.

5 Change the settings in Bath 2 to 1 minute.

6 Place some barcoded bloods in a cassette and process the cassettes.

7 Keep the stained slides to show facilitator.

Page 144: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-46 of 48

You reached the goal!

Please continue to the Skill Check on the next page.

Page 145: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

TB-47 of 48

OPTIMIZE STAIN PROTOCOL SKILL CHECK

1. Show facilitator the changed protocol you created.

2. Coloration on a slide is primarily controlled by the ______

of the __________.

3. What are the 2 Default Stain Protocols in the Slidemaker Stainer?

____________________________________

____________________________________

4. Show facilitator the stained slides.

5. What was the probable cause of why these slides look the way they do and what would be your recommended action?

________________________________________________________

________________________________________________________

Page 146: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules Beckman Coulter Professional Development PN: A96923AA.2 (June 2014)

TB-48 of 48

Page 147: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-1 of 44

COURSE MAP – ADDITIONAL TOPICS

These modules are elective. You may perform them in any order desired. If a topic is not applicable to your lab situation, you do not need to do it at all.

History Logs

Operators & Roles

Configure Auto Stop Settings

Enabling Batching

Smear Quality & Stain Quality

Mentor Trainer

Page 148: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-2 of 44

PERSONAL PROGRESS SUMMARY

Additional Topics Modules Date Initial Facilitator

History Logs

Operators & Roles

Configure Auto Stop Settings

Enabling Batching

Smear Quality & Stain Quality

Mentor Trainer

Page 149: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-3 of 44

HISTORY LOGS

OBJECTIVES Given an operating DxH Slidemaker Stainer

Locate the History Logs (Event) Alert Status Icon

State the four types of History Logs

Show how to review (clear) a red or yellow icon

Find the Details of an event

Add a comment to an event

Print a log

Export a log

Why Is It Important?

This module introduces History (Event) Logs. These logs record every event that occurs on the system, divided into separate categories based on the type of error. You must review any event that causes a red icon so that the system may continue running.

Resources

DxH Slidemaker Stainer System

Storage media for exporting logs

Page 150: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-4 of 44

Page 151: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-5 of 44

INFORMATION/PRACTICE The History (Event) logs have a variety of types. You access them by selecting the Event Status Icon . Learn more about this area by using the System Help.

Step Action

1 Select at the top of any screen to display the History Logs screen.

2

Select a log name from the Log Type drop-down list to view that log.

Review

There are four log types: Event Logs, Data Summary Logs, Audit Log and Maintenance Logs.

Whenever the Event Status Icon is yellow or red, select it to review the message. This always takes you to the General tab. This log records all entries for System problems/warnings. You must review any

entry with a or . The signifies information only. Selecting the Review button causes the Event Status Icon to return to a neutral color. Select the Details Button to see more information about the entry. Just as in the Daily Checks area, you may want to add a comment to the entry to explain why you reviewed and accepted the message.

The Patients tab records information about individual specimens, e.g. a no read of a barcode.

The Supplies tab records information associated with the reagents, when a supply is depleted or switched.

The System Status tab records information on reports and LIS activity.

Page 152: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-6 of 44

The Daily Checks tab records the details of each Daily Checks test run in the past.

The Supplies tab records the details of each reagent container change.

The Audit tab records information such as when a configuration was changed.

The Maintenance tab records status messages and allows the lab to add New Entries to record cleaning or other procedures performed. It can be thought of as an Action Log. Try adding an entry now!

The Service tab records any diagnostic procedures performed. It can only be accessed and commented on by someone with a Service access level. The lab may view and print this log.

Details

Most of the logs give you access to the Details button to find out more about the message posted. Highlight any entry in a log and select the Details button.

Comments

Any log allows you to attach a comment. Highlight an entry and select the Comment button. You may enter any text you wish. You may also amend, add to or delete previous comments. Whoever is logged in at the time is recorded as the comment author along with the date and time.

Add a comment now! You may also automatically add this to the Maintenance Log by selecting the checkbox “Log Comment/Event in Maintenance Log”.

Delete

Some logs allow a Delete. We do not recommend deleting any entries in the logs.

Page 153: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-7 of 44

Print

You may print logs manually or automatically.

Manual

Open a log, select a couple of entries and select the manual print button from the utility area. A dialog box opens giving you options to print the entire log, by filters or selected entries. You may optionally choose Details as well.

Choose Selected Entries and OK.

Auto Print

If you want to print all or certain logs on a regular basis, choose the Auto Print button.

This allows you to schedule printing of each category of the logs, weekly or monthly, at a certain day and time, with or without details.

Export

Selecting the Export button gives you the option of exporting the logs you choose to a CD or flash drive.

If you choose to save to a flash drive, you must first save temporarily to E:\, then drag and drop to your flash drive. You may then want to delete that temporary file.

The logs exported are converted to a .csv format that you can read on any computer that has a spreadsheet program on it (not on the System Manager).

You may perform this now if you think you may use this method in your lab.

Step Action

1 Click on the Event Alert Status icon.

2 Select one of the logs and select a message.

3 Click on the message and find the details pertaining to it.

4 Select another message and add a comment to the message.

Page 154: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-8 of 44

You reached the goal!

Please continue to the Skill Check on the next page.

Page 155: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-9 of 44

HISTORY LOGS SKILL CHECK

1. Point to the Event Alert Status Icon.

2. State the four types of History Logs. ________________________________________________________________________________________________________________________________________________________________________________________________________________________________

3. Point to the Review button, used to clear a red or yellow Event icon.

4. Highlight an entry in a log and display the Details of an event.

5. Add a comment to an event.

6. Export a log using a CD or flash drive and show the file to the facilitator.

Page 156: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-10 of 44

Page 157: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-11 of 44

OPERATORS AND ROLES

OBJECTIVES Given an operating DxH Slidemaker Stainer and access to System HELP:

Create new operator logins for a Level I, Level II and Level III access

Edit an existing operator account

Reset an operator password

Change your own password

Set a Password Expiration and enable Auto – Lockout (optional)

Why Is It Important?

As a DxH Slidemaker Stainer operator with Level III (highest customer) privileges, you will help decide the levels of security access for those working with you, based on their laboratory responsibilities.

Resources

DxH Slidemaker Stainer System

Page 158: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-12 of 44

Page 159: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-13 of 44

INFORMATION/PRACTICE

During the classroom exercises, you are at a Level III access. These next steps will show you how to manage and create the other two customer levels. For more information, see System HELP Appendix D, Operator Access Chapter 6.

Step Action

1 Select Menu>Setup>Operators and Roles.

2 On the Operators and Roles screen, view the list of various operators who are set up to use the system.

Note that the Level III access (aka Lab Administrator) is the only one of the levels that can set up or edit other operator levels.

TIP Ultimately, the Operator ID and Password will be the information entered on the login screen. Remember that BOTH are case-sensitive.

TIP The initial password entered at this point is considered a “temporary” password. Simply use the word temp for now. Write down the information you enter for each level below:

3 Create a new Level III login for yourself as well as a Level I and a Level II login. You may be creative with your entries!

4

On the Operators and Roles Screen, Select New Operator from the local navigation bar. On the New Operator dialog box, enter the operator information and password, select an access level using the drop-down list and select an account status. Select OK to confirm you selection or cancel to go back to previous screen or select HELP for more information.

L I

First Name _____________ Last Name _____________ Operator ID ____________ Password __temp________ Confirm Password _temp__

L II

First Name _____________ Last Name _____________ Operator ID ____________ Password __temp________ Confirm Password _temp__

Page 160: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-14 of 44

L III

First Name _____________ Last Name _____________ Operator ID ____________ Password __temp________ Confirm Password _temp__

5 Once you complete the steps above, select Log off .

Log on at Your New Levels

Step Action

1 At the Log on screen, type in your new Level I Operator ID and the Password temp.

2 A message appears telling you that the password is a temporary one and to enter a new password. Select OK.

3 Type in temp for the old password.

4 Type in the new password you want to use. You will need to do this for Password and Confirm Password.

5 Record your password for Level I here ___________________.

6 Explore the different screen areas noting the differences (for example, the areas to which you no longer have access).

7 Once you complete steps 1-6 above, select Log off .

8 Now repeat steps 1-7 for the Level II log on.

9 Record your password for Level II here ___________________.

10 Now repeat steps 1-7 for the Level III log on.

11 Record your password for Level III here ___________________.

Page 161: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-15 of 44

Editing an Existing Operator

You must be logged in at Level III access to perform this exercise. One reason for needing to perform this editing is to change an existing operator’s access, e.g. from Level I to Level II.

Step Action

Follow the steps and choose the Level II account you previously set up. Change it to Level 1 and make it Inactive.

Note that you may also edit the first and last name.

1 Select Menu>Setup>Operator and Roles.

2 On the Operators and Roles screen, select Edit Operator from the local navigation bar.

3 On the Edit Operator dialog box, change the operator information, select an access level using the drop-down list, and select an account status.

4 Select OK to save the changes.

Resetting an Operator’s Password

The Level III Operator may reset the password of another operator. This may be necessary if operators forget their own password.

Step Action

Follow the steps and select the Level I Operator ID you previously set up.

1 Select Menu>Setup>Operator and Roles.

2 On the Operators and Roles screen select Reset Password from the local navigation bar.

3 Change the password. _______ Select OK to save the changes.

4 Log off and Log on to try it out.

Page 162: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-16 of 44

Changing Your Password

Any operator may change his or her own password at any time. Select the Change Password button from the Operators and Roles screen.

Try this now and change back to your original password for Level I.

Timeout Settings - Setting the Password Expiration and Enabling Auto-Lockout

You must be at Level III access to do this exercise.

The Password Expiration defaults to 60 days. You may change this to 30 days or 90 days or Disable it.

Note: The Auto-Lockout feature is initially disabled. You may enable it and set the number of minutes between 5 and 60. This is the amount of time of non-use that requires the user to log in again.

Operator Roles Access

In System HELP, go to Appendix D and expand Operator Access.

There are ten tables that list the various operator functions by Level I, II and III.

Review this for your information.

You reached the goal!

Please continue to the Skill Check on the next page.

Page 163: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-17 of 44

OPERATORS AND ROLES SKILL CHECK

1. Show facilitator the three access levels you created.

2. Demonstrate editing an account.

3. Demonstrate resetting a password.

4. Demonstrate changing your own password.

5. Demonstrate the timeout features.

Page 164: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-18 of 44

Page 165: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-19 of 44

CONFIGURE AUTO STOP SETTINGS

OBJECTIVES Using an operating DxH Slidemaker Stainer and System HELP:

Demonstrate how to set up Auto Stop Settings

Why Is It Important?

You can use the Auto Stop Settings when you want the instrument to stop operations when a criteria is triggered per the Auto Stop Configuration.

Resources

DxH Slidemaker Stainer

System HELP

Page 166: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-20 of 44

Page 167: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-21 of 44

INFORMATION/PRACTICE

Configure Auto Stop Settings

Step Action

1 Select System Status>More.

2 Select Auto Stop from the local navigation bar.

3

On the Configure Auto Stop Settings dialog box, select from the following options:

4 Select OK to confirm your selection.

You reached the goal!

Please continue to the Skill Check on the next page.

Page 168: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-22 of 44

Page 169: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-23 of 44

CONFIGURE AUTO STOP SKILL CHECK

Show the facilitator the where you set-up the Auto Stop.

Page 170: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-24 of 44

Page 171: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-25 of 44

ENABLING BATCHING

OBJECTIVES Using an operating DxH Slidemaker Stainer and System HELP:

Enable Batching

Why Is It Important?

You can use batching when you want to analyze all your specimens with the same default test panel (for example, when your LIS is down).

Resources

DxH Slidemaker Stainer

System HELP

Blood specimens collected in EDTA

Cassettes

Page 172: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-26 of 44

Page 173: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-27 of 44

INFORMATION/PRACTICE

Enable Batching

Step Action

Note The system must be offline and no staining should be in progress in order to perform this procedure.

1 Select to display the System Status screen.

2 Select the Details Status from the local navigation bar.

3 Select the Batching from the local navigation bar.

4

Select the checkbox to enable Batching.

Note All results are labeled as Default Slide Order.

5 Select OK.

You reached the goal!

Please continue to the Skill Check on the next page.

Page 174: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-28 of 44

Page 175: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-29 of 44

ENABLING BATCHING SKILL CHECK

Show your facilitator the screen where you Enable Batching.

Page 176: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-30 of 44

Page 177: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-31 of 44

SMEAR QUALITY AND STAIN QUALITY

OBJECTIVES Use the System HELP > Appendix B and Chapter 4

Overall Stain Quality

Stain Characteristics

Smear Quality

Why Is It Important?

A well-made stained blood smear will yield interesting and significant information when reviewing results. The time taken in setting up these specifications will be rewarded later when the smears are scanned.

Resources

DxH Slidemaker Stainer System

System HELP

Page 178: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-32 of 44

Page 179: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-33 of 44

INFORMATION/PRACTICE When you troubleshoot problems that may occur the Diagnostic features available to you through the software will help you to solve the problem and resume operation. In your lab, you may do these procedures at the direction of a BCI representative.

Step Action

1 Open System HELP .

2 Expand Chapter 4 Sample Processing.

3 Read the section on Checking Smear Quality to familiarize yourself on Slide Quality and the Characteristics of a Good Smear.

4 Also, select Appendix B to expand

5 Review the section Overall Stain Quality.

6 Review Stain Characteristics.

You reached the goal!

Please continue to the Skill Check on the next page.

Page 180: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-34 of 44

Page 181: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-35 of 44

SMEAR & STAIN QUALITY SKILL CHECK

1. What are probable causes of a stained smear that is Too Light?

________________________________________________________________________________________________________________________________________________________________________

2. What actions would you take if you have a smear with precipitate?

________________________________________________________________________________________________________________________________________________________________________

3. What are probable causes of a stained smear that is Too Blue?

________________________________________________________________________________________________________________________________________________________________________

Page 182: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-36 of 44

Page 183: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-37 of 44

MENTOR / TRAINER

OBJECTIVES Given an operating UniCel DxH Slidemaker Stainer Coulter Analysis System, training modules, and appropriate supplies for administering modules:

Format a training plan for your laboratory

Verify skill performance for each trainee

Have a course Facilitator and/or Applications Specialist verify your performance as a mentor/trainer

Why Is It Important?

In order to have competent Slidemaker Stainer operators in your laboratory, it is important that they be grounded in the basics of instrument operation. As the mentor/trainer for your laboratory, it is your responsibility not only to train, but also to mentor your colleagues as they progress through the skills required to become a competent, Slidemaker Stainer operator.

Resources

Slidemaker Stainer Training Modules with Skill Checks

Training Checklist

System HELP

Page 184: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-38 of 44

Page 185: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-39 of 44

INFORMATION / PRACTICE SECTION

Hints Before You Start

Step Action

1 Have a plan!

2 Let trainees know that you expect them to become competent Slidemaker Stainer operators.

3 Have supplies ready: copies of modules, Skill Checks, etc.

4 Provide practice, practice, and more practice for trainees!

5 Watch trainees, but let them do the “hands-on” stuff!

6 It’s OK for trainees to make a mistake! Don’t jump in too quickly, unless there is a real danger involved!

7 Offer encouragement and positive reinforcement!

8 Have trainees tell you what they learned about the topics.

9 Training is an ongoing process, not a one-time event!

10 Encourage pride in a job well done.

Page 186: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-40 of 44

Create a Training Plan

Step Action

1 Create a written training plan outline for your facility. Include basic training topics (for all) as well as more advanced training topics (selected individuals).

Basic topics should include: System Overview, Daily Checks, Sample Processing, Shutdown and Supplies.

You may also want to consider Using Help as part of basic topics. Any other topics to include are up to you.

More advanced topics for selected individuals are up to

you but you may want to consider: Staining Protocols.

2 List the topics you plan to cover in each section.

3 Describe how you will do the training for each topic. List the resources you will need and the techniques (verbal, module, practice, demo, summaries, etc) you plan to use.

4 List the important points that need to be covered in each topic.

5 Describe how you will verify that a trainee learned the topics listed (observation, skill check, etc).

6 Describe the documentation you will use to show completed training for an individual (checklist provided or your own sign off sheet).

Page 187: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-41 of 44

Train a Person in the Basics

Step Action

1 Choose a person to train in all the basic topics listed.

2 Following your training plan, provide the trainee with any supplies and materials needed.

3 Using the technique(s) you outlined, have the trainee learn all the topics on your list.

4 Verify each topic learned according to your outline.

5 Document your training on a checklist.

You reached the goal!

Please continue to the Skill Check.

Page 188: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-42 of 44

Page 189: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training Modules PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development

AT-43 of 44

MENTOR / TRAINER SKILL CHECK

Review of Training Plan

Training plan contains a list of the basic topics to teach to everyone.

_______________________________________________________

Training plan includes a list of more advanced topics to teach to selected individuals.

_______________________________________________________

Training plan contains a list of the important points to cover in each topic.

_______________________________________________________

Training plan describes how a trainee will be verified.

_______________________________________________________

There is a checklist and sign off sheet to document each individual’s training.

_______________________________________________________

Continued on back >>>>>

Page 190: UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System … · 2015-08-14 · UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System UniCel DxH Slidemaker Stainer Training

UniCel DxH SLIDEMAKER STAINER Coulter Cellular Analysis System

UniCel DxH Slidemaker Stainer Training PN: A96923AA.2 (June 2014) Beckman Coulter Professional Development AT-44 of 44

Observation of Training

Trainer followed their training plan.

________________________________________________________

Trainer used the techniques outlined in their plan.

________________________________________________________

Trainer had all supplies on hand.

________________________________________________________

Trainer coached the trainee offering corrective action, as needed.

________________________________________________________

Trainer offered encouragement and positive reinforcement.

________________________________________________________

Trainer verified each topic learned according to their plan.

________________________________________________________

Trainer documented the training using their checklist.

________________________________________________________