vet abc troubleshooting. changing the pack service menu change pack remove the old pack etc....
TRANSCRIPT
CHANGING THE PACKCHANGING THE PACK
• Service menu Change Pack Remove the old pack etc. (on-screen instructions)
• Remove the Red caps from the new pack• Install new pack ensuring arrows have disappeared• Prime the new pack into an open vessel• Re-install waste pipe onto new pipe• Write on the pack the installation date
Operational ProblemsOperational Problems
• Basic understanding of how the ABC works • Function of the main components• How to spot potential problems• How to carry out minor repairs
““STARTUP FAILED CHECK STARTUP FAILED CHECK REAGENTS”REAGENTS”
Blank values high: PLT high, pack can be contaminated, reagent too cold
Perform concentrated cleaning or change pack
Electromagnetic noise (X-rays) Isolate Vet ABC and run start-up
““STARTUP FAILED CHECK STARTUP FAILED CHECK REAGENTS”REAGENTS”
If other parameters are OK, photometer blank failed Check photometer (see below)
Pack was shaken while in transport (bubbles) After Transport by a carrier, wait 1-2 hours before
installation
Tube on valve, needle, chamber or dispenser block is loose Check tube connections and repair if possible
““SENSOR DEFECT OR REAGENT EMPTY”SENSOR DEFECT OR REAGENT EMPTY”
• Pack wasn’t installed properly• Re-seat the pack ensuring arrows have disappeared
• Valve on base of pack has collapsed inwards• Adjust the position of the valve
• Pack is empty• Change the pack
““SENSOR DEFECT OR REAGENT EMPTY”SENSOR DEFECT OR REAGENT EMPTY”
• Pack is distended with excess air• “Burp” the pack
• Red caps were not removed• Remove the Red caps
• Waste tube is blocked or bent• Check tube is free of obstructions
““ERROR: MOTOR PRESSURE SYRINGE”ERROR: MOTOR PRESSURE SYRINGE”• Pack is distended with excess air• “Burp” the pack
• Piston is pushed out of chamber by high pressure• Re-install the piston
• Piston is sticking• Check movement – Service menu Mechanic
Pressure Syringe.
““ERROR: LIQUID SYRINGE MOTOR” ERROR: LIQUID SYRINGE MOTOR”
• Piston or gear is damaged• Check movement – Service menu Mechanic Liquid
Syringe
• Dilution Block is damaged• Check for leaks – Call Woodley Equipment.
““START-UP FAILED HGB REFERENCE FAILURE”START-UP FAILED HGB REFERENCE FAILURE”
• WBC Chamber is contaminated• Run concentrated clean
• Reagent is contaminated• Change pack
““START-UP FAILED HGB REFERENCE FAILURE”START-UP FAILED HGB REFERENCE FAILURE”
• No reagent is available• Prime the pack – check that reagent appears in
chambers
• Photometer is out of adjustment• Run concentrated clean; Check photometer
adjustment – Service menu Technician Adjustments