without reference, identify principles relating to central gas systems as they pertain to facility...
TRANSCRIPT
Without reference, identify principles relating to central gas systems as they pertain to facility and equipment interface with at
least 70 percent accuracy.
CENTRAL GAS SYSTEMS
Primary objective of suppliers of medical gas distribution systems is to provide Creative technical service Quality products that keep pace with changing
needs Central Gas Systems
Supply medical gases in life support situations It is essential that the reliability and integrity of
the system be the responsibility of all professions • Manufacturing, Design, Installation, Testing and the
Operation
CENTRAL GAS SYSTEMS
Medical Gas Systems The main features of medical gas systems are the
types of gases and the means of their delivery Types of medical gases • Oxygen
Oxygen is stored as a compressed gas at room temperature or refrigerated as a liquid
Primarily used for respiratory therapy and anesthesia Constant supply and immediate availability through out
medical facility is essential Small hospitals often store oxygen in banks of cylinders
connected by a manifold to provide a “central gas supply” Should be designed to provide 60 psig at the outlets Most anesthesia machines accommodate one or two cylinders
to be used in emergencies
CENTRAL GAS SYSTEMS
Nitrous oxide Primarily used for OR area Is almost always stored by hospitals in large high-
pressure cylinders connected by a manifold Should be designed to provide 60 psig at the outlets Bulk liquid storage of nitrous oxide is economical
only in large institutions Most anesthesia machines accommodate one or two
cylinders to be used in emergencies
CENTRAL GAS SYSTEMS
Medical air (compressed air) Primarily used for respiratory therapy in areas like
nurseries, emergency rooms, intensive care areas and OR’s
CENTRAL GAS SYSTEMS
The use of air is becoming more frequent in anesthesiology due to the hazards of high concentrations of oxygen
Dehumidified but non sterile air is provided to the hospital pipeline by compression pumps• Compression pumps should be designed to provide
60 psig at the outlets • Must be free from oil and moisture • The compressor will not add contaminants in the
form of particulate matter, odor or other gases
CENTRAL GAS SYSTEMS
The medical air compressors intake will be • Located outdoors • Above roof level • A minimum distance of 10 feet from any door,
window, exhaust, other intake, or opening in the building
• A minimum distance of 20 feet above the ground The intake will be turned down and screened or
otherwise be protected against the entry of vermin or water
CENTRAL GAS SYSTEMS
Compressor alternation Compressors will automatically alternate in operation
to allow division of operating time If automatic alternation is not provided, the facility
will arrange a proper schedule for manual alternation A backup medical compressor will be available to
automatically activate if the unit in operation is: Incapable of adequately maintaining pressure Inoperative
CENTRAL GAS SYSTEMS
Compressors will be alarmed appropriately for Conditions that affect air quality Failure
Vacuum Not a medical gas, but an essential part of
medical gas system Widely used throughout medical facilities in
Lab, patient treatment areas, surgical recovery and intensive care areas
CENTRAL GAS SYSTEMS
Uses
To remove fluid from incisions and assist in vital post-operative drainage
In active scavenging systems Used to remove excess waste gases from the patient
circuit of anesthesia units Routes gases out of the operating room and towards a
safe area outside of medical facility
CENTRAL GAS SYSTEMS
Vacuum system • Normally designed to provide 15 - 20 in
Hg of negative pressure at the furthest point from central vacuum • Types of systems
Wet systems will remove all waste through the service inlet
Dry systems will only remove air-gas through the service inlet
• Types of waste removed Liquid, Air-gas and/or Solids
CENTRAL GAS SYSTEMS
Vacuum pumps
Two or more pumps will be installed that serve the system alternately or simultaneously on demand
Each pump will be large enough to serve peak demand with the other pump inoperative
Station inlets for vacuums will be legibly labeled “Suction” or “Vacuum”
CENTRAL GAS SYSTEMS
Common Medical Gases Labels Air – air Carbon dioxide – CO2 Carbon monoxide – CO Helium – He Hydrogen – H2 Nitrogen – N2 Nitrous oxide – N2O Oxygen – O2
CENTRAL GAS SYSTEMS
Delivery of Medical Gases Gases are delivered from their central supply
source to the operating room through a piping network.
Usually constructed of seamless copper tubing Internal contamination of the pipelines with
dust, grease, or water must be avoided. The delivery system terminates in the
operating room as • Hose drops • Gas columns • Elaborate articulating arms
CENTRAL GAS SYSTEMS
Medical equipment is connected to the gas system through quick coupler mechanisms Each coupler is color coded
O2 – greenAir – yellowN2O – blueVacuum – white
CENTRAL GAS SYSTEMS
Designed to fit a specific gas PISS – pin index safety systems
This system is used with small gas cylinders Accomplished by 2 metal pins on the yoke that
correspond to two holes in the cylinder head There are six pin positions, each of which are specific to
a particular gas Some gas mixtures have 2 pin positions assigned to
differentiate different gas proportionsOne cylinder gasket must be used Multiple gaskets can defeat the PISS
CENTRAL GAS SYSTEMS
DISS – diameter indexing safety system This system is used with central pipeline hoses and
large gas cylinders Accomplished by using various diameter male to
female connectors specific to each gas
CENTRAL GAS SYSTEMS
Factors in Supplying Medical Gases Location Number of outlets required Pipeline routing and size of pipes Medical gas supply source Intermediate pipeline controls Gas service outlets Critical care area and equipment System specifications
CENTRAL GAS SYSTEMS
Medical Gas Alarms Master alarm • Monitors the operation and condition of the source of
supply, the reserves and the pressure of the main lines of all medical gas piping systems
• The master alarm system consists of two or more alarm panels located in two separate locations
Area alarm • Provided for each medical gas piping system
supplying anesthetizing locations and other vital life support and critical care areas
• Usually located at the nurse’s station
CENTRAL GAS SYSTEMS