WMO - TECO 200 St, Petersburg, Russia, 27-29 November 2008 Universi ty of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA ITALIAN AIR FORCE METEOROLOGICAL SERVICE The WMO Field Intercomparison of RI gauges was started in September 2007 in Vigna di Valle, Rome (Italy). Installation of the instruments in the field was preceded by the laboratory calibration of all submitted catching type rain gauges at the University of Genoa. Periodic testing of these gauges by means of dynamic calibration is now performed throughout the measurement campaign, using a portable calibration device.

Upload: mason-hiner

Post on 30-Mar-2015

219 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

The WMO Field Intercomparison of RI gauges was started in September 2007 in

Vigna di Valle, Rome (Italy). Installation of the instruments in the field

was preceded by the laboratory calibration of all submitted catching type rain gauges at the University of Genoa.

Periodic testing of these gauges by means of dynamic calibration is now performed

throughout the measurement campaign, using a portable calibration device.

Page 2: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Laboratoryof the University of Genoa

Preliminary tests in the laboratory(see WMO Lab

Intercomparison procedure)

Page 3: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

ID Model Nation Measuring principle

1 7499020BoMV2/RIMCO AUSTRALIA Tipping bucket

2 AP23/PAAR AUSTRIA Tipping bucket

3 R01 3070/PRECIS-MECANIQUE FRANCE Tipping bucket

4 PT 5.4032.35.008/THIES GERMANY Tipping bucket

5 R 102 ETG ITALY Tipping bucket

6 DQA031/LSI LASTEM ITALY Tipping bucket

7 T-PLUV UM7525/I/SIAP-MICROS ITALY Tipping bucket

8 PM B2 CAE ITALY Tipping bucket

9 RAIN COLLECTOR II7852 DAVIS USA Tipping bucket

10 15188/LAMBRECHT GERMANY Tipping bucket

11 PP040/MTX ITALY Tipping bucket

12 ARG100/ENV. MEAS. Lmt. BRASIL / UK Tipping bucket

13 MRW500 METEOSERVIS CZECH REP. Weighing gauge

14 VRG101/VAISALA FINLAND Weighing gauge

15 PLUVIO/OTT GERMANY Weighing gauge

16 T-200B / GEONOR NORVAY Weighing gauge

17 TRwS/MPS SVOLACK REP. Weighing gauge

18 ANS 410-H/EIGENBRODT GERMANY Pressure sensor

19 Electrical rain gauge/KNMI NETHERLAND Level sensor

List of catching-type instruments involved

Page 4: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Box-plots indicate the mean (solid line), median (thin line), 25-75th percentiles (box limits), 10-90th percentiles (whisker caps) and outliers (black circles).

Page 5: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Tipping-bucket (pit gauge)Aggregation : 1 min

-10

-5

0

5

10

0 100 200 300

Ir [mm/h]

Error [%]

Weighing gaugeAggregation : 1 min

-10

-5

0

5

10

0 100 200 300

Ir [mm/h]

Error [%]

Novel approach:tests are performed atone-minute resolution.“Instantaneous” RI rather than long term averages.

Page 6: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Box-plots indicate the mean (solid line), median (thin line), 25-75th percentiles (box limits), 10-90th percentiles (whisker caps) and outliers (black circles).

Best performing tipping-bucket rain gauge

Page 7: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Box-plots indicate the mean (solid line), median (thin line), 25-75th percentiles (box limits), 10-90th percentiles (whisker caps) and outliers (black circles).

Best performing weighing gauge

Page 8: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Box-plots indicate the mean (solid line), median (thin line), 25-75th percentiles (box limits), 10-90th percentiles (whisker caps) and outliers (black circles).

Typical (left) and bad (below) tipping-bucket rain gauges

Page 9: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

0

100

50

0

10050 75

75

Iref [mm h-1]

Imeas [mm h-1]

200100

300

250150

100

150

200

250

300 Iref [mm h-1]

Imeas [mm h-1]

Range: 0-100 mm·h-1 Range: 100-300 mm·h-1

Note: both axes of the graphs are distorted by a power factor of three.

All catching type rain gauges

Page 10: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Range: 0-100 mm·h-1 Range: 100-300 mm·h-1

Note: both axes of the graphs are distorted by a power factor of three.

All tipping-bucket rain gauges

0

100

50

0

10050 75

75

Iref [mm h-1]

Imeas [mm h-1]

200100

300

250150

100

150

200

250

300

Iref [mm h-1]

Imeas [mm h-1]

Page 11: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Range: 0-100 mm·h-1 Range: 100-300 mm·h-1

Note: both axes of the graphs are distorted by a power factor of three.

Tipping-bucket rain gauges with correction

0

100

50

0

10050 75

75

Iref [mm h-1]

Imeas [mm h-1]

200100

300

250150

100

150

200

250

300 Iref [mm h-1]

Imeas [mm h-1]

Page 12: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Range: 0-100 mm·h-1 Range: 100-300 mm·h-1

Note: both axes of the graphs are distorted by a power factor of three.

Weighing rain gauges

0

100

50

0

10050 75

75

Iref [mm h-1]

Imeas [mm h-1]

200100

300

250150

100

150

200

250

300 Iref [mm h-1]

Imeas [mm h-1]

Page 13: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

A portable device was developed at the University of Genoa with the aim of providing the on-site capability of performing the same kind of tests that were preliminarily used for calibration of all submitted catching type rain gauges under controlled conditions in the laboratory.

Page 14: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

The principle exploited by this portable device is that of preserving a constant

hydraulic head over a given orifice area by ensuring the automatic and continuous

pressure adaptation of the air/water contained inside a closed container.

The transit time of the water level between two fixed limits is the only variable to be

measured to complete the test at any reference rainfall intensity.

In order to reduce the sampling error, with reference to e.g. a tipping-bucket rain gauge

having a resolution of 0,2 mm (bucket volume of 20 g) and a collector’s area of 0,1

m2, the container should be filled in with at least 2 litres of water, so that at least 100 tips

of the buckets will occur. Patent n° 102006A000868 dated 07/12/2006

Page 15: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Application in the field test site

WMO procedure for field verification …

Page 16: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

ConclusionsThe laboratory phase of the RI Field Intercomparison proved to be very useful in providing basic information on the behaviour of the catching type instruments involved, in order to single out counting errors from the assessment of the overall accuracy of the instruments, with specific reference to their catching errors and uncertainties due to the operating conditions (dynamic performances, wind and atmospheric conditions, etc.). In this view the laboratory calibration tests were performed at the resolution of one minute, so that the spreading of the errors around their average value already investigated during the previous Laboratory Intercomparison could be also evaluated.

The development of a portable Field Calibration Device to be used for checking the instruments as installed – with no need to return them to the laboratory where standard calibration tests can be performed – was very effective in keeping the behaviour of the instruments under control throughout the measurement campaign. The operational procedure is that of comparing on a regular basis the calibration data of each instrument at a few calibration points (constant flow rates) with the full calibration curves preliminarily obtained in the laboratory, and this is currently used as a criterion to assess the correct behaviour of that particular instrument.

Page 17: WMO - TECO 2008 St, Petersburg, Russia, 27-29 November 2008 University of Genoa PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING

WMO - TECO 2008St, Petersburg, Russia, 27-29 November 2008

University of Genoa

PRELIMINARY LABORATORY CALIBRATION AND VERIFICATION OF THE GAUGES USING A FIELD CALIBRATION DEVICE LANZA, L.G., STAGI, L., VUERICH, E. and F. MALASPINA

ITALIAN AIR FORCEMETEOROLOGICAL SERVICE

Future perspectives

The calibration in the field using a portable device allows investigating the achievable uncertainty in the field – filling the gap between precise (though far from the real world) laboratory tests and the necessary metrological confirmation in the field.

The experience gained at the laboratory of the University of Genoa and at the field site in Vigna di Valle (Rome) were very fruitful and allowed to develop a consistentspecific know-how on rainfall intensity measurements and the related instruments.

It is important that after completion of this RI intercomparison the acquired know-how and the professional capacities developed are not abandoned. Specific operational structures dedicated to rainfall intensity measurements so as to ensure the quality, continuity, homogeneity and archiving/accessibility of the related information could be easily built upon such capacities.This same structures will exploit the above capacity and technical-scientific know-how as well as operational capacities in verification, calibration, metrological confirmation and performance testing on prototype and commercial instruments.