x-rite exact calibration neostampa 8.1.5 guide 170612 · when the last patch is passed release the...

11
User Manual neoMatch of 1 11 Version: 8.1.5 Last update: 12/06/2017 neoStampa X-Rite eXact Guide for neoStampa

Upload: hadan

Post on 05-Nov-2018

232 views

Category:

Documents


2 download

TRANSCRIPT

User Manual neoMatch of 1 11

Version: 8.1.5 Last update: 12/06/2017

neoStampa

X-Rite eXact Guide for neoStampa

!

Table of ContentsINFORMATIONCopyright Support Material About the Documentation Technical requirements

2 2 2 2

1/ INTRODUCTION 4

Overview 4

2/ DEVICE CONFIGURATION 5

Driver Connection Configuration in Calibration Wizard

5 7

3/ PERFORMANCE 8

Single Ink Cut/Linearization chart measurement Profiler target chart measurement

8 11

neoStampa - X-Rite eXact of 2 11

!

Copyright © Inèdit Software all rights reserved. The software described herein is property of the manufacturer, for the exclusive use of the registered users. The total or partial reproduction of the program is prohibited, as well as all its support materials, by electronic, mechanical, photographic or any other, means without the express written authorization of the © Inèdit Software. All trademark mentioned in this manual is Copyright of its owners.

Support Material In our web http://www.inedit.com/en/downloads you will find further support materials such as video tutorials, apart from all the information of the programs and solutions of Inèdit. For technical support and phone support information, please visit http://www.inedit.com/. ︎ ︎If you think you can help us improve manual

contents, you can send your suggestions to [email protected].

About the Documentation In this User Manual you will find descriptions of the options that the program offers, and extensive explanations of various processes and aspects involved in this process.

Tips and Notes will help you make the most of the software’s possibilities.

Warnings will help you avoid making mistakes in the workflow’s process.

Technical requirements Windows Microsoft Windows XP (Service Pack 3), Microsoft Windows 7, Microsoft Windows 8.1, Microsoft Windows 10

4GB of RAM (16GB recommended) 500GB of hard disk

neoStampa - X-Rite eXact of 3 11

!

1/ Introduction

Overview This guide introduce the usage of X-Rite eXact portable spectrophotometer in combination with neoStampa 8.1 Calibration Wizard. The full process of Calibration Wizard is described in the separated user manual. Please refer to http://www.inedit.com/en/downloads.

neoStampa - X-Rite eXact of 4 11

!

2/ Device Configuration

Driver Connection

The device driver will be installed by Windows automatically when is connected to the computer and switched on.

If you have trouble with automated installation of the driver, follow the description below.

Connect the device to the computer using the supplied USB cable and switch the eXact on.

Open Control Panel | Device Manager. In the list, open Other devices and select X-Rite eXact.

Click right and select Update Driver Software…

neoStampa - X-Rite eXact of 5 11

!

From new window chose to Browse my computer for driver software to locate the driver.

Browse to and select C:\Program Files (x86)\Inedit\neoStampa 8\required\eXact. Click on Next to continue.

The installation of the driver starts and when is completed, click on Close.

neoStampa - X-Rite eXact of 6 11

!

Configuration in Calibration Wizard

Start Calibration Wizard. Click on Configure… bottom of the window.

Select Barbieri device form the list.

Select Illumination (D65 or D50) and Observer (2 or 10).

If you need to use UV-Filter in enable the option from Configuration window.

neoStampa - X-Rite eXact of 7 11

!

3/ Performance

Start the Calibration as usual until the point to measure printed Single Ink Cut chart, Linearization chart and Printer profiling target chart.

Single Ink Cut/Linearization chart measurement

Before starting using the device, it must be calibrated. The device has an integrated calibration plaque and will calibrate automatically when required. The calibration plaque is located directly below the optics when the device is in the open position (not locked).

Click on Calibrate button to make the action and follow the steps on the device display. Locate the device on a flat surface and tap the START button on display. It will display a countdown screen and then perform the calibration.

Print the charts as Calibration Wizard requires in the manual.

neoStampa - X-Rite eXact of 8 11

!

Press Measure… button to read the chart. It opens the measurement dialog and is ready to be measured.

Position the ruler with the device on the media above the first patch of row #1 of your chart. Lower the device to the target base and hold steady until the measurement is complete.

In one continuous motion, roll the device across the color bar without stopping. Use the black scribe line on the edge of the chassis to align device during the scan, and view the device display for scan speed indication. One of the circles appears on the device to provide feedback on your scanning speed.

Scanning the color bar too fast or too slow for an extended period of time may produce inaccurate measurement results. Always try to maintain a steady scanning speed with the white circle showing as much as possible for best results.

neoStampa - X-Rite eXact of 9 11

!

When the last patch is passed release the device.

Position the ruler with the device in front of the next row of your test chart and measure the next row.

Simultaneously in Calibration Wizard the lines are completed with each measures line. When done, click on OK in the measurement dialog and continue with the calibration until you reach the printed profiler target chart.

neoStampa - X-Rite eXact of 10 11

!

Profiler target chart measurement

©2017 Inedit Software S.L.

Using the large target chart of 1516 patches.

Press Measure… button to read the chart. It opens the measurement dialog and is ready to be measured.

Remember to calibration the device as described above before starting the measurement.

As seen above, position the ruler with the device on the media above the first patch of row #1 of your chart. Lower the device to the target base and hold steady until the measurement is complete. When the last patch is passed release the device.

Simultaneously in Calibration Wizard the lines are completed with each measures line. When done, click on OK in the measurement dialog and finish the calibration.

Print the charts as Calibration Wizard requires in the manual. Once again repeat the device calibration as described above.

neoStampa - X-Rite eXact of 11 11