O 2 4700(X) ppb module Instruction manual Order number: 52 121 222

Upload: others

Post on 05-Aug-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

O2 4700(X) ppb moduleInstruction manual

Order number: 52 121 222

70321_E_4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1

Page 2: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

TA-201.063-MTE04 101005 Software version: 2.2

Registered trademarksThe following registered trademarks are used in this instruction manualwithout further marking

SensocheckSensofaceVariPower

SMARTMEDIA®

is a registered trademark of Toshiba Corp., Japan

InPro®

is a registered trademark of Mettler-Toledo GmbH, Switzerland

WarrantyDefects occurring within 1 years from delivery date shall be remedied free of charge at our plant(carriage and insurance paid by sender). ©2005 Subject to change without notice

Mettler-Toledo GmbH, Process Analytics, Industrie Nord,CH-8902 Urdorf, Tel. +41 (44) 736 22 11 Fax +41 (44) 736 26 36 Subject to technical changes. Mettler-Toledo GmbH, 10/05. Printed in Germany.

Return of products under warrantyPlease contact your local Mettler-Toledo representative. Ship the cleaned device to the address you havebeen given. If the device has been in contact with process fluids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certificate, for the health andsafety of our service personnel.

Disposal Please observe the applicable local or national regulations concerning the disposal of “waste electri-cal and electronic equipment”.

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 2

Page 3: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 3

Page 4: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 4

Page 5: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 5

Page 6: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

4 O2 4700(X) ppb

Warranty ......................................................................................................2Registered trademarks ..................................................................................2EC Declaration of Conformity ......................................................................3Intended use ............................................................................................... 8Safety information....................................................................................... 9Software version ...................................................................................... 10

Modular concept and instruction manuals .......................................... 11Short description: M 700 FRONT ...............................................................12Short description: Menu structure ..............................................................13Short description: M 700 BASE...................................................................15Parameter tables (Excel) at www.mtpro.com ............................................. 16

Terminal plate ......................................................................................... 17Inserting the module ................................................................................. 19

Wiring examples..................................................................................... 20

Quick start: Menu selection/structure ........................................................................... 22Passcode entry .......................................................................................... 23Configuring the measurement display ....................................................... 24Setting diagnostics messages as favorite.....................................................26

Calibration / adjustment........................................................................ 28Common combination: process variable / calibration mode ........................29Selecting a calibration method ...................................................................30

Automatic calibration in water.................................................................32Automatic calibration in air......................................................................34Product calibration: Saturation ................................................................36Product calibration: Concentration...........................................................38Data entry of premeasured sensors .........................................................40Zero point correction ...............................................................................41Adjustment..............................................................................................42

Contents

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 4

Page 7: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

5

Contents

Maintenance ............................................................................................43

Setting the module parametersOperating levels...................................................................................... 44Enable / lock functions ........................................................................... 45Sensor data, Cal preset values, Pressure/Salt correction........................... 47Messages................................................................................................ 48

Setting the logbook parameters (System control)Logbook ..................................................................................................50Factory setting .........................................................................................50

Setting the system control and the outputs (BASE)Setting the current output .......................................................................51Output filter (time constant) ....................................................................54NAMUR signals (current outputs) - Behavior during messages - ...............55NAMUR signals (relay contacts):failure, maintenance request, function check...........................................56Relay contacts: Protective wiring..............................................................57Setting the relay contacts ........................................................................58Limit value, hysteresis, contact type .........................................................59

Setting the OK1, OK2 inputs (BASE)Optocoupler inputs (BASE): Usage and switching level.............................60Switching parameter sets via OK2 (system control) ..................................61

Calculation Blocks (System control)Calculation of new variables from measured variables .............................62

Diagnostics functions ..............................................................................65Message list.............................................................................................65Logbook ..................................................................................................65Device description....................................................................................66Module FRONT ........................................................................................66Module BASE...........................................................................................66Module diagnostics..................................................................................67Sensor monitor ........................................................................................67

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 5

Page 8: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

6 O2 4700(X) ppb

Cal record................................................................................................68Sensor network diagram..........................................................................68Statistics ..................................................................................................68Sensoface ................................................................................................69

Specifications .......................................................................................... 71

Appendix................................................................................................. 75Minimum measuring spans for current outputs ......................................... 75Dissolved oxygen measurement in carbonated beverages ...........................76(Additional function SW 700-011)

Parameter setting menu (Overview of all setting possibilites).....................................................77

Index........................................................................................................ 82

Overview of icons ...................................................................................89

Overview of menu selection ..................................................................90

Contents

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 6

Page 9: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

7

Contents

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 7

Page 10: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

8

Intended use

The module is used to measure traces of dissolved oxygen in liquids andgases using Mettler-Toledo sensors of the InPro 6900 Series.The module permits simultaneous measurement of oxygen partial pressure,barometric pressure and temperature. In addition, oxygen saturation or con-centration can be calculated and displayed.

The O2 4700X ppb module is intended for operation in locations subject toexplosion hazards which require equipment of Group II, device category 2(1),gas/dust.

Conformity with FDA 21 CFR Part 11In their directive “Title 21 Code of Federal Regulations, 21 CFR Part 11,Electronic Records; Electronic Signatures” the US American health agencyFDA (Food and Drug Administration) regulates the production and processingof electronic documents for pharmaceutical development and production.This results in requirements for measuring devices used for correspondingapplications. The following features ensure that the M 700(X) modularprocess analysis system meets the demands of FDA 21 CFR Part 11:

Electronic SignatureAccess to the device functions is regulated and limited by individuallyadjustable codes – “Passcodes”. This prevents unauthorized modification ofdevice settings or manipulation of the measurement results. Appropriate useof these passcodes makes them suitable as electronic signature.

Audit Trail LogEvery change of device settings can be automatically recorded and docu-mented in the Audit Trail Log on the SmartMedia card. The recording can beencoded.

O2 4700(X) ppb

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 8

Page 11: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

9

Safety information

Caution!Never try to open the module! If a repair should be required, return themodule to our factory.

If the specifications in the instruction manual are not sufficient for assessingthe safety of operation, please contact the manufacturer to make sure thatyour intended application is possible and safe.

Be sure to observe during installation:• Switch off power supply before replacing or inserting a module.• Protect the signal inputs of the modules against electrostatic discharge.• Before commissioning it must be proved that the device may be

connected with other equipment.• Observe correct shielding: To avoid interferences, the cable shielding must

be completely covered by the ESD shielding cap.

Application in hazardous locations: O2 4700X ppb moduleWhen using the O2 4700X ppb module, the stipulations for electrical installa-tions in hazardous areas (EN 60079-14) must be observed. When installing the device outside the range of applicability of the 94/9/ECdirective, the appropriate standards and regulations in the country of usemust be observed. The module has been developed and manufactured incompliance with the applicable European guidelines and standards.

Compliance with the European Harmonized Standards for use in hazardouslocations is confirmed by the EC-Type-Examination Certificate. Compliance with the European guidelines and standards is confirmed by theEC Declaration of Conformity.

There is no particular direct hazard caused by the operation of the device inthe specified environment.

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 9

Page 12: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

10 O2 4700(X) ppb

Software versionO2 4700(X) ppb module

Device software M 700The O2 4700(X) ppb module is supported by software version 4.0 or higher.

Module software O2 4700(X) ppbSoftware version 2.1 Nov. 17, 2003Software version 2.2 May 12, 2004

Menu Display Device description

Provides information about all mod-ules installed: Module type and func-tion, serial number, hardware andsoftware version and device options. Select the different modules (FRONT,BASE, slots 1 - 3) using the arrowkeys.

Device description

Return

22.7 °C83 % Air

Operating panel M 700Module FRONT M 700-011

Hardware: 2, Software: 6.2Serial number: 0000815

BASEModule FRONT

Query actual device/module softwareWhen the analyzer is in measuring mode:Press menu key, open Diagnostics menu.

Options

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 10

Page 13: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

11

Modular concept and instruction manualsInstruction manuals for basic unit, measuring module, additional functions.

• The instruction manual for the M 700(X) describes how to install, commission and operate the basic unit.

• The instruction manual for the measuring or communication module describes all functions required for commissioning and working with the respective measuring or communication module.

• Additional functions are supplied with a function description.

Latest product information and instruction manuals are available viawww.mtpro.com.

The M 700(X) is an expandable modular process analysis system.The basic unit (M 700 FRONT and BASE) provides three slots which can beequipped by the user with any combination of measuring or communicationmodules. The software capabilities can be expanded by additional functions(options). Additional functions must be ordered separately. They are suppliedwith a device-specific TAN for function release.

3 module slotsfor free combination of measuring and communication modules

M 700(X) modular process analysis system

retrofita

ble

Op

tion Measuring module• pH / ORP / Temp (also ISM)• 02/Temp (also ISM)• Noncontacting conductivity / Temp• Contacting conductivity / Temp

Additional functionsActivation via device-specific TANFor an overview, see www.mtpro.com

SmartMedia cardData recordingFor an overview, see www.mtpro.com

Communication modules• Out (additional switching and

current outputs)• PID (analog and digital controller)• Profibus PA

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 11

Page 14: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

12 O2 4700(X) ppb

Short description: M 700 FRONTM 700Modular hardware and software system for liquid analysis.

Transflective LC graphic display(240 x 160 pixels)white backlighting, high resolutionand high contrast.

Red LEDsignals failure (On) ormaintenance request/function check(flashing) according to NE 44.

4 captive scewsfor opening the analyzer(Caution! Make sure that the gasket between FRONT and BASE is properly seated and clean!)

Green LEDVoltage supply okay

5 self-sealing cable glandsM20 x 1.5for entry of voltage supply and signal lines

2 softkeyswith context-sensitive functions.

User interfacewith plaintext menus as recom-mended by NAMUR. Menu texts can be switched to:German, English, French, Italian,Swedish, and Spanish.Intuitively acquirable menu logic,based on Windows standards.

Control panel3 function keys(menu, meas, enter)and 4 arrow keys for menu selectionand data entries

Measurement displayFor parameter setting, see Pg 24

Secondary displaysFor parameter setting, see Pg 26

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 12

Page 15: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

13

Measuring

Short description: Menu structureBasic functions: Calibration, maintenance, parameter setting, diagnostics

Calibration Maintenance Parameter setting Diagnostics

Menu groups

1

2

3

1147 2958 1246Operator level

1989Admin. level

Passcode:

Selection offurthermenu items:

Module 1Module 2Module 3

BASEModule 1Module 2Module 3

SYSTEMFRONTBASEModule 1Module 2Module 3

Message listPoint of measdescriptionLogbookDevice description

FRONTBASEModule 1Module 2Module 3

4

5

6

Legend:(1) Pressing the menu key accesses menu selection(2) Pressing the meas key returns to measurement(3) Menu groups are selected using the arrow keys(4) Press enter to confirm, enter passcode(5) Further menu items are displayed(6) Selected functions of the Diagnostics menu can be

recalled via softkey even when in measuring mode (Pg 26)

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 13

Page 16: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

14 O2 4700(X) ppb

Short description: M 700 FRONTM 700View into the open device (M 700 FRONT)

Terminal platesof “hidden” modulesEach module comes with an adhesivelabel containing the contact assignments.This label should be sticked to the innerside of the front (as shown).Then, the terminal assignments remainvisible even if further modules areinserted.

Slot for SmartMedia card• Data recording

The SmartMedia card expands the measurement recorder capacity to > 50000 records.

• Exchange of parameter sets5 parameter sets can be stored on the SmartMedia card, 2 of them can be loaded to the M 700 and switched by remote control. Configurations can be transferred from one M 700 to the other.

• Function expansionsare possible with additional softwaremodules which are released using transaction numbers (TAN).

• Software updates

The circumferential sealingguarantees IP 65 protection and allows spraycleaning / disinfection.Caution! Keep clean!

Replacing the front modulePull off power cord and ground wire. To separate the M 700 FRONT from the M 700 BASE, turn the retaining screwsof the pivot hinge by 90°.

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 14

Page 17: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

15

Short description: M 700 BASEM 700View into the open device (M 700 BASE, 3 function modules installed)

M 700 BASE2 current outputs (free assignment ofprocess variable) and 4 relay contacts, 2 digital inputs. VariPower broad-range power supplyunit, 20 ... 265 V AC/DC, suitable for allpublic mains supplies in the world.

Power supply units, IS version:100 ... 230 V AC or24 V AC/DC

Important note concerning SmartMedia cardThe SmartMedia card may be inserted or replaced with the power supplyswitched on. Before a memory card is removed, it must be “closed” in themaintenance menu. When closing the device, make sure that the sealing isproperly seated and clean.

Module equipmentModule identification: Plug & PlayUp to 3 modules can be combined asdesired. Several input and communicationmodules are available

Warning! Do not touch the terminal compartment, there may bedangerous contact voltages!

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 15

Page 18: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

16 O2 4700(X) ppb

Parameter tables (Excel):www.mtpro.com

Parameter tables (Excel)2 complete parameter sets can be stored in the Premium TransmitterM 700(X). You can document the parameter settings of your complete point of measurement in an Excel table that can be downloaded from ourwebsite.

The complete documentation is available in the download area of ourwebsite www.mtpro.com.

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 16

Page 19: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

17

Terminal plate O2 4700(X) ppb module

Attaching the terminal platesThe terminal plates of the lower modules can besticked to the inner side of the door. This facilitates maintenance and service.

Terminal plate O2 4700 ppb module:

Terminal plate O2 4700X ppb module:

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 17

Page 20: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

18 O2 4700(X) ppb

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 18

Page 21: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

19

Inserting the moduleNote: Be sure to connect the shielding properly!

Make sure that the cableglands are tightly closed toprotect against humidity.

Terminals 2 and 8 are coveredby an ESD shield. To connect the sensor cable, justpull it back.

Switch off power supplyOpen the device (loosen the 4 screws at the front)Place module in slot (D-SUB connector)Tighten fastening screws of the moduleOpen ESD shielding cap (covering terminals 2 and 8)Connect sensor cable. To avoid interferences, the cable shielding must be completelycovered by the ESD shielding cap.Close ESD shielding cap (covering terminals 2 and 8)Close device, tighten screws at the frontSwitch on power supplySet parameters (Pg 47)

1.2.3.4.5.6.

7.8.9.

10.

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 19

Page 22: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

20 O2 4700(X) ppb

Wiring example 1Mettler-Toledo InPro 6900 trace sensor, VP cable

red

whi

te

gree

n

O2 4700 ppb

gn/y

w

gray

tran

spar

ent

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 20

Page 23: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

21

Wiring example 2Mettler-Toledo InPro 6800 sensor, VP cable

O2 4700 ppb

red

whi

te

gree

n

tran

spar

ent

Blue, gray wires not connected

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 21

Page 24: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

22 O2 4700(X) ppb

12

16.8%Airman

25.6 °C09.06.05

Menu selection

Return to meas

16.8 %Air25.6 °C

Select: [enter]

Lingua

Menu selection

1 Pressing the menu key accesses menu selection.2 Pressing the meas key returns to measurement.

After switching on, the analyzer performs an internal test routine and auto-matically detects the number and type of modules installed.Then, the analyzer goes to measuring mode.

Measuring

Menu structure

Calibration Maintenance Parameter setting Diagnostics

Menu groups (Select using arrow keys)

1

2

3

1147 2958 1246Operator level

1989Administrator level

Passcode(as delivered)

Selection of furthermenu items:

3 Arrow keys for selecting a menu group4 enter key for confirming a selection

1

2

3

4

70321_E_4700xppb.QXD 13.10.2005 13:28 Uhr Seite 22

Page 25: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

23

Changing a passcode:“Passcode entry” menuWhen this menu is opened, theanalyzer displays a warning (Fig.).Passcodes (factory settings): Calibration (cal) 1147Maintenance (maint) 2958Operator level (opl) 1246Administrator level (adm) 1989CautionIf you lose the Administrator pass-code, system access is locked! Pleaseconsult our technical support!

To change a passcode, select “On”using the arrow keys. Confirm with enter.Select the position using theleft/right keys, then edit thenumber using the up/down keys. When all numbers have beenentered, confirm with enter.

To enter a passcode:Select the position using the left/right keys,then edit the number using the up/down keys. When all numbers have been entered, confirm with enter.

To change a passcode• Open the menu selection (menu key)• Select parameter setting• Administrator level, enter passcode• Select System control: Passcode entry

Passcode entry

Menu Display System control:Passcode entry

Passcode entry (Administrator)

OK

25.6 °C16.8 %Air

Passcode entry (Administrator)

25.6 °C16.8 %Air

cal Calibration On Off

maint Maintenance On Off

Change passcode

opl Operator level

1147

On Off

Close

cal Calibration On

maint

opl

Off

If you lose youradm passcode, systemaccess will be locked!

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 23

Page 26: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

24 O2 4700(X) ppb

Configuring the measurement displaySelect menu: Parameter setting/Module FRONT/Measurement display

Pressing meas (1) returns the M 700 to the measuring mode from anyfunction. All process variables coming from the modules can be displayed. The tableon the next page describes how to configure the measurement display.

Secondary displaysAdditional values, also date andtime, can be displayed dependingon the modules installed. Theyare selected using the softkeys.

Measurement displayTypical display for 2 measuredvalues.

16.924.0

%Air

°C10.3 mA 03.09.05

1

SoftkeysIn measuring mode, the softkeysallow selection of values for thesecondary displays or control offunctions (user defined).

70321_E_4700xppb.QXD 13.10.2005 13:28 Uhr Seite 24

Page 27: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

25

Menu Display Configure measurement display

Configure measurement displayPress menu key to select menu.Select parameter setting using arrowkeys, confirm with enter. Select:“Administrator level”: Passcode 1989

Parameter setting:Select “Module FRONT”

FRONT module:Select “Measurement display”

Measurement display:Set the number of primary values(large display) to be displayed

Select process variable(s) to bedisplayed and confirm with enter.

Pressing the meas key returns tomeasurement.

Menu selection

Return to meas

16.8 %Air25.6 °C

Select: [enter]

Lingua

Parameter setting (Administrator)System control

Return

25.6 °C

Module BASE M 700-021Module O2 4700 ppb

16.9 %Air

Module PH 2700Module Cond Ind 7700

Module FRONT M 700-011 (Administrator)Languages

Return

25.6 °C

Measurement recorder

16.8 %Air

EnglishMeasurement display

Module FRONT M 700-011

25.6 °C16.9 %Air

1 primary value1st primary value

Viewing angle

2 primary values

Abort OK

Measurement display (Administrator)

25.6 °C16.9 %Air

1st primary value

Viewing angle

Main display

Abort K

Main display

mg/l

%Air%O2

mg/lppmmbarpH

values

2nd primary value °C

Measurement display (Administrator)

2nd primary value

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 25

Page 28: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

26 O2 4700(X) ppb

Setting diagnostics messages as favoriteSelect menu: Parameter setting/System control/Function control matrix

Example:“Favorites” to be selected with“Right softkey”

To select a softkey function:Select desired function usingarrow keys,press “Connect” softkeyand confirm with enter.

Deselect function:Press “Disconnect” softkey,confirm with enter.

25.6 °C

Input OK2

7.00 pH

Return Connect

Function control matrix (Administrator)

Left softkey

Right softkeyProfibus DO 2

ParSet KI rec. Fav–

FavoritesSelected Diagnostics functions can be calledup directly from the measuring mode using asoftkey. The following table (Pg 27) explains how toselect favorites.

Secondary displays (1)Here, additional values are displayed in themeasuring mode according to the factorysetting. When the respective softkey (2) ispressed, the process variables measured bythe modules plus date or time are displayed.

In addition, you can use the softkeys (2)to control functions. Too assign a function toa softkey, select

• Parameter setting/System control/• Function control matrix (Fig.):

Function which can be controlled by soft-keys:• Parameter set selection• KI recorder Start/Stop• Favorites• EC 400 (fully automated probe controller)

1

2

70321_E_4700xppb.QXD 13.10.2005 13:28 Uhr Seite 26

Page 29: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

27

Set/delete favorite:“Set favorite” allows activation ofthe selected diagnostic functiondirectly from the measuring mode viasoftkey. The respective function ismarked with a heart icon (seeSoftkey usage, Pg 26).

Pressing the meas key returns tomeasurement. When the softkey hasbeen assigned to “Favorites”,“Favorites menu” is read in the sec-ondary display (see “Function controlmatrix”, Pg 26).

Note:When one of the softkeys has been assigned to the “Favorites menu” func-tion, diagnostic functions which have been set as “Favorite” can be directlycalled up from the measuring mode.

Menu Display Select favorites

Favorites menuDiagnostics functions can be calledup directly from the measuring modeusing a softkey. The “Favorites” are selected in theDiagnostics menu.

Select favoritesPress menu key to select menu.Select diagnostics using arrow keys,confirm with enter. Then selectmodule and confirm with enter.

98.1240

%Air

mbar24.3°C Favorites menu

98.1240

%Air

mbar24.3°C Favorites menu

Menu selection

Return to meas

16.8 %Air25.6 °C

Select: [enter]

Lingua

Return

25.0 °C

Set favorite

16.8 %Air

Sensor monitorCal recordSensor network diagram OXYStatistics

Module diagnostics

Module O2 4700 ppb

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 27

Page 30: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

28 O2 4700(X) ppb

Calibration / adjustmentNote: Function check activeCurrent outputs and relay contacts behave as configured

• Calibration: Detecting deviations without readjustment• Adjustment: Detecting deviations with readjustment

Caution:Without adjustment every dissolved oxygen meter delivers an imprecise or wrong output value! After replacing the sensor, the electrolyte, or the sensormembrane, you must perform a calibration!The resulting values must be taken over by an adjustment for calculating themeasured variables (measured value display, output signals)!

ProcedureEvery dissolved oxygen sensor has its individual slope and zero point. Bothvalues are altered, for example, by aging. For sufficiently high accuracy ofoxygen measurement, the meter must be regularly adjusted for the sensordata (calibration + adjustment).

Sensor replacement (First Calibration)After replacement of the sensor, electrolyte or sensor membrane, a “FirstCalibration” should be performed. During First Calibration, the sensor dataare stored as reference values for the statistics. The ”Statistics” menu of Diagnostics (Pg 68) shows the deviations of zero,slope, calibration temperature, calibration pressure and response time of thelast three calibrations with respect to the reference values of the FirstCalibration. This allows evaluation of the drift behavior and aging of thesensor.

Calibration/adjustment methods• Automatic calibration in water/air• Product calibration (saturation/concentration)• Data entry• Zero point correction

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 28

Page 31: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

29

Recommendations for calibrationIt is always recommended to calibrate in air. Compared to water, air is a cali-bration medium which is easy to handle, stable, and thus safe. In the mostcases, however, the sensor must be dismounted for a calibration in air. When dealing with biotechnological processes which require sterile condi-tions, the sensor cannot be removed for calibration. Here, calibration mustbe performed with aeration directly in the process medium (e.g. aftersterilization). In the field of biotechnology often saturation is measured andcalibration is performed in the process medium (water) for reasons of sterility. For other applications where concentration is measured (water, waste water,etc.), calibration in air has proved to be useful.

Common combination: process variable / calibration modeMeasurement CalibrationSaturation WaterConcentration Air

If there is a temperature difference between the calibration medium and themeasured medium, the sensor must be kept in the respective medium forseveral minutes before and after calibration in order to deliver stable mea-sured values. The type of calibration pressure detection is preset duringparameter setting (Pg 47).

Adjustmentmeans that the values determined by a calibration are taken over.The values determined for zero and slope are entered in the calibrationrecord. (Cal record can be called up in the Diagnostics menu for the module,see Pg 68). These values are only effective for calculating the measured variables whenthe calibration has been terminated with an adjustment.A passcode ensures that an adjustment can only be performed by anauthorized person (Administrator).The Operator can check the current sensor data by a calibration and informthe Administrator when there are deviations.You can use the additional function SW 700-107 for granting access rights(passcodes) and for AuditTrail (continuous data recording and backupaccording to FDA 21 CFR Part 11).

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 29

Page 32: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

30 O2 4700(X) ppb

Calibration / adjustmentSelecting a calibration method

Measuring

Calibration Maintenance Parameter setting Diagnostics

Menu groups

1

2

3

1147Passcode:

Select O2module:

Module 1Module 2Module 3

4

5

To calibrate the O2 4700 ppb module: Select calibration method(1) Press the menu key to access menu selection(2) Pressing the meas key returns to measurement from every level(3) Select Calibration menu group using the arrow keys(4) Press enter to confirm, enter passcode(5) Select “Module O2 4700 ppb”, confirm with enter(6) Select calibration method

6

Module O2 4700 ppb

Return

25.6°C

Automatic - Air

Product calibration ConcData entry

Product calibration Sat

Zero correction

83.0%Air

Automatic - Water

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 30

Page 33: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

31

Menu Display Select calibration method

Call up calibrationPress menu key to select menu. Select calibration using arrow keys,confirm with enter, passcode 1147(To change passcode: Parameter set-ting/System control/Passcode entry)After passcode entry, the system is infunction check mode: Current out-puts and relay contacts behave asconfigured (BASE, Out, PID) until theCalibration menu is exited.

Calibration:Select “Module O2 4700 ppb”.

Menu selection

Return to meas

83.0%Air25.6 °C

Select: [enter]

Lingua

Calibration

Return

25.6°C

Module pH 2700Module Cond 7700

Info

83.0%Air

Module O2 4700 ppb

Module O2 4700 ppb

Return

25.6°C

Automatic - Air

Product calibration ConcData entry

Product calibration Sat

Zero correction

-83.0%Air

Automatic - Water

Select a calibration method:• Automatic - water (Pg 32)• Automatic - air (Pg 34)• Product calibration saturation (Pg 36)

• Product calibration concentration (Pg 38)• Data entry (Pg 40)• Zero point correction (Pg 41)

When you open the Calibration menu,the analyzer automatically proposesthe previous calibration method. If you do not want to calibrate, ”Return” with the left softkey.

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 31

Page 34: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

32 O2 4700(X) ppb

Calibration / adjustmentAutomatic calibration in water

Automatic calibration in waterThe calibration value is always the oxygen saturation (referred to air).First, the slope is corrected using the 100 % value.

During calibration the analyzer is in function check mode.Current outputs and relay contacts behave as configured (BASE, Out, PID).

Caution!Ensure sufficient medium flow to the sensor (see Specifications of dissolvedoxygen sensors). The calibration medium must be in equilibrium with air.Oxygen exchange between water and air is very slow. Therefore, it takes arelatively long time until water is saturated with atmospheric oxygen. If thereis a temperature difference between calibration medium and measuredmedium, the sensor must be kept in the respective medium for severalminutes before and after calibration.

Menu Display Select calibration mode

Select module: O2 4700 ppbThe analyzer is in function checkmode. Current outputs and relaycontacts behave as configured (BASE,Out, PID). Confirm with enter.

Select calibration method“Automatic - Water”Remove sensor and immerse it in calmedium (air-saturated water), ensuresufficient medium flow to the sensor.Confirm with enter.

Calibration

Return

25.6°C

Module pH 2700Module Cond 7700

Info

80.3%Air

Module O2 4700 ppb

Module O2 4700

Return

25.6°C

Automatic - Air

Product calibration ConcData entry

Product calibration Sat

Zero correction

80.3%Air

Automatic - Water

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 32

Page 35: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

33

Menu Display Automatic calibrationin water

Drift check.Display during calibration• Sensor current• Calibration temperature• Calibration pressure• Response timeWaiting time can be reduced bypressing enter (without drift check:reduced accuracy of calibrationvalues!). From the response time,you see how long it takes the sensorto deliver a stable signal. If thesignal or the measured temperaturefluctuate greatly, the calibrationprocedure is aborted after 2 min.Calibration must be re-started. Ifsuccessful, place sensor in process,end calibration with softkey orenter.

Automatic - Water

Cal medium: Air-sat. water

25.6 °C80.3%Air

When changing sensors performFirst cal for statistics!

Start

Automatic - Water

Drift check running

25.6 °C80.3%Air

Sensor currentSlope correction

-60.8nACalibration temp +025.6°CCalibration pressure 1013mbarResponse time 0002s

Calibration data record

Cal mode

25.6 °C80.3%Air

Response timeSlope

Automatic - Water

+059.3 nA

End Adjust

0070 s

End

Return

Display of selected calibrationmedium (Air-sat. water)Sensor replacement: Pg 28Enter cal pressure if “manual” hasbeen configured (Pg 47)Start with softkey or enter

AdjustmentPress “Adjust” to take over thevalues determined during calibrationfor calculating the measuredvariables. See Page 42.

Sensor replacementInput cal pressure 1013 mbars

Calibration

Zero point

09/12/05 15:20

+0.030 nA

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 33

Page 36: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

34

Calibration / adjustmentAutomatic calibration in air

Automatic calibration in airThe calibration value is always the oxygen saturation (referred to air).The slope value is corrected. If you do not know the exact value of the rela-tive humidity of the air used, you can take the following standard values fora sufficiently precise calibration:• Ambient air: 50 % (average)• Bottled gas (synthetic air): 0 %

During calibration the analyzer is in function check mode.Current outputs and relay contacts behave as configured (BASE, Out, PID).Caution!The sensor membrane must be dry. Be sure to keep temperature and pres-sure constant during calibration. If there is a temperature difference betweencalibration medium and measured medium, the sensor must be kept in therespective medium for several minutes before and after calibration.

O2 4700(X) ppb

Menu Display Select calibration mode

Select module: O2 4700 ppbThe analyzer is in function checkmode. Current outputs and relaycontacts behave as configured (BASE,Out, PID). Confirm with enter.

Select calibration method“Automatic - Air”Remove sensor and place it in air.

Confirm with enter.

Calibration

Return

25.6°C

Module pH 2700Module Cond 7700

Info

80.3%Air

Module O2 4700 ppb

Module O2 4700 ppb

Return

25.6°C

Product calibration ConcData entry

Product calibration Sat

Zero correction

80.3%Air

Automatic - WaterAutomatic - Air

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 34

Page 37: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

35

Menu Display Automatic calibrationin air

Drift check.Display during calibration• Sensor current, calibration temp,

cal pressure and response time.Waiting time can be reduced bypressing “End” (without drift check:reduced accuracy of calibrationvalues!). From the response time,you see how long it takes the sensorto deliver a stable signal. If thesignal or the measured temperaturefluctuate greatly, the calibrationprocedure is aborted after about2 min. Calibration must be re-started. If successful, place sensor inprocess, end calibration with softkeyor enter

Automatic - Air

Cal medium: Air-sat. water

25.6 °C80.3%Air

When changing sensors performFirst cal for statistics!

Start

Input cal pressureRelative humidity 0050%

Return

Automatic - Air

Drift check running

25.6 °C80.3%Air

Sensor currentSlope correction

-60.8nACalibration temp +025.6°CCalibration pressure 1001mbarResponse time 0002s

1013 mbars

End

Cal medium: AirSelect: First Calibration (Pg 28)Enter relative humidity, e.g.:• Ambient air: 50 %• Bottled gas: 0 %Enter cal pressure if “manual” hasbeen configured (Pg 47)Start with softkey or enter

Sensor replacement

Calibration data record

Cal mode

25.6 °C80.3%Air

Response timeSlope

Automatic - Air

+059.3 nA

End Adjust

0070 s

AdjustmentPress “Adjust” to take over thevalues determined during calibrationfor calculating the measuredvariables. See Page 42.

Calibration

Zero point

09/12/05 15:20

+0.030 nA

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 35

Page 38: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

36 O2 4700(X) ppb

Calibration / adjustmentProduct calibration: Saturation

Product calibration: Saturation(Calibration with sampling)When the sensor cannot be removed – e.g. for sterility reasons – its slopecan be determined with “sampling”. To do so, the currently measuredsaturation value of the process is stored by the measuring module. Directlyafterwards, a reference value is determined using a portable meter, forexample. The reference value is entered into the measuring system. From the difference between measured value and reference value, theanalyzer calculates the sensor slope. With low saturation values, the analyzercorrects the zero point, with high values the slope.During calibration the analyzer is in function check mode.Current outputs and relay contacts behave as configured (BASE, Out, PIDmodules).Caution!The reference value must be measured at temperature and pressure condi-tions similar to those of the process.

Menu Display Product calibration: Saturation

Select module: O2 4700 ppbThe analyzer is in function checkmode. Current outputs and relaycontacts behave as configured (BASE,Out, PID). Confirm with enter.

Select calibration method“Product calibration-Sat”

Confirm with enter.

Calibration

Return

25.6°C

Module pH 2700Module Cond 7700

Info

80.3%Air

Module O2 4700 ppb

Module O2 4700 ppb

Return

25.6°C

Product calibration ConcData entryZero correction

80.3%Air

Automatic - WaterAutomatic - AirProduct calibration Sat

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 36

Page 39: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

37

Menu Display Product calibration: Saturation

Product calibration SatProduct calibration is performedin 2 steps.Prepare reference measurement (e.g. with portable meter),start with softkey or enter

Step 2Lab value has been measured.When you open the Product calibra-tion menu again, the display shownon the left appears:Enter reference value (“Lab value”).Confirm with “OK”.

Step 1Take sample.Store measured value and tempera-ture at the moment of sampling(“Save” softkey or enter)Press meas to return to measure-ment.

Exception:Sample value can be measured onthe site and be entered immediately.To do so, press “Input” softkey.

Product calibration Sat

Cal medium: product

25.6 °C80.3%Air

Cal by taking sampleand entering saturation

StartReturn

Product calibration Sat

Step 1: Sampling‘Save’ the sample value

25.6 °C80.3%Air

‘Input’ lab valueSaturation 80.3%AirPressure 1014mbarTemperature +25.6°C

SaveInput

Product calibration Sat

Step 2: Lab valueInput sample lab value

25.6 °C80.3%Air

When changing sensors performFirst cal for statistics!

Lab value 80.0%Air

Abort OK

Calibration data record

Cal mode

25.6 °C80.3%Air

Slope

Product calibration

+059.3 nA

End Adjust

AdjustmentPress “Adjust” to take over thevalues determined during calibrationfor calculating the measured vari-ables. See Page 42.

Calibration

Zero point

09/12/05 15:20

+0.030 nA

Sensor replacement

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 37

Page 40: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

38 O2 4700(X) ppb

Calibration / adjustmentProduct calibration: Concentration

Product calibration: Concentration(Calibration with sampling)When the sensor cannot be removed – e.g. for sterility reasons – its slopecan be determined with “sampling”. To do so, the currently measuredconcentration value of the process is stored by the measuring module.Directly afterwards, a reference value is determined using a portable meter,for example. The reference value is entered into the measuring system. From the difference between measured value and reference value, theanalyzer calculates the sensor slope. With low concentration values, the analyzer corrects the zero point, with high concentrations the slope.During calibration the analyzer is in function check mode.Current outputs and relay contacts behave as configured (BASE, Out, PIDmodules).Caution!The reference value must be measured at temperature and pressure condi-tions similar to those of the process.

Menu DisplayProduct calibration:Concentration

Select module: O2 4700 ppbThe analyzer is in function checkmode. Current outputs and relaycontacts behave as configured (BASE,Out, PID). Confirm with enter.

Select calibration method“Product calibration Conc”

Confirm with enter.

Calibration

Return

25.6°C

Module pH 2700Module Cond 7700

Info

80.3%Air

Module O2 4700 ppb

Module O2 4700 ppb

Return

25.6°C

Data entryZero correction

80.3%Air

Automatic - WaterAutomatic - AirProduct calibration SatProduct calibration Conc

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 38

Page 41: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

39

Menu Display Product calibration:Concentration

Product calibration ConcProduct calibration is performedin 2 steps.Prepare reference measurement (e.g. with portable meter),start with softkey or enter

Step 1Take sample.Store measured value and tempera-ture at the moment of sampling(“Save” softkey or enter)Press meas to return to measure-ment.

Exception:Sample value can be measured onthe site and be entered immediately.To do so, press “Input” softkey.

Product calibration Conc

Cal medium: product

25.6 °C80.3%Air

Calibration by samplingand entering concentration

StartReturn

Product calibration Conc

Step 1: Sampling‘Save’ the sample value

25.6 °C80.3%Air

‘Input’ lab valueConcentration 0014µg/lPressure 1013mbarTemperature +25.6°C

SaveInput

Step 2Enter reference value (“Lab value”).When you open the Product calibra-tion menu again, the display shownon the left appears: Enter reference value (“Lab value”).Confirm with “OK”.

Product calibration Conc

Step 2: Lab valueInput sample lab value

25.6 °C80.3%Air

When changing sensors performFirst cal for statistics!

Return

Lab value 0017µg/l

OK

Sensor replacement

Calibration data record

Cal mode

25.6 °C80.3%Air

Slope

Product calibration

+059.3 nA

End Adjust

AdjustmentPress “Adjust” to take over thevalues determined during calibrationfor calculating the measured vari-ables. See Page 42.

Calibration

Zero point

09/12/05 15:20

+0.030 nA

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 39

Page 42: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

40 O2 4700(X) ppb

Calibration / adjustmentData entry of premeasured sensors

Data entry of premeasured sensorsEntry of values for slope and zero point of a sensor, related to 25 °C, 1013 mbars.During calibration the analyzer is in function check mode.Current outputs and relay contacts behave as configured (BASE, Out, PIDmodules).

Slope = Sensor current at 100 % atmospheric oxygen, 25 °C, 1013 mbars

Menu DisplayData entry ofpremeasured sensors

Select module: O2 4700 ppbThe analyzer is in function checkmode. Current outputs and relaycontacts behave as configured (BASE,Out, PID). Confirm with enter.

Select calibration method“Data entry”

Confirm with enter.

Enter the values for• Slope• Zero pointof premeasured sensorConfirm with “OK”.

Calibration

Return

25.6°C

Module pH 2700Module Cond 7700

Info

80.3%Air

Module O2 4700 ppb

Module O2 4700 ppb

Return

25.6°C

Zero correction

80.3%Air

Automatic - WaterAutomatic - AirProduct calibration SatProduct calibration ConcData entry

Data entry

Slope converted to100% O2-Air/25°C/1013mbar

25.6 °C80.3%Air

Zero point

Return

+0.000nA+050.0nASlope

OK

Sensor replacement

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 40

Page 43: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

41

Zero correctionThe InPro6900 sensors have a very low zero current. Therefore, a zero cali-bration is only required for measurement of oxygen traces.If a zero point correction is performed, the sensor should remain for at least10 to 30 minutes in the calibration medium in order to obtain stable, non-drifting values. During zero point correction, a drift check is not performed.Zero point current of a properly functioning sensor is notably less than 0.5 %of air current.

Calibration / adjustmentZero correction

Menu Display Zero correction

Select module: O2 4700 ppbThe analyzer is in function checkmode. Current outputs and relaycontacts behave as configured (BASE,Out, PID). Confirm with enter.

Zero point correction:Display of measured sensor current• Enter input current

for zero pointConfirm with “OK”.Adjustment: see Page 42

Select calibration method“Zero correction”

Confirm with enter.

Calibration

Return

25.6°C

Module pH 2700Module Cond 7700

Info

80.3%Air

Module O2 4700 ppb

Module O2 4700 ppb

Return

25.6°C-80.3%Air

Automatic - WaterAutomatic - AirProduct calibration SatProduct calibration ConcData entryZero correction

Zero correction

Direct displacementof sensor zero point

25.6 °C80.3%Air

Measured sensor current

Abort

-0.009nA

Zero point +0.000nA

OK

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 41

Page 44: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

Operator (without Administrator rights)

After calibration, change to measur-ing mode. Inform Administrator.When opening the menu(Calibration, respective module), theAdministrator sees all data of the lastcalibration and can take over thevalues or perform a new calibration.

Calibration data record

Cal mode

25.6 °C80.3%Air

Slope

Product calibration

+059.3 nA

End Adjust

Calibration

Zero point

09/12/05 12:44

+0.030 nA

42 O2 4700(X) ppb

AdjustmentTaking over the determined calibration parameters for calculating the measured value

Adjustmentmeans that the values determined by a calibration are taken over.The values determined for zero and slope are entered in the calibrationrecord (Cal record can be called up in the Diagnostics menu for the module). These values are only effective for calculating the measured variables whenthe calibration has been terminated with an adjustment.A passcode ensures that an adjustment can only be performed by anauthorized person (Administrator).The Operator can check the current sensor data by a calibration and informthe Administrator when there are deviations.You can use the additional function SW 700-107 for granting access rights(passcodes) and for Audit Trail (data recording and backup according toFDA 21 CFR Part 11).

Menu Display Adjustment after calibration

AdministratorWith the corresponding access rights,the device can immediately beadjusted after calibration. Thecalibration values are taken over forcalculating the measured variables.

Module O2 4700 ppb

Return

25.6°C

View/adjust calibration data record

80.3%Air

Start new calibration

Stored calibration data recordCalibration 09/12/05 12:44

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 42

Page 45: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

43

MaintenanceSensor monitor, temperature probe adjustmentNote: Function check active (Parameter setting: BASE, Out, PID modules)

Menu Display Maintenance

Call up MaintenanceFrom the measuring mode:Press menu key to select menu.Select Maintenance (maint) using arrowkeys, confirm with enter.Passcode 2958 (To change passcode,select: Parameter setting/System control/Passcode entry) Then select “Module O2 4700 ppb”.

Temp probe adjustmentThis function allows you to compensatefor the individual temperature probetolerance and the influence of the leadresistances to increase accuracy of tem-perature measurement. Adjustmentmay only be carried out when theprocess temperature is precisely mea-sured using a calibrated referencethermometer! The measurement errorof the reference thermometer should beless than 0.1 °C. Adjustment without precise measure-ment might result in considerable devia-tions of the measured value display!

Sensor monitorDuring maintenance, the sensor monitorallows validation of the sensor byimmersing it in a known solution, forexample, and checking the values mea-sured.

Sensor monitor

Return

25.6°C80.7%Air

Sensor currentSensor current (25°C)

Air pressureRTDTemperature

-60.2 nA-58.5 nA1014 mbars22.0 kΩ25.1 °C

Adjust temp probe

Probe tolerance and lead adjustmentEnter measured process temp

25.6°C16.4%Air

Installation adjustment

Abort

OffProcess temperature

On+025.0°C

OK

Menu selection

Return to meas

25.6°C

Select: [enter]

Lingua

80.7%Air

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 43

Page 46: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

44 O2 4700(X) ppb

Menu Display Viewing level, Operator level,Administrator level

Call up parameter settingFrom the measuring mode:Press menu key to select menu.Select parameter setting using arrowkeys, confirm with enter.

Administrator levelAccess to all functions, also passcodesetting.Releasing or blocking function foraccess from the Operator level.

Parameter setting: Operating levelsViewing level, Operator level, Administrator levelNote: Function check active (Parameter setting: BASE, Out, PID modules)

Functions which can be blocked forthe Operator level are marked withthe “lock” symbol.The functions are released orblocked using the softkey.

Operator levelAccess to all functions which havebeen released at the Administratorlevel. Blocked functions are displayedin gray and cannot be edited (Fig.).

Viewing levelDisplay of all settings.No editing possible!

Menu selection

Return to meas

25.6 °C83.2 %Air

Select: [enter]

Lingua

Parameter setting

Return

25.6 °C83.2 %Air

Viewing level

Administrator level (All Data) adm(Operation Data) opl(All Data) view

Operator level

Module FRONT M 700-011 (Administrator)

Return

25.6 °C83.2 %Air

LanguagesMeasurement display

KI recorder

English

Measurement recorder

Release

Module FRONT M 700-011 (Administrator)

Return

25.6 °C83.2 %Air

Measurement display

KI recorder

English

Measurement recorder

Languages

70321_E_4700xppb.QXD 10.10.2005 14:01 Uhr Seite 44

Page 47: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

45

Menu Display Administrator level:Enable / lock functions

Example: Blocking access to the calibrationadjustments (O2 4700 ppb module) from theOperator level

Call up parameter settingSelect Administrator level.Enter passcode (1989).Select “Module O2 4700 ppb”, confirm with enter.

Parameter setting: Lock functionsAdministrator level: Enable / lock functions for Operator levelNote: Function check active (Parameter setting: BASE, Out, PID modules)

Select “Cal preset values” using arrow keys.“Block” with softkey.

Now, the “Cal preset values” line ismarked with the “lock” icon. Thisfunction cannot be accessed from theOperator level any more. The softkeyfunction changes to “Release”.

Call up parameter settingSelect Operator level, passcode (1246).Select “Module O2 4700 ppb”. Now, the locked function is displayedin gray and marked with the “lock”icon.

Module O2 4700 ppb (Administrator)25.0°C

Return

Sensor dataCal preset values

16.7%Air

Salt correctionMessages

Pressure correction

Input filter

Block

Parameter setting (Administrator)25.0°C

Return

Module FRONT M 700-011Module BASE M 700-021

16.5%Air

System control

Module pH 2700Module Cond 7700

Module O2 4700 ppb

Module O2 4700 ppb (Administrator)25.0°C

Return

Sensor dataCal preset values

16.7%Air

Salt correctionMessages

Pressure correction

Input filter

Release

Module O2 4700 ppb25.0°C

Return

Sensor dataCal preset values

16.7%Air

Salt correctionMessages

Pressure correction

Input filter

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 45

Page 48: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

46 O2 4700(X) ppb

Setting the module parametersCall up parameter settingNote: Function check active (Parameter setting: BASE, Out, PID modules)

Menu Display Parameter setting

Call up parameter settingFrom the measuring mode:Press menu key to select menu.Select parameter setting using arrowkeys, confirm with enter. Passcode 1989

Select “Module O2 4700 ppb”.Confirm with enter.

Select parameter using arrow keys,confirm with enter.Module O2 4700 (Administrator)

25.0°C

Return

Sensor dataCal preset values

16.5%Air

Salt correctionMessages

Pressure correction

Input filter

Block

Parameter setting (Administrator)25.0°C

Return

Module FRONT M 700-011Module BASE M 700-021

16.5%Air

System control

Module pH 2700Module Cond 7700

Module O2 4700 ppb

Menu selection

Return to meas

25.0°C

Select: [enter]

Lingua

16.5%Air

During parameter setting the analyzer is in function checkmode. Current outputs and relay contacts behave as configured (BASE, Out, PID).

Parameter tables (Excel)2 complete parameter sets can be stored in the Premium TransmitterM 700(X). You can document the parameter settings of your complete point of measurement in an Excel table that can be downloaded from ourwebsite: www.mtpro.com

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 46

Page 49: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

Liquids, Gases (Vol%), Gases (ppm)A Standard, B Trace sensor (with guard) *, C Trace sensor (without guard), OthersNTC 30 kOhm, NTC 22 kOhmWith guard, Without guardOff, Onxxxx mV (entry)Off, failure, maintenance request

47

Parameter Default Selection / Range

Off, OnInput filter• Pulse suppression Off

Sensor data• Measure in• Sensor type

• Temperature probe **

• Sensor **

• Reference electrode **

• Polarization voltage• Sensocheck

LiquidsA Standard

NTC 22kOhmsWithout guardOff0675 mVOff

%Airµg/l, mg/l, ppb, ppmxxxx h (entry)

Cal preset values• Cal saturation• Cal concentration• Calibration timer

%AIRmg/l0000 h

Auto, Manual (default value 1013 mbars)Auto, Manual (default value 1013 mbars)

Pressure correction• Pressure during meas• Pressure during cal

AutoAuto

Salinity, Chlorinity, Conductivity(00.00 g/kg or 0.000 µS/cm, depending onselection)

Salt correction• Input Salinity

* Sensocheck not possible for trace sensor with guard, therefore disabled** Can only be set with “Sensor type Others” selected

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 47

Page 50: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

48 O2 4700(X) ppb

Device limits• Device limits max.: Maximum measurement range of device• Variable limits: Range limits specified

Variable limits

Device limits max.

Parameter settingMessages: Default settings and selection rangeNote: Function check active

Off, device limits max., variable limits*Off, device limits max., variable limits*Off, device limits max., variable limits*Off, device limits max., variable limits*Off, device limits max., variable limits*Off, device limits max., variable limits*

*) With “Variable limits” selected, the following parameters can be edited:

• Failure Limit Lo• Warning Limit Lo• Warning Limit Hi• Failure Limit Hi

Messages• Saturation %SAT• Saturation %O2• Concentration• Partial pressure• Temperature• Air pressure

Parameter Default Selection / Range

Limits maxOffOffOffLimits maxOff

0.000 %Air

199.9 %Air

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 48

Page 51: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

49

Messages

Parameter settingMessagesNote: Function check active

Menu Display

MessagesAll parameters determined by themeasuring module can generatemessages.• Device limits max.:Messages are generated when theprocess variable (e.g. %Air) is out-side the measurement range. The“Failure” icon is displayed, theNAMUR failure contact is activated(BASE module, factory setting:contact K4, N/C contact). Thecurrent outputs can signal a 22 mA message (user defined).

• Variable limits:For the “failure” and “warning”messages you can define upper andlower limits for message generation.

• Message icons:

Messages (Administrator)22.3 °C

Return

16.5%Air

Messages %SAT (Administrator)22.3 °C

Abort

Monitoring

16.5%Air

Variable limits

OK

Device limits max.Off

Messages Partial pressure

Messages % Air

Messages % SAT

Messages Temperature

Message Concentration

Messages (Administrator)20.1 °C16.5%Air

Warning Limit Hi

MonitoringFailure Limit Lo

Failure Limit Hi

Warning Limit Lo280%Air

Variable limits

290%Air

230%Air200%Air

Abort OK

Message list22.3 °C

Fail ppm rangeFail Meas. processing

16.5%Air

Fail Temperature range

Abort

Diagnostics menuIf the “Maintenance” or “Failure”icons are flashing in the display, youshould call up the Diagnostics menu.The messages are displayed in the“Message list”.

Failure (Failure limit Hi/Lo)

Maintenance (Warning limit Hi/Lo)

Message Air pressure

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 49

Page 52: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

50 O2 4700(X) ppb

Factory setting, logbookParameter setting/System control/LogbookNote: Function check active

Menu Display Logbook, factory setting

LogbookSelect which messages are to belogged in the logbook. The last 50events are recorded with date andtime. This permits quality managementdocumentation to ISO 9000.

The logbook can be called up fromthe diagnostics menu (Fig.).

Additional function SW 700-104:Extended logbook for recording dataon SmartMedia card (TAN).

Logbook (Administrator)24.8 °C83.0 %Air

Return

Erase logbook

NoLog failure

Log warning

Yes

NoYes

Logbook

Return

24.8 °C83.0 %Air

13/12/04 09:50 Measurement active

12/12/04 17:52 Measurement active13/12/04 17:44 Parameter setting active

13/12/04 09:36 Parameter setting active

12/12/04 17:04 Measurement active12/12/04 17:40 Wrong passcode

12/12/04 16:53 Diagnostics active

Factory settingAllows resetting the parameters totheir factory setting. When thismenu is opened, the analyzer dis-plays a warning (Fig.).

Factory setting (Administrator)24.8 °C83.0 %Air

Return

The factory setting erasesall your set parameters!

Set factory setting NoYes

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 50

Page 53: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

51

Current outputs, contacts, OK inputsSelect menu: Parameter setting/M 700 BASENote: Function check active

Menu Display Parameter setting M 700 BASE

Configure current output• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Output current ...”

• Select measured variable

• Select Characteristic, e.g. “Linear”: The measured variable is representedby a linear output current curve. The desired range of the measuredvariable is specified by the values for“Start” and “End”.Minimum span: Pg 75

BASE module (Administrator)19.2°C

Return

Contact K2 (NAMUR Function check)

Contact K4 (NAMUR Failure)Contact K3 (NAMUR Maintenance)

16.5%Air

Output current I2

Contact K1 (Limit)Inputs OK1, OK2

Block

Output current I1 (Administrator)

OK

19.0 °C16.5%Air

Measured variable

Output

Abort

Characteristic Linear°C

TrilinearFunctionTable

Output current I1 (Administrator)

OK

19.0 °C16.5%Air

CharacteristicOutput

Behavior during messages

Start

Abort

Measured variable

°C%O2S/cm°Cg/kg

Off%Air

EndStart

End

Output filter

Output current

[%Air]

20

49080

Assignment of measured values: Start (4 mA) and end (20 mA)

[%Air]

20

48785

[mA]

Example 1: Range %Air 80 ... 90

Example 2: Range %Air 85 ... 87 Advantage: Higher resolution in range of interest

[mA]

85 87

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 51

Page 54: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

52 O2 4700(X) ppb

[%]

100

Start End

0

Output currentY

XMeasured variable

[%]

100

Start End

0

Output currentY

XMeasured variable

2nd corner Y

1st corner Y

2nd corner X

1st corner X

• Linear characteristicThe measured variable is represented by a linear output current curve.

• Note: Bilinear characteristicFor a bilinear characteristic, identical parameters are entered for the two corner points (1st corner, 2nd corner).

• Trilinear characteristicTwo additional corner points must be entered:

Current outputs: Current output curvesSelect menu: Parameter setting/M 700 BASE

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 52

Page 55: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

53

[%]

Entry for “50 % point”:e.g.: 10 % of measured value

End

0

Output currentY

XMeasured variable

• Function characteristicNonlinear output current characteristic: allows measurements over several decades, e.g. for measuring very low values with a high resolutionand high values with a low resolution.Required: Entering a value for 50 % output current.

100

50

Equation

Output current (4 to 20 mA) = 16 mA + 4 mA(1+K)x

1+KxE + I - 2 * X50%

X50% - IK =

M - I

E - Ix =

I: Initial value at 4 mAX50%: 50% value at 12 mA (output current range 4 to 20 mA)E: End value at 20 mAM: Measured value

Logarithmic output curve over one decade: I: 10 % of maximum valueX50%: 31.6 % of maximum valueE: Maximum value

Logarithmic output curve over two decades I: 1 % of maximum valueX50%: 10 % of maximum valueE: Maximum value

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 53

Page 56: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

54 O2 4700(X) ppb

Output filterTime constant.

Time constant of output filterTo smoothen the current output, a low-pass filter with adjustable timeconstant can be switched on. When there is a jump at the input (100 %),the output level is at 63 % after the time constant has been reached.The time constant can be set from 0 to 120 sec. If the time constant is set to 0 s, the current output follows the input.

Note:The filter only acts on the current output and the current value of thesecondary display, not on the measurement display, the limit values, or thecontroller!

Time constant 0 to 120 sec

Measured variable 0 (4) - 20 mA

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 54

Page 57: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

55

NAMUR signals: Current outputsBehavior during messages: Function check, 22 mA signal

Behavior during messages

Output current[mA] “Fixed 22 mA”

Function check

22

4

“Last usable value”

Function check

Depending on the parameter setting(“Messages”), the current outputs switch to:• Currently measured value• Last measured value (HOLD function)• Fixed value (22 mA)

In the case of a fault a 22 mA signal can begenerated for the selected process variable (1st primary value).

Function check

“Current meas.”

Behavior during messages

OK

19.0 °C83.1 % Air

22mA message

Abort

Function check

Fixed 22 mA

Current meas.Last usable value

Message when the current range is exceededAs delivered, the “Maintenance request” (Warn) message is generated whenthe current range is exceeded (< 3.8 mA or > 20.5 mA).This setting can be changed in the Parameter setting menu of the respectivemeasuring module at “Messages”.

To generate a “Failure” message, the limit value monitoring must be set to“Variable limits”:Parameter setting - <measuring module> - Messages - Variable limits -Failure limit ...

Enter the same values for the failure limits as for the current output:Parameter setting - Module BASE - Output current - Variable Start / End.

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 55

Page 58: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

56 O2 4700(X) ppb

NAMUR signals: Relay contactsFailure, maintenance request, function check

Failure is active when a value has exceeded (or fallen below, resp.) a preset ”Failure Limit Hi”or ”Failure Limit Lo”, when the measured value is out of range or in theevent of other failure messages. That means that the equipment no longeroperates properly or that process parameters have reached a critical value.Failure is disabled during “Function check”.

Maintenance request is activewhen a value has exceeded (or fallen below, resp.) a preset “Warning LimitHi” or “Warning Limit Lo”, or when other warning messages have been acti-vated. That means that the equipment is still operating properly but shouldbe serviced, or that process parameters have reached a value requiring inter-vention.Warning is disabled during “Function check”.

Function check is active: • during calibration• during maintenance (current source, meas. point maintenance)• during parameter setting at the Operator level and the Administrator level• during an automatic rinsing cycle.

As delivered, the floating relay outputs of the M 700 BASE are assigned tothe NAMUR signals:• Failure: Contact K4, normally closed contact (signaling current failure)• Maintenance request: Contact K3, normally open contact• Function check: Contact K2, normally open contact

NAMUR signals: Factory setting of contacts• Select parameter setting• Select Administrator level• Select M 700 BASE (Fig.)You can define a delay time for “Maintenance request”and “Failure”, resp. If an alarm message is released, the contact will only beactivated after expiry of this delay time.

M odule BASE (Administrator)Output current I1

Return Block

83.1 % Air

Contact K4Output current I2

Contact K3

19.0 °C

(NAMUR Failure)

Contact K2Contact K1

(NAMUR maintenance)(NAMUR function check)(Limit)

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 56

Page 59: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

57

Relay contacts: Protective wiring

1

23

1 2

3

Typical AC applicationswith inductive load

1 Load2 RC combination, e.g. RIFA PMR 209

Typical RC combinationse.g.:Capacitor 0.1 µF, Resistor 100 Ohms / 1 W

3 Contact

Warning!Make sure that the maximum ratings of the relay contacts arenot exceeded even during switching!

Protective wiring of relay contacts

Relay contacts are subjected to electrical erosion. Especially with inductiveand capacitive loads, the service life of the contacts will be reduced. For suppression of sparks and arcing, components such as RC combinations,nonlinear resistors, series resistors and diodes should be used.

Information concerning relay contactsAs delivered, the relay contacts are suitable for low signal currents (down toapprox. 1mA). If currents above approx. 100 mA are switched, the goldplating is destroyed during the switching process. After that, the contactswill not reliably switch low currents.

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 57

Page 60: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

58 O2 4700(X) ppb

Relay contactsParameter setting/M 700 BASE/Relay contacts

The M 700 BASE provides 4 relay contacts(max. AC/DC rating 30 V / 3 A each).Contact K4 is provided for failure message.The switching behavior (normally open ornormally closed), as well as a switch-on orswitch-off delay can be defined.

Default settings of the user-definable relaycontacts of the M 700 BASE:K3: NAMUR maintenance requestK2: NAMUR function checkK1: Limit value

The contact assignment K1 - K3 is user defined(“Usage”):• NAMUR maintenance request• NAMUR function check• Limit value• Rinse contact• Parameter set 2 active• USP output (Cond module only)

Menu Display Setting the relay contacts

Configure current output• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Output current I1”

Relay contacts, usage• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Contact ...”• “Usage” (Fig.)

M 700 BASE

Alarm (K4)

K1

K2

K3

Contact K1 (Administrator)19.2°C

Return

Hysteresis

Measured variablLimit value

83.1 % Air

Usage

Effective directionContact type

Limit value

NAMUR maintenanceNAMUR function check

Rinse contactParameter set 2 activeUSP output

Contact assignment: See terminal plate M 700 BASE

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 58

Page 61: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

59

Limit value, hysteresis, contact typeParameter setting/M 700 BASE/Relay contacts/Usage

Menu Display Usage as limit value

Configure current output• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Output current I1”

Relay output: Limit value• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Contact ...”• “Usage: Limit value” (Fig.)

Contact K1 (Administrator)19.2°C

Return

Hysteresis

Measured variablLimit value

83.1 % Air

Usage

Effective directionContact type

Limit value

NAMUR maintenanceNAMUR function check

Rinse contactParameter set 2 activeUSP output

Limit valueEffective direction min

Hysteresis +

Limit value

Contact

0

1

Signal

Limit valueEffective direction max

Hysteresis -

Signal

Limit value

Contact

0

1

HysteresisTolerance band around the limit value, within which the contact is not actu-ated. Serves to obtain appropriate switching behavior at the output and suppress slight fluctuations of the measured variable (Fig.)

Contact typeSpecifies whether the active contact is closed (N/O) or open (N/C).

Icons in the measurement display:Measured value falls below limit:Measured value exceeds limit:

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 59

Page 62: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

60 O2 4700(X) ppb

OK1, OK2 inputs: Specify levelParameter setting/M 700 BASE/Inputs OK1, OK2Note: Function check active

Menu Display Setting the OK inputs

Configure current output• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Output current I1”

OK1 usage• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Inputs OK1/OK2”• Select “OK1 usage”

Configure current output• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Output current I1”

OK1/OK2 switching level• Call up parameter setting• Enter passcode• Select M 700 BASE• Select “Inputs OK1/OK2”• Specify active switching level

The M 700 BASE provides 2 digital inputs(OK1, OK2). The following functions (depend-ing on the parameter setting) can be startedvia a control signal:

• OK1: “Off” or “Function check”• OK2: System control menu /

Function control matrix. (“Off”, “Parameter set A/B”, “Start KI recorder”)

The switching level for the control signal must be specified:(active 10...30 V or active > 2 V).

M 700 BASE

OK1

OK2

OK1

OK2

OK

Inputs OK1, OK2 (Administrator)19.2°C

Return

OK2 usage

83.1 % Air

OK1 usage Off

Function check

For OK2 usage see“Function control matrix”

Inputs OK1, OK2 (Administrator)19.2°C

Return

OK1 usage

83.1 % Air

Input OK1 active 10 ... 30 V

active < 2 V

For OK2 usage see“Function control matrix”

Function check

Contact assignment:See terminal plate M 700 BASE

Input OK2

OK

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 60

Page 63: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

61

Switching parameter sets via OK2Parameter setting / System control / Function control matrixNote: Function check active

Parameter sets2 complete parameter sets (A, B) can be stored in the M 700. You can switch between the parameter sets using the OK2 input. The currently activated set can be signaled by a relay contact . An icon in themeasurement display shows which parameter set is active: or

Menu Display Parameter sets

Signal active parameter setvia relay contact• Call up parameter setting• M 700 BASE• Select contact• Usage: “Parameter set ...”.

Select parameter set (A, B) via input OK2• Call up parameter setting• System control• Function control matrix• Select “OK2”• Connect “Parameter set A/B”

Contact K3 (Administrator)

Abort

24.8 °C

Contact type

OK

NAMUR maintenanceUsage

ON delayNAMUR function checkLimit valueRinse contactParameter set B activeUSP output

83.1 % Air

OFF delay

NoteThe selection has no effect when working on SmartMedia card with SW 700-102.

24.8 °C

Left softkey

83.1 % Air

Return Connect

Function control matrix (Administrator)

Input OK2

Right softkeyProfibus DO 2

ParSet KI rec. Fav–

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 61

Page 64: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

O2 4700(X) ppb62

Calculation BlocksSelect menu: Parameter setting/System control/Calculation BlocksCalculation of new variables from measured variables

Calculation BlocksTwo measuring modules with all their measured values serve as input for thecalculation block. In addition, the general device status (NAMUR signals) istaken into account. The analyzer calculates the differences between theexisting values.These output variables are then available in the system and can be assignedto the outputs (current, limit values, display ...)

Module I

e.g.:OXY module

Inputs:Sensors, terminals

OutputsMeas. values / Status

Oxy

Temp

%Air%O2ppm°Cg/lmbarnA

Measuring module

Module I + II

OXY / OXY

Inputs:Outputs of 2 measuring modules

OutputsMeas. values / Status

Module

I

%Air diff%O2 diff°C diff

CalculationBlock

Module

II

Functionality of measuring module

Functionality of Calculation Block

g/l diffppm diffVol% diff

for mea-surementsin gases

°FVol%I input

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 62

Page 65: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

63

Parameter setting (Administrator)

Return

25.6 °C83.1 % Air

Module FRONT M 700-011

Module O2 4700 ppbModule O2 4700 ppb

Module BASE M 700-021

System control

Calc OXY/OXY

Combining measuring modulesWith three measuring modules the following Calculation Block combinationsare possible:

Up to two Calculation Blocks can be activated.All current outputs can be set to output the new process variables formed bythe Calculation Blocks.All new process variables can be displayed as primary or as secondary value.Controller functions are not supported.

I + II , I + III , II + III

Activating Calculation BlocksSelect menu: Parameter setting/System control/Calculation BlocksCombining measuring modules to Calculation Blocks

Menu Display Activate Calculation Blocks

Calculation Blocks• Call up parameter setting• System control• Select “Calculation Blocks”

• Depending on modules installed, the possible combinationsfor Calculation Blocks are offered

Return

25.6 °C83.1 % Air

Point of meas description

Software update

Factory setting

Release of options

Calculation Blocks

Logbook

System control (Administrator)

Return

25.6 °C83.1 % Air

Block 2Block 1

Calculation Blocks (Administrator)

OffCalc OXY/OXY

During parameter setting theCalculation Blocks are displayed likemodules.

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 63

Page 66: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

64 O2 4700(X) ppb

Configuring a Calculation BlockSelect menu: Parameter setting/System control/Select Calculation BlockSetting the process variable to be calculated

Parameter setting (Administrator)

Return

25.6 °C83.1 % Air

Module FRONT M 700-011

Module O2 4700 ppbModule O2 4700 ppb

Module BASE M 700-021

System control

Calc OXY/OXY

Menu Display Configuring a Calculation Block

Select Calculation Block• Call up parameter setting• System control• Select module

• Depending on modules installed, the possible combinationsfor Calculation Blocks are offered

MessagesYou can activate messages for theselected variables.

Variables which have been set as“Off” cannot be processed further.

The measured values which shallrelease a message are set using thearrow keys(left/right: select position, up/down:edit number). Confirm with enter.

Calc OXY/OXY (Administrator)

Return

25.6 °C83.1 % Air

Difference %O2

MessagesDifference °C

Difference %Air

Messages (Administrator)

Return

25.6 °C83.1 % Air

Messages %O2 diffMessages Temp diff

Messages %Air diff

Messages %Air diff (Administrator)

Return

25.6 °C83.1 % Air

Failure Limit Lo

Warning Limit HiWarning Limit Lo

Monitoring Variable limits

Failure Limit Hi

-5.0 % Air-2.0 % Air

+2.0 % Air+5.0 % Air

–Off

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 64

Page 67: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

65

Menu Display Diagnostics functions

Call up diagnosticsFrom the measuring mode:Press menu key to select menu.Select diagnostics using arrow keys,confirm with enter.

The “Diagnostics” menu gives anoverview of all functions available.Functions which have been set as“Favorite” can be directly accessedfrom the measuring mode (see Pg 27).

Diagnostics functionsGeneral status information of the measuring systemSelect menu: Diagnostics

Message listShows the currently activated warn-ing or failure messages in plain text.

LogbookShows the last 50 events with dateand time, e.g. calibrations, warningand failure messages, power failureetc. This permits quality managementdocumentation to ISO 9000.Extended logbook:SmartMedia card (SW 700-104)

Menu selection

Return to meas

83.1 % Air23.7 °C

Select: [enter]

Lingua

Diagnostics

Return

23.7 °C

Set favorite

83.1 % Air

Logbook

Message list

Device description

Point of meas. description

Module FRONTModule BASE

Message list

Return

23.7 °C83.1 % Air

WarnCurrent IV1 load

1 messg.

Logbook

Return

23.7 °C83.1 % Air

13/06/05 15:00 Diagnostics active

11/06/05 09:50 Parameter setting active11/06/05 10:05 Measurement active

10/06/05 14:33 Measurement active11/06/05 09:42 Wrong passcode

10/06/05 14:11 Parameter setting active

13/06/05 15:17 Measurement active

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 65

Page 68: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

66 O2 4700(X) ppb

Menu Display Diagnostics functions

Device descriptionSelect module using arrow keys:Provides information about all mod-ules installed: Function, serial num-ber, hardware and software versionand device options.

Module FRONTThe module the display and keypadcontrol. Test possibilities:• Module diagnostics• Display test• Keypad test

Module BASEThe module generates the standardoutput signals. Test possibilities:• Module diagnostics• Input/output status

Example:Module BASE, input/output status.

Device description

Return

22.7 °C83.1 % Air

Input for O2 and °CModule O2 4700 ppb

Hardware: 1, Software: 2.2Serial number: 0002483

BASEFRONT

Module FRONT

Return

22.7 °C83.1 % Air

Display testKeypad test

Module diagnostics

Module BASE

Return

22.7 °C83.1 % Air

Input/output status

Set favorite

Module diagnostics

Input/output status

Return

22.7 °C83.1 % Air

Current load I1

K1Current load I2 okContact

Input OK1

ok

Input OK2

K3InactiveInactive

K2K4

Options

Module

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 66

Page 69: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

67

Menu Display Module diagnostics, sensormonitor

Call up diagnosticsFrom the measuring mode:Press menu key to select menu.Select diagnostics using arrow keys,confirm with enter. Then select“Module O2 4700 ppb”.

The Diagnostics menu gives anoverview of all diagnostics functionsavailable. Messages set as “Favorite”can be called up directly from themeasuring mode using a softkey. To configure, select: Parameter setting / System control /Function control matrix.

DiagnosticsModule diagnostics, sensor monitor, cal record, sensor network diagram, statistics

Module diagnosticsFunction test of internal components:- Internal device communication- Check of firmware (module)- Factory settings, measured value

processing

Sensor monitorShows the current directly measuredby the sensor, the barometric pres-sure, and temperature. Importantfunction for diagnostics and valida-tion!

Menu selection

Return to meas

23.0 °C

Select: [enter]

Lingua

3.4 %O2

Module O2 4700 ppb

Return

23.0 °C3.4 %O2

Sensor monitorCal recordSensor network diagramStatistics

Set favorite

Module diagnostics

Module diagnostics

Return

23.0 °C3.4 %O2

CommunicationFlash check sumMeas. processing

okokok

Sensor monitor

Return

23.0 °C3.4 %O2

Sensor currentSensor current (25°C)Air pressure

08.03 nA08.03 nA1014 mbars

TemperatureImpedance (25°C)

+025.0 °C12.5 kΩ

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 67

Page 70: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

68 O2 4700(X) ppb

Menu DisplayCal record, sensor networkdiagram, statistics

Cal recordData of last adjustment/calibration,suitable for documentation toISO 9000 and GLP/GMP(Date, time, calibration method, sensor zero andslope, rel. humidity for calibration in air)

Sensor network diagram• Slope• Zero point• Sensocheck• Response time• Calibration timer

The measured values are continuous-ly monitored during the measure-ment process. The sensor networkdiagram provides at-a-glance infor-mation about critical parameters. If a tolerance limit has beenexceeded, the respective parameteris flashing.Values in gray: Monitoring switchedoff. To switch on, refer to “Info”.

StatisticsIndication of sensor data for the FirstCalibration and the last 3 calibra-tions. (Date and time of First Calibration, sensor zero andslope, temperature, pressure, response time)

Cal record

Return

23.0 °C3.4 %O2

Last calibration

Cal modeZero point +0.010 nASlope -050.0 nA

Return

2 - Zero point1 - Slope

4 - Response time3 - Sensocheck

5 - Cal timer

3.4 %O2

Statistics

Return

22.7 °C3.4 %O2

Zero point1st CalDiff

+0.000nA 07/01/05 10:20+0.010nA 07/11/05 12:34

DiffDiff

–0.020nA 07/12/05 11:35+0.900nA 07/18/05 10:42

Slope

Critical range“inner circle”Value out of tolerance

“Outer circle”Value within tolerance

Info

Sensor network diagram22.8 °C

Automatic - Water

07/04/05 10:29

00150313Serial numberSensor model InPro6900

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 68

Page 71: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

69

SensofaceGraphical indication of sensor conditionSensocheck must have been activated during parameter setting

Sensocheck – Sensor monitoringAutomatic monitoring of membrane and elec-trolyte

The “smileys” provide information on wear and required maintenance of thesensor (“friendly” - “neutral” - “sad”).

Menu Display Activate Sensocheck

Open menu selectionSelect parameter settingEnter passcode (Administrator)

Select “Module O2 4700 ppb”

Confirm with enter.

Select “Sensor data”.Confirm with enter.Then select “Sensocheck”(Fig.)Assign function and confirm withenter.

Menu selection

Return to meas

Select: [enter]

Lingua

3.4 %O225.6°C

25.6°C

Return

3.4 %O2

Parameter setting (Administrator)

System controlModule FRONT M 700-011Module BASE M 700-021

25.6°C

Return

3.4 %O2

Sensor data (Administrator)

Maintenance request

OffFailure

OK

16.924.0

%Air

°C10.3 mA 09.06.05

Sensor typeTemperature probeSensor

Sensocheck

A StandardNTC 22 kOhmWithout guard

Polarization voltagReference electrod

Module pH 2700Module O2 4700 ppb

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 69

Page 72: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

70 O2 4700(X) ppb

Sensor network diagram“Diagnostics / Module O2 4700 ppb/ Sensor network diagram”. Tolerance limit violations can be seen at aglance. Critical parameters are flashing.Parameters displayed in gray have beendisabled during parameter setting or do notapply to the currently selected sensor.

Sensoface is “sad” ...The “Diagnostics / Module O2 4700 ppb/ Sensor network diagram ” shows all current sensor parameters in a graphic diagram.

Sensoface criteria

Return Info

0.3*R or > 3.5*Rbut always R < 20 kΩ or > 4 MΩ

> 600 s

80 % elapsed

Parameter

Slope*

Zero

Sensocheck (Ref. impedance)

Response time

Cal timer

Type A sensor

< 30 nA or > 110 nA

< -0.6 nA or > 0.6 nA

Type B sensor

< 225 nA or > 525 nA

< -1 nA or. > 1 nA

* Value at oxygen saturation, 25°C and normal pressure 1013 mbars

Critical range

2 - Zero point1 - Slope

4 - Response time3 - Sensocheck

5 - Cal timer

3.4 %O2

Sensor network diagram22.8 °C

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 70

Page 73: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

71

SpecificationsO2 4700(X) ppb module

Specificationsl–––––––––––––––––––––––––––– l––––––––––––––––––––––––––––––––––––––––––––––

Oxy input Trace measurement with the Mettler-Toledo InPro 6900 (EEx ia IIC) sensorsMeasuring current 0 ... 600 nA, resolution 10 pASaturation (-10 ... 80 °C) 0.0 ... 199.9%Air

0.0 ... 29.9 %O2

Measurement error** < 0.5 % meas.val. + 0.1 %Concentration (-10 ... 80 °C) 0000 ... 9999 µg/l (overrange up to 19.99 mg/l)

0000 ... 9999 ppb (overrange up to 19.99 ppm)Measurement error** < 0.5 % meas.val. + 0.005 mg/l or 0.005 ppmPolarization voltage 0 ... –1000 mV, default –675 mV, Ri ≈ 10 kΩPartial pressure 0 ... 2000 mbarsAir pressure 700 ... 1100 mbars

Manual 0 ... 9999 mbarsSalt correction 0.0 ... 45.0 g/kgAdm. guard current ≤ 20 µARef voltage ±500 mV (voltage across ref connection and anode)l–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Measurement in gases 0 ... 2000 mbars0 ... 9999 ppm0.00 ... 29.9 / 30.0 .... 120.0 Vol% (display only)0.00 ... 120.0 Vol% (current, limit values)(1 Vol% = 10,000 ppm)

Current start / end As desired within rangeCalibration methods Automatic - Air

- with the following default settings:rH = 50 %, p 0 measured barometric pressure, calibration medium air (dry air = 20.95 Vol%)

Product calibration (select ppm or Vol%)Data entryZero correction

l–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Sensor monitoring * Sensocheck (disabled for sensors with guard)Monitoring of membrane and electrolyte

l–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Sensoface Provides information on the sensor condition:Sensor network diagram Zero, slope, response time, cal timer,

Sensocheck

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 71

Page 74: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

O2 4700(X) ppb72

l–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Sensor monitor Direct display of measured values from sensor for validation sensor current / barometric pressure / temperature

l–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Calibration * Operating modes- Automatic calibration in air-saturated water- Automatic calibration in air- Product calibration: Saturation- Product calibration: Concentration- Data entry zero/slope- Zero point correction

Calibration record/statistics Recording of:Zero, slope, response time, calibration method,with date and time of the last three calibrations and the First Calibration

ll–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Temperature input(EEx ia IIC)Temperature probe * NTC 22 kΩ / NTC 30 kΩ

2-wire connection, adjustableRange –20 ... +150 °CResolution 0.1 °CMeasurement error** 0.2 % meas.val. + 0.5 K

* User-defined** To IEC 746 Part 1, at nominal operating conditions, ± 1 count, plus sensor error

SpecificationsO2 4700(X) ppb module

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 72

Page 75: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

73

SpecificationsO2 4700(X) ppb module

General data l–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Explosion protection Rating plate: KEMA 04 ATEX 2056(Module O2 4700X ppb only) II 2 (1) GD EEx ib [ia] IIC T4l–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

EMC NAMUR NE 21 andEN 61326 VDE 0843 Part 20 /01.98EN 61326/A1 VDE 0843 Part 20/A1 /05.99

Emitted interference Class BImmunity to interference Industryl–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Lightning protection EN 61000-4-5, Installation Class 2l–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Nominal operating Ambient temperature -20 ... +55 °C (Ex: max. +50 °C)conditions Relative humidity 10 ... 95 % not condensingl–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Transport/ -20 ... +70 °CStorage temperaturel–––––––––––––––––––––– l–––––––––––––––––––––––––––––––––––––––––––––––––

Screw clamp connection Single wires and flexible leads up to 2.5 mm2 (AWG 14)

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 73

Page 76: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

74 O2 4700(X) ppb

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 74

Page 77: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

75

Appendix:Minimum measuring spans for current outputs

The O2 4700(X) ppb module is a measuring module. It does not provide cur-rent outputs. Current outputs are provided by the M 700 BASE (basic device)or by communication modules (e.g. Out). The corresponding parametersmust be set there.

The minimum current span shall prevent that the resolution limit of themeasurement technology (± 1 count) is seen in the current.

O2 4700(X) ppb module%Air 10.0%O2 2.0°C 10.0mg/l 10 %ppm 10 %mbar 20.0 (partial pressure)Vol% 2.0ppm 1000°F 10.0

Calculation Block OXY/OXYDiff %Air 10.0Diff %O2 2.0Diff mg/l 10 % min. 20.0 µg/lDiff ppm 10 % min. 20.0 ppbDiff °C 10.0Diff Vol% 2.0Diff ppm 1000 (= 0.1 Vol%)

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 75

Page 78: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

76 O2 4700(X) ppb

Dissolved oxygen measurement in carbonated beverages(SW 700-011)Application-specific additional function for breweries

This additional function simplifies parameter setting since all steps notrequired for dissolved oxygen measurement in carbonated beverages areomitted. It simultaneously acts on all installed O2 modules (module software version 2.2 and higher).

Function principle:The following processes are automated by the additional function, i.e. allparameters required for the respective program step are set automatically.

During the filling process, for example, it must be ensured that as little oxy-gen as possible is dissolved in the beer to extent its shelf life.During oxygen trace measurement the sensor is operated with a very lowpolarization voltage (-500 mV). This results in low cross-sensitivity to CO2.

For a calibration in air, this polarization voltage is too low.It must be set to -675 mV and afterwards be reduced again to -500 mV formeasuring in the trace range.Be sure to wait long enough for the sensor to stabilize.

Opening and closing of valves causes pressure variations in the beer pipeswhich momentarily falsify the O2 signal. Therefore the input signal must beattenuated correspondingly to suppress transient interferences.

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 76

Page 79: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

77

Parameter settingActivated from measuring mode: Press menu key to selectmenu.Select parameter setting using arrow keys, confirm with enter.Administrator levelAccess to all functions, also passcode setting. Releasing or block-ing function for access from the Operator level.Operator levelAccess to all functions which have been released at theAdministrator level. Blocked functions are displayed in gray and cannot be edited.Viewing levelOnly display, no editing possible!

Select menu

Return to meas

Select: [enter]

Lingua

Parameter setting

Return

Viewing level

Administrator level (All Data) adm(Operation Data) opl(All Data) view

Operator level

System control

Parameter sets• Save• Load

Parameter setting menu

2 parameter sets (A,B) are available in the analyzer. The currently active parameter set is read on the display.Parameter sets contain all settings except:Sensor type, Options, System control settingsUp to 5 parameter sets (1, 2, 3, 4, 5) are available when a SmartMedia card (Option) is used.

Memory card (Option)• Register Logbook• Register recorder• Decimal separator• Card full• Formatting

Menu only appears with SmartMedia Card inserted. Make sure that it is a memory card,not an update card.Commercially available SmartMedia cards must be formattedbefore they can be used as memory card.

Copy configuration The complete configuration of an analyzer can be written on aSmartMedia card. This allows transferring all device settings - exceptoptions and passcodes - to other devices with identical equipment.

Function control matrix• Input OK2• Left softkey• Right softkey

Selecting the control element for the following functions:- Parameter set selection- KI recorder (Start/Stop)- Favorites menu (selected diagnostics functions)- EC 400 (fully automated probe controller)

Time/datePoint of meas descriptionRelease of optionsSoftware updateLogbookFactory settingPasscode entry

Selecting the display format, entryCan be called up in the diagnostics menu.A TAN is required to release an Option.Software update from SmartMedia card (update card)Selecting events to be recordedResetting all parameters to factory settingEditing the passcodes

25.6 °C83.2 %Air

25.6 °C83.2 %Air

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 77

Page 80: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

78 O2 4700(X) ppb

Display settings: M 700 FRONT

Measurement recorder• Time base• Zoom function• Min/Max display

Parameter setting menu

Option. 2-channel, selection of process variable, start and end

Languages

Measurement display• Main display• Display format• Viewing angle

Representation of measured values on the display:- Selecting the number of primary values displayed (one or two)- Decimal places

Output current I1, I2• Process variable• Characteristic• Output (0/4 - 20 mA)• Output filter• Behavior during messages

- Function check.--- Currently meas. value--- Last meas. value--- Fixed 22 mA

-22 mA message

2 current outputs, separately adjustable

KI recorder Option. See more detailed “Options” manual

Signal outputs and inputs, contacts: M 700 BASE

Contact K4• Contact type• ON delay• OFF delay

NAMUR Failure

Contacts K3, K2, K1• Usage

- Maintenance request- Function check

- Limit value (adjustable) - Rinse contact (adjustable)- Parameter set B active- USP output- KI recorder active

• Contact type• ON delay• OFF delay

Factory setting:K3: Maintenance request, K2: Function check, K1: Limit value

- Variable, limit value, hysteresis, effective direction, ...- Rinsing interval, lead times, rinse duration, logbook entry, ...

Inputs OK1, OK2• OK1 usage

- Signal level

Optocoupler - signal inputs Off, function checkactive level switchable from 10 to 30 V or < 2 V, resp.For OK2 see System control/Function control matrix

Output current[mA]

“Fixed22 mA”

Function check

22

4

“Last usablevalue”

Function check

Function check

“Currentmeas.”

Behavior during messages

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 78

Page 81: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

79

O2 4700(X) ppb module

Parameter setting menu

Input filter

Sensor data• Sensor type• Temperature probe• Sensor• Reference electrode• Sensor polarization• Polarization voltage• Sensoface• Details

- Slope- Zero- Sensocheck- Response time

Representation of measured values on the display:- Select- Selection for Measurement / Calibration

Cal preset values• Cal saturation• Cal concentration

- mg/l- µg/l- ppm- ppb

• Calibration timer

Salinity correction• Eintry

- Salinity- Chlorinity- Conductivity

• Salinity

Messages• Saturation %Air• Saturation %O2• Concentration• Partial pressure• Temperature• Air pressure

Pressure correction• Pressure during meas• Pressure during cal

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 79

Page 82: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

80 O2 4700(X) ppb

O2 4700(X) ppb module

Calibration menu

Automatic - WaterAutomatic - Air Productcalibration SatProduct calibration ConcData entryZero correction

Current source Output current definable 0 ... 22 mA

Message listPoint of meas descriptionLogbookDevice description

List of all warning and failure messages

Hardware version, Serial no., (Module) Firmware, Options

Maintenance menu

M 700 BASE

O2 4700(X) ppb module

Sensor monitorTemp probe adjustment

Sensor current, air pressure, RTD, temperature, impedanceCompensating for lead length

Diagnostics menu

M 700 FRONT

Module diagnosticsDisplay testKeypad test

M 700 BASE

Module diagnosticsInput/output status

O2 4700(X) ppb module

Module diagnosticsSensor monitorCal recordSensor network diagramOxyStatistics

Internal function testShows the values currently measured by the sensorData of last adjustment / calibrationGraphical representation of the sensor parametersDisplays first calibration and deviations of last 3 calibrations

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 80

Page 83: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

81

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 81

Page 84: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

AAdjustment 29, 42Administrator level 44Audit Trail log 8Automatic calibration in air 34Automatic calibration in water 32

BBehavior during messages 55

CCal preset values 47Cal record 68Calculation Blocks 62Calibration / adjustment 28Calibration methods 30Calibration with sampling 36, 38Configuring the module 46Conformity with FDA 21 CFR Part 11 8Connecting a sensor 20-21Contact type 59Current outputs: Characteristics 52Current outputs: Settings 51

DData entry of premeasured sensors 40Declaration of Conformity 3Device description 10, 66Device limits 48Diagnostics functions 65Disposal 2Dissolved oxygen measurement in carbonated beverages 76

EEC Declaration of Conformity 3Electronic Signature 8

Index

O2 4700(X) ppb82

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 82

Page 85: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

83

EMC 73Enable / lock functions 45Explosion protection 73

FFactory settings 50Failure 49, 56Favorites 26FDA 21 CFR Part 11 8First calibration 28Function check 56Function control matrix 26Functionality of Calculation Block 62

HHazardous-area application 9Hysteresis 59

IIcons 89InPro 6900, VP cable 20InPro 6800, VP cable 21Input filter 47Inserting the module 19Intended use 8

LLimit value 59Locking a function 45Logbook 50, 65

MM 700 BASE 15M 700 FRONT 12, 14Maintenance 43, 49Maintenance request 56

Index

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 83

Page 86: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

84

Measurement display settings 24Menu selection – quick access 90Menu selection 22Menu structure 13, 22Message icons 49Message list 49, 65Messages 48, 64Minimum measuring spans for current outputs 75Modular concept 11Module diagnostics 66-67

NNAMUR signals: Current outputs 55NAMUR signals: Relay contacts 56

OOK inputs 60OK1,OK2 inputs 60Operator level 44Output filter 54Overview of parameter setting 77

PParameter sets 61Parameter setting: Locking a function 45Parameter setting: Messages 48Parameter setting: Operating levels 44Parameter tables (Excel) 16Passcode entry 23Pressure correction 47Product calibration: Concentration 38Product calibration: Saturation 36

RRelay contacts: NAMUR signals 56Relay contacts: Protective wiring 57

Index

O2 4700(X) ppb

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 84

Page 87: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

85

Relay contacts: Usage 58Return of products 2

SSafety information 9Salinity correction 47Secondary displays 24, 26Sensocheck - Sensor monitoring 69Sensoface 69Sensoface criteria 70Sensor connection 20-21Sensor data 47Sensor monitor 43, 67Sensor network diagram 68, 70Sensor replacement (First Calibration) 28Setting diagnostics messages as favorite 26Short description 12Slope 40SmartMedia card 14-15Softkeys 24, 26Software version 10Specifications 71Statistics 68

TTemp probe adjustment 43Terminal plate 17Time constant of output filter 54Trace sensor 20Trademarks 2

VViewing level 44VP cable 20

Index

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 85

Page 88: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

86

WWarranty 2Wiring examples 20-21

ZZero point correction 41

Index

O2 4700(X) ppb

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 86

Page 89: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

87

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 87

Page 90: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

88 O2 4700(X) ppb

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 88

Page 91: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

89

Icon Explanation of icons important for this module

The analyzer is in measuring mode.

The analyzer is in calibration mode. Function check is active.

The analyzer is in maintenance mode. Function check is active.

The analyzer is in parameter setting mode. Function check is active.

The analyzer is in diagnostics mode.

Function check. The NAMUR “function check” contact is active (factory setting: BASEmodule, contact K2, N/O contact). Current outputs as configured: • Currently measured value: The currently measured value appears at the current output• Last measured value: The last measured value is held at the current output• Fix 22 mA: The output current is at 22 mA

Failure: The NAMUR “failure” contact is active (factory setting: BASE module, contactK4, N/C contact). To view error message, call up: Diagnostics menu/Message listMaintenance request. The NAMUR contact is active (factory setting: BASE module,contact K2, N/O contact). To view error message, call up: Diagnostics menu/Message list

NA

MU

R s

ign

als

Calibration is performed

Calibration - Step 1 of product calibration has been executed. The analyzer is waiting for the sample value

Temperature detection by manual input

In the plaintext display in front of a menu line:Access to next menu level with enter

In the plaintext display in front of a menu line when it has been blocked by theAdministrator against access from the Operator level.

Designates the module slot (1, 2 or 3), allowing the clear assignment of measured-value/parameter displays in the case of identical module types.

Indicates the active parameter set (The analyzer provides two parameter sets A and B. Up to 5 sets can be added using additional functions and SmartMedia card.)

man

Limit indication: Lower / upper range exceeded

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 89

Page 92: 70321 E 4700xppb.QXD 10.10.2005 14:00 Uhr Seite 1 O2 4700 ... · The module permits simultaneous measurement of oxygen partial pressure, barometric pressure and temperature. In addition,

Menu selectionO2 4700(X) ppb module

Calibration and adjustment.........................................28Automatic - Water................................................................32Automatic - Air.....................................................................34Product calibration: Saturation..............................................36Product calibration: Concentration........................................38Data entry ............................................................................40

Messages..............................................................................48

Maintenance ....................................................................43

Temp probe adjustment ........................................................43

Parameter setting...........................................................44Input filter ............................................................................47Sensor data ..........................................................................47Cal preset values ..................................................................47Pressure correction................................................................47Salt correction ......................................................................47

Diagnostics .......................................................................67Module diagnostics...............................................................67Sensor monitor .....................................................................67Cal record.............................................................................68Sensor network diagram.......................................................68Statistics ...............................................................................68

Sensor monitor .....................................................................43

Zero correction .....................................................................41Adjustment...........................................................................42

Set favorite ...........................................................................26

70321_E_4700xppb.QXD 10.10.2005 14:02 Uhr Seite 90