A Michelson-Morley Experiment for the Advanced Lab Charles Rogers and Richard Selv aggi Texas A&M Univ ersity - Commerce Abstract The famous 1887 Michelson-Morley experiment measured the propagation of light in a rotating interferometer in the physics department at Case Western Reserve University. For more than one hundred and twenty years, the data obtained from this experiment have been examined by many professionals interested in this experiment. These investigations have raised numerous questions concerning the analysis and interpretation of the originally published data. With interest in replicating this experiment, we have designed a new rotating interferometer of the same scale as the original experiment. This apparatus is composed on a standard four foot by six foot optical breadboard supported on a cylindrical float inside a concentric cylindrical tank. The interferometer consist of sixteen adjustable front surface mirrors, an aluminized beamsplitter, and a compensation plate arranged in the same geometry as used in the original experiment. The interferometer is rotated smoothly by the equivalent of a 2-meter diameter brushless motor. The light source is a helium-neon laser and a wireless optical sensor. Our students learn to align the optics, collect realtime data using a LabVIEW vi, and analyze their data using both original and current theories of operation of this instrument. We will present the details of this experiment and preliminary data collected. References 1. A. A. Michelson, “The Relative Motion of the Earth and the Luminiferous Ether”, Am. J. Sci. 22, 120 (1881). 2. A. A. Michelson and E. W. Morley, “On the Relative Motion of the Earth and the Luminiferous Ether”, Am. J. Sci. 34, 333-345 (1887). 3. D. C. Miller, “The Ether-Drift Experiment and the Determination of the Absolute Motion of the Earth”, Rev. Mod. Phys. 5, 203-242 (1933). 4. M. A. Handschy, “Re-examination of the 1887 Michelson-Morley experiment”, Am. J. Phys. 50, 987-990 (1982). 5. R. A. Schumacher, “Special relative and the Michelson-Morley interferometer”, Am J. Phys. 62, 609-612 (1994). 6. M. Sato, “Interperation of the slight periodic displacement in the Michelson-Morley experiments”, arXiv:physics/0605067, 7. R. T. Cahill and K. Kitto, “Michelson-Morley Experiments Revisited and the Cosmic Background Radiation Preferred Frame”, arXiv:physics/0205065v1, 23 May 2002. 8. H. Munera, “Observation of a Non-conventional Influence of the Earth's Motion on the Velocity of Photons, and Calculation of the Velocity of Our Galaxy”, PIERS Proceedings, 23 March 2009. 9. T. Roberts, “What is the experimental basis of Special Relativity Theory?”, Usenet Physics FAQ, October 2007. HeNe laser Light source CCD light sensor Beam splitter and one set of mirrors One set of mirrors with temperature sensor Fringe pattern at the CCD Adjusting the CCD position Optical breadboard balanced on the support Concentric water tanks and optical support Thermal shroud and half of the enclosure Calibration Methods 1. Michelson and Morley 2 2. Miller 3 3. Cahill and Kitto 7 t = Ln 3 v 2 c 3 t = 2 Ln n 2 1 v 2 c 3 cos 2 t = 2 k 2 L v p 2 c 3 cos A L L V = c/n CCD C B A 2 B L C v A 1 V = c/n L Introduction The Michelson-Morley experiment was designed to measure relative motion between the experimental apparatus and the “Luminiferous Ether”. The failure to measure the relative motion of the orbital or larger velocities has since perplexed the world’s scientific community. The reasons for replicating this experiment include confirmation of the original results with more accurate monitoring and measuring tools, to develop a easy to replicate experimental apparatus that can be produced in other advanced physics labs, to use new techniques such as a stepper motor to rotate the apparatus at different rotational velocities, to perform additional experimental applications including those recommended by Michelson and Morley, to confirm theoretical explanations of the measured results, to develop a better understanding of the factors affecting the oscillation of the interference pattern and the unidirectional drift of the interference fringes. Description of the replica M-M Interferometer Photographs taken during the assembly of the Michelson-Morley interferometer are shown in the lower left corner of this poster. An optical breadboard is supported on a wooden table within concentric cylindrical tanks. The optical components for the interferometer are positioned on the breadboard with a 12 meter path length for each arm. A stabilized HeNe laser is used for the light source and a CCD sensor is used to detect the interference pattern. Estimated mass of the final rotating platform is 800 kilograms. Fringe movement data is recorded as a time- lapse video file and is currently analyzed using Matlab scripts. A microcontroller with 10-bit ADC are used to digitize output from the temperature and pressure sensors and communicate these data. Another microcontroller provides timing and control of the platform rotation. We also record the air temperature on each arm of the interferometer and the ambient air temperature and atmospheric pressure. We plan to replace the temperature sensors with 14-bit temperature and relative humidity sensors. Summary and Conclusion Changes in atmospheric pressure appears to be the most important environmental variable affecting the interference pattern for this interferometer. Turning off the air conditioning fans within the laboratory reduced pressure fluctuations and improved the stability of the fringe pattern. Temperature of the air within the interferometer remained constant. The unidirectional drift is present in all previously described one-way and two-way light experiments. In both Michelson-Morley and Miller experiments the fringe drift was removed before analyzing the data. The drift of the interference fringes has been observed by others with considerable speculation as to their cause. We believe that corrections for environmental variations can be applied to the data to accurately remove this effect. Further experimental scenarios include the apparatus remaining stationary for twenty-four hours or rotating at one-quarter, one-half, one, and two revolutions per minute. The accuracy of the apparatus can be increased by lengthening the light’s pathway and by increasing the number of pixels per fringe. Reduction in the baseline oscillation artifacts can be achieved by placing the apparatus in a sealed remote subterranean room without heating or cooling needs. The Michelson-Morley experiment is one of the most important experiment in physics. It should be taught in the advanced laboratory. This experiment offers an excellent hands on experience with optical interferometers and various signal processing and laboratory instrumentation and automation techniques. Rotor magnets, two drive coils, and Hall sensor Aligning the magnetic drive Data acquisition display Michelson and Morley's 1887 Experiment 2 Data Collection The following figures show initial data collected from the stationary interferometer to survey environmental conditions within the laboratory. Date Time Temp1 Temp2 Temp3 Temp4 Temp5 Pressure Additional images of this apparatus and videos of the rotating MM interferometer replica and recorded fringe pattern movement will be shown at the conference. CCD

Upload: others

Post on 12-Aug-2020

13 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A Michelson-Morley Experiment for the Advanced Lab · A Michelson-Morley Experiment for the Advanced Lab Charles Rogers and Richard Selvaggi Texas A&M University - Commerce Abstract

A Michelson-Morley Experiment for the Advanced Lab Charles Rogers and Richard Selvaggi Texas A&M University - Commerce

AbstractThe famous 1887 Michelson-Morley experiment measured the propagation of light in a rotating interferometer in the physics department at Case Western Reserve University. For more than one hundred and twenty years, the data obtained from this experiment have been examined by many professionals interested in this experiment. These investigations have raised numerous questions concerning the analysis and interpretation of the originally published data. With interest in replicating this experiment, we have designed a new rotating interferometer of the same scale as the original experiment. This apparatus is composed on a standard four foot by six foot optical breadboard supported on a cylindrical float inside a concentric cylindrical tank. The interferometer consist of sixteen adjustable front surface mirrors, an aluminized beamsplitter, and a compensation plate arranged in the same geometry as used in the original experiment. The interferometer is rotated smoothly by the equivalent of a 2-meter diameter brushless motor. The light source is a helium-neon laser and a wireless optical sensor. Our students learn to align the optics, collect realtime data using a LabVIEW vi, and analyze their data using both original and current theories of operation of this instrument. We will present the details of this experiment and preliminary data collected.

References1. A. A. Michelson, “The Relative Motion of the Earth and the Luminiferous Ether”, Am. J. Sci. 22, 120 (1881).

2. A. A. Michelson and E. W. Morley, “On the Relative Motion of the Earth and the Luminiferous Ether”, Am. J. Sci. 34, 333-345 (1887).

3. D. C. Miller, “The Ether-Drift Experiment and the Determination of the Absolute Motion of the Earth”, Rev. Mod. Phys. 5, 203-242 (1933).

4. M. A. Handschy, “Re-examination of the 1887 Michelson-Morley experiment”, Am. J. Phys. 50, 987-990 (1982).

5. R. A. Schumacher, “Special relative and the Michelson-Morley interferometer”, Am J. Phys. 62, 609-612 (1994).

6. M. Sato, “Interperation of the slight periodic displacement in the Michelson-Morley experiments”, arXiv:physics/0605067,

7. R. T. Cahill and K. Kitto, “Michelson-Morley Experiments Revisited and the Cosmic Background Radiation Preferred Frame”, arXiv:physics/0205065v1, 23 May 2002.

8. H. Munera, “Observation of a Non-conventional Influence of the Earth's Motion on the Velocity of Photons, and Calculation of the Velocity of Our Galaxy”, PIERS Proceedings, 23 March 2009.

9. T. Roberts, “What is the experimental basis of Special Relativity Theory?”, Usenet Physics FAQ, October 2007.

HeNe laser Light source CCD light sensor

Beam splitter and one set of mirrors One set of mirrors with temperature sensor

Fringe pattern at the CCD Adjusting the CCD position

Optical breadboard balanced on the support

Concentric water tanks and optical support

Thermal shroud and half of the enclosure

Calibration Methods

1. Michelson and Morley 2

2. Miller 3

3. Cahill and Kitto 7

t=Ln3 v2

c3

t=2 Lnn2−1 v2

c3cos2

t=2 k2L∣v p2∣c3

cos−

A

L

L

V = c/n

CCD

C

B

A2

B

L C

v

A1

V = c/n

L

Introduction The Michelson-Morley experiment was designed to measure relative motion between the experimental apparatus and the “Luminiferous Ether”. The failure to measure the relative motion of the orbital or larger velocities has since perplexed the world’s scientific community. The reasons for replicating this experiment include confirmation of the original results with more accurate monitoring and measuring tools, to develop a easy to replicate experimental apparatus that can be produced in other advanced physics labs, to use new techniques such as a stepper motor to rotate the apparatus at different rotational velocities, to perform additional experimental applications including those recommended by Michelson and Morley, to confirm theoretical explanations of the measured results, to develop a better understanding of the factors affecting the oscillation of the interference pattern and the unidirectional drift of the interference fringes.

Description of the replica M-M Interferometer Photographs taken during the assembly of the Michelson-Morley interferometer are shown in the lower left corner of this poster. An optical breadboard is supported on a wooden table within concentric cylindrical tanks. The optical components for the interferometer are positioned on the breadboard with a 12 meter path length for each arm. A stabilized HeNe laser is used for the light source and a CCD sensor is used to detect the interference pattern. Estimated mass of the final rotating platform is 800 kilograms. Fringe movement data is recorded as a time-lapse video file and is currently analyzed using Matlab scripts. A microcontroller with 10-bit ADC are used to digitize output from the temperature and pressure sensors and communicate these data. Another microcontroller provides timing and control of the platform rotation. We also record the air temperature on each arm of the interferometer and the ambient air temperature and atmospheric pressure. We plan to replace the temperature sensors with 14-bit temperature and relative humidity sensors.

Summary and Conclusion Changes in atmospheric pressure appears to be the most important environmental variable affecting the interference pattern for this interferometer. Turning off the air conditioning fans within the laboratory reduced pressure fluctuations and improved the stability of the fringe pattern. Temperature of the air within the interferometer remained constant. The unidirectional drift is present in all previously described one-way and two-way light experiments. In both Michelson-Morley and Miller experiments the fringe drift was removed before analyzing the data. The drift of the interference fringes has been observed by others with considerable speculation as to their cause. We believe that corrections for environmental variations can be applied to the data to accurately remove this effect. Further experimental scenarios include the apparatus remaining stationary for twenty-four hours or rotating at one-quarter, one-half, one, and two revolutions per minute. The accuracy of the apparatus can be increased by lengthening the light’s pathway and by increasing the number of pixels per fringe. Reduction in the baseline oscillation artifacts can be achieved by placing the apparatus in a sealed remote subterranean room without heating or cooling needs. The Michelson-Morley experiment is one of the most important experiment in physics. It should be taught in the advanced laboratory. This experiment offers an excellent hands on experience with optical interferometers and various signal processing and laboratory instrumentation and automation techniques. Rotor magnets, two drive coils, and Hall sensor Aligning the magnetic drive Data acquisition display

Michelson andMorley's 1887Experiment 2

Data Collection The following figures show initial data collected from the stationary interferometer to survey environmental conditions within the laboratory.

Date Time Temp1 Temp2 Temp3 Temp4 Temp5 Pressure

Additional images of this apparatus and videos of the rotating MM interferometer replica and recorded fringe pattern movement will be shown at the conference.