al moretti , apc, fermilab

15
1 Al Moretti, APC, Fermilab MAP- Winter Meeting February 28 - March 4, 2011 TJNAF Newport News, VA

Upload: lela

Post on 15-Feb-2016

62 views

Category:

Documents


0 download

DESCRIPTION

Al Moretti , APC, Fermilab MAP- Winter Meeting February 28 - March 4, 2011 TJNAF Newport News, VA. Discussion of the Magnetically Insulated Cavity The orthogonal Box Cavity. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Al  Moretti , APC, Fermilab

1

Al Moretti, APC, Fermilab MAP- Winter Meeting February 28 - March 4, 2011 TJNAF Newport News, VA

Page 2: Al  Moretti , APC, Fermilab

Discussion of the Magnetically Insulated Cavity The orthogonal Box Cavity

2

Palmer and D. Stratakis developed as theory that a large magnetic field at right angle to the RF electric would magnetically insulate a vacuum RF cavity from electric breakdown.

It was then decided to construct a build cavity to test this theory.

A low cost design was chosen in the shape of a boxTwo cavities of this type have been built and only one has

been tested.We need to decide what to do with the second cavity. It has

been suggested that this cavity be coated with TiN and tested in the MTA?

There is in storage material to built 2 more cavities. It has been suggested that the material be used to built the parallel B and E configuration cavities?

Page 3: Al  Moretti , APC, Fermilab

3

Original LBL Waveguide RF power Coupler section; ½ height Standard WG section

Rectangular Coupling aperture with rounded edges

Orthogonal E and H Cavity Coupled to the Waveguide

HFSS Model RF cavity and WG coupler

¼ height Coupling Cell. This is built in two sections with WG; flanges not shown.

Pickup Ports

Sapphire Viewing Port

E

B

Page 4: Al  Moretti , APC, Fermilab

4

Calculated Parameters of the cavity and cavity dimensions

HFSS normalizes all parameters to 1 W of input power to the waveguide coupler and solves for the frequency, gradient, coupling factor, Qo (in cavity mode) etc.

1W produces a gradient of 25 kV/m by scaling:

25 MV/m would take 1 MW ideal.

The Impedance across the center of the cavity is

Imp =9.5 MΩ.

This is the resistance across the center of the cavity given by

Imp = (gap Voltage)²/1 W.

This uses the peak voltage and is in agreement with SuperFish and most published accelerator designs.

Qo = 27,400.

250 mm

276.50 mm

123.82 mm

Aperture Fields about a factor 4 lower than central max. field.

E

B

Page 5: Al  Moretti , APC, Fermilab

5

Orthogonal Cavity Electric Field Simulation Measurement Results

Peak Surface field on-axis (Eps)

Peak Surface Field on Aperture (Epap)

Ratio Eps/ Epap= 3.5 for orthogonal case.

Network analyzer measurements:

Fo = 805.33 MHz test lab; Simulation Fo=806.2 MHz

Qo= 27,400; Simulation Qo= 27,400

β= 0.96 coupling factor; Simulation β= 1.06.

These values are preliminary and may change when attached to the LBL stepped WG coupler and mounted in the magnet. Even with a coupling factor of β= 0.96, 99.95 % of the transmitted RF power go into the cavity.

Page 6: Al  Moretti , APC, Fermilab

6

End ¼ height Waveguide Coupling Piece

Compound Tin Seal Location

Water cooling tubes

Pickup ports

Sapphire Viewing port location

Box Cavity in A0 Test Lab

Page 7: Al  Moretti , APC, Fermilab

7Y. Torun

Pick-up and Sapphire ports

Water cooling tubes

Box Cavity in the MTA Magnet on its Rail Support

Page 8: Al  Moretti , APC, Fermilab

8Y. Torun

Calibration Adapter WG to type N Vacuum Pump

Ceramic window for Vacuum

Calibration Setup befor connection to 12 MW Supply WG

WG Flex Section for Angle Changing

Page 9: Al  Moretti , APC, Fermilab

9

0.00E+005.00E+051.00E+061.50E+062.00E+062.50E+063.00E+063.50E+064.00E+060

5

10

15

20

25

30

35

40

45

Spark Gradient vs. number of pulses at 0 deg and 3 T

Number of RF pulses

Gra

dien

t in

MV/

m

Large YOYO Effect starts here

Page 10: Al  Moretti , APC, Fermilab

10

0.00E+00 5.00E+05 1.00E+06 1.50E+06 2.00E+06 2.50E+06 3.00E+06 3.50E+06 4.00E+060

5

10

15

20

25

30

35

40

45

Spark Gradient vs. Angle between E and B at 3 T

Series1

Series3

Series5

Series7

Series9

Number of Pulses

Gra

dien

t in

MV/

m

= 90 deg

= 89 deg

=87 deg

=91 deg

=86 deg

Page 11: Al  Moretti , APC, Fermilab

11

85 86 87 88 89 90 91 920

5

10

15

20

25

30

35

40

Box Cavity Gradient vs Angle Between E &B at 3 T

RF field to DC Magnetic Field Angle

Gra

dien

t in

MV/

m

Gradient Line for Very Low Sparking Rate.

The cavity was also tested at magnetic field levels from 0.5 to 3.5 T at 0 Deg. There was very little drop off observed in the break down levels.

There was no X-Ray’s observed with the Magnetic field on

There may have been single surface mult-pactiring observed in the early phase of the RF commissioning with the magnetic field

The vacuum burst were 10 to 100 times larger during a spark with the magnetic field than without.

Page 12: Al  Moretti , APC, Fermilab

1212

Pictures of Damaged Area in the Box Cavity

The inside of the cavity was very clean with not many spark pits visible if any. There were blistered areas in appearance in the top and bottom corners

along the axis of the cavity both the upstream and downstream ends.These are pictures of the cavity at about ½ the running time. At the end of the

experiment, the areas appeared darker and spread out further and this was also after the 50 MV/m RF commissioning at 0 T.

Blistered AreasCu to SS pickup hole surface.

Page 13: Al  Moretti , APC, Fermilab

13

The SPDT RF Switch has been installed in the MTA

This gives us 2 RF test location in the MTA Hall.

Page 14: Al  Moretti , APC, Fermilab

14

Parallel E and H field Design:

E Field pick-ups

Optical View port

¼ height Coupling Cell. This is built in two sections with WG flanges not shown.

Original LBL Waveguide RF power Coupler section; ½ height Standard WG section

E & B

Page 15: Al  Moretti , APC, Fermilab

15

• We have completed the Box cavity Study ( magnetic field insulation effect) at 91,90, 89, 87 and 86 Degrees at 3. Tesla. Before each run we commissioned the cavity upto 33 MV/m at 0. T except for the first 0 Degree run commissioned to 23 MV/m. In some of the later runs, we commissioned the cavity to 50 MV/m at 0 T.

• The results have been shown: from 89 to 91 deg we reached a low sparking rate at a average gradient of 33 MV/m

at 87 deg a average gradient of 25 MV/m and at 86 deg a average gradient of 22.5 MV/m.• We need to decide if the second orthogonal box cavity should be

tested with a TiN coating or tested at all?• We also need to decide if we need to build a parallel Box for

testing in the MTA from the copper in storage?• The second RF test station is now ready for RF commissioning

cavities that do not require a magnetic field.

Summary Of the Box cavity