bridge al trapianto cardiaco luigi martinelli sc cardiochirurgia ao niguarda ca’ granda milano...
TRANSCRIPT
Bridge al trapianto cardiaco
Luigi MartinelliSC Cardiochirurgia
AO Niguarda Ca’ GrandaMilano
GENOVA, 9 novembre 2012GENOVA, 9 novembre 2012
Trattamento dello scompenso Trattamento dello scompenso cardiaco refrattario: cardiaco refrattario: trend dell’ultimo trend dell’ultimo
decennio decennio
Trapianto di cuore in Italia Trapianto di cuore in Italia 1992-2012 1992-2012
Source: National Transplant Center, Italy
2012 forecast
Waiting list in Italy - Year 2011Waiting list in Italy - Year 2011
Pts on January 1Pts on January 1stst 711711
Pts on December 31Pts on December 31stst 733733
New patientsNew patients +432+432
HTxHTx 278278
Mean waiting timeMean waiting time 2.5ys2.5ys
Mortality on waiting listMortality on waiting list 94 (8.2%)94 (8.2%)
HTx/pts on waiting listHTx/pts on waiting list 24.3% 24.3%
HTx vs LVAD: trend dal 1995 HTx vs LVAD: trend dal 1995 AO Niguarda Ca’ Granda AO Niguarda Ca’ Granda
Pulsatile LVADPulsatile LVAD Continuous flow LVADContinuous flow LVAD HTxHTx
Kirklin JK, J Heart Lung Transplant 2012
1990 2000Shock cardiogeno refrattario
Recupero del danno d’organo
VAD paracorporeo a breve termine (giorni-settimane)
Shock cardiogeno imminente
Prevenzione del danno d’organo
LVAD a lungo termine, bridge al
TC (1-2 anni)
Impianto elettivo
Regressione dell’ipertensione polmonare
LVAD permanente (Destination Therapy)
LVAD: evoluzione di indicazioni e tecnologia
LVAD: recente evoluzione tecnologica (2000-12)
Thoratec - HeartMate II
Micromed - DeBakey
Berlin Heart - Incor
LVAD intracorporeo a flusso continuo (2a generazione)
LVAD senza valvole
LVAD senza valvole
0
2
4
6
8
10
12
14
16
1992
1993
1994
1995
1996
1997
1998
1999
2000
2001
2002
2003
2004
2005
2006
2007
2008
2009
2010
2011
2012
HeartWareHM IINewCorTechIncorDeBakeyNovacor
LVAD – A.O. Niguarda Ca’ Granda
Confronto tra 2a e 3a generazione di LVAD
Thoratec - HeartMate II HeartWare HVAD
Inizio sviluppo
Primo impianto umano
N° totale impianti
Peso
Tipo di pompa
Flusso massimo
Driveline percutanea
Durata batterie
1995
2006
1’000
160 g
centrifuga
10 L/min
Sì ( 4,2 mm)
fino a12 ore
1991
2000
7’000
289 g
assiale
10 L/min
Sì ( 6,3 mm)
fino a 10 ore
Thoratec - HeartMate II
HeartWare
Confronto tra 2a e 3a generazione di LVAD
Stewart and Stevenson, Circulation 2011
In attesa di REVIVE-IT
Trial
Attuali indicazioni alla VAD
Kirklin JK, J Heart Lung Transplant 2012
Bridge to candidacy
Ipertensione polmonare severa
con resistenze fisse
Regressione dell’ipertensione polmonare con
l’unloading del VS
Bridge to candidacy: ipertensione polmonare
Kirklin JK, J Heart Lung Transplant 2012
Permanent LVAD
(30% of pts!)
Bridge to HTx
Frazier OH, JTCVS 2001; Miller LW, NEJM 2007; Lietz K, JACC 2007; Kirklin JK, J Heart Lung Transplant 2010
BTT – HeartMate II (2007)
On waiting list, Status 1, no LVAD (Lietz, JACC 2007)
BTT – INTERMACS (2012)
BTT HeartMate VE (2001)
LVAD a flusso assiale: bridge al TCLVAD a flusso assiale: bridge al TC
Kirklin JK, J Heart Lung Transplant 2012
More than 20% More than 20% still waiting…still waiting…
Follow-up resultsFollow-up results
FUP durationFUP duration 274 patient-months (1-25 months)274 patient-months (1-25 months)
Mean FUP durationMean FUP duration 10.1 months10.1 months
Bridged to Bridged to urgenturgent HTx HTx 88
Still on LVADStill on LVAD 1717
ThromboembolismThromboembolism 00
Major GI bleedingMajor GI bleeding 11
Driveline infectionDriveline infection 66
Broken drivelineBroken driveline 2 2
LVAD failureLVAD failure 22
DeathsDeaths 22
HeartMate II LVAS HeartMate II LVAS Niguarda Hospital 2008-2012Niguarda Hospital 2008-2012
(LVAD failures in end-stage cancer)
Broken driveline 1Inflow cannula misalignment 1
Driveline abscess 4Recurrent GI
bleeding 1
Tempo di attesa in lista on LVAD Tempo di attesa in lista on LVAD Niguarda Hospital 2002-2011Niguarda Hospital 2002-2011
209,2 ± 214,1 days
380,0 ± 188,4 days
De facto permanent LVAD: De facto permanent LVAD: già un’alternativa già un’alternativa
al trapianto? al trapianto?
ESC Guidelines 2012 ESC Guidelines 2012 - Recommendations for surgical implantation of LVADs in patients with - Recommendations for surgical implantation of LVADs in patients with systolic heart failuresystolic heart failure
Heart Failure 201219-22 May 2012, Belgrade - Serbia
Key evidenceVentricular assist devices may ultimately become a more general alternative to transplantation, as current 2- to 3-year survival rates in carefully selected patients receiving the latest continuous flow devices are much better than with medical therapy only.Patients receiving these devices also have a post-transplant survival rate similar to those not requiring bridging
ESC/HFA Guidelines 2012
LVAD: alternativa al TC?
• Per poter essere considerata una reale
alternativa
• sopravvivenza comparabile
• qualità della vita comparabile
• adottabile in caso di controindicazioni al TC
• disponibilità pari o superiore
Changes in global health satisfaction: HTx vs LVADChanges in global health satisfaction: HTx vs LVAD
Kugler C, J Heart Lung Transplant 2011
Kugler C, J Heart Lung Transplant 2011
Kirklin JK, J Heart Lung Transplant 2011
Slaughter MS, NEJM 2009
Controindicazioni al TC e a LVAD
Criteri di esclusione a LVAD
• insufficienza renale severa• insufficienza epatica severa• insufficienza respiratoria severa
ma anche…• prognosi quoad vitam limitata• controindicazioni alla TAO• aorta a porcellana• scarso supporto socio-familiare• problemi psicologici• tossicodipendenza, alcolismo
Molte controindicazioni al
trapianto sono controindicazioni anche
per LVAD!
Richiesta potenziale di DT in Italia
Mortalità in lista Mortalità in lista d’attesa per TCd’attesa per TC
100 pz100 pzImpianti effettivi di Impianti effettivi di LVAD in Italia/annoLVAD in Italia/anno
100 pz100 pz
Potenziali Potenziali candidati a LVADcandidati a LVAD
ca 6’000 pzca 6’000 pz
Based on Miller LW, Circulation 2011
Trapianti Trapianti in Italia/anno in Italia/anno
300 pz300 pz
LVAD: quanto costa?
SC Cardiochirurgia AO Niguarda Ca’ Granda
Consumi sanitari 2011
LVAD: 1’188’054 € (29%)1’188’054 € (29%)
14 pazienti (1.7%)14 pazienti (1.7%)
Protesi valvolariProtesi valvolari: : 1’045’696 € (25%)1’045’696 € (25%)
440 pazienti (53%)440 pazienti (53%)
(% della spesa totale)(% della spesa totale)
(% dei pazienti totali)(% dei pazienti totali)
Chi avviare Chi avviare al supporto meccanico long- al supporto meccanico long-term term
in un’epoca in un’epoca di scarsità di organi di scarsità di organi
Scompenso Scompenso refrattariorefrattario
Controindicazioni Controindicazioni a LVAD a LVAD (piccola BSA, (piccola BSA,
disfuzione biventricolare)disfuzione biventricolare)
TRAPIANTO TRAPIANTO DI CUORE DI CUORE
Assenza di Assenza di controindicazioni a LVADcontroindicazioni a LVAD
Lungo tempo di attesa Lungo tempo di attesa (gruppo AB0, grande BSA)(gruppo AB0, grande BSA)
Trapianto ad alto rischio Trapianto ad alto rischio (alte PVR)(alte PVR)
Rischio di deterioramento Rischio di deterioramento emodinamico emodinamico (INTERMACS level < 3)(INTERMACS level < 3)
LVADLVAD
Destination TherapyDestination Therapy
Età > 60-65 anniEtà > 60-65 anni
Disponibilità di un organo idoneo
No organo idoneoSI
NO LISTA LISTA per TCper TC
Permanent LVADPermanent LVAD
INTERMACS 1-2
INTERMACS 5-7
INTERMACS 3-4
Troppo Troppo tarditardi
Mortalità Mortalità eccessivaeccessiva
Troppo Troppo
precoceprecoce
Condizioni Condizioni deterioratedeteriorate
Ampio Ampio vantaggio vantaggio
prognosticoprognostico
Eccesso di Eccesso di complicanzecomplicanze
Scarso Scarso vantaggio vantaggio
prognosticoprognostico
ATTUALE ATTUALE CRITERIO DI CRITERIO DI
INDICAZIONEINDICAZIONE
INTERMACS 1-2
INTERMACS 5-7
INTERMACS 3-4
Troppo Troppo tarditardi
Mortalità Mortalità eccessivaeccessiva
Troppo Troppo
precoceprecoce
ATTUALE ATTUALE CRITERIO DI CRITERIO DI
INDICAZIONEINDICAZIONE
Survival benefit margin
Stewart and Stevenson, Circulation 2011
INTERMACS 1INTERMACS 1
INTERMACS 7INTERMACS 7
Conclusioni• Nell’ultimo decennio si è assistito ad una riduzione del numero di donatori di cuore e ad un miglioramento sensibile dei risultati degli LVAD intracorporei
• La scarsità di organi ha determinato anche una riduzione del numero dei trapianti nei pazienti portatori di LVAD, trasformando in molti casi il bridge al trapianto in impianto permanente
• L’indicazione a LVAD deve essere valutata precocemente, e periodicamente riconsiderata, nei pazienti in lista d’attesa per trapianto di cuore
Conclusioni• L’impianto di LVAD deve essere seriamente considerato come una terapia affidabile dello scompenso refrattario e deve essere proposto tempestivamente prima del deterioramento clinico
• La principale limitazione all’impianto di LVAD rimane l’elevato costo economico