cleaning procedure

Upload: monir61

Post on 14-Apr-2018

241 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/30/2019 cleaning procedure

    1/38

    Cleaning procedures and testing

    Challenging the process

    Peter NewsonAbu Dhabi Patient Safety

    15 October 2012

  • 7/30/2019 cleaning procedure

    2/38

    Objectives

    How clean is clean ?

    Manual cleaning

    EN ISO 15883 and the need for soil removal efficacy and protein

    testing

    Ultrasonics

    Standard instrument cleaning systems

    Complex and lumen instruments

    Conclusion

  • 7/30/2019 cleaning procedure

    3/38

    Learning Objectives

    3

    Discuss how to manage and control the risks associated with

    decontamination reprocessing

    Identify best practices for evaluating the effective cleaning of

    reusable instruments

    Present the difference between soil removal and protein

    residue tests

  • 7/30/2019 cleaning procedure

    4/38

    Cleaning Processes

    Instrument cleaning is the first and arguably the most

    important step in the decontamination process.

    Microscopic quantities of blood, skin, mucous etc. left on

    the instruments by an ineffective washing process can

    seriously compromise the overall sterility of the

    instruments.

  • 7/30/2019 cleaning procedure

    5/38

  • 7/30/2019 cleaning procedure

    6/38

    This norm regulates the validation of processing sequences in

    washer-disinfectors for medical devices as well as revalidation and

    routine monitoring of these processes.

    Since the coming into force of EN ISO 15883-1, only washer-

    disinfectors that have been subjected to a type test as per ISO

    15883-1 may be purchased.

    EN ISO 15883

  • 7/30/2019 cleaning procedure

    7/38

    Manual Cleaning

    The most important step in all instrument including Endoscope

    reprocessing and is defined as

    "Manual cleaning" refers to the physical task, performed

    by hand, of removing secretions and contaminants from

    the endoscope with appropriate brushes, cloths, detergents and

    water.

  • 7/30/2019 cleaning procedure

    8/38

    Manual Cleaning

    There are two purposes of manual cleaning

    As a pre-cleaning prior to using an automated Washer

    Disinfector

    As a cleaning method where other methods are not availableor appropriate.

    Two methods of manual cleaning

    Immersion

    Non immersion

  • 7/30/2019 cleaning procedure

    9/38

    Manual cleaning

    Manual cleaning can be done using a range of cleaning aids:

    Brushes: external and internal

    A wide range of brushes are available for manual cleaning, suitable for

    individual cleaning tasks.

    External brushes: for cleaning the outside surfaces of objects. They may have

    hard or soft hairs.

    Internal brushes: for cleaning of hollow instruments. A range of diameters and

    lengths are available for any type/size of hollow instruments and materials.

    steel brushes are not recommended for cleaning , as they damage the

    protective layer on stainless steel and aluminium instruments!

  • 7/30/2019 cleaning procedure

    10/38

    Examples

  • 7/30/2019 cleaning procedure

    11/38

    Manual cleaning

    Spray gun

    A spray gun is essential for rinsing/flushing of hollow instruments. Variousnozzles are available for a range of specific cleaning applications.

    Hand shower

    The hand shower can be used for an initial rinse of instruments.

    Use a deep basin, which will help to prevent splashing sideways. Make surethat the water pressure is not too high in order to limit splashing. Only usecold water for blood removal!

  • 7/30/2019 cleaning procedure

    12/38

    Manual Cleaning

    Hand shower

    Personal protection during manual cleaning - a range of

    materials are available including - Gloves , plastic aprons,

    mouth masks , goggles, visors and splash screens

  • 7/30/2019 cleaning procedure

    13/38

    What leads to successful

    Manual cleaning To be performed by a person trained in cleaning techniques and fully

    conversant with the instrument they are cleaning

    To be undertaken as soon as possible after the procedure to avoid soil and

    debris drying.

    To follow the stated unit protocols , which using the appropriate detergents

    and cleaning equipment allows for all internal and external instrument

    surfaces to be cleaned.

    To be followed by thorough rinsing to ensure that all debris and detergent

    residues are removed prior to the next step in the decontamination process.

  • 7/30/2019 cleaning procedure

    14/38

    Mechanical cleaning

  • 7/30/2019 cleaning procedure

    15/38

    - Part 1: General requirements, terms and definitions and tests

    - Part 2: Requirements and tests for washer-disinfectors employing thermal disinfectionfor surgical instruments, anesthetic equipment, bowls, dishes, receivers, utensils,

    glassware, etc.

    - Part 3: Requirements and tests for washer-disinfectors employing thermal disinfection

    for human waste containers

    - Part 4: Requirements and tests for washer-disinfectors employing chemical

    disinfection for thermolabile endoscopes

    - Part 5: Test soils and methods for demonstrating cleaning efficacy (Technical

    Specification)

    EN ISO 15883

  • 7/30/2019 cleaning procedure

    16/38

    What are we validating ?

    The washer

    Soil removal including the removal of protein

    The cycle - including all parameters incorporated within the washing

    cycle.

    Impingement / spray arms, complex channel pressure

    The load and the impact this load has on the effectiveness of the cycle.

    This includes :

    Type of instruments more challenging than beforeLevel of soil

    Load configuration.

  • 7/30/2019 cleaning procedure

    17/38

    EN ISO 15883

    6.10 Tests of cleaning efficacy

    6.10.1 General

    During tests of cleaning efficacy, the cycle shall be run withouta disinfection stage. The drying stage may also be omitted if

    this is necessary to facilitate the detection of residualcontamination or test soil

    6.10.2 Cleaning efficacy test 1

    6.10.2.1 General

    The tests for cleaning efficacy shall be carried out using theappropriate test method(s) and test soil(s) as described inISO/TS 15883-5 by taking into consideration the correspondingcategory of load.

  • 7/30/2019 cleaning procedure

    18/38

    Routine Monitoring: In General

    Every load

    Independent from type and age of the WD

    Documentation of the results

    The results must reflect the demands

  • 7/30/2019 cleaning procedure

    19/38

    Minimum for Routine Monitoring

    Loading pattern

    as described in standards

    and dependent on the process andinstruments being cleaned

    Cleaning efficacy

    daily optical control of cleaning

    daily optical view of drying

    cleaning indicator

    protein test

  • 7/30/2019 cleaning procedure

    20/38

    So is it clean ??

  • 7/30/2019 cleaning procedure

    21/38

    Soil removal test and

    Protein test

    Soil removal a visual test

    Protein test protein is invisible so a swab plus chemicalinterpretation is required

  • 7/30/2019 cleaning procedure

    22/38

    Preference must be given to the use of standardised,

    quantitatively assessable test models as they yield

    reproducible results.

    ( i.e. The tests need to be consistent)

    Guidelines - Soil Testing

  • 7/30/2019 cleaning procedure

    23/38

    Cleaning Efficacy / soil testing

    National test soils are described in TS EN ISO 15883 5 where

    a total of 19 different test soils are described.

    The challenge is a combination of the test soil and the barrier

  • 7/30/2019 cleaning procedure

    24/38

    Test Soils - 15883 5Load type Country Constituents

    Annex

    Surgical

    instruments (

    including rigid

    endoscopes)

    NL Bovine serum albumin fraction 5,porcine gastric mucin type 3,

    bovine fibrinogen fraction 1,

    bovine trombin Annex K

    SE Citrated cattle blood coagulatedwith calcium chloride

    Annex M

    UK Defibrinated horse/sheep blood,egg yolk, dehydrated hog mucin

    Annex N

    DE Sheep blood, E. faeciumbEgg yolk, E. faeciumb

    Semolina, butter, sugar, milkpowder, E. faeciumb

    Annex G

    US Protein/organic soil (userpreference), B. atrophaeus

    endospores

    Annex S

  • 7/30/2019 cleaning procedure

    25/38

    Test Soils - 15883 5Load type Country Constituents Annex

    Flexible

    endoscopes

    AT Nigrosin, wheat flour,hens egg, E.Faecium

    Annex E

    DE Blood E Faecium Annex I

    NL Bovine serum albumin,porcine mucin,bovin

    fibrinogen, bovin

    thrombin

    Annex L

    UK Glycerol, dehydratedhog mucin,

    horse serum,unbleached plain

    flour, aqueous safranine

    solution,

    water

    Annex R

  • 7/30/2019 cleaning procedure

    26/38

    EN ISO 15883 Protein testing

    EN ISO 15883-1:2004(E) 10

    NOTE 2 Additional verification of attainment of the required

    cleaning efficacy during operational testing may be

    provided by the use of one of the methods for the detection and

    assessment of residual protein given in 10.2.3 and

    Annex C.

  • 7/30/2019 cleaning procedure

    27/38

    Protein testing - EN ISO 15883-1

    The test method for performance qualification of cleaning

    efficacy is described in 6.10.3 and shall include the use of

    one of the methods for the detection and assessment of

    residual proteinaceous contamination given in Annex C.

  • 7/30/2019 cleaning procedure

    28/38

    EN ISO 15883 -1 Protein testing

    EN ISO 15883-1:2004(E) 10

    NOTE 4 The three test methods for protein residue testingin Annex C are not equally sensitive. The ninhydrinmethod

    (C.1) and Biuret method (C.3) have similar sensitivities but

    should be regarded as a limit test and a semi-quantitativetest

    respectively. The OPA method (C.2) is more sensitive butrequires the use of laboratory facilities.

  • 7/30/2019 cleaning procedure

    29/38

    So how do we protect the

    patient, the healthcare workerand the hospital?

    EN ISO 15883, and numerous national and international

    standards and recommendations all reflect the need to

    routinely monitor the Soil removal efficacy of your

    process. In general this is undertaken on a daily basis.

    Added to this is the routine need to test protein residues

    up to an accuracy of a minimum 2ug.

  • 7/30/2019 cleaning procedure

    30/38

    Surrogate devices standard

    Representative of standard hospital instrumentation such as

    box joints , burr baskets hinged instruments etc.

    Most made from stainless steel offering various challenges to

    the cleaning process.

    Universally applicable and in most cases usable in Ultrasonic

    machines.

  • 7/30/2019 cleaning procedure

    31/38

    Surrogate devices - standard

  • 7/30/2019 cleaning procedure

    32/38

    Surrogate devices

    TS ISO 15883 Annex E E.6.1 Tubes for use as test pieces ( Austria )

    (PTFE) tubes having a length of 2 m and inner diameters of 1,0 mm and

    2,0 mm respectively.

    A.6.2 Instruments for minimal invasive surgery

    As a surrogate for rigid endoscopes dummies made of stainless steel

    tubing should be used with a wall

    thickness of approximately 1 mm and:

    length of 150 mm, inner diameter of 8 mm;

    length of 300 mm, inner diameter of 4 mm and 6 mm

    respectively.

  • 7/30/2019 cleaning procedure

    33/38

  • 7/30/2019 cleaning procedure

    34/38

    Surrogate devices

    TS ISO 15883 Annex R 6 Test pieces ( UK )

    A surrogate device for investigation of cleaning and disinfectionmay be constructed from two 1,5 m lengths of PTFE tube having

    an inner diameter of 2mm and one 1,5 m length of PTFE tube

    having an inner diameter of 1 mm. These should be bound

    together with adhesive tape at intervals of approximately 150

    mm.

  • 7/30/2019 cleaning procedure

    35/38

    Surrogate devices

  • 7/30/2019 cleaning procedure

    36/38

    Conclusion

    In conclusion the challenge to those in the CSSD is

    The more complex the instrument the more vital it is that weare assured that it has been effectively cleaned. Not every now

    and again but consistently hour by hour , day by day whether

    manually or through an automated process.

    It is therefore critical that we are constantly challenging that

    process using the tools provided in international standards in

    whatever form is most suitable to your needs and practices.

  • 7/30/2019 cleaning procedure

    37/38

    Thanks for your time !

  • 7/30/2019 cleaning procedure

    38/38

    References

    EN ISO standards

    EN ISO 15883 1

    TS EN ISO 15883 - 5

    National and regional guidelines

    38