Page 1: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”
Page 2: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Consistency of the Planck Data

C. R. Lawrence, JPL

On behalf of the Planck Collaboration

Consistency of the Planck data Planck Collaboration—2 Ferrara, 2014 December 01

Page 3: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Outline

• Comparison of maps

• Calibration

• More maps

• Comparison of power spectra

Consistency of the Planck data Planck Collaboration—3 Ferrara, 2014 December 01

Page 4: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

70GHz

Consistency of the Planck data Planck Collaboration—4 Ferrara, 2014 December 01

Page 5: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

100GHz

Consistency of the Planck data Planck Collaboration—5 Ferrara, 2014 December 01

Page 6: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

143GHz

Consistency of the Planck data Planck Collaboration—6 Ferrara, 2014 December 01

Page 7: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

217GHz

Consistency of the Planck data Planck Collaboration—7 Ferrara, 2014 December 01

Page 8: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

100 � 70

Consistency of the Planck data Planck Collaboration—8 Ferrara, 2014 December 01

Page 9: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

143 � 70

Consistency of the Planck data Planck Collaboration—9 Ferrara, 2014 December 01

Page 10: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

143 � 100

Consistency of the Planck data Planck Collaboration—10 Ferrara, 2014 December 01

Page 11: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Photometric Calibration

Important changes in 2014:

• Primary calibration signal for both LFI and HFI changed from the

“Solar dipole” to the “orbital dipole”

• Both LFI and HFI made improvements in calibration, taking

advantage of improved beam models and better control of

systematic effects

Consistency of the Planck data Planck Collaboration—11 Ferrara, 2014 December 01

Page 12: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

DipolesSolar dipole⌘dipole induced by the motion of the Solar system barycenter wrt the CMBOrbital dipole⌘modulation of the Solar dipole induced by the orbital motion of Planck around

the SS barycenter

• Solar dipole depends on TCMB

motion of SS barycenter wrt CMB– In 2013, Planck calibrated on the WMAP7 Solar dipole

• Orbital dipole is differential, and the velocity of Planck around the

SS barycenter is known with exquisite accuracy.

– Amplitude 240µK

– In 2014, LFI and HFI have systematics under control at a level that can

measure such a small signal with precision.

Removes the error in the WMAP Solar dipole as a common source of error

for LFI and HFI. Gives a shift of +0.28 % in gain calibration for both LFI and HFI.

PreliminaryConsistency of the Planck data Planck Collaboration—12 Ferrara, 2014 December 01

Page 13: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Improvements in Calibration— LFI• Use full 4⇡ beam to convolve with dipole (solar + orbital + relativistic

quadrupole), rather than “pencil beam” as in 2013

– Improves both null tests and the quality of the polarization maps

– Gives average shifts of +0.15%, +0.035%, and +0.17% in gain calibration

at 70, 44, and 30 GHz, respectively.

• Reduce the effects of 1/f noise with an improved destriping code.

Remove Galactic sidelobe contamination

Improve calibration smoothing algorithm — better SNR with less

susceptibility to transients

– Give better null test results, and a shift of +0.4% in gain calibration

PreliminaryConsistency of the Planck data Planck Collaboration—13 Ferrara, 2014 December 01

Page 14: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Improvements in Calibration— HFI

• Improved determination and handling of near and far sidelobes

• Improved ADC non-linearity correction

• Improved handling of very long time constants

– Give an average shift of +0.5% in gain calibration

PreliminaryConsistency of the Planck data Planck Collaboration—14 Ferrara, 2014 December 01

Page 15: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Comparisons of Calibration

• Look at consistency of LFI and HFI Solar dipole estimates

• Compare measurements of first acoustic peak after component

separation

Consistency of the Planck data Planck Collaboration—15 Ferrara, 2014 December 01

Page 16: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Compare Dipoles

Amplitude

Experiment [µK] Latitude Longitude

LFI................................ 3365.5± 3 48.�26 264.�01

HFI DPC ...................... 3364.1± 2 48.�23± 0.�1 263.�96± 0.�03

Planck 2014 nominal ..... 3364.5± 2a 48.�24± 0.�1 264.�00± 0.�03

WMAP.......................... 3355b ± 8 48.�26± 0.�03 263.�99± 0.�14

a 0.06%b 0.28% lower than Planck 2014 measurement

The change to orbital dipole calibration gives a common +0.28% shift in gain

calibration for both LFI and HFI.

PreliminaryConsistency of the Planck data Planck Collaboration—16 Ferrara, 2014 December 01

Page 17: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Compare Amplitudes During Component Separation

• Both SMICA and Commander calculate amplitude scaling factors

applied to the input frequency maps that minimize residuals in the

CMB component fitted to the data.

• These amplitude scaling factors provide an assessment of the

relative calibration accuracy across frequencies

Consistency of the Planck data Planck Collaboration—17 Ferrara, 2014 December 01

Page 18: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Compare Amplitudes During Component SeparationIntercalibration Factors [%]

SMICA Commander

Frequency [GHz] 100 ` 700 25 ` 100

30 .............................. . . . �0.3± 0.1

44 .............................. . . . 0.3± 0.1

70 .............................. �0.13± 0.04 0.0± 0.1

100 .............................. 0.08± 0.02 0.1± 0.02

143 .............................. Reference frequency Reference frequency

217 .............................. 0.29± 0.03 Det. 1 reference frequency

353 .............................. 0.80± 0.16 0.6± 0.1

• The SMICA and Commander calculations of relative calibration factors cover different ` rangesand handle residual monopoles and dipoles somewhat differently. We don’t expect exactequality.

• From 30–217 GHz, agreement is a few tenths of a percent. Preliminary

Consistency of the Planck data Planck Collaboration—18 Ferrara, 2014 December 01

Page 19: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

North Ecliptic Pole: 70, 100, Diff

60� patches

Consistency of the Planck data Planck Collaboration—19 Ferrara, 2014 December 01

Page 20: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

North Ecliptic Pole: 70

Consistency of the Planck data Planck Collaboration—20 Ferrara, 2014 December 01

Page 21: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

North Ecliptic Pole: 100

Consistency of the Planck data Planck Collaboration—21 Ferrara, 2014 December 01

Page 22: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

North Ecliptic Pole: Diff

Consistency of the Planck data Planck Collaboration—22 Ferrara, 2014 December 01

Page 23: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

South Ecliptic Pole: 70, 100, Diff

Consistency of the Planck data Planck Collaboration—23 Ferrara, 2014 December 01

Page 24: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

South Ecliptic Pole: 70

Consistency of the Planck data Planck Collaboration—24 Ferrara, 2014 December 01

Page 25: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

South Ecliptic Pole: 100

Consistency of the Planck data Planck Collaboration—25 Ferrara, 2014 December 01

Page 26: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

South Ecliptic Pole: Diff

Consistency of the Planck data Planck Collaboration—26 Ferrara, 2014 December 01

Page 27: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

LFI-HFI Temperature Ratio

!

100HR1 ⇥ 100HR2/(100HR2 ⇥ 70)

143HR1 ⇥ 143HR2/(143HR2 ⇥ 70)

PreliminaryConsistency of the Planck data Planck Collaboration—27 Ferrara, 2014 December 01

Page 28: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

LFI-HFI-WMAP Temperature RatioWMAP is V band

LFI70 / WMAP

HFI100 / WMAP

LFI70/ WMAP (2013)

HFI100/ WMAP (2013)

Changes in LFI and HFI since 2013 from better control of (independent) systematics and beams

Planck vs WMAP

PRELIMINARY Te

mpe

ratu

re r

atio

PreliminaryConsistency of the Planck data Planck Collaboration—28 Ferrara, 2014 December 01

Page 29: Consistency of the Planck Data - European Space Agency€¦ · • Primary calibration signal for both LFI and HFI changed from the “Solar dipole” to the “orbital dipole”

Summary

• Overall photometric calibration has changed by +0.83% for LFI

+0.78% for HFI

– Orbital dipole calibration and the elimination of the Solar dipole error

accounts for 0.28%.

– Better understanding of beams and various systematics, mostly

different for LFI and HFI, accounts for the rest.

• Agreement between LFI, HFI, and WMAP over the first peak is now

at a level of a few tenths of a percent.

• Maps and map differences between frequencies provide spectacular

confirmation that Planck sees the same CMB at all frequencies.

Consistency of the Planck data Planck Collaboration—29 Ferrara, 2014 December 01