The only IMS readily coupled to an existing API-MS without modifying the MS Mobility resolution > 50 Transmission > 50 % Convert an existing mass spectrometer into a powerful hybrid Ion Mobility-Mass Spectrometry instrument Uniform Mobility peak of the monomer (THA + ) of tetraheptylammonium bromide (THABr) at increasing DMA sheath air speeds, with resolutions increasing from 40 to 80 at the highest speeds 1000 5000 0 2 4 6 8 10 Signal (V) DMA Voltage (V) THA+ peak at different seath air speeds The best ion mobility spectrometer to couple to your MS 0 5 10 15 20 25 30 35 40 0 10 20 30 40 50 DMA at Mach~0.17 DMA at Mach~0.10 Transmission (%) q in / q out (%) Flows per unit length. 3D effects at the outlet not considered Sheath gas inlet Sheath gas outlet Ions outlet. Sampled q out to mass spectrometer Ions inlet Ionization source Upper electrode Lower electrode Isolator box Inlet slit of charged particles of several mobilities flow at velocity U Δ Exit hole of particles at mobility Z = 2 / LV DMA Electric Field E = V DMA / Δ L Seadm’s DMA

Upload: others

Post on 26-Sep-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Convert an existing mass spectrometer into a powerful hybrid Ion … DMA MS 2.pdf · 2010. 5. 4. · spectrometer to couple to your MS 0 5 10 15 20 25 30 35 40 0 10 20 30 40 50 DMA

The only IMS readily coupled to an existing API-MS without modifying the MS Mobility resolution > 50

Transmission > 50 %

Convert an existing mass spectrometer into a powerful hybrid Ion

Mobility-Mass Spectrometry instrument

Uniform

Mobility peak of the monomer (THA+)of tetraheptylammonium bromide (THABr) at increasing DMA sheath air speeds, with resolutions increasing from 40 to 80 at the highest speeds

1000 50000

2

4

6

8

10

Sig

nal (

V)

DMA Voltage (V)

THA+ peak at different seath air speeds

The best ion mobility spectrometer to

couple to your MS

0 5 10 15 20 25 30 35 400

10

20

30

40

50

DMA at Mach~0.17 DMA at Mach~0.10

Tra

nsm

issi

on (

%)

qin/ q

out (%)

Flows per unit length. 3D effects at the outlet not considered

Sheath gas inlet

Sheath gas

outletIons outlet.

Sampled qout to mass spectrometer

Ions inlet

Ionization source

Upper electrode

Lower electrode

Isolator box

Inlet slit of charged particles of several mobilities

Uniform flow at velocity U

Exit hole of particles at mobility Z = U∆2/ LVDMA

Electric Field E = VDMA / ∆

L

Seadm’s DMA

Page 2: Convert an existing mass spectrometer into a powerful hybrid Ion … DMA MS 2.pdf · 2010. 5. 4. · spectrometer to couple to your MS 0 5 10 15 20 25 30 35 40 0 10 20 30 40 50 DMA

MS systems successfully upgraded with Seadm’s IMS capability:

Sciex API 365Sciex API 3000Sciex API 5000Sciex QStarShimadzu LCMS2010EV

Tandem mass-mobility spectrum of electrosprayed Polydimethyl-siloxane (PDMS) ions where charged induced conformational changes in ion structure are clearly discerned. Courtesy of C. Larriba.

Tandem mass-mobility spectrum of Lysozyme monomer and aggregate ions formed by electrospray ionization under non-denaturing conditions. The DMA shows that these low charge state protein ions are relatively compact

DMA-MS spectrum of GroEL tetradecamers (MW~800kDa) with unprecedented resolving power (FWHM 2.6-2.9%)

Selected bibliography:-Differential mobility analysis of molecular ions and nanometer particles. Fdez. de la Mora; de Juan; Eichler; Rosell. Trends in An. Chem. 1998, 17, 328-339.-A planar DMA coupled to a MS for tandem IMS-MS separation at high transmission, with IMS resolution approaching 100. Juan Rus; Francisco Estévez; Juan Fernández de la Mora. ASMS07 Session: Ion Mobility – 106.-Evaluation of the Analytical Characteristic of a Differential Mobility Analysis coupled to a Triple Quadrupole System (DMA-MSMS). Hassan Javaheri ; Yves Le Blanc; Bruce A. Thomson; Juan Fernández de la Mora; Juan Rus; Juan A. Sillero. ASMS08, Session: Ion Mobility – 061.-Observation of Neutral Molecule (Ion-Pair) Evaporation from Ionic Liquid Nanodroplets by Tandem Differential Mobility Analysis-Mass Spectrometry (DMA-MS). Christopher J. Hogan; Juan Fernandez de la Mora. ASMS09 Oral Session: Applications of Ion Mobility Spectrometry - 10:10 am.-Tandem Ion Mobility-Mass Spectrometry (IMS-MS) Study of Ion Evaporation from Ionic Liquid-Acetonitrile Nanodrops. C. J. Hogan Jr. & J. Fernández de la Mora; Phys. Chem. Chem. Phys., 2009, 11, 8079-8090.-IMS-MS studies based on coupling a DMA to commercial API-MS systems. J. Rus; D. Moro; J.A. Sillero; J. Royuela, A. Casado, J. Fernández de la Mora, submitted to Int. J. Mass Spectrom. Special IMS-MS edition, J. Scrivens and R. Yost Special Editors, Dec. 2008. Review completed, pending minor changes.-Ion Mobility Measurements of Non-Denatured 12-150 kDa Proteins and Protein Multimers by Tandem Differential Mobility Analysis-Mass Spectrometry (DMA-MS). C. J. Hogan Jr. & J. Fernández de la Mora, to be submitted to J. Am. Soc. Mass Spectr., 2009.-True Mobility measurements of the 0.8 MDa GroEL tetradecamer show an anomalously compact gas phase structure. C. J. Hogan, B. Ruotolo, C. Robinson, J. Fernandez de la Mora; Second draft completed.

SEADM is a Spanish SME established in 2005 with the aim of developing a new generation of analytical instruments able to detect trace elements at the

subfemtogram level, both in the atmosphere or within the body fluids.For product information & sales contact [email protected]

Or visit www.seadm.com

SEADM DMA-MS instruments available at:MDS Sciex Shimadzu Research Lab. Europe Ltd.University of Minnesota US Dept. of Energy molecular-sciences lab (through Batelle)Yale’s facility open to visitors: http://www.eng.yale.edu/DMAMSfacility/

power (FWHM 2.6-2.9%) at m/z > 8000. Multiple tetradecamer conformers are observed.

DMA-MS spectrum of spherical ionic liquid (Ethyl-methyl imidazolium+ tris (trifluoromethylsulfonyl)

methide- ) nanometer sized clusters.