developing a new separation paradigm for intact protein lc...
TRANSCRIPT
PO
93
920
312
_W_5
Developing a New Separation Paradigm for Intact Protein LC/MS Analysis us-ing Next Generation Wide-Pore Core-Shell Media
Michael McGinley, Jeff Layne, and Jason Anspach
Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA
Analyzing intact proteins by reversed phase often pushes LC/MS to the limits of its capabilities. Reversed phase chromatography struggles to separate minor differences in post-translational modified proteins; electrospray mass spectrometers generate arrays of differentially charged forms requiring deconvolution software to identify minor differences. These limitations are especially evident when analyzing monoclonal IgG therapeutics. Identifying and quantitating by LC/MS minor glycoforms, glycations, as well as structural isoforms common with antibodies, are limited by complexity of the sample as well as poor recovery.
Recently introduced wide-pore core-shell media fundamentally changed the chromatography of large intact proteins. Analysis conditions are required to be dramatically altered versus the highly specialized isopropanol-based mobile phases previously reported with other intact antibody methods. Lower retentivity requires lower initial organic concentration, but also results in higher recovery of hydrophobic proteins.
Efforts were undertaken to explain the rationale for particle design as well as present some parameters to consider when developing methods using such media.
Introduction
Mouse IgG and EGF were purchased from ProSpec (New Brunswick, NJ); myoglobin, bovine IgG, and other chemicals were purchased from Sigma (St. Louis, MO). Solvents were purchased from EMD (San Diego, CA). Core-shell columns were packed at Phenomenex (Torrance, CA) and fully porous columns were purchased from column manufacturers (various sources).
Myoglobin and other protein samples were partially degraded by incubation at room temperature for up to a week in dilute acid. For refolding analysis, samples were reduced with DTT; reduced/non-reduced mixtures were generated by spiking different ratios of the native to the
reduced sample prior to injection on HPLC. The various protein samples were analyzed on an Agilent® 1200 HPLC system (Palo Alto, CA). Detection was monitored by either UV or MS (API 3000™ triple quadrapole MS, ABSCIEX™ Foster City, CA) in full-scan MS mode. Mass reconstruction was performed using ProMass software (Novatia, Monmouth Junction, NJ). Mobile phases used were either 0.1 % TFA in water (A) and 0.085 % TFA in acetonitrile (B) or formic acid in similar concentrations. Different gradients, flow rates, and column temperatures are listed with corresponding chromatograms.
Materials and Methods
Figure 1. Protein Diffusion vs. Size
Shown is a linear graph of protein diffusion rates through porous particles based onGutenwick et al. (JCA 2004) and Davies et al. (1989). While diffusion rates are a minor factor for small molecule chromatographic efficiency, forproteins larger than 25 kDa diffusion in and out of a porous column becomesthe limiting factor in efficiency of reversed phase separations.
0
0.5
1
1.5
2
2.5
3
3.5
0 50 10 0 1 50 2 00
Diffusion Coefficient
Protein MW (kDa)
Diff
usio
n C
oef
fiec
ient
Figure 2. Resolution vs. Shell Thickness
Lysozyme is compared on two different core-shell prototypes with varying porous shellthickness. The thin shell media provided narrower peak widths and increasedefficiency compared to the thicker shell particle. Optimal reversed phase protein separationis a balance between shell thickness, pore permeability, and loading.
min13 14 15
mAU
0
50
100
150
min13 14 15
mAU
0
50
100
150
Thick Shell (0.35 µm) 3.6 µm Media Thin Shell (0.15 µm) 3.6 µm Media
Ap
p ID
20
691
Column: Proprietary MaterialDimensions: 100 x 4.6 mm ID
Part No.: 00B-4252-B0 Mobile Phase: A: 0.1 % TFA in Water
B: 0.1 % TFA in AcetonitrileGradient: Time (sec) B %
0 3 20 6
Flow Rate: 1.5 mL / minTemperature: 40 ºC
Comparative separations may not be representative of all applications.
Figure 3. Loading vs. Peak Width
Peak width for Aeris WIDEPORE core-shell compared to a fully porous 300 Å high surfacearea media. Note the narrower peak width at analytical loads for lysozyme. As loadingincreases peak width for both columns increase, shell thickness was optimized to providesignificantly better performance compared to fully porous particles of the same diameter atloadings up to 20 µg.
0
0 .0 1
0 .0 2
0 .0 3
0 .0 4
0 .0 5
0 .0 6
0 .0 7
0 .0 8
0 .0 9
0 .1
0 .1 1 1 0 1 0 0
Peak Width
Lysozyme LoadingAeris™ WIDEPORE XB-C18 vs. Fully Porous 5 µm C18
Load (µg on column)
Aeris WIDEPORE 3.6 µm XB-C18
Fully Porous 5 µm C18
Pea
k W
idth
Comparative separations may not be representative of all applications.
Figure 4. Core-Shell vs. Fully Porous Media
Plates per column efficiency of fully porous 100 Å particles compared to core-shell particlesof a similar size. Across all particle sizes core-shell columns provided significantly higherefficiency over fully porous columns. Running conditions: 50 x 2.1 mm, 0.5 mL/min, mobile phase 50:50 acetonitrile/water, sample naphthalene.
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
1.7 2.6/2.5 3.6/3.5Particle Size (µm)
Plates
Core-Shell Fully Porous
Figure 5. Retention Comparison
Retention of a core-shell Aeris WIDEPORE column and fully porous 5 µm 300 Å column fordegraded myoglobin. The core-shell column was significantly less retentive than thefully porous column, yet provided dramatically improved resolution of closely related modified proteins. Optimizing methods for wide-pore core-shell column should use lowerinitial organic and shallower gradients to compensate for the reduced retention of the column. See Figure 6 for column conditions.
min7.5 10 12.5 15 17.5 20 22.5 25
100
200
Aeris WIDEPORE 3.6 µm XB-C18
Fully Porous 5 µm C18
Ap
p ID
20
69
0
Figure 6. Aeris WIDEPORE 3.6 µm Core-Shell vs. Fully Porous 1.7 µm
Substantially lower backpressure of the 3.6 µm core-shell column allow the use of longercolumns (in this case 250 x 2.1 mm) to get increased resolution of closely eluting proteins.
250 x 2.1 mm
Conditions for all columns:Column: Aeris WIDEPORE 3.6 µm XB-C18
Fully Porous 300 Å 1.7 µm C18Dimensions: as noted
Mobile Phase: A: Water with 0.1 % TFAB: Acetonitrile
Gradient: A/B (90:10) for 5 min to A/B (50:50) over 45 min
Flow Rate: 0.2 mL/minTemperature: 22 ºC
Injection Volume: 20 µLInstrument: Agilent® 1200SL
Detection: UV @ 210 nmSample: Degraded Myoglobin
min32 34 36 38 40 42
mAU
0
50
100
150
200
19824
160 bar
Even Greater Resolution on Your
UHPLC System
150 x 2.1 mm
Moving to a Longer Column
min32 34 36 38 40 42
mAU
0
50
100
150
200
AppID 19822
120 bar
Increased Resolution Using Core-Shell
Technology
Aeris WIDEPORE 3.6 µm XB-C18
Aeris WIDEPORE 3.6 µm XB-C18
150 x 2.1 mm
400 bar
Fully Porous 300 Å 1.7 µm C18
Ap
p ID
19
822
Ap
p ID
19
823
Ap
p ID
19
824
Comparative separations may not be representative of all applications.
Figure 7. Intact IgG Comparison
Separation of intact mouse monoclonal IgG antibody. Note the increased resolution andrecovery of IgG isomers for the core-shell column compared to a fully porous media.
min0 2
AppID 19852
4 6 8 10 12 14
mAU
02.5
57.510
12.515
17.520
Ap
p ID
19
852
Aeris WIDEPORE 3.6 µm XB-C18
min0 2 4 6 8 10 12 14
mAU
0
10
20
30
40
19841
Ap
p ID
19
841
Column: As noted.Dimensions: 100 x 4.6 mm ID
Mobile Phase: A: 0.1 % TFA in Water B: 0.085% TFA in Acetonitrile
Gradient: Time (sec) B % 0 10 1 20 15 40
Flow Rate: 1 mL/min Temperature: 80 ºC
Detection: UV @ 280 nm (ambient)
Fully Porous 300 Å 5 µm C18
Comparative separations may not be representative of all applications.
Figure 8. Intact IgG Separation of Core-Shell Media
Separation of bovine IgG on Aeris WIDEPORE 3.6 µm XB-C18. Good resolution and recoveryof IgG isoforms are obtained using relatively low column temperature (40 ºC) andacetonitrile-only organic mobile phase using the wide-pore core-shell column.
min14 15 16 17 18 19
mAU
0
10
20
30
40
Bovine serum IgG on Aeris WIDEPORE 3.6 µm 150 x 4.6 mm
Column: Aeris WIDEPORE 3.6 µm
Dimensions: 150 x 4.6 mmPart No.: 00F-4282-E0
Mobile Phase: A: Water with 0.1 % TFAB: Acetonitrile with 0.1 % TFA
Gradient: A/B (97:3) for 3 min to A/B (35:65) over 30 minFlow Rate: 1.2 mL/min
Temperature: 40 ºCInjection Volume: 100 µL
Instrument: Agilent® 1200Detection: DAD @ 214 nm
Sample: 1 mg/mL of bovine serum IgG in 0.1% TFA in water
Previous reports have shown that diffusion rates of proteins are inversely proportional to the log of the molecular weight of a protein; data confirms that slow diffusion of large proteins in and out of a porous particle is the limiting factor in minimizing peak width of a protein in reversed phase chromatography similar to previous results (Figure 1). The newest generation of core-shell media utilizes thin porous layers to minimize peak dispersion of large proteins while maintaining strong shape selectivity; separation of closely related modified proteins demonstrate similar separation with a concomitant reduction in protein peak width based on the shell thickness of the core-shell particle (Figure 2).
Optimizing shell thickness is a balance between loading and performance; a thinner shell has a shorter diffusion path with lower surface area, loading, and retention. A thicker shell material, or in this case fully porous material, allows for greater loading and retention but reduces performance due to the longer diffusion path (Figure 3).
Setting a uniform thickness of 0.2 µm seems an acceptable balance in the development of a optimal column for “analytical” where one would typically load in the 1 to 20 µg range (for a 4.6 mm ID column).
However, even when a thicker shell is used as is typically used for small molecule separations, the performance of core-shell columns outperform fully porous media columns when equal particle sized columns are compared
using small molecule probes (Figure 4).
Results show that the lower retentivity of the thin shell product requires significant reduction in organic content during gradient methods (resulting in subsequent reduction in electrospray efficiency); however, the increased resolution and efficiency of wide-pore core-shell columns far outweighs such losses with an increase in MS sensitivity for evaluated protein samples (Figure 5).
A n a d d i t i o n a l e x a m p l e i s s h o w i n Figure 6 where the w ide -pore core -she l l column is compared against a ful ly porous 1.7 µm 300 Å column showing improved resolution. However, since the wide-pore core-shell column utilizes a 3.6 µm particle size, the significantly lower backpressure of the core shell column allows for the use of longer columns allowing for additional resolution of similar components when run time is not a factor.
Optimized methods for IgG antibodies demonstrate separation of major disulfide-related isoforms with significant increases in protein recovery (Figures 7 and 8). The reduced retention of these media allow the use of acetonitrile as the organic mobile phase (versus IPA mixtures for fully porous media) as well as lower column temperatures which allow for different selectivity and resolution of post-translational modifications.
Results and Discussion
Conclusions
The development of wide-pore core-shell media specifically designed to overcome the dif fusion characteristics of large proteins allows for improved recovery and resolution of intact proteins. However, key considerations must be made in developing successful methods for intact protein analysis. Initial conditions and gradient slopes must be optimized to take into account the reduced hydrophobicity of wide-pore core-shell media. In addition, the media design necessitates its use for analytical applications specifically, as high loading
adversely impacts performance of both core-shell and fully porous media.
High resolution and recovery of intact proteins using wide-pore core-shell media allow for enhanced structural information of large proteins, especially recombinant antibody therapeutics where peptide mapping application potentially do not fully address all folding and structural heterogeneity of IgG based therapeutic drugs.
References
1. Mellors J.S., Jorgenson J.W; Anal. Chem., 76 (2004), 5441–5450
2. Thompson J.W., Mellors S., Jorgenson J.W., Eschelbach J.W; LC/GC North America, April 2006, 16
3. van Deemter J.J., Zuiderweg F.J., Klinkenberg A; Chem. Eng. Sci., 5 (1956), 271-289
4. Gritti, F., Leonardis, I., Shock, D.,Stevenson, P., Shalliker, A., Guiochon, G; J.Chromatogr A, 1217 (2010) , 1589-1603
5. Gutenwik J, Nilsson B, Axelsson A.; J. Chromatogr. A, 1048 (2004), 161-172
6. Davies P.A.; J. Chromatogr, 483 (1989), 221-237
7. McGinley M.D., Jarrett D., Layne J., Chitty M., Farkas T.; Chromatography Today 4 (2011) 33-36
8. Fekete S., Berky R, Fekete J., Veuthey J-L., Guillarme D.; J.Chromatogr A 1236 (2012) 177-188
TrademarksAeris is a trademark of Phenomenex. Agilent is a registered trademark of Agilent Technologies, Inc. AB SCIEX 4000 and API 3000 are trademarks of AB SCIEX Pte Ltd.AB SCIEX™ is being used under license.
Phenomenex is not affiliated with Agilent Technologies, Inc. Comparative separations may not be representative of all applications.
© 2012 Phenomenex, Inc. All rights reserved.