ELECTROPHORESISBASIC CONCEPTS,
INSTRUMENTATION, TECHNIQUE, TYPES AND APPLICATIONS
By
Dr. Basil, B – MBBS (Nigeria),Department of Chemical Pathology/Metabolic Medicine,
Benue State University Teaching Hospital, Makurdi.February, 2015.
INTRODUCTION
• ELECTROPHORESIS – migration of charged solutes or particles in a liquid medium under the influence of an electric field
• The aim of carrying out electrophoresis include:– To determine the number, amount and mobility of
components in a given sample or to separate them.– To obtain information about the electrical double layers
surrounding the particles.– Determination of molecular weight of proteins and DNA
sequencing.
BASIC CONCEPTS
• Iontophoresis – migration of small ions• Ionized species move towards the anode or cathode depending
on their charges• Ampholytes become +vely charged in a soln more acidic than
its isoelectric point (pI), and –vely charged in a more alkaline soln
• Rate of migration depends on:– Net electrical charge of the molecule– Size and shape of the molecule– Electrical field strength– Properties of the supporting medium– Temperature of operation
• Electrophoretic mobility (µ) – rate of migration (cm/s) per unit field strength (volts/cm)
µ = Q/6∏rȠWhere:µ = electrophoretic mobility in cm2/(V)(s)Q = net charge on the ionr = ionic radius of the soluteȠ = viscosity of the buffer soln
• Other factors affecting mobility:– Endosmotic flow: This is the preferential movement of water in
one direction through an electrophoresis medium due to selective binding of one type of charge on the surface of the medium
– Wick flow: Movement of water from the buffer reservoir towards the center of an electrophoresis gel or strip to replace water lost by evaporation
INSTRUMENTATION/REAGENTS
Power Supply:• Electrophoresis is done under conditions of constant
voltage, current or power.• Current (I) flow through the medium > heat production
due to resistance (R) > increase in thermal agitation of dissolved ions > decrease in resistance and increase in current flow >> more heat and evaporation of water from the buffer > increase in ionic concentration of the buffer
• Migration rate is kept constant by using a constant-current power supply.
• Since, EMF = IR, and increase in R will result in and increase in EMF (V) at constant current (I), thus, no appreciable heat production
Buffers:• Function include:
– Carries the applied current– Establishes the pH of the system– Determines the electrical charge on the solute
• The ionic strength influences:– Conductance of the support– Thickness of the ionic cloud (buffer and non-buffer ions)
surrounding a charged molecule– Rate of its migration– Sharpness of the electrophoretic zone
• Features of an ideal buffer:– Does not interfere with the ability to detect the analytes of interest– Maintains solubility of the analytes– Maintains buffering capacity through the analysis– Produce the desired separation
• Commonly used buffers include:– Low pH (acidic): phosphate, acetate, formate, citrate– High pH (basic): tris, tricine, borate, CAPS (N-cyclohexyl-3-
aminopropanesulfonic acid)• Higher the ionic strength (and concentration) > higher size of
ionic cloud > lower mobility of the particle• Also, higher ionic strength > sharper protein-band separation
and increased heat production > denaturation of heat-labile proteins
• Buffers used are made of monovalent ions because their valencies (ionic strength) and molality are equal.
• They are good culture media for microorganisms and should be refrigerated when not in use,
• Also cold buffer reduces evaporation and improves resolution• Small volumes should be discarded after use, but large
volumes can be reused up to 4 times
Support Media:• Insoluble gels e.g sheets, slabs or columns of starch, agarose
or polyacrylamide• Membranes (paper) of cellulose acetate• Pure buffer soln in a capillaryStarch Gel:• Separate macromolecules on basis of surface charge and
molecular size• Obsolete because preparation of reproducible starch gel is
difficultCellulose Acetate:• made by treating cellulose with acetic anhydride• They need to be soaked in buffer to soften them before use• Require clearing before densitometry• Hardly used in routine clinical application
Agarose:• Used in AGE for the separation of Serum, Urine or CSF
proteins, Hb variants, Isoenzymes, Lipoproteins etc.• Separation is based only on charge-to-mass ratio – pore
size is large enough for all proteins to pass• Separate proteins into only 5 fractions (zones): albumin,
alpha-1, alpha-2, beta- and gamma-globulins• Advantages:– Permits excellent densitometry – lower affinity for proteins
(migration is not affected) and native clarity after drying– Little Endosmosis – free of ionizable groups (neutral)
• Disadvantage:– DNA recovery is affected by inhibitors
Polyacrylamide:• thermostable, transparent, durable and relatively
chemically inert• No endosmosis – uncharged• Pore size does not allow larger proteins like fibrinogen,
B1-lipoprotiens, Y-globulins etc, to migrate• Separation is based on both charge-to-mass ratio and
molecular mass – molecular sieving• Carcinogenic – CAUTION when handling• Accommodates a large amount of sample in a single
sample slot• DNA recovered is pure with no inhibitors unlike Agarose.• Separates proteins into 20 or more fractions• Used to study individual proteins (e.g Isoenzymes)
Equipment used for Gel Electrophoresis:
TECHNIQUESeparation:• Blot the hydrated support medium to remove
excess buffer• The sample is added to the support for about 5min• The support is then placed into the electrophoretic
chamber in contact with the buffer• Apply a constant-current or constant-voltage power
for a specified time• The support is then removed and placed in a fixative
or rapidly dried to prevent diffusion of the sample
Staining:• The support is then stained and dried after washing out
the excess dye• Amido Black B or members of the Coomassie Brillant
Blue series are the commonest dyes• The amount of dye taken up is dependent on the type
of protein, degree of denaturation by the fixing agent and quality of the dye
• To visualize isoenzymes, incubate the gel in contact with a solution of substrate, which is linked structurally or chemically to a dye, before fixing.
• Typical stain solution may be used several times• When the stain solution becomes faulty, protein zones
will appear too lightly stained• It must be stored tightly covered to prevent evaporation
Detection and Quantification:• Detection can be achieved using UV light, but
quantification is by Densitometry• Densitometers integrate the area under a peak,
and the result is printed as percentage of the total• Reliable quantification of stained zones using
densitometry requires:– Light of an appropriate wavelength– Linear response from the instrument– Transparent background in the strip being scanned
• Mass spectrometers – determine the molecular weights of proteins and their cleavage products, and for peptide sequencing.
Blotting Techniques:• Southern Blotting – widely used in molecular biology:– DNA or DNA fragments separated by AGE– Strip of nitrocellulose or a nylon membrane is laid over the
gel blotting the DNA or DNA fragments onto it by capillary, electro-, or vacuum blotting
– Detection and identification is by hybridization with a labeled complementary nucleic acid probe.
• Northern Blotting:– Separates and detects RNAs– Uses labeled RNA probe for hybridization
• Western Blotting:– Separates and detects proteins– The membrane is reacted with a reagent containing an
antibody raised against the protein of interest
Common Problems Encountered:
TYPES OF ELECTROPHORESIS
• Zone Electrophoresis• Slab Gel Electrophoresis• Disc Electrophoresis• Isoelectric Focusing Electrophoresis• 2-Dimensional Electrophoresis• Capillary Electrophoresis• Microchip Electrophoresis
Zone Electrophoresis:• Produce zones of proteins that are heterogeneous
and physically separated from one another• Classified according to type and structure of the
support material e.g AGE, CAE, PAGE etc
Slab Gel Electrophoresis:• Use of a rectangular gel regardless of the thickness• Main advantage – ability to simultaneously separate
several samples in one run• Primary method used in clinical chemistry lab• Gels (usually agarose) may be cast on sheet of plastic
backing or completely encased within a plastic walled cell allowing horizontal or vertical electrophoresis and submersion for cooling, if needed.
• May be cast with additives like:– Ampholytes which create a pH gradient, or– Sodium dodecyl sulfate (SDS) that denatures protiens
Schematic representation of Slab gel electrophoresis
• Useful in the separation of serum proteins, isoenzymes, lipoproteins, hemoglobins, and fragments of DNA and RNA
Common problems encountered in SGE include:• Discontinuities in sample application – dirty applicator;
cleaned by agitating in water• Unequal migration of samples across the width of the gel –
dirty electrodes (uneven application of electric field), uneven wetting of the gel
• Distorted protein zones – bent applicator, incorporation of air bubble during sample application, over application of sample, excessive drying of support
• Unusual bands – artefacts, other causes e.g hemolysed samples > increased b-globulin band
• Atypical bands – denatured protein from deteriorated serum, also rule out a true paraprotein
Disc Electrophoresis:• 3-gel system:– small-pore separating gel (running gel),– a larger-pore spacer gel (stacking gel), – and a thin layer of large-pore monomer solution (sample gel) –
containing about 3 µL of serum• The different composition cause discontinuities in the
electrophoresis matrix• During electrophoresis, all proteins migrate easily through
the large-pore gels and stack up on the separation gel in a very thin zone
• This improves resolution and concentrates protein components at the border (or starting zone)
• Separation occurs at the bottom separation gel by the molecular sieve phenomenon
Schematic representation of Disc Electrophoresis
Isoelectric Focusing Electrophoresis:• Separates amphoteric compounds with increased resolution
in a medium possessing a stable pH gradient• The protein migrate to a zone in the medium where the pH of
the gel matches its pI• At this point, the charge of the protein becomes zero and its
migration ceases – it become “focused”.• Regions associated with a given pH are very narrow – enough
to separate proteins that differ in their pI values by only 0.02pH units
• A high voltage power source is needed because carrier ampholytes are used in relatively high concentrations
• Thus, the electrophoretic matrix must be cooled• IEF is used in neonatal screening programs to test for variant
Hb
2-Dimensional Electrophoresis:• Uses charge-dependent IEF (first dimension) and molecular
weight-dependent electrophoresis (in the second)• 1st dimension – carried out in a large-pore medium like
agarose or large-pore PAG; to which ampholytes are added to yield a pH gradient
• 2nd dimension is often polyacrylamide in a linear or gradient format
• It achieves the highest resolving power for the separation of DNA fragments– 1st dimension – normal AGE– 2nd dimension – ethidium bromide is added to the gel to open up
the fragments and cause changes in their mobility• Method of choice when complex samples need to be
arrayed for characterization, as in proteomics.
Schematic representation of 2D Electrophoresis
Capillary Electrophoresis:• Separation in narrow-bore fused silica capillaries (inner
diameter 25 – 75µm) filled with buffer – gel media can also be used
• Sample is loaded after filling capillary with buffer, and electric field applied
• Borate is a classic CE buffer that generates relatively low current and heat, even with a high ionic strength
• Electro-osmotic flow (EOF) controls the amount of time solutes remain in the capillary (also Electroendosmostic flow or endosmosis)– EOF = bulk flow of liquid towards the cathode upon application of an
electric field and it is superimposed on electrophoretic migration• Cations migrate fastest due to EOF and electrophoretic
attraction towards the cathode
• Neutral molecules are all carried by EOF but not separated from each other
• Anions move slowest because EOF is slightly greater than the attraction twds the anode and repulsion from cathode
• Used for determination of MWts of proteins and peptides, analysis of PCR products, inorganic ions, organic acids, pharmaceuticals, optic isomers, and drugs of abuse in serum and urine.
• When using a new capillary or changing buffer, the capillary must be equilibrated with the buffer – conditioning– Particularly important when phosphate-containing buffer is involved– This should be done for at least 4hrs before electrophoresis is
started• The capillary surface must be regenerated or reconditioned to
remove any material adsorbed onto the wall– Done by following each run with 3- to 5-column volume rinse using
NaOH, and flushing with 5- to 8-column volumes of fresh buffer.
Microchip Electrophoresis:
• Similar in principle to CE, but differs in that the separation channels, sample injection channels and reservoirs are all fabricated into the same planar substrate using photolithographic processes.
• More so, sample preparationand/or precolumn or postcolumn reactors, detectors, and excitation sources are intergrated into the chip
QUALITY CONTROL
• A Clear SOP must be adhered to during each run to ensure reproducibility
• Control serum must be included in each electrophoretic run to evaluate its quality and that of the densitometer.
APPLICATIONS• DNA Analysis:
– specific DNA sequences can be analyzed, isolated and cloned. – The analyzed DNA may be used in forensic investigations and
paternity tests.• Protein Analysis:
– In the diagnosis of conditions where levels of specific proteins or total protein is low or higher than normal e.g monoclonal gammopathy etc.
• Antibiotic Analysis:– Synthesis of new antibiotics– Analysis of bacteria response to antibiotics and determining
antibiotic-resistance• Vaccine Analysis:
– Purification, processing and analysis of vaccines e.g influenza vaccine, hepatitis vaccine, polio vaccine
REFERENCES:• Tietz Textbook of Clinical Chemistry and Molecular
Diagnostics, fifth ed., by Burtis et al.• Clinical Chemistry; Principles, Techniques and
Correlations, 7th ed., by Bishop et al.• http://www.slideshare.net/jyots23/electrophoresis-
presentation• http://www.bio-rad.com/en-us/applications-
technologies/protein-electrophoresis-methods• Landers JP. Molecular diagnostics on electrophoretic
microchips. Anal Chem 2003;75:2919-27• St. claire iii RL. Capillary electrophoresis. Anal Chem
1996;379-423
THANK YOU FOR LISTENING