ELECTROPHORESISBASIC CONCEPTS,

INSTRUMENTATION, TECHNIQUE, TYPES AND APPLICATIONS

By

Dr. Basil, B – MBBS (Nigeria),Department of Chemical Pathology/Metabolic Medicine,

Benue State University Teaching Hospital, Makurdi.February, 2015.

Page 2: Electrophoresis

INTRODUCTION

• ELECTROPHORESIS – migration of charged solutes or particles in a liquid medium under the influence of an electric field

• The aim of carrying out electrophoresis include:– To determine the number, amount and mobility of

components in a given sample or to separate them.– To obtain information about the electrical double layers

surrounding the particles.– Determination of molecular weight of proteins and DNA

sequencing.

Page 3: Electrophoresis

BASIC CONCEPTS

• Iontophoresis – migration of small ions• Ionized species move towards the anode or cathode depending

on their charges• Ampholytes become +vely charged in a soln more acidic than

its isoelectric point (pI), and –vely charged in a more alkaline soln

• Rate of migration depends on:– Net electrical charge of the molecule– Size and shape of the molecule– Electrical field strength– Properties of the supporting medium– Temperature of operation

Page 4: Electrophoresis

• Electrophoretic mobility (µ) – rate of migration (cm/s) per unit field strength (volts/cm)

µ = Q/6∏rȠWhere:µ = electrophoretic mobility in cm2/(V)(s)Q = net charge on the ionr = ionic radius of the soluteȠ = viscosity of the buffer soln

• Other factors affecting mobility:– Endosmotic flow: This is the preferential movement of water in

one direction through an electrophoresis medium due to selective binding of one type of charge on the surface of the medium

– Wick flow: Movement of water from the buffer reservoir towards the center of an electrophoresis gel or strip to replace water lost by evaporation

Page 5: Electrophoresis

INSTRUMENTATION/REAGENTS

Page 6: Electrophoresis

Power Supply:• Electrophoresis is done under conditions of constant

voltage, current or power.• Current (I) flow through the medium > heat production

due to resistance (R) > increase in thermal agitation of dissolved ions > decrease in resistance and increase in current flow >> more heat and evaporation of water from the buffer > increase in ionic concentration of the buffer

• Migration rate is kept constant by using a constant-current power supply.

• Since, EMF = IR, and increase in R will result in and increase in EMF (V) at constant current (I), thus, no appreciable heat production

Page 7: Electrophoresis

Buffers:• Function include:

– Carries the applied current– Establishes the pH of the system– Determines the electrical charge on the solute

• The ionic strength influences:– Conductance of the support– Thickness of the ionic cloud (buffer and non-buffer ions)

surrounding a charged molecule– Rate of its migration– Sharpness of the electrophoretic zone

• Features of an ideal buffer:– Does not interfere with the ability to detect the analytes of interest– Maintains solubility of the analytes– Maintains buffering capacity through the analysis– Produce the desired separation

Page 8: Electrophoresis

• Commonly used buffers include:– Low pH (acidic): phosphate, acetate, formate, citrate– High pH (basic): tris, tricine, borate, CAPS (N-cyclohexyl-3-

aminopropanesulfonic acid)• Higher the ionic strength (and concentration) > higher size of

ionic cloud > lower mobility of the particle• Also, higher ionic strength > sharper protein-band separation

and increased heat production > denaturation of heat-labile proteins

• Buffers used are made of monovalent ions because their valencies (ionic strength) and molality are equal.

• They are good culture media for microorganisms and should be refrigerated when not in use,

• Also cold buffer reduces evaporation and improves resolution• Small volumes should be discarded after use, but large

volumes can be reused up to 4 times

Page 9: Electrophoresis

Support Media:• Insoluble gels e.g sheets, slabs or columns of starch, agarose

or polyacrylamide• Membranes (paper) of cellulose acetate• Pure buffer soln in a capillaryStarch Gel:• Separate macromolecules on basis of surface charge and

molecular size• Obsolete because preparation of reproducible starch gel is

difficultCellulose Acetate:• made by treating cellulose with acetic anhydride• They need to be soaked in buffer to soften them before use• Require clearing before densitometry• Hardly used in routine clinical application

Page 10: Electrophoresis

Agarose:• Used in AGE for the separation of Serum, Urine or CSF

proteins, Hb variants, Isoenzymes, Lipoproteins etc.• Separation is based only on charge-to-mass ratio – pore

size is large enough for all proteins to pass• Separate proteins into only 5 fractions (zones): albumin,

alpha-1, alpha-2, beta- and gamma-globulins• Advantages:– Permits excellent densitometry – lower affinity for proteins

(migration is not affected) and native clarity after drying– Little Endosmosis – free of ionizable groups (neutral)

• Disadvantage:– DNA recovery is affected by inhibitors

Page 11: Electrophoresis

Polyacrylamide:• thermostable, transparent, durable and relatively

chemically inert• No endosmosis – uncharged• Pore size does not allow larger proteins like fibrinogen,

B1-lipoprotiens, Y-globulins etc, to migrate• Separation is based on both charge-to-mass ratio and

molecular mass – molecular sieving• Carcinogenic – CAUTION when handling• Accommodates a large amount of sample in a single

sample slot• DNA recovered is pure with no inhibitors unlike Agarose.• Separates proteins into 20 or more fractions• Used to study individual proteins (e.g Isoenzymes)

Page 12: Electrophoresis

Equipment used for Gel Electrophoresis:

Page 13: Electrophoresis

TECHNIQUESeparation:• Blot the hydrated support medium to remove

excess buffer• The sample is added to the support for about 5min• The support is then placed into the electrophoretic

chamber in contact with the buffer• Apply a constant-current or constant-voltage power

for a specified time• The support is then removed and placed in a fixative

or rapidly dried to prevent diffusion of the sample

Page 14: Electrophoresis

Staining:• The support is then stained and dried after washing out

the excess dye• Amido Black B or members of the Coomassie Brillant

Blue series are the commonest dyes• The amount of dye taken up is dependent on the type

of protein, degree of denaturation by the fixing agent and quality of the dye

• To visualize isoenzymes, incubate the gel in contact with a solution of substrate, which is linked structurally or chemically to a dye, before fixing.

• Typical stain solution may be used several times• When the stain solution becomes faulty, protein zones

will appear too lightly stained• It must be stored tightly covered to prevent evaporation

Page 15: Electrophoresis

Detection and Quantification:• Detection can be achieved using UV light, but

quantification is by Densitometry• Densitometers integrate the area under a peak,

and the result is printed as percentage of the total• Reliable quantification of stained zones using

densitometry requires:– Light of an appropriate wavelength– Linear response from the instrument– Transparent background in the strip being scanned

• Mass spectrometers – determine the molecular weights of proteins and their cleavage products, and for peptide sequencing.

Page 16: Electrophoresis

Blotting Techniques:• Southern Blotting – widely used in molecular biology:– DNA or DNA fragments separated by AGE– Strip of nitrocellulose or a nylon membrane is laid over the

gel blotting the DNA or DNA fragments onto it by capillary, electro-, or vacuum blotting

– Detection and identification is by hybridization with a labeled complementary nucleic acid probe.

• Northern Blotting:– Separates and detects RNAs– Uses labeled RNA probe for hybridization

• Western Blotting:– Separates and detects proteins– The membrane is reacted with a reagent containing an

antibody raised against the protein of interest

Page 17: Electrophoresis

Common Problems Encountered:

Page 18: Electrophoresis

TYPES OF ELECTROPHORESIS

• Zone Electrophoresis• Slab Gel Electrophoresis• Disc Electrophoresis• Isoelectric Focusing Electrophoresis• 2-Dimensional Electrophoresis• Capillary Electrophoresis• Microchip Electrophoresis

Page 19: Electrophoresis

Zone Electrophoresis:• Produce zones of proteins that are heterogeneous

and physically separated from one another• Classified according to type and structure of the

support material e.g AGE, CAE, PAGE etc

Page 20: Electrophoresis

Slab Gel Electrophoresis:• Use of a rectangular gel regardless of the thickness• Main advantage – ability to simultaneously separate

several samples in one run• Primary method used in clinical chemistry lab• Gels (usually agarose) may be cast on sheet of plastic

backing or completely encased within a plastic walled cell allowing horizontal or vertical electrophoresis and submersion for cooling, if needed.

• May be cast with additives like:– Ampholytes which create a pH gradient, or– Sodium dodecyl sulfate (SDS) that denatures protiens

Page 21: Electrophoresis

Schematic representation of Slab gel electrophoresis

Page 22: Electrophoresis

• Useful in the separation of serum proteins, isoenzymes, lipoproteins, hemoglobins, and fragments of DNA and RNA

Common problems encountered in SGE include:• Discontinuities in sample application – dirty applicator;

cleaned by agitating in water• Unequal migration of samples across the width of the gel –

dirty electrodes (uneven application of electric field), uneven wetting of the gel

• Distorted protein zones – bent applicator, incorporation of air bubble during sample application, over application of sample, excessive drying of support

• Unusual bands – artefacts, other causes e.g hemolysed samples > increased b-globulin band

• Atypical bands – denatured protein from deteriorated serum, also rule out a true paraprotein

Page 23: Electrophoresis

Disc Electrophoresis:• 3-gel system:– small-pore separating gel (running gel),– a larger-pore spacer gel (stacking gel), – and a thin layer of large-pore monomer solution (sample gel) –

containing about 3 µL of serum• The different composition cause discontinuities in the

electrophoresis matrix• During electrophoresis, all proteins migrate easily through

the large-pore gels and stack up on the separation gel in a very thin zone

• This improves resolution and concentrates protein components at the border (or starting zone)

• Separation occurs at the bottom separation gel by the molecular sieve phenomenon

Page 24: Electrophoresis

Schematic representation of Disc Electrophoresis

Page 25: Electrophoresis

Isoelectric Focusing Electrophoresis:• Separates amphoteric compounds with increased resolution

in a medium possessing a stable pH gradient• The protein migrate to a zone in the medium where the pH of

the gel matches its pI• At this point, the charge of the protein becomes zero and its

migration ceases – it become “focused”.• Regions associated with a given pH are very narrow – enough

to separate proteins that differ in their pI values by only 0.02pH units

• A high voltage power source is needed because carrier ampholytes are used in relatively high concentrations

• Thus, the electrophoretic matrix must be cooled• IEF is used in neonatal screening programs to test for variant

Hb

Page 26: Electrophoresis
Page 27: Electrophoresis

2-Dimensional Electrophoresis:• Uses charge-dependent IEF (first dimension) and molecular

weight-dependent electrophoresis (in the second)• 1st dimension – carried out in a large-pore medium like

agarose or large-pore PAG; to which ampholytes are added to yield a pH gradient

• 2nd dimension is often polyacrylamide in a linear or gradient format

• It achieves the highest resolving power for the separation of DNA fragments– 1st dimension – normal AGE– 2nd dimension – ethidium bromide is added to the gel to open up

the fragments and cause changes in their mobility• Method of choice when complex samples need to be

arrayed for characterization, as in proteomics.

Page 28: Electrophoresis

Schematic representation of 2D Electrophoresis

Page 29: Electrophoresis

Capillary Electrophoresis:• Separation in narrow-bore fused silica capillaries (inner

diameter 25 – 75µm) filled with buffer – gel media can also be used

• Sample is loaded after filling capillary with buffer, and electric field applied

• Borate is a classic CE buffer that generates relatively low current and heat, even with a high ionic strength

• Electro-osmotic flow (EOF) controls the amount of time solutes remain in the capillary (also Electroendosmostic flow or endosmosis)– EOF = bulk flow of liquid towards the cathode upon application of an

electric field and it is superimposed on electrophoretic migration• Cations migrate fastest due to EOF and electrophoretic

attraction towards the cathode

Page 30: Electrophoresis
Page 31: Electrophoresis

• Neutral molecules are all carried by EOF but not separated from each other

• Anions move slowest because EOF is slightly greater than the attraction twds the anode and repulsion from cathode

• Used for determination of MWts of proteins and peptides, analysis of PCR products, inorganic ions, organic acids, pharmaceuticals, optic isomers, and drugs of abuse in serum and urine.

• When using a new capillary or changing buffer, the capillary must be equilibrated with the buffer – conditioning– Particularly important when phosphate-containing buffer is involved– This should be done for at least 4hrs before electrophoresis is

started• The capillary surface must be regenerated or reconditioned to

remove any material adsorbed onto the wall– Done by following each run with 3- to 5-column volume rinse using

NaOH, and flushing with 5- to 8-column volumes of fresh buffer.

Page 32: Electrophoresis

Microchip Electrophoresis:

• Similar in principle to CE, but differs in that the separation channels, sample injection channels and reservoirs are all fabricated into the same planar substrate using photolithographic processes.

• More so, sample preparationand/or precolumn or postcolumn reactors, detectors, and excitation sources are intergrated into the chip

Page 33: Electrophoresis
Page 34: Electrophoresis

QUALITY CONTROL

• A Clear SOP must be adhered to during each run to ensure reproducibility

• Control serum must be included in each electrophoretic run to evaluate its quality and that of the densitometer.

Page 35: Electrophoresis

APPLICATIONS• DNA Analysis:

– specific DNA sequences can be analyzed, isolated and cloned. – The analyzed DNA may be used in forensic investigations and

paternity tests.• Protein Analysis:

– In the diagnosis of conditions where levels of specific proteins or total protein is low or higher than normal e.g monoclonal gammopathy etc.

• Antibiotic Analysis:– Synthesis of new antibiotics– Analysis of bacteria response to antibiotics and determining

antibiotic-resistance• Vaccine Analysis:

– Purification, processing and analysis of vaccines e.g influenza vaccine, hepatitis vaccine, polio vaccine

Page 36: Electrophoresis

REFERENCES:• Tietz Textbook of Clinical Chemistry and Molecular

Diagnostics, fifth ed., by Burtis et al.• Clinical Chemistry; Principles, Techniques and

Correlations, 7th ed., by Bishop et al.• http://www.slideshare.net/jyots23/electrophoresis-

presentation• http://www.bio-rad.com/en-us/applications-

technologies/protein-electrophoresis-methods• Landers JP. Molecular diagnostics on electrophoretic

microchips. Anal Chem 2003;75:2919-27• St. claire iii RL. Capillary electrophoresis. Anal Chem

1996;379-423