electrospray and atmospheric pressure chemical ionization quadrupole ms for small molecules

22
Electrospray and Electrospray and Atmospheric Pressure Atmospheric Pressure Chemical Ionization Chemical Ionization Quadrupole MS for Small Quadrupole MS for Small Molecules Molecules Jonathan A. Karty, Ph.D. Jonathan A. Karty, Ph.D.

Upload: ziarre

Post on 07-Feb-2016

50 views

Category:

Documents


0 download

DESCRIPTION

Electrospray and Atmospheric Pressure Chemical Ionization Quadrupole MS for Small Molecules. Jonathan A. Karty, Ph.D. LC-MS Picture. General Sample Guidelines. Purify analyte if possible Analyte should be 1-100 mg/L in concentration Colored samples should be translucent - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Electrospray and Atmospheric Electrospray and Atmospheric Pressure Chemical Ionization Pressure Chemical Ionization

Quadrupole MS for Small MoleculesQuadrupole MS for Small Molecules

Jonathan A. Karty, Ph.D.Jonathan A. Karty, Ph.D.

Page 2: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

LC-MS PictureLC-MS Picture

Page 3: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

General Sample GuidelinesGeneral Sample Guidelines Purify analyte if possiblePurify analyte if possible

Analyte should be 1-100 mg/L in concentrationAnalyte should be 1-100 mg/L in concentration Colored samples should be translucentColored samples should be translucent

Sample must have no particulates or precipitatesSample must have no particulates or precipitates

Use only volatile solvents/buffersUse only volatile solvents/buffers MeOH, HMeOH, H22O, acetone, CHO, acetone, CH33CN, THF, CHClCN, THF, CHCl33, etc., etc.

HOAc, HCOOH, NHHOAc, HCOOH, NH44OAc, NHOAc, NH33, etc. (weak, volatile), etc. (weak, volatile) Ionic strength < 20 mM is best (e.g. 0.1% v/v HOAc)Ionic strength < 20 mM is best (e.g. 0.1% v/v HOAc) 1 – 5 1 – 5 μμM NaOAc for acid/base labile samplesM NaOAc for acid/base labile samples

Need at least 50 Need at least 50 µµL for loop injectionL for loop injection If you need non-polar solvent for APCI, see Dr. If you need non-polar solvent for APCI, see Dr.

Karty firstKarty first

Page 4: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Agilent 6130 System DescriptionAgilent 6130 System Description Inlet is an Agilent 1200 HPLC SystemInlet is an Agilent 1200 HPLC System

Binary gradient (4 solvent bottles, only 2 can be used in an Binary gradient (4 solvent bottles, only 2 can be used in an analysis (A1 or A2 with B1 or B2, not A1 and A2)analysis (A1 or A2 with B1 or B2, not A1 and A2)

50-2,500 uL/min flow rates 50-2,500 uL/min flow rates Autosampler 0.5-100 Autosampler 0.5-100 μμL injection volumeL injection volume

6-column selection valve (loop or 1 of 5 HPLC columns)6-column selection valve (loop or 1 of 5 HPLC columns) CC1818 reversed phase is default column (H reversed phase is default column (H22O-CHO-CH33OH solvents)OH solvents)

Agilent 1200 DAD detectorAgilent 1200 DAD detector 190-800 nm UV-VIS detector190-800 nm UV-VIS detector Can record entire UV-VIS spectrum throughout a runCan record entire UV-VIS spectrum throughout a run

Electrospray Ionization-Atmospheric Pressure Chemical Electrospray Ionization-Atmospheric Pressure Chemical Ionization/Quadrupole (ESI/APCI-Q) mass spectrometerIonization/Quadrupole (ESI/APCI-Q) mass spectrometer 50 – 3,000 m/z range50 – 3,000 m/z range Can make ions by ESI, APCI, or bothCan make ions by ESI, APCI, or both Can alternate positive and negative ion modes during a runCan alternate positive and negative ion modes during a run

Page 5: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Air Sensitive MSAir Sensitive MS Source on 6130 MS is flushed constantly with Source on 6130 MS is flushed constantly with

nitrogen from a dewarnitrogen from a dewar Enables walk-up air senstive ESI/APCI MSEnables walk-up air senstive ESI/APCI MS

A syringe pump with air-tight “Sample-Lock” A syringe pump with air-tight “Sample-Lock” syringe can be used for air-sensitive analysessyringe can be used for air-sensitive analyses Requires special training from Dr. KartyRequires special training from Dr. Karty A separate syringe is needed for each sampleA separate syringe is needed for each sample

A “dummy” sample must still be placed in the A “dummy” sample must still be placed in the autosamplerautosampler

Remember to keep concentrations around 20 mg/LRemember to keep concentrations around 20 mg/L Colored solutions should be translucentColored solutions should be translucent

Page 6: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

HPLC-MS vs. Loop InjectionHPLC-MS vs. Loop Injection MSF uses the 1200 autosampler for nearly all MSF uses the 1200 autosampler for nearly all

samplessamples Must inject something, even if syringe pump is usedMust inject something, even if syringe pump is used

Sample can go through an empty loopSample can go through an empty loop Solution must contain only sample and volatile Solution must contain only sample and volatile

componentscomponents Allows 2.5 minute analysis timeAllows 2.5 minute analysis time

Sample can also be separated by HPLCSample can also be separated by HPLC Up to 5 columns are availableUp to 5 columns are available Currently only CCurrently only C1818 reversed phase available for general reversed phase available for general

useuse

Page 7: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Electrospray Ionization (ESI)Electrospray Ionization (ESI) Dilute solution of analyte (1 mg/L) infused Dilute solution of analyte (1 mg/L) infused

through a fine needle in a high electric fieldthrough a fine needle in a high electric field

Very small, highly charged droplets are createdVery small, highly charged droplets are created

Solvent evaporates, droplets split and/or ions Solvent evaporates, droplets split and/or ions ejected to lower charge/area ratio ejected to lower charge/area ratio Warm nebulizing gas accelerates dryingWarm nebulizing gas accelerates drying Basic sites are eventually protonated (M+H)Basic sites are eventually protonated (M+H)++

Ketones, esters, and alcohols can sodiate (M+Na)Ketones, esters, and alcohols can sodiate (M+Na)++

Free ions are directed into the vacuum chamberFree ions are directed into the vacuum chamber

Page 8: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

ESI PictureESI Picture

http://newobjective.com/images/electro/spraytip_bw.jpg

Page 9: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Advantages of ESIAdvantages of ESI

Gentle ionization processGentle ionization processHigh chance of observing intact molecular ionHigh chance of observing intact molecular ionVery labile analytes can be ionizedVery labile analytes can be ionizedNon-covalent complexes can be studied by ESINon-covalent complexes can be studied by ESI

Molecule need not be volatileMolecule need not be volatileProteins/peptides easily analyzed by ESIProteins/peptides easily analyzed by ESISalts can be analyzed by ESISalts can be analyzed by ESI

Easily coupled with HPLCEasily coupled with HPLC Both positive and negative ions can be Both positive and negative ions can be

generated by the same sourcegenerated by the same source

Page 10: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Characteristics of ESI IonsCharacteristics of ESI Ions ESI is a thermal process (1 atm in source)ESI is a thermal process (1 atm in source)

Little fragmentation due to ionization (cf EI) Little fragmentation due to ionization (cf EI) Solution-phase ions are often preservedSolution-phase ions are often preserved

e.g. organometallic saltse.g. organometallic salts ESI ions are generated by ion transferESI ions are generated by ion transfer

(M+H)(M+H)++, (M+Na), (M+Na)++, or (M-H), or (M-H)--, rarely M, rarely M++•• or M or M--••

ESI often generates multiply charged ionsESI often generates multiply charged ions (M+2H)(M+2H)2+2+ or (M+10H) or (M+10H)10+10+

Most ions are 500-1500 m/zMost ions are 500-1500 m/z ESI spectrum x-axis must be mass/charge (m/z or Th, ESI spectrum x-axis must be mass/charge (m/z or Th,

not amu or Da)not amu or Da)

Page 11: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

ESI DisadvantagesESI Disadvantages Analyte must have an acidic or basic siteAnalyte must have an acidic or basic site

Hydrocarbons and steroids not readily ionized by ESIHydrocarbons and steroids not readily ionized by ESI

Analyte must be soluble in polar, volatile solventAnalyte must be soluble in polar, volatile solvent ESI is less efficient than other sourcesESI is less efficient than other sources

Most ions don’t make it into the vacuum systemMost ions don’t make it into the vacuum system

ESI is very sensitive to contaminantsESI is very sensitive to contaminants Solvent clusters can dominate spectraSolvent clusters can dominate spectra

Distribution of multiple charge states can make Distribution of multiple charge states can make spectra of mixtures hard to interpretspectra of mixtures hard to interpret e.g. polymer mass spectrae.g. polymer mass spectra

Page 12: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

What Samples Can Be Analyzed by What Samples Can Be Analyzed by ESI-QMS?ESI-QMS?

Non-volatile organic moleculesNon-volatile organic molecules Should be soluble in polar, volatile solventsShould be soluble in polar, volatile solvents Molecule must be ionizable (MS detects ions)Molecule must be ionizable (MS detects ions)

R-NHR-NH22, R-CO, R-CO22H, R-HSOH, R-HSO33, R-OH, R-H, R-OH, R-H22POPO33 work best for ESI work best for ESI Aromatics, ketones, protected heteroatoms work well by APCIAromatics, ketones, protected heteroatoms work well by APCI

Organometallic complexesOrganometallic complexes Organometallic salts work especially wellOrganometallic salts work especially well Source is flooded with NSource is flooded with N22 gas from a dewar gas from a dewar

HPLC-MSHPLC-MS With appropriate mobile phasesWith appropriate mobile phases

Page 13: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

What Samples Are Inappropriate What Samples Are Inappropriate for ESI-QMS analysis?for ESI-QMS analysis?

Samples in non-polar or non-volatile Samples in non-polar or non-volatile solvent:solvent:Hexane, benzene, CHHexane, benzene, CH22ClCl22, DMSO, etc., DMSO, etc.Run these in APCI or mixed modeRun these in APCI or mixed mode

Buffer systems incompatible with ESIBuffer systems incompatible with ESI6M urea, 10% glycerol, 0.1 M NaH6M urea, 10% glycerol, 0.1 M NaH22POPO44, TBAF, , TBAF,

(involatile)(involatile)Strong acid solutions (too conductive)Strong acid solutions (too conductive)detergentsdetergents

Molecules that have no ionizable groupsMolecules that have no ionizable groups

Page 14: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

O

O

O

O

OTBDPS

C

O

O

Me

OTIPS

OMOM

Me

Me

OTIPS

OTBS

Me

Na+

O

O

O

HO

OTBDPS

C

O

O

Me

OTIPS

OMOM

Me

Me

OTIPS

OTBS

Me

Na+

OH

Chemical Formula: C77H136NaO13Si4+

Exact Mass: 1403.9m/z: 1404.9 (100.0%), 1403.9 (94.7%), 1405.9 (67.1%), 1406.9 (33.1%), 1407.9 (13.1%),

1408.9 (4.3%), 1409.9 (1.2%)

Chemical Formula: C77H134NaO12Si4+

Exact Mass: 1385.9m/z: 1386.9 (100.0%), 1385.9 (94.7%), 1387.9 (66.9%), 1388.9 (32.8%), 1389.9 (12.9%), 1390.9 (4.2%), 1391.9 (1.2%)

H2O

ESI ExampleESI Example

O

O

O

HO

OTBDPS

C

O

O

Me

OTIPS

OMOM

Me

Me

OTIPS

OTBS

Me

Na+

OH

O

O

O

O

OTBDPS

C

O

O

Me

OTIPS

OMOM

Me

Me

OTIPS

OTBS

Me

Na+

78%22%

Page 15: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Atmospheric Pressure Chemical Atmospheric Pressure Chemical Ionization (APCI)Ionization (APCI)

APCI uses a corona discharge to generate acidic APCI uses a corona discharge to generate acidic solvent cations from a vaporsolvent cations from a vapor Ionizing reagent often CHClIonizing reagent often CHCl33++ or CH or CH33OHOH++

These solvent cations can protonate hydrophobic These solvent cations can protonate hydrophobic species not amenable to ESIspecies not amenable to ESI APCI can be done from hexane or THFAPCI can be done from hexane or THF Often used to study lipids and steroidsOften used to study lipids and steroids In MSF, completely protected macrocycles are routinely In MSF, completely protected macrocycles are routinely

studied by APCIstudied by APCI

APCI is a little harsher than ESIAPCI is a little harsher than ESI Labile molecules might fragmentLabile molecules might fragment

Page 16: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

APCI DiagramAPCI Diagram

http://imaisd.usc.es/riaidt/masas/imagenes/apci1.jpg

Page 17: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

APCI ExampleAPCI Example

Page 18: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Agilent 6130 Multi-mode SourceAgilent 6130 Multi-mode Source

http://www.chem.agilent.com/Library/Images1/MMS_schematic_300dpi_039393.jpg

Page 19: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Cluster and Background IonsCluster and Background Ions The low m/z (<300) region of the mass spectrum is The low m/z (<300) region of the mass spectrum is

often dominated by solvent cluster ionsoften dominated by solvent cluster ions These are distinguished by low mass defects due to NaThese are distinguished by low mass defects due to Na++

and Kand K++ incorporation incorporation Masses are XXX.0 or XXX.9Masses are XXX.0 or XXX.9

Compounds from previous samples can coat the Compounds from previous samples can coat the source leading to persistent “background” ionssource leading to persistent “background” ions e.g. 242.1 for tetrabutylammoniume.g. 242.1 for tetrabutylammonium Might see large ion from previous person’s runMight see large ion from previous person’s run

To confirm an ion is “background” try injecting a To confirm an ion is “background” try injecting a blank of ESI solution with your solvent systemblank of ESI solution with your solvent system

Page 20: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

About the Fragmentor…About the Fragmentor… There is a region in the source where ion-There is a region in the source where ion-

nitrogen collisions can cause fragmentationnitrogen collisions can cause fragmentation This occurs between the end of the heated capillary This occurs between the end of the heated capillary

and the first skimmer coneand the first skimmer cone

The voltage difference between capillary and The voltage difference between capillary and skimmer is called the “Fragmentor” voltageskimmer is called the “Fragmentor” voltage

Fragmentor voltage has 2 postive main effectsFragmentor voltage has 2 postive main effects It increases the number of ions that make it to the quadIt increases the number of ions that make it to the quad It can knock non-covalent attached neutralsIt can knock non-covalent attached neutrals

However, a high fragmentor value can cause However, a high fragmentor value can cause labile groups to dissociatelabile groups to dissociate Phospho-diesters can be rather vulnerable to thisPhospho-diesters can be rather vulnerable to this

Page 21: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

A Word About QuantificationA Word About Quantification LC-MS is a quantitative techniqueLC-MS is a quantitative technique Use only 1 m/z when quantifying a compoundUse only 1 m/z when quantifying a compound Intensity is proportional to concentrationIntensity is proportional to concentration

I I αα [X] [X] α is unique to each compoundα is unique to each compound The more two compounds differ chemically, The more two compounds differ chemically,

the more careful one must be when comparing the more careful one must be when comparing their intensitiestheir intensities

Ideally a calibration curve is constructed using Ideally a calibration curve is constructed using multiple solutions of pure analyte at varying multiple solutions of pure analyte at varying concentrationsconcentrations

Page 22: Electrospray  and Atmospheric Pressure Chemical Ionization  Quadrupole  MS for Small Molecules

Hands-on TrainingHands-on Training

Training starts AFTER 11/7/10Training starts AFTER 11/7/10Groups of no more than threeGroups of no more than threeOne hour or so to completeOne hour or so to completeNo charge for first sessionNo charge for first sessionAfter training, students must demonstrate After training, students must demonstrate

competency by running their own samples competency by running their own samples prior to being granted after-hours accessprior to being granted after-hours access