emg lecture

Upload: keri-gobin

Post on 02-Jun-2018

241 views

Category:

Documents


1 download

TRANSCRIPT

  • 8/10/2019 emg lecture

    1/26

    7th LectureDimitar Stefanov

  • 8/10/2019 emg lecture

    2/26

    RecappingThree types electrodes are used for sensing of EMG signals:

    1. indwelling (intramuscular) electrodes (single fiber electrodes, monopolar

    electrodes, concentric electrodes)

    2. Wire electrodes

    3. surface electrodesnon-invasive recordings

    Potential of surface electrode (V)

    Differential voltage waveform

  • 8/10/2019 emg lecture

    3/26

    Velocity of propagation of the m.a.p.4 m/s

    There is a delaybetween the EMG and muscle contraction (30-80 milliseconds).

    In case of isometric muscle tension, a linear dependencybetween the muscle tensionand the rectified EMG output is observed.

    Important parameters of the EMG amplifiers:1. Gain and dynamic range

    2. Input impedance

    3. Frequency response4. Common mode rejection.

    EMG signal:

    contains certain level of noises

    has specific spectral densityfunction.

    Fatigue

    (1) If we assume that the EMG is stimulation rate remains constant then the muscle

    tension deceases in case of fatigue.

    (2) The shape of the m.a.p. is altered in case of fatigue.(3) tremor occurs.

  • 8/10/2019 emg lecture

    4/26

    Problem with the electrodes: polarization

    The electric conductivity of the body involves ions as charge carrier.

    Electrodes can be considered as electrical conductors in contact with the aqueous

    ionic solutions of the body.The interaction between electrons in the electrodes and ions in the body can affect

    the EMG signal

    Half -cell potential(HCP) is called thepotential differencebetween the metal of

    the electrode and the bulk of the electrolyte.

    HCP depends on the ionic concentration

    HCP can be measured when no electric current flows between an

    electrode and the electrolyte

  • 8/10/2019 emg lecture

    5/26

    Problem with the electrodes: polarization

    Polarizationarises in case when current flows between the electrodeand the solution.

    Perfectly polarizable electrodesno actual current crosses the

    electrode- electrolyte interface

    Nonpolarized electrodesallow the current to pass freely in

    electrode-electrolyte interface.

    Silversilver chloride electrode (Ag/AgCl)it possesses characteristics which

    are similar to a perfect nonpolarizable electrode.

  • 8/10/2019 emg lecture

    6/26

    AgCl film

    insulated lead wire

    Ag metal

    Ag lead wiresintered Ag and AgCl

    (Ag and AgCl powder mechanically pressed)

    greater mechanical stability

    Silversilver chloride electrodes

    Low noise electrodes

  • 8/10/2019 emg lecture

    7/26

    Equivalent circuit of a biopotential electrode

    Ehchalf-cell potential

    Rdand Cdrepresent the impedance associated with the electrode-electrolyte

    interfaceRsseries resistance.

    Biopotential electrode impedance

    as a function of frequency

  • 8/10/2019 emg lecture

    8/26

    EMG amplifiers

    Amplifier gainthe ratio of the output voltage to the input voltage

    Surface EMG electrodes- maximum amplitude of 5 mV peak-to-peak

    Indwelling electrodesamplitude of up to 10 mV

    Single m.a.p. electrodesamplitude of 100 mV

    Amplitudes of the EMG signal :

    Noise level of the amplifieris the amplitude of the higher frequency random

    signal on the output of the amplifier when the electrodes are shorten together.

    Noise level of the amplifier should not exceed 50 mV,

    (preferably 20mV).

  • 8/10/2019 emg lecture

    9/26

    Input impedance of an amplifier of biosignals

    The resistance of the electrode-skin interface depends on:

    thickness of the skin layer,

    the cleaning of the skin prior to the attachment of the electrodes,the area of the electrode surface,

    temperature.

    Electrode pastedecreases the resistance

    between the electrode

    and the skin.

  • 8/10/2019 emg lecture

    10/26

    Input impedance of an amplifier of biosignals

    The capacitancebetween the electrode and the skin causes

    frequency distortions.

    EMG amplifiers should possess high input resistance

  • 8/10/2019 emg lecture

    11/26

    Frequency response of the EMG amplifier

    Frequency bandwidth

    All frequencies presentin the EMG should be

    amplified at one and the

    same level.

    Bandwidththe difference between upper cutoff frequencyf2and the lower cutoff

    frequencyf1.The gain of the amplifier atf1andf2is 0.707 from the gain of the gain in the mid-

    frequency region (half-power).

    Amplifier gain:

    Example: linear gain 1000, or 60 dB; gain at the cutoff frequencies

    57 dB (3dB less than that atthe mid-frequencies).

  • 8/10/2019 emg lecture

    12/26

    The EMG amplifier should amplify equally all EMG frequency components.

    Most of the EMG signalsare concentrated in the band

    between 20 and 200 Hz.

  • 8/10/2019 emg lecture

    13/26

    Recommended range of the EMG amplifiers:

    from 10 Hz to 1000 Hzwhen the signal is collected with surface

    electrodes;

    from 20 Hz to 2000 Hzwhen the signal is collected withindwelling electrodes.

    Interferences:

    Hum from power line (60 Hz in the USA and 50 Hz in Europe)-in the middle of the EMG spectrum

    Movement artifactstheir frequency lies in the 0 to 10 Hz range

    dont cause big problems

    Noise from low quality cabling systemsinterfere with thebaseline of the EMG signal; can be eliminated by good low

    frequency filtering (by setting of f1 to about 20 Hz).

  • 8/10/2019 emg lecture

    14/26

  • 8/10/2019 emg lecture

    15/26

    Single-ended amplifier

    Differential amplifier

    A perfect subtraction

    never occurs.

  • 8/10/2019 emg lecture

    16/26

    Common mode rejection ratio (CMRR)

    CMRR is measured in dB.

    In good quality EMG amplifiers CMRR should be 10,000 (80 dB) or higher.

  • 8/10/2019 emg lecture

    17/26

    Processing of EMG

    Example:

    1. Half of full-wave rectification (absolute value)

    2. Linear envelope (low-pass filtering of the rectified signal)

    main decision here is the choice of the low pass filter!

    3. Integration of the signal from (2)over the period of the musclecontractionarea under the curve

    4. Integration of the signal from (2) for a fixed time, reset to zero,

    and repeating the integration cyclesuch scheme represents the

    trend of the EMG amplitude with time

    5. Integration of the signal from (2) to a present level, reset to zero,

    and repeating the integration cyclerepresents the level of the

    muscle activity (high or low muscle activity).

  • 8/10/2019 emg lecture

    18/26

    Diagram of several common EMG processing systems and the processing results

  • 8/10/2019 emg lecture

    19/26

    Biopotential amplifiers

    Basic amplifier requirements:1. The physiological process to be monitored should not be influenced in any way

    by the amplifier

    2. The measured signal should be not distorted

    3. The amplifier should provide the best possible separation of signal and

    interferences

    4. The amplifier should offer protection of the patient from any hazard and electric

    shock

    5. The amplifier should be protected against damages due to high input voltages.

    The input signal to the amplifier consists of 5 components:

    1. Desired biopotential

    2. Undesired biopotentials

    3. A power line interference signal and its harmonics

    4. Interference signals generated by the tissue-electrode interface

    5. Noise.

  • 8/10/2019 emg lecture

    20/26

    Block diagram of a biopotential amplifier

    FET transistorsGalvanic

    decoupling of

    the patient

    Motion artifactsthe contact between the electrode and the tissue changes

    during the relative motions between the electrodes and the tissue.

    Measures for decreasing the motion artifacts:

    High input resistance of the amplifier

    Usage of non-polarized electrodes (Ag/AgCl)

    Reduction of the source impedance by usage of electrode gel.

    Artifactsdue to electric and magnetic fieldsExample.

  • 8/10/2019 emg lecture

    21/26

    Amplitude/frequency characteristics of the

    bioamplifiers used in different applications

    S i l i i hi h b il h bi i l lifi

  • 8/10/2019 emg lecture

    22/26

    Special circuits which built the biopotential amplifier

    Instrumentation amplifiers

    DCinstrumentation amplifiers

  • 8/10/2019 emg lecture

    23/26

    AC instrumentation amplifiers

    AC amplifiers eliminate the electrode offset potential, permit high gain and

    permits higher CMRR.

    The capacitors between the electrodes and the input stage of the amplifier causecharging effects from the input bias current.

  • 8/10/2019 emg lecture

    24/26

    Isolation amplifier

    Isolation is realized in the following technologies:Transformer isolation

    Opto-isolation.

    Isolation provides a complete galvanic separation between the input stage

    (patient) and the other part of the measure equipment.

  • 8/10/2019 emg lecture

    25/26

    Surge protection of the bioamplifiers

    Protection of the amplifier from damage due to surge input potentials.

    Diodes

    Zener diodes

    Gas-discharge tubes

  • 8/10/2019 emg lecture

    26/26

    Input guarding

    Instrumentation

    amplifier providing

    input guarding

    Technique for increase both the input impedance of the amplifier of biopotentials and

    the CMRR

    Dr iven-r ight-leg circui t

    reducing common-mode

    interference