ا�ل�ط�ب�ي ا�ل�ت�ص�و�ي�ر�MEDICAL IMAGING

ع�ب�د� ا�ل�د�ي�ن� ع�ل�م� ا�ل�م�ه�ن�د�س�ا�ل�د�ج�ا�ن�ي ا�ل�س�م�ي�ع�

[email protected]

Page 3: Introduction of imaging dept

 

Techniques/ Application

ANATOMY

تشريحيPHYSIOLOGY

وظيفيMETABOLISM

االيضMOLECULAR

جزيئات

االشعاعيRadiation

38% 

الصوتي الفوقUltrasonic

38% 

المغناطيسيMagnetic Fields 82%  

النووي Nuclear   94%

الضوئي Optical   88%  

MEDICAL IMAGING IMAGING FIELDS & TECHNIQUES IN

HUMAN BODY

Eng.Alam Dajani 00966592601260

Page 4: Introduction of imaging dept

WORK FLOW :(PATEINTS,TECHNICIAN,SERVICE..)

FLEXBILITY :(DESIGN & FUTURE PLANS) EQUIPMENT: (TECHNOLOGY,SPECIFICATIONS, INSTALLATION…)PROTECTION (SAFETY ) :

HUMAN,ENVIREMENT,EQUIPMENT.DATA FLOW & SOFTWARE:

(ACQUIRE,TRANSFER,STORE …) PACS,LAN,WAN,HL7,HIS,RIS…

DEPARTMENT DESIGN CONSEDRATIONS MEDICAL IMAGING DEPARTMENT

Eng.Alam Dajani 00966592601260

Page 5: Introduction of imaging dept

DI ( central)

ER

HOSP.MAIN ENTRANCEOPD

IN PAT .

LOCATION : IN THE HEART OF THE HOSPITAL DEPARTMENT DESIGN CONSEDRATIONS

Eng.Alam Dajani 00966592601260

Page 6: Introduction of imaging dept

FRONT ROUTINE/SHORT DURATION

          

                         

BACK  

COMPLEX / LONG DURATION

EQUIPMENT DISTRIBUTION DEPARTMENT DESIGN CONSEDRATIONS

OPD INPAT.

INTERVENTIONAL RAD. ANGI.O+LITHOTRIPSY MRI

Eng.Alam Dajani 00966592601260

Page 7: Introduction of imaging dept

Eng.Alam Dajani 00966592601260

Page 8: Introduction of imaging dept

DATA FLOW DEPARTMENT DESIGN CONSEDRATIONS

Eng.Alam Dajani 00966592601260

Page 9: Introduction of imaging dept

PRE INSTALLATIONS SCOPE : MODILITIES PROVIDORES WITH COORDINATION WITH CLEINT(PMO & CONSLTANT) & MAIN CONTRACTOR

CEILINGS : NOT USED FOR ELECTRICAL OR MECHNICALS SYSTEMS (HVAC,PLUMBING ..)

WALLES & CEILING : SOUND INSULATION esp.(MRI) CLADDING FLOORS : VINAL USING WELDING PAPERS IN THE JOINTS . DOORS : app. WIDE 240cm & THICKNESS 4.45cm . EQUIPMENT WEIGHT : HEAVY SOME OF THEM AROUND TONES SHEILDING & LEADING : SHOULD BE DEPONDS ON THE TYPE ,

STRENGTH & TECHNOLOGY USED FEAR VOBIA esp. IN MRI & CT WINGS : FIXED SCREEN IN CEILING ,

SOFT MUSIC , AMASING FINISHING ….) . WALL HEIGHT : MIN. 3 m.

GENERAL HINTS

MEDICAL IMAGING DEPARTMENT

PRE INSTALLATION CONSEDRATIONS

Eng.Alam Dajani 00966592601260

Page 10: Introduction of imaging dept

1.GENERAL X-RAY , MOBILE , PANORAMA , BMD & MAMOGRAPH

2.RADIOLOGY &FLOUROSCOPY (R&F)

3.ANGIOGRAPHY & IR (INTERVENTIONAL RADIOLOGY)

4.CT – Computed Tomography5. U/S - ULTRASOUND 6.MRI- Magnetic Resonance Imaging

MEDICAL IMAGING DEPARTMENT EQUIPMENT & FACILITIES

Eng.Alam Dajani 00966592601260

Page 11: Introduction of imaging dept

7.GAMMA CAMERA

8. SPECT ( SINGLE PHOTON EMMISION COMPUTED TOMOGRAPHY)

9. PET (POSITRON EMMISION TOMOGRAPHY(Positron (+ charge) = Anti Electron)

MEDICAL IMAGING DEPARTMENT EQUIPMENT & FACILITIES

Eng.Alam Dajani 00966592601260

Page 12: Introduction of imaging dept

10. PET MRI (POSITRON EMMISION TOMOGRAPHY W/ MAGENTIC RESONANCE IMAGING

11. PET CT (POSITRON EMMISION TOMOGRAPHY W/COMPUTED TOMOGRAPHY)

MEDICAL IMAGING DEPARTMENT EQUIPMENT & FACILITIES

Eng.Alam Dajani 00966592601260

Page 13: Introduction of imaging dept

GENERAL X-RAY

Eng.Alam Dajani 00966592601260

Page 14: Introduction of imaging dept

GENERAL X-RAY

1. USING & SPECIFICATIONS

For imaging the body organs, especially the chest and skull and other parts of the skeleton (fractures and injuries).

Ionized rays used in imaging operations where X is meted into force dependent on the member "of different density and the amount of permeability and absorption.

These devices are used in addition to the emergency department radiography.

For imaging very few (seconds). There are risks for some patients, especially the pregnant (the

fetus).You do not need imaging operations for special preparations

or to any other devices help.Eng.Alam Dajani 00966592601260

Page 15: Introduction of imaging dept

The device consists of: Generator is characterized by high energy of up to 80 000-watt and high-frequency ; CRT (Cathode Ray Tube) &The detector which are received after the X-ray permeability of the body of the patient.

All parts are suspended on ceiling with supports or props. Imaging by two cases : using bed ( lying down ) and standing

up using the board so that it is receiving the scan window therein. Device location needs special equipment and install the sealing

of the pillars to carry the device and ease of movement directions (X, Y, Z) and rotation (TUBE).

GENERAL X-RAY2. USING & SPECIFICATIONS

Eng.Alam Dajani 00966592601260

Page 16: Introduction of imaging dept

GENERAL X-RAY

Eng.Alam Dajani 00966592601260

Page 17: Introduction of imaging dept

ا�ال�ش�ع�ا�ع�ي� ا�ل�ت�ص�و�ي�ر� أ�ج�ه�ز�ة�GENERAL X-RAY

ENGR. ALAM DAJANI

Eng.Alam Dajani 00966592601260

Page 18: Introduction of imaging dept

FACILITY SPACES

MODILITY SPACES - METER SQR

STND PR.RM EQ.RM CN.RM

Radiography General (standard) 17-28 17-28  X 14

Mammography (standard) 15  X  X X 

GENERAL X-RAY

Eng.Alam Dajani 00966592601260

Page 19: Introduction of imaging dept

GENERAL X-RAY

PROTECTION

WARNING SIGNS

ON X RAY DOOR

YES FOR PERGNANT

MODILITY LEADING

WALLS DOORS CONTROL WINDOW

Radiography General YES YES YES

Mammography NO NO X

Eng.Alam Dajani 00966592601260

Page 20: Introduction of imaging dept

ELETRICAL NEEDS

MODILITY ELECTRICAL REQUIR.

LIGHTING POWER TYPE

Radiography General (standard) FLUORECENT 50CF & DIMMER TO 5 CF 380-3P 150A CB EMERGENCY

Mammography (standard) FLUORECENT 40CF 250-2P 30A CB EMERGENCY

GENERAL X-RAY

Eng.Alam Dajani 00966592601260

Page 21: Introduction of imaging dept

MECHNICAL NEEDS 1

MODILITY HVAC TEMP HUM AC AIR

CHANGE/H PRESS. EXH.

Radiography General 21-24 30-60 CENTRAL 6 POSITIVE NO

Mamograph 21-24 30-60 CENTRAL 6 POSITIVE NO

GENERAL X-RAY

Eng.Alam Dajani 00966592601260

Page 22: Introduction of imaging dept

MECHNICAL NEEDS 2

MODILITY PLUMBING M.GASES

WATER(C/H) DRAIN O2 VAC AIR

Radiography General √ √ √ √ √

Mammography √ √ NO NO NO

GENERAL X-RAY

Eng.Alam Dajani 00966592601260

Page 23: Introduction of imaging dept

RADIO &FLUOROSCOPY

Eng.Alam Dajani 00966592601260

Page 24: Introduction of imaging dept

USING & SPECIFICATION

RADIO &FLUOROSCOPY

For imaging the body organs, especially the gut the top and bottom, bladder and esophagus.

Imaging member while in motion and directly (real time - no recording). Directly and fixed are shed rays to be continuous imaging process. The patient is injected contrast media to show the user to be filmed or

through ingestion. Imaging for longer and at greater than normal radiographic radiation. There are risks for some patients, especially the pregnant (the fetus).

Imaging need for special operations and preparations to some assistive devices as organs injection and monitoring of vital signs.

Eng.Alam Dajani 00966592601260

Page 25: Introduction of imaging dept

IR & ANGIOGRAPHY

Eng.Alam Dajani 00966592601260

Page 26: Introduction of imaging dept

USING & SPECIFICATION

IR & ANGIOGRAPHY

These devices are used in operations Cardiac Catheterization and Angiography and some operations that need put stends or Catheters ..., where the device uses a multi-purpose format (C-ARM) with bilateral levels (both sides and axial), in addition to the fact that comics and vivid.

A key device in such operations and be one of the devices such as electronic monitoring devices and anesthesia ....

This type of operations have radiation risk and need special preparations .

Eng.Alam Dajani 00966592601260

Page 27: Introduction of imaging dept

R/F ,IR & ANGIOGRAPHY

Eng.Alam Dajani 00966592601260

Page 28: Introduction of imaging dept

FACILITIES SPACES

R/F ,IR & ANGIOGRAPHY

MODILITY SPACES - METER SQR

STND PR.RM EQ.RM CN.RM

RADIOGRAPHIC /Fluoroscopy (R/F) 30 PATIENT TOILET W/NEGATIVE PRESSURE

ANGIO / IR /HYBRID OR 47 47 12 12

Eng.Alam Dajani 00966592601260

Page 29: Introduction of imaging dept

PROTECTION

WARNING SIGNS

ON X RAY DOOR

YES FOR PERGNANT

R/F ,IR & ANGIOGRAPHY

MODILITY LEADING

WALLS DOORS CONTROL WINDOW

RADIOGRAPHIC /Fluoroscopy (R/F) (standard) YES YES YES

ANGIO / IR /HYBRID OR YES YES YES

Eng.Alam Dajani

Page 30: Introduction of imaging dept

ELECTRICAL NEEDS

R/F ,IR & ANGIOGRAPHY

MODILITY ELECTRICAL REQUIR.

LIGHTING POWER TYPE

RADIOGRAPHIC /Fluoroscopy(R/F) (standard) FLUORECENT 100CF & DIMMER 2CF 380-3P 150A CB EMERGENCY

ANGIO / IR /HYBRID OR FLUORECENT 150CF & DIMMER TO 0.7 CF 380 - 3P 150A CB EMG +UPS

All the systems used to be designed to high efficiency and low cost of consumption

Supplies the distribution unit and loads of the device by vendor and contractor in coordination with the pre-installation sub contractor .

And fixing distribution points ... and all the electrical work, all within the pipes to prevent noise of radio waves .

Eng.Alam Dajani 00966592601260

Page 31: Introduction of imaging dept

MECHNICAL NEEDS 1

R/F ,IR & ANGIOGRAPHY

MODILITY HVAC

TEMP HUM AC AIR CHANGE/H PRESS. EXH.

RADIOGRAPHIC /Fluoroscopy(R/F) (standard) 21-24 30-60 CENTRAL 6 NEGATIVE YES

ANGIO / IR /HYBRID OR 21-24 40-60 INDPENENT 15 POSITIVE NO

Eng.Alam Dajani 00966592601260

Page 32: Introduction of imaging dept

MECHNICAL NEEDS 2

MEDICALGASESO2 VAC AIR√ √ √

PLUMBINGWATER DRAIN

√ √

R/F ,IR & ANGIOGRAPHY

Eng.Alam Dajani 00966592601260

Page 33: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)COMPUTERIZED AXIAL

TOMOGRAPHY(CAT)

Eng.Alam Dajani 00966592601260

Page 34: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

USING & SPECIFICATION 1

The use of X-ray technology ionized and then penetrate the patient's body and then detect X-piercing process by the detector in all directions.

It is taken anatomical cross-sections of the body (organ) to be filmed and then and through special programs (soft ware ) for these sections are assembled to form a three-dimensional image ( re constructions image )

CT is non invasive (means that any information is taken from the body from outside the body and not by placing a telescope or other).

The total duration of imaging almost short ( 5) minutes (duration of exposure to radiation of no more than 30 seconds.

Eng.Alam Dajani 00966592601260

Page 35: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

Eng.Alam Dajani

Page 36: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

USING & SPECIFICATION 2

The most widely used that accuracy and speed, especially in many cases :

1.HYDROSEPHALUS 2. INFLAMMATION & BLEEDING 3.CYSTS & TUMOERS 4.BONE INJURY 5.SKULL & SPINE 6.BRAIN PROBLEM

Eng.Alam Dajani 00966592601260

Page 37: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

USING & SPECIFICATION 3

Injection or inhalation of substance to show some blood clearly may be used, so you must make some tests for allergies or kidney function prior to the shooting.

Painless and non-security high amount of radiation that must advance warning "and not to use it for pregnant women.

Eng.Alam Dajani 00966592601260

Page 38: Introduction of imaging dept

ا�ل�م�ق�ط�ع�ي� ا�ل�ت�ص�و�ي�ر� ج�ه�ا�ز�COMPUTED TOMOGRAPHY (CT)

Eng.Alam Dajani

Page 39: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

USING & SPECIFICATION 4 The Modality consists of: 1. GANTRY: which is a model of circular open from the middle (donut shape). 2. Generator and detector and be within Gantry and rotate very quickly around the body. 3. The bed and who holds the patient where the subject is inserted into the Gantry . 4. Image display devices.    Image resolution depends on the amount of radiation and the

duration of the survey (test) and efficiency of the component parts of the device and the software responsible for collecting information and rebuilt.

  Where it is to differentiate between models : 1. The number of slides (32, 64 128 ....) 2. Detector type 3. Gantry speed ..Eng.Alam Dajani 00966592601260

Page 40: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

Eng.Alam Dajani

Page 41: Introduction of imaging dept

ا�ل�م�ق�ط�ع�ي� ا�ل�ت�ص�و�ي�ر� ج�ه�ا�ز�COMPUTED TOMOGRAPHY (CT)

Eng.Alam Dajani

Page 42: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

SPACES - METER SQR

STND PR.RM EQ.RM CN.RM

STANDARD 40  40 12 14

FACILITY SPACES

Eng.Alam Dajani 00966592601260

Page 43: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

Eng.Alam Dajani

PROTECTION

WARNING SIGNS

ON X RAY DOOR

YES FOR PERGNANT

LEADINGWALLS DOORS CONTROL

WINDOW

YES YES YES

Page 44: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

Eng.Alam Dajani

ELECTRICAL NEEDS

ELECTRICAL REQUIR.LIGHTING POWER TYPE

FLUORECENT 50CF & DIMMER TO 5 CF 380-3P 150A CB EMERGENCY

The supplying of distribution unit and loads by the equipment supplier in coordination with the pre-installation contractor , fix distribution points ... and all the electrical work, all within the pipes to prevent noise of radio waves (interference )

All the systems used to be designed to high efficiency and low cost of consumption

Page 45: Introduction of imaging dept

COMPUTERIZED TOMOGRAPHY (CT)

MECHNICAL NEEDS

HVAC TEMP HUM AC AIR

CHANGE/H PRESS. EXH.

21-24 30-60 CENTRAL 6 POSITIVE NO

Eng.Alam Dajani 00966592601260

Page 46: Introduction of imaging dept

ULTRASOUND (U/S)

Eng.Alam Dajani 00966592601260

Page 47: Introduction of imaging dept

ULTRASOUND (U/S)

USING & SPECIFICATIONS 1

Eng.Alam Dajani

Ultrasonography device (sonar) is a secur device of the radiation is used where the ultrasonic waves with non-ionized & reflected features.

Used in the study of the body's organs of anatomical terms (size, shape & movement).

Widely used: 1. Follow-up stages of embryo development - Obstetrics and Gynecology clinics (OBGYNE DEPT ) 2. The study of the heart and blood flow - Heart Clinics( CARDIOLOGY & VASCULAR) 3. General uses : study of the shape of kidney, liver, and in the non-surgical Lithotripsy - Imaging Division

Page 48: Introduction of imaging dept

ULTRASOUND (U/S)

Eng.Alam Dajani 00966592601260

Page 49: Introduction of imaging dept

ULTRASOUND (U/S)

Eng.Alam Dajani

Page 50: Introduction of imaging dept

ا�ل�ص�و�ت�ي�ة ف�و�ق� ل�أل�م�و�ا�ج� ا�ل�ت�ص�و�ي�ر� ج�ه�ا�ز�ULTRASOUND (U/S)

Eng.Alam Dajani

Page 51: Introduction of imaging dept

Eng.Alam Dajani

Page 52: Introduction of imaging dept

ULTRASOUND (U/S)

USING & SPECIFICATIONS 2

Eng.Alam Dajani

The wave Band device uses ultrasound frequencies ranging from (1 MHz – up to 12 MHz) ;Sound band width is (4-20 kHz) human can hear.

The sound Speed inside the body is 1540 m / secThese waves are issued through the quartz crystal and which Pizo electric

(pressure to generate an electric current produces waves and vice versa in the reception stage.

The probe is a transmitter and receiver, as the outgoing waves are reflected from both parts of the body, "according to the depth (distance) and intensity (power) signal.

Reflected waves are processed in the electronic circuits and display.It is to be the image on the three forms (two-dimensional, three-dimensional

and four dimensions).Device gives continuous waves feature (continuous survey).

Page 53: Introduction of imaging dept

ULTRASOUND (U/S)

USING & SPECIFICATIONS 3

A device (the main unit , probes , displays , recording devices and printers normal injection or laser)

Wave (Arrays) Types: - LINEAR ARRAY - CURVED ARRAY - PHASED ARRAY

Eng.Alam Dajani 00966592601260

Page 54: Introduction of imaging dept

ULTRASOUND (U/S)

Eng.Alam Dajani 00966592601260

Page 55: Introduction of imaging dept

ULTRASOUND (U/S)

FACILITY SPACES

SPACES - METER SQR

STND PR.RM EQ.RM CN.RM

STANDARD 16 PATIENT TOILET W/NEGATIVE PRESSURE

Eng.Alam Dajani 00966592601260

Page 56: Introduction of imaging dept

ULTRASOUND (U/S)

PROTECTION

LEADING

WALLS DOORS CONTROL WINDOW

NO NO X

Eng.Alam Dajani 00966592601260

Page 57: Introduction of imaging dept

ULTRASOUND (U/S)

ELECTRICAL NEEDS

ELECTRICAL REQUIR.

LIGHTING POWER TYPE

FLUORECENT 40CF 250-2P 30A CB EMERGENCY

Eng.Alam Dajani 00966592601260

Page 58: Introduction of imaging dept

ULTRASOUND (U/S)

MECHNICAL NEEDS 1

HVAC

TEMP HUM AC AIR CHANGE/H PRESS. EXH.

21-24 30-60 CENTRAL 8 POSITIVE  NO

Eng.Alam Dajani 00966592601260

Page 59: Introduction of imaging dept

ULTRASOUND (U/S)

MECHNICAL NEEDS 2

M.GASESO2 VAC AIR

√ √ √ PLUMBING

WATER(C/H) DRAIN

√ √

Eng.Alam Dajani 00966592601260

Page 60: Introduction of imaging dept

MRI MAGNETIC RESONANCE IMAGING

Eng.Alam Dajani 00966592601260

Page 61: Introduction of imaging dept

USING & SPECIFICATIONS 1

Eng.Alam Dajani

MRI MAGNETIC RESONANCE IMAGING

Depends on the presence (mount) of hydrogen ions are the most prevalent in the body.

The magnetic field (fixed and variable) is used in the imaging process. (0.7, 3 , 9 , ..Tesla) ... greater than the magnetic field of the earth (of 8000 - 30000) ; (1 Tesla = 10,000 Gauss) ; Its magnitude at the Earth's surface ranges from 25 to 65 microteslas (0.25 to 0.65 gauss)..

Doughnuts shape (round cake open from center) , weight and size of the senior.

Two types of form modality : open and close system. It requires constant cooling, and uses liquid helium in (temperature 269 C

below zero) to the presence of (SUPER CONDUCTOR)The design of the device in a manner so as to minimize the effect of the

magnetic field by using (ACTIVE SHIELD)

Page 62: Introduction of imaging dept

USING & SPECIFICATIONS 2

MRI MAGNETIC RESONANCE IMAGING

Safe for the patient and the user where there is no radiation, especially (for pregnant women and children).

Procedure Time : per case is (30-45 minutes).Sometimes the patient needs (anesthesia) and a contrast media

(intravenous).It used in imaging the brain and soft tissues and ligaments and spine.Painless and harmless.Three-dimensional (with three views) ; SAGITTAL (R / L), CORONAL (F / B),

TRANSVERSE (U / D)It needs some of the accessories (for patient) to care , such as pictures

(screen) and sounds, etc. The possible use of glasses, headsets because the device also issues noise.

Eng.Alam Dajani 00966592601260

Page 63: Introduction of imaging dept

MRI MAGNETIC RESONANCE IMAGING

Eng.Alam Dajani 00966592601260

Page 64: Introduction of imaging dept

USING & SPECIFICATIONS 3

MRI MAGNETIC RESONANCE IMAGING

MR suit is a four protection zones: 1.No restricted,2. restricted , 3. High restricted ; all equal to or outside the five Gauss line And 4. Ultra restricted ( 5 Gauss line ; 1 Gauss = 0.1 m Tesla ).

The using Radio frequencies .. so it must be protected with RF shielding and be a device far from the noise sources as lifts or broadcast and other communications devices.

The presence of liquid helium in the cooling process requires the presence of the exits of helium gas (ports outside the hospital building) in cases of liquid heating and that may lead to a severe explosion condition they do not exist (turning the liquid into a gas) ( 1 liter liquid turn into a 700-liter gas). Scientifically called QUENCH process

Eng.Alam Dajani 00966592601260

Page 65: Introduction of imaging dept

USING & SPECIFICATIONS 4

MRI MAGNETIC RESONANCE IMAGING

All medical and non-medical furniture of the MR suit (ultra restricted area ) , such as syringes , vital signs and a blood pressure monitors ,and tables must not be of ferrous metals, all of which must be comply w /resonance device call : MRI COMPATABLE furniture .

All pipes , accessories, building materials .. must be NON FERRO MAG.

All devices used must not be issued radio interference (RF INTERFERNCES).

Eng.Alam Dajani 00966592601260

Page 66: Introduction of imaging dept

MRI MAGNETIC RESONANCE IMAGING

Eng.Alam Dajani

Page 67: Introduction of imaging dept

FACILITY SPACES

SPACES - METER SQR

STND PR.RM EQ.RM CN.RM

STANDARD 30-58 30-58 14-35 9

MRI MAGNETIC RESONANCE IMAGING

Eng.Alam Dajani 00966592601260

Page 68: Introduction of imaging dept

PROTECTION

RF SHEILDING

WALLS DOORS

YES YES

LEADING

WALLS DOORS

NO NO

FERROMAGNETIC DETECTOR

ON MRI DOOR

YES RESTRICETED ACCESS

MRI MAGNETIC RESONANCE IMAGING

Eng.Alam Dajani 00966592601260

Page 69: Introduction of imaging dept

ELECTRICAL NEEDS

ELECTRICAL REQUIR.

LIGHTING POWER TYPE

INDIRECT 380 - 3P UPS

Eng.Alam Dajani

MRI MAGNETIC RESONANCE IMAGING

The supplying of distribution unit and loads by the equipment supplier in coordination with the pre-installation contractor , fix distribution points ... and all the electrical work, all within the pipes to prevent noise of radio waves (interference )

All the systems used to be designed to high efficiency and low cost of consumption

Page 70: Introduction of imaging dept

MECHNICAL NEEDS 1

MEDICALGASESO2 VAC AIR√ √ √

PLUMBINGWATER DRAIN

√ √

MRI MAGNETIC RESONANCE IMAGING

Eng.Alam Dajani 00966592601260

Page 71: Introduction of imaging dept

MECHNICAL NEEDS 2

HVAC TEMP HUM AC AIR

CHANGE/H PRESS. EXH.

21-24 C 40-60 INDPENENT 12 POSITIVE

NO EMER. CAS

MRI MAGNETIC RESONANCE IMAGING

Eng.Alam Dajani 00966592601260

Page 72: Introduction of imaging dept

Eng.Alam Dajani

MRI MAGNETIC RESONANCE IMAGING

Page 73: Introduction of imaging dept

NMNUCLEAR MEDICINE

The use of radioisotopes in the medical field in two forms: 1. Therapeutic: the treatment of tumors by extrusion of the tumor by radiation of the isotopes (high activity )2. Diagnostic: imaging member while performing his job by infusion or swallowed or inhaled radioactive isotopes after configured for human use and be (by medium and low activity )

INTRODUCTION

Eng.Alam Dajani 00966592601260

Page 74: Introduction of imaging dept

IMAGING BY USING ISOTOPES

Here and after injection or swallowing or inhalation of radioactive materials by patient, the body becomes the source of radiation; these rays resulting from the decay of radioactive isotope.The advantage of that so that the safe of this radioactive isotope out of the body with time .Radioactive isotope modality is divided into two parts, according to reactivity, namely: 1. PET : high activity isotopes ( Positron emission tomography) 2. GAMMA and SPECT : low activity isotopes .

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 75: Introduction of imaging dept

RADIOISOTOPES MODALITY APPLICATION

I125 & I131 THYROIDE UPTAKE SYSTEM

It calculates the amount of thyroid

hormone (Thyroxin) absorption

Xe133 GAMMA CAMERA /SPECT A radiation survey of the work of the lungs

Tc99 GAMMA CAMERA /SPECT Liver ,spleen and brain

FDG Flouro. Dioxide Glucose PET TUMERS

ISOTOPES & NUCLIDES

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 76: Introduction of imaging dept

NMNUCLEAR MEDICINE

1 .ISOTOPES PRODUCTION STAGE

2 .NUCLEAR IMAGING STAGE

3. WASTE & DISPOSABLES MANAGEMENT

USING ISOTOPES IN MEDICAL IMAGING

Eng.Alam Dajani 00966592601260

Page 77: Introduction of imaging dept

NMNUCLEAR MEDICINE

1 .ISOTOPES PRODUCTION STAGE

Cyclotron Hot LabChemical Lab Quality Control Lab

Eng.Alam Dajani 00966592601260

Page 78: Introduction of imaging dept

PET CENTER

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 79: Introduction of imaging dept

NMNUCLEAR MEDICINE

2 .NUCLEAR IMAGING STAGE

Hot Lab Modification Room Imaging modalities

Eng.Alam Dajani 00966592601260

Page 80: Introduction of imaging dept

HOT LAB DWG

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 81: Introduction of imaging dept

NMNUCLEAR MEDICINE

3. WASTE & DISPOSABLES MANAGEMENT Radioactive waste decontamination

Eng.Alam Dajani

Storage unit for radioactive decay

Biological sewage purification unit

Data command-control-acquisition-analysis-

archiving unit

Page 82: Introduction of imaging dept

MODILITY HEIGHT

min.

SPACES - METER SQR

STND

Injection Room 2.4 m 12

Hot lab 2.4 m 15

SPACES

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 83: Introduction of imaging dept

MODILITY TECHNIQUESLEADING

WALLS DOORS CONTROL WINDOW

Injection Room

NUCLEAR UTILITIES

0.18 YES 0.635

Hot lab 0.18 YES 0.635

PROTECTION

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 84: Introduction of imaging dept

ELECTRICAL NEEDS

MODILITY ELECTRICAL REQUIR.

LIGHTING PWR TYPE

Injection Room 220V,1P-60A X

Hot lab 220V,1P-60A X

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 85: Introduction of imaging dept

MECHNICAL NEEDS 1

NMNUCLEAR MEDICINE

MODILITY

HVAC

TEMP HUM % ACAIR

CHANGE/H

PRESS. EXH.

Injection Room 21-24 C 45-80 Y 6 NEG. NO

Hot lab 21-24 C 45-80 Y 6 NEG. YES

Eng.Alam Dajani 00966592601260

Page 86: Introduction of imaging dept

MECHNICAL NEEDS 2

NMNUCLEAR MEDICINE

MODILITY

PLUMBING M.GASES

WATER(C/H) DRAIN O2 VAC AIR

Injection Room √ √ X X X

Hot lab √ √ X X X

Eng.Alam Dajani 00966592601260

Page 87: Introduction of imaging dept

Thyroid Uptake System

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 88: Introduction of imaging dept

GAMMA CAMERA

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 89: Introduction of imaging dept

PET SCANNER

NMNUCLEAR MEDICINE

Eng.Alam Dajani 00966592601260

Page 90: Introduction of imaging dept

NMNUCLEAR MEDICINE

PET/CT Hybrid modality

Eng.Alam Dajani 00966592601260

Page 91: Introduction of imaging dept

NMNUCLEAR MEDICINE

PET /MRI = mMR Multi modalities Integrated

Eng.Alam Dajani 00966592601260

Page 92: Introduction of imaging dept

PET /MRI = mMR Multi modalitiesTandem

Eng.Alam Dajani

Page 93: Introduction of imaging dept

NMNUCLEAR MEDICINE

MODILITY HEIGHT

min.SPACES - METER SQR

STND PR.RM EQ.RM CN.RM

SPECT 3 m 37 37 x x

SPECT-CT 3 m 42 38 11.6 14

PET 3 m 37 37 x 14PET-MRI 3 m 50 44 11.6 14PET-CT 3 m 42 42 x x

PEM (Positron Emission Mammography) 3 m 14 38 11.6 14

FACILITY SPACES & HEIGHT

Eng.Alam Dajani 00966592601260

Page 94: Introduction of imaging dept

NMNUCLEAR MEDICINE

MODILITY ELECTRICAL REQUIR.

PWR TYPE

SPECT 480V,3P-100A UPSSPECT-CT 480V,3P-150A UPS

PET 480V,3P-100A UPS

PET-MRI 480V,3P-150A UPSPET-CT 480V,3P-150A UPS

PEM (Positron Emission Mammography) 480V,3P-100A X

ELECTRICAL NEEDS

Eng.Alam Dajani 00966592601260

Page 95: Introduction of imaging dept

NMNUCLEAR MEDICINE

MODILITY SHEILDING & LEADING

WALLS DOORS CONTROL WINDOW

SPECT 0.18 YES 0.635

SPECT-CT YES YES YES

PET YES YES YES

PET-MRI YES YES YES

PET-CT YES YES YES

PEM (Positron Emission Mammography) YES YES YES

PROTECTION

Eng.Alam Dajani 00966592601260

Page 96: Introduction of imaging dept

NMNUCLEAR MEDICINE

MODILITY

HVAC

TEMP HUM % AC AIR CHANGE/H PRESS. EXH.

SPECT 21-24 C 45-80 Y 6 NEG. YES

SPECT-CT 21-24 C 45-80 Y 6 NEG. YES

PET 21-24 C 45-80 Y 6 NEG. YES

PET-MRI 21-24 C 45-80 Y 6 NEG. YES

PET-CT 21-24 C 45-80 Y 6 NEG. YES

PEM(Positron Emission Mammography) 21-24 C 45-80 Y 6 NEG. YES

MECHNICAL NEEDS 1

Eng.Alam Dajani 00966592601260

Page 97: Introduction of imaging dept

NMNUCLEAR MEDICINE

MODILITY PLUMBING M.GASES

WATER(C/H) DRAIN O2 VAC AIR

SPECT √ √ √ √ √SPECT-CT √ √ √ √ √

PET √ √ √ √ √PET-MRI √ √ √ √ √PET-CT √ √ √ √ √

PEM (Positron Emission Mammography) X X X X X

MECHNICAL NEEDS 2

Eng.Alam Dajani 00966592601260

Page 98: Introduction of imaging dept

SHEILDING & LEADING EQUATION

Eng.Alam Dajani 00966592601260

Page 99: Introduction of imaging dept

GMP (GOOD MANFAC. PRACT.

  MANUFACTURERES PRODUCTS 1 SIEMENS MRI , CT ,US, G-X RAY2 GE MRI , CT ,US, G-X RAY3 PHILIPS MRI , CT ,US, G-X RAY4 TOSHIBA MRI , CT ,US, G-X RAY5 HITACHI CT ,US, G-X RAY6 AGFA CT ,US, G-X RAY7 HOLOGIC BMD,MAMMAOGRAPH,X RAY8 VARIAN MRI

Eng.Alam Dajani 00966592601260

Page 100: Introduction of imaging dept