medcard technical features medcard · the new global 12 leads with average and st measurements...

2
The new global 12 leads with average and st measurements PC/ECG SOLUTIONS Rest & exercice testing Full real time monitoring Arrhytmia detection & screening Networking & web server Technical Features Your official dealer : MediSoft s.a . se réserve le droit de changer les spécifications reprises sur cette notice sans avertissement préalable. MEDCARD ® GENERAL SPECIFICATIONS Dimensions : (L x W x H) : 110 x 85 x 17 mm Weight : 114 gr Powered by USB port or Battery Bluetooth version Leakage currents max. : < 5µA Protection against defibrillation shocks (built-in system) Electrical safety : opto-coupling isolation Bandwidth : 0.05 Hz to 150 Hz Time constant > to 3.2 sec CMRR : 130 dB Resolution : 2,54 µ Volt Detection : Faulty electrode contact : automatic detection with visualisation Gain : 10 to 50 mm/mV Time base : 10, 20, 50 mm/sec Sampling frequency : 2000 Hz/channel Hardware : analog QRS trigger signal output (option) and Pacemaker Wave detection (option) Multi Software languages FILTERING Muscle Tremor Filter : 20, 40, 100 Hz Filter rejection 50 or 60 Hz Baseline stabilisation : - B-spline - Complex average mode - High pass filter AMBIENT CONDITIONS Temperature 10 to 40°C Humidity 30 to 85% uncondensed ELECTRICAL SAFETY Conform to the standard electrical safety 60601-2-25 REQUIRED PC CONFIGURATION Microsoft Windows Latest Processor Spec for Peak Performance Minimum 1024 MB RAM SVGA monitor (1280/1024) Printer : laser or inkjet color or b/w VERSIONS & OPTIONS USB or Bluetooth Version for Rest only Rest & exercice testing Computer and periphericals Trolley Suction electrodes Cycle ergometer Treadmill Spirometry Ergospirometry (Ergocard) NIBP NIBP with QRS synchonisation SpO 2 Networking Pulmonary Function Integration Integrated into one single module : ECG amplificator with electrical protections, digital conversion and PC communication. The MediSoft facilities is a state of the art modern facility with clinical research, precision engineering and computer design departments. MEDCARD ® Medisoft s.a. reserves the right to change and improve the above specifications without prior notice 0029 Cert. N°: 09/BE/1814 ISO 13485 (2003) BELGIUM HEADQUARTER P.A.E. de Sorinnes Route de la Voie Cuivrée 1 B-5503 Sorinnes – Belgium Tel: +32 (0)82 22 30 20 Fax: +32 (0)82 22 33 34 [email protected] www.medisoft.be MEDISOFT FRANCE Rue du Maréchal de Lattre de Tassigny, 8 59000 Lille Tel: + 33 (0)825 280 030 Fax: +33 (0)320 787 609 [email protected] www.medisoft.fr MEDISOFT CHINA West Tower, Shun Tak Centre 168 Connaught Road Central Hong Kong Tel: +852 2267 8933 Fax: +852 2267 8983 [email protected] MEDISOFT ITALIA Via Germania, 14 35127 CAMIN (PD) ITALY Tel: +39 049 87 03 355 Fax: +39 049 87 03 388 [email protected] www.medisoft-ram.it MEDISOFT GERMANY GmbH Universitätsstraße, 136 D-44799 Bochum GERMANY Tel.: +49 (0) 234 976-58410 Fax: +49 (0) 234 976-69507 [email protected] www.medisoft-pft.de 120515_Medcard_EN

Upload: others

Post on 27-Sep-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MEDCARD Technical Features MEDCARD · The new global 12 leads with average and st measurements PC/ECG SOLUTIONS • Rest & exercice testing • Full real time monitoring • Arrhytmia

The new global 12 leads with average and st measurements PC/ECG SOLUTIONS• Rest & exercice testing• Full real time monitoring• Arrhytmia detection & screening• Networking & web server

Technical Features

Your official dealer :

MediSoft s.a . se réserve le droit de changer les spécifications reprises sur cette notice sans avertissement préalable.

MEDCARD®

General specifications

Dimensions : (L x W x H) : 110 x 85 x 17 mmWeight : 114 grPowered by USB port or Battery Bluetooth versionLeakage currents max. : < 5µAProtection against defibrillation shocks (built-in system)Electrical safety : opto-coupling isolationBandwidth : 0.05 Hz to 150 HzTime constant > to 3.2 secCMRR : 130 dBResolution : 2,54 µ VoltDetection : Faulty electrode contact : automatic detection with visualisationGain : 10 to 50 mm/mVTime base : 10, 20, 50 mm/secSampling frequency : 2000 Hz/channelHardware : analog QRS trigger signal output (option) and Pacemaker Wave detection (option)Multi Software languages

filterinG

Muscle Tremor Filter : 20, 40, 100 HzFilter rejection 50 or 60 Hz Baseline stabilisation : - B-spline - complex average mode - High pass filter

amBient conditions

Temperature 10 to 40°cHumidity 30 to 85% uncondensed

electrical safety

Conform to the standard electrical safety 60601-2-25

required pc confiGuration

Microsoft WindowsLatest Processor Spec for Peak PerformanceMinimum 1024 MB RAM SVGA monitor (1280/1024)Printer : laser or inkjet color or b/w

Versions & options

USB or Bluetooth Version for Rest onlyRest & exercice testingComputer and periphericalsTrolleySuction electrodesCycle ergometerTreadmillSpirometryErgospirometry (Ergocard)NIBPNIBP with QRS synchonisationSpO2NetworkingPulmonary Function Integration

Integrated into one single module : ECG amplificator with electrical protections, digital conversion and PC communication.

The MediSoft facilities is a state of the art modern facility with clinical research, precision engineering and computer design departments.

MEDCARD®

Medisoft s.a. reserves the right to change and improve the above specifications without prior notice

0029

Cert. N°: 09/BE/1814ISO 13485 (2003)

BELGIUM HEADQUARTERp.a.e. de sorinnesroute de la Voie cuivrée 1B-5503 sorinnes – Belgiumtel: +32 (0)82 22 30 20fax: +32 (0)82 22 33 [email protected]

MEDISOFT FRANCErue du maréchal delattre de tassigny, 859000 lilletel: + 33 (0)825 280 030fax: +33 (0)320 787 [email protected]

MEDISOFT CHINAWest tower, shun tak centre168 connaught road centralHong Kongtel: +852 2267 8933fax: +852 2267 [email protected]

MEDISOFT ITALIAVia Germania, 1435127 camin (pd)italytel: +39 049 87 03 355fax: +39 049 87 03 [email protected]

MEDISOFT GERMANY GmbHuniversitätsstraße, 136d-44799 BochumGermany tel.: +49 (0) 234 976-58410 fax: +49 (0) 234 [email protected] 12

0515

_Med

card

_EN

Page 2: MEDCARD Technical Features MEDCARD · The new global 12 leads with average and st measurements PC/ECG SOLUTIONS • Rest & exercice testing • Full real time monitoring • Arrhytmia

The most compact light-weight PC electrocardiograph !Close coupled digital interface reduces lead artefact

MEDCARD

p At private office to full hospital clinical use

p From basic resting ECG to the most complete stress ECG capability

p Real-time printing & user friendly setting up

p High definition optimized graphics screens

p Full integration compatibility for spirometry & ergospirometry combinations

Advanced & Powerfull

1- rest version2- stress test with treadmill3- cardio-pulmonary exercice testing with cyclo-ergometer4- trolley with suction electrodes

1

2

3

4

global functions

Selection of :

• Trace speed • Screen freeze of last 60 sec. recorded • Sensitivity (Gain) • Event marker • Filtering mode • Reference complex • Baseline stabilisation

Multiple ECG displays with 3 stored formats :

• 12 leads • Multiple user configurations • 2 columns 6 lead + rhythm

Real time immediate printing (15 seconds for a full page)

Automatic faulty electrode contact detection

Configuration :

• Lead colours (real-time & average complex)• Background colour • Trend screens : ST Slope and Amplitude, HR, NIBP, Load ( speed & slope)

functions at rest

> HR, INTERVALS, AXIS> Editing of QRS measurement points> Pacemaker wave detection> Rest interpretation (option)

functions for stress test in real time

> Full design and storage of exercise protocols for ergocycle and tread-mill

• Automatic protocol creation software based on max load setting desired • Step/ ramp graph of the exercise protocol (load/time & speed, slope/time) • Automatic calculation of predicted max load • Standard protocols (Bruce, Naughton,...) • Introduce Comment

> 7 frames can be configured to display :

• 3, 6 or 12 leads in 1 or 2 columns + rhythm • ST segment amplitude histogram on all 12 leads, ST slope and amplitude

trends on selected lead • HR & BP trends against load or speed and slope • Graph of the reference average complex superimposed on the average com-

plexes every 10 sec for 1, 2, 3 or all leads • Arrhythmia history • Graph of the detected arrhythmias displayed by type

> Recalculate Complexes References on real time> On line introduction of the BP or automatic readings (NIBP option)> On line introduction of the SpO2 or integrated oximeter (option)> User configurable settings of the measuring points for the QRS complex

and the J point.

> Control of the stress protocol • Pause, lock • Next step • Modification of the load or speed & slope • Manual or automatic recovery step • Monitoring & printing available after recovery phase • Post recovery monitor 12 leads and printing

arrhythmia detection & classification (optional)

> Arrhythmia management allows arrhythmia :

• Recognition • Classification • Restoration within the time axis with post review capabilities • History

post recording functions

> Post testing displays show all the key information

> The average complex can be reviewed at the various steps of the exam:

• Reference • Each step • HR Max • Recovery • Stress max • At any time

> Graph of the changes in ST segment amplitude andslope throughout the full test on all leads

> Display mode of ST changes, selected lead and 3D image

• 3D view of the average complex of the selected lead to show changes in the depression of the ST slope

• Display of the reference average complex and the lead average complex cho-sen by the user from the 3D graph

• Adjustment and measurement of the J point • HR trend graph and ST amplitude and slope graph

> Full disclosure of the exercise test and event reviewing

powerful database

> With Expair Software, all tests are stored on the computer with the possibility to backup or export accross a network

printing reports configuration

> Choice of standard preformatted reports or customisablereports