Optimization of Microfluidic “Lab-on-a-Chip” Devices for Capillary Electrophoresis Separations A. Brown* and F.A. Gomez Department of Chemistry and Biochemistry California State University, Los Angeles, 5151 State University Dr., Los Angeles, CA 90032 Abstract Introduction Discussion Conclusion Acknowledgements Microfluidic devices (MD’s) are powerful tools for performing an array of applications. The nanoliter volumes and parallel sample processing are advantages of MD’s that make them ideal for chemical analysis, high-throughput screening, and other limited reagent scenarios. A major challenge associated with the desired scales in MD’s is to simultaneously reduce the number of pipetting steps needed to load the devices while amortizing the sample volume over several reactions or separation steps. Capillary electrophoresis (CE) is a technique that has shown great promise when coupled to microfluidic devices. Although CE has gained widespread use because of its versatility there is still the need to prepare samples at variable concentrations which inherently slows down the analysis rate. Using multilayer soft lithography (MSL), fluid and control channels are fabricated to allow for manipulation of material on the device without the need of time-consuming pipetting steps. Subsequent electrophoresis using CE affords separation of materials. Herein, we describe our work on coupling affinity CE (ACE) to “lab-on-a-chip” devices using as a model system the binding of arylsulfonamides to carbonic anhydrase B (CAB, EC 4.2.1.1). The development of new molecular biological techniques has provided for a myriad of biological interactions. One technique that has shown great promise in quantifying receptor–ligand interactions in the recent past is capillary electrophoresis (CE). In a typical form of CE a sample of receptor and a non-interacting standard(s) are injected into the capillary column and are electrophoresed in an increasing concentration of ligand in a running buffer generating an electropherogram readout. A change in migration time of the receptor relative to the standard(s) is induced upon the formation of the receptor-ligand complex. This change in the migration time is then used for Scatchard analysis. Based on the Scatchard analysis, an equation is utilized to calculate a binding constant value (K b ) on a relative scale using the non-interacting standard(s). In the past, instances when quantities of materials were limited, CE was the ideal technique to be utilized to measure receptor-ligand interactions. More recently, MD’s have been fabricated and shown to be capable of running comparable CE experiments to those previously done on large electrophoresis machines, on a smaller scale and at a fraction of the cost. Micro- scale quantities and cost efficiency as well as reproducibility of both Microfluidic devices and the analytical setup that have made the fabrication and experimentation of MD’s desirable. We have demonstrated that affinity capillary electrophoresis (ACE) coupled to a microfluidic format can be utilized in the estimation of binding constants between a receptor and a ligand. Using the model system carbonic anhydrase B (CAB) and its binding to an arylsulfonamide we have shown proof-of-concept that a microfluidic format can readily be used to examine biomolecular non-covalent interactions. Further work will focus on optimizing the conditions for assay, developing a high-throughput system to assay, many potential drug targets, and to develop a computer interface that will allow for a completely automated drug assay system. The authors gratefully acknowledge financial support for this research by grants from the National Science Foundation Research in Undergraduate Institutions (DMR-0080065, CHE-0136724, CHE- 0515363), Partnership for Research and Education in Materials (PREM) (DMR-0351848) Programs,and the National Institutes of Health (R15 AI055515-01 and 1 R15 AI065468-01). MD’s created during the course of this research were made using Multilayer Soft Lithography (MSL) utilizing poly(dimethylsiloxane) (PDMS), a silicone rubber using 2-component elastomer materials (GE RTV or Sylgard). An AutoCAD design was developed onto a mold by incorporating photolithography SU-8 (neg) photoresist. There are two major components necessary for Microfluidic devices: two separate layers created in a 4:1 (thick layer: thin layer) ratio, which are representative of the “flow layer” where reagents are deposited and manipulated and the “control layer” where air is utilized to control the flow of reagents respectively. Design features that make this device unique include a mixing circle, off and on chip CE regions, and a bypass region to allow for the off chip CE region to be bypassed to use the on chip CE region making this a multifunctional device. The off chip CE region of this device includes two separate hole punches, the first for the insertion of an electrode which is connected to a voltage meter controlled by a LabView 7.0 program and the second for the insertion of a capillary which is connected to a UV-vis detector that allows for the detection of reagent flow to generate electropherogram's for analysis. The on chip CE region of the device allows for florescent CE. It consists of a long channel which acts as the capillary and a hole punch for the insertion of an electrode. Using the ACE technique coupled to MD’s, we examined the interaction between Cab and 4-carboxybenzenesulfonamide (Ligand 1). In theses studies, a plug of increasing concentrations of 1 was pressure injected into the microfluidic device for 10 sec, followed by buffer for 10 sec, then by a sample containing DMF and CAB. DMF is a non- interacting standard which does not interact with 1 or CAB. The samples are then pushed past the anode and into the capillary. Upon electrophoresis, the sample flows into the zone of 1 where a dynamic equilibrium between 1 and CAB is established. The complex that begins to form between CAB and 1 is more negatively charged then CAB uncomplexed and, therefore, the peak for the complex (CAB-1) shifts to a greater migration time on increasing the concentration of 1 in buffer. A Scatchard plot of the data for CAB and 1 is shown. In this form of analysis K b is estimated using a single-marker form of analysis based on the receptor’s electrophoretic mobility (μ). Analysis of the magnitude of the change in μ (μ) as a function of the concentration of ligand ([L]) yields a value for the binding constant. A K b of 0.679 x 106 L mol 1 was obtained for the interaction between CAB and 1 which is comparable to our previous results using ACE (0.69 x 10 6 L mol 1 ) and to values obtained using other assay techniques. A correlation coefficient (R 2 ) of 0.996 was obtained for the fitness of the Scatchard plot. Off chip CE channel: Florescence Mixing circle Bypass: For on chip CE Hole punched: Insert electrode Hole punched: Insert Capillary for off chip CE Picture of mixing circle. Features range from 2-200 µm. Push Up Architecture: Deep Channel Option Flow Layer: For applications that require suspensions of large particles. Rounded dimensions 20µm wide x 45 (+/- 3) µm high AZ100 Control layer: 100 µm wide x 11 µm high (un-rounded) Schematic of a Flow-Through Partial-Filling Capillary Electrophoresis experiment. Valve Structure: Control and Flow layer Microvalve with control layer on top. S O O NH2 O HO Ligand 1:Structure of 4-carboxybenzenesulfonamide (CBSA) A representative set of electropherograms of Carbonic anhydrase B (CAB) in 0.192 M glycine- 0.025M Tris Buffer (pH 8.3). Scatchard plot of the data for carbonic anhydrase B in conjunction with 1. Simplified AutoCAD chip design. AutoCAD chip design.

Upload: others

Post on 25-Feb-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Optimization of Microfluidic “Lab-on-a-Chip” Devices for ...Optimization of Microfluidic “Lab-on-a-Chip” Devices for Capillary Electrophoresis Separations A. Brown* and F.A

Optimization of Microfluidic “Lab-on-a-Chip” Devices for Capillary Electrophoresis SeparationsA. Brown* and F.A. Gomez

Department of Chemistry and Biochemistry California State University, Los Angeles, 5151 State University Dr., Los Angeles, CA 90032

Abstract

Introduction

Discussion

Conclusion

Acknowledgements

Microfluidic devices (MD’s) are powerful tools for performingan array of applications. The nanoliter volumes and parallel sampleprocessing are advantages of MD’s that make them ideal forchemical analysis, high-throughput screening, and other limitedreagent scenarios. A major challenge associated with the desiredscales in MD’s is to simultaneously reduce the number of pipettingsteps needed to load the devices while amortizing the samplevolume over several reactions or separation steps. Capillaryelectrophoresis (CE) is a technique that has shown great promisewhen coupled to microfluidic devices. Although CE has gainedwidespread use because of its versatility there is still the need toprepare samples at variable concentrations which inherently slowsdown the analysis rate. Using multilayer soft lithography (MSL), fluidand control channels are fabricated to allow for manipulation ofmaterial on the device without the need of time-consuming pipettingsteps. Subsequent electrophoresis using CE affords separation ofmaterials. Herein, we describe our work on coupling affinity CE(ACE) to “lab-on-a-chip” devices using as a model system thebinding of arylsulfonamides to carbonic anhydrase B (CAB, EC4.2.1.1).

The development of new molecular biological techniques hasprovided for a myriad of biological interactions. One technique thathas shown great promise in quantifying receptor–ligand interactionsin the recent past is capillary electrophoresis (CE). In a typical form ofCE a sample of receptor and a non-interacting standard(s) areinjected into the capillary column and are electrophoresed in anincreasing concentration of ligand in a running buffer generating anelectropherogram readout. A change in migration time of the receptorrelative to the standard(s) is induced upon the formation of thereceptor-ligand complex. This change in the migration time is thenused for Scatchard analysis. Based on the Scatchard analysis, anequation is utilized to calculate a binding constant value (Kb) on arelative scale using the non-interacting standard(s).

In the past, instances when quantities of materials werelimited, CE was the ideal technique to be utilized to measurereceptor-ligand interactions. More recently, MD’s have beenfabricated and shown to be capable of running comparable CEexperiments to those previously done on large electrophoresismachines, on a smaller scale and at a fraction of the cost. Micro-scale quantities and cost efficiency as well as reproducibility of bothMicrofluidic devices and the analytical setup that have made thefabrication and experimentation of MD’s desirable.

We have demonstrated that affinity capillary electrophoresis(ACE) coupled to a microfluidic format can be utilized in the estimationof binding constants between a receptor and a ligand. Using the modelsystem carbonic anhydrase B (CAB) and its binding to anarylsulfonamide we have shown proof-of-concept that a microfluidicformat can readily be used to examine biomolecular non-covalentinteractions. Further work will focus on optimizing the conditions forassay, developing a high-throughput system to assay, many potentialdrug targets, and to develop a computer interface that will allow for acompletely automated drug assay system.

The authors gratefully acknowledge financial support for thisresearch by grants from the National Science Foundation Research inUndergraduate Institutions (DMR-0080065, CHE-0136724, CHE-0515363), Partnership for Research and Education in Materials (PREM)(DMR-0351848) Programs,and the National Institutes of Health (R15AI055515-01 and 1 R15 AI065468-01).

MD’s created during the course of this research were made usingMultilayer Soft Lithography (MSL) utilizing poly(dimethylsiloxane)(PDMS), a silicone rubber using 2-component elastomer materials (GERTV or Sylgard). An AutoCAD design was developed onto a mold byincorporating photolithography SU-8 (neg) photoresist. There are twomajor components necessary for Microfluidic devices: two separatelayers created in a 4:1 (thick layer: thin layer) ratio, which arerepresentative of the “flow layer” where reagents are deposited andmanipulated and the “control layer” where air is utilized to control theflow of reagents respectively.

Design features that make this device unique include a mixingcircle, off and on chip CE regions, and a bypass region to allow for theoff chip CE region to be bypassed to use the on chip CE region makingthis a multifunctional device. The off chip CE region of this deviceincludes two separate hole punches, the first for the insertion of anelectrode which is connected to a voltage meter controlled by a LabView7.0 program and the second for the insertion of a capillary which isconnected to a UV-vis detector that allows for the detection of reagentflow to generate electropherogram's for analysis. The on chip CE regionof the device allows for florescent CE. It consists of a long channelwhich acts as the capillary and a hole punch for the insertion of anelectrode.

Using the ACE technique coupled to MD’s, we examined theinteraction between Cab and 4-carboxybenzenesulfonamide (Ligand 1).In theses studies, a plug of increasing concentrations of 1 was pressureinjected into the microfluidic device for 10 sec, followed by buffer for 10sec, then by a sample containing DMF and CAB. DMF is a non-interacting standard which does not interact with 1 or CAB. The samplesare then pushed past the anode and into the capillary. Uponelectrophoresis, the sample flows into the zone of 1 where a dynamicequilibrium between 1 and CAB is established.

The complex that begins to form between CAB and 1 is morenegatively charged then CAB uncomplexed and, therefore, the peak forthe complex (CAB-1) shifts to a greater migration time on increasing theconcentration of 1 in buffer. A Scatchard plot of the data for CAB and 1is shown. In this form of analysis Kb is estimated using a single-markerform of analysis based on the receptor’s electrophoretic mobility (μ).Analysis of the magnitude of the change in μ (∆μ) as a function of theconcentration of ligand ([L]) yields a value for the binding constant. A Kbof 0.679 x 106 L mol−1 was obtained for the interaction between CABand 1 which is comparable to our previous results using ACE (0.69 x 106

L mol−1) and to values obtained using other assay techniques. Acorrelation coefficient (R2) of 0.996 was obtained for the fitness of theScatchard plot.

Off chip CE channel:Florescence

Mixing circle

Bypass:For on chip CE

Hole punched: Insert electrode

Hole punched:Insert Capillary for off chip CE

Picture of mixing circle. Features range from 2-200 µm.

Push Up Architecture: Deep Channel OptionFlow Layer: For applications that require suspensions oflarge particles. Rounded dimensions 20µm wide x 45 (+/- 3)µm high AZ100Control layer: 100 µm wide x 11 µm high (un-rounded)

Schematic of a Flow-Through Partial-FillingCapillary Electrophoresis experiment.

Valve Structure: Control and Flow layer

Microvalve with control layer on top.

S

O

O

NH2

O

HO

Ligand 1:Structure of 4-carboxybenzenesulfonamide (CBSA)

A representative set of electropherograms of Carbonic anhydrase B (CAB) in 0.192 M glycine-0.025M Tris Buffer (pH 8.3).

Scatchard plot of the data for carbonic anhydrase B in conjunction with 1.

Simplified AutoCAD chip design.AutoCAD chip design.