research article performance investigation of cmm...

12
Research Article Performance Investigation of CMM Measurement Quality Using Flick Standard Salah H. R. Ali Engineering and Surface Metrology Department, National Institute for Standards (NIS), Giza 12211-136, Egypt Correspondence should be addressed to Salah H. R. Ali; [email protected] Received 6 May 2014; Accepted 17 June 2014; Published 17 July 2014 Academic Editor: Shey-Huei Sheu Copyright © 2014 Salah H. R. Ali. is is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. Quality of coordinate measuring machine (CMM) in dimension and form metrology is designed and performed at the NIS. e experimental investigation of CMM performance is developed by using reference Flick standard. e measurement errors of corresponding geometric evaluation algorithm (LSQ, ME, MC, and MI) and probe scanning speed (1, 2, 3, 4, and 5mm/s) are obtained through repeated arrangement, comparison, and judgment. e experimental results show that the roundness error deviation can be evaluated effectively and exactly for CMM performance by using Flick standard. Some of influencing quantities for diameter and roundness form errors may dominate the results at all fitting algorithms under certain circumstances. It can be shown that the 2 mm/s probe speed gives smaller roundness error than 1, 3, 4, and 5 mm/s within 0.2 : 0.3 m. It ensures that measurement at 2 mm/s is the best case to satisfy the high level of accuracy in the certain condition. Using Flick standard as a quality evaluation tool noted a high precision incremental in diameter and roundness form indication. is means a better transfer stability of CMM quality could be significantly improved. Moreover, some error formulae of data sets have been postulated to correlate the diameter and roundness measurements within the application range. Uncertainty resulting from CMM and environmental temperature has been evaluated and confirmed the quality degree of confidence in the proposed performance investigation. 1. Introduction e advanced electromechanical systems like coordinate measuring machine (CMM) is one of the most diversified fields of mechanical engineering due to the promising tech- nology cooperation with modern industrial strategies. CMM is generally better optimized in microcoordinate metrology. Microcoordinate metrology is an important branch of quality assurance. CMM is typically numerical control system that can be used for dimensional inspection of complex 3D geometry product surfaces. Surface quality is an important parameter in the modern manufacturing, aerospace, and automotive. e influence of dimensional and geometrical form errors is one of the most issues of interest regarding the manufacturing accuracy and surfaces quality. Out of tolerance can produces number of problems like vibration, wear, and noise [1, 2]. Studying the influence of the geo- metric fitting algorithms and probe scanning speeds on the measurement error can be an interesting topic for a future research, especially with the growing use of CMM machines [25]. In addition, the development of a new CMM machine with accuracy less than a nanometer is necessary, because it is close to the industrial actual needs. For these reasons, most available verification and calibration of CMM requires reference standards artifacts. Flick standard is embodiment of outer surface. New published work in parallel with my research confirmed that the Flick standard is a suitable tool used to calibrate the spindle motion error and the probe which equips cylindrical measuring machines to ensure such a level of uncertainty, both stability and performance [5]. e resultant measurement quality of CMM is limited by deviation and some uncertainties due to different factors. e measurement deviations in CMM metrology can be related to the operator performance quality, environmental interaction, and CMM performance accuracy. It can be assumed that some important factors such as operator behavior and CMM soſtware accuracy have effective reaction on the measurement quality. Uncertainty in measurement is the most important Hindawi Publishing Corporation Journal of Quality and Reliability Engineering Volume 2014, Article ID 960649, 11 pages http://dx.doi.org/10.1155/2014/960649

Upload: others

Post on 15-Aug-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

Research ArticlePerformance Investigation of CMM Measurement Quality UsingFlick Standard

Salah H. R. Ali

Engineering and Surface Metrology Department, National Institute for Standards (NIS), Giza 12211-136, Egypt

Correspondence should be addressed to Salah H. R. Ali; [email protected]

Received 6 May 2014; Accepted 17 June 2014; Published 17 July 2014

Academic Editor: Shey-Huei Sheu

Copyright © 2014 Salah H. R. Ali. This is an open access article distributed under the Creative Commons Attribution License,which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Quality of coordinate measuring machine (CMM) in dimension and form metrology is designed and performed at the NIS.The experimental investigation of CMM performance is developed by using reference Flick standard. The measurement errorsof corresponding geometric evaluation algorithm (LSQ, ME, MC, and MI) and probe scanning speed (1, 2, 3, 4, and 5mm/s)are obtained through repeated arrangement, comparison, and judgment. The experimental results show that the roundness errordeviation can be evaluated effectively and exactly for CMMperformance by using Flick standard. Some of influencing quantities fordiameter and roundness form errors may dominate the results at all fitting algorithms under certain circumstances. It can be shownthat the 2mm/s probe speed gives smaller roundness error than 1, 3, 4, and 5mm/s within 0.2 : 0.3𝜇m. It ensures that measurementat 2mm/s is the best case to satisfy the high level of accuracy in the certain condition. Using Flick standard as a quality evaluationtool noted a high precision incremental in diameter and roundness form indication. This means a better transfer stability of CMMquality could be significantly improved. Moreover, some error formulae of data sets have been postulated to correlate the diameterand roundness measurements within the application range. Uncertainty resulting from CMM and environmental temperature hasbeen evaluated and confirmed the quality degree of confidence in the proposed performance investigation.

1. Introduction

The advanced electromechanical systems like coordinatemeasuring machine (CMM) is one of the most diversifiedfields of mechanical engineering due to the promising tech-nology cooperation with modern industrial strategies. CMMis generally better optimized in microcoordinate metrology.Microcoordinatemetrology is an important branch of qualityassurance. CMM is typically numerical control system thatcan be used for dimensional inspection of complex 3Dgeometry product surfaces. Surface quality is an importantparameter in the modern manufacturing, aerospace, andautomotive. The influence of dimensional and geometricalform errors is one of the most issues of interest regardingthe manufacturing accuracy and surfaces quality. Out oftolerance can produces number of problems like vibration,wear, and noise [1, 2]. Studying the influence of the geo-metric fitting algorithms and probe scanning speeds on themeasurement error can be an interesting topic for a future

research, especially with the growing use of CMM machines[2–5]. In addition, the development of a new CMMmachinewith accuracy less than a nanometer is necessary, becauseit is close to the industrial actual needs. For these reasons,most available verification and calibration of CMM requiresreference standards artifacts. Flick standard is embodimentof outer surface. New published work in parallel with myresearch confirmed that the Flick standard is a suitable toolused to calibrate the spindle motion error and the probewhich equips cylindrical measuring machines to ensure sucha level of uncertainty, both stability and performance [5].The resultant measurement quality of CMM is limited bydeviation and some uncertainties due to different factors.Themeasurement deviations inCMMmetrology can be related tothe operator performance quality, environmental interaction,and CMM performance accuracy. It can be assumed thatsome important factors such as operator behavior and CMMsoftware accuracy have effective reaction on themeasurementquality. Uncertainty in measurement is the most important

Hindawi Publishing CorporationJournal of Quality and Reliability EngineeringVolume 2014, Article ID 960649, 11 pageshttp://dx.doi.org/10.1155/2014/960649

Page 2: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

2 Journal of Quality and Reliability Engineering

concept in roundness metrology according to the GUM andattracts wide attention in the world to confirm the confidencein measurement result [6].

In this paper, measurements for transverse circle loca-tion of certified Flick standard have been carried out. Themeasurement results of dimension and geometrical form atdifferent CMM strategies have been studied.The influence ofgeometric fitting algorithms equipped with probe scanningspeeds as metrological parameters on both diameter (D) androundness form errors (RON) is investigated and discussed.The modifications and analysis of software are employedexperimentally. The statistical study based on design ofexperiment was accomplished in order to evaluate the mea-surement result and estimate the expanded uncertainties.Therefore, the main objective of this research is projectedto study the influence of four fitting algorithms through fivedifferent probe scanning speeds for transverse circle locationof carrying out signals to:

(a) develop the CMM software using closed loop con-trol for increase the accuracy and precision of newmachines,

(b) increase CMM operator skills and reduce the opera-tion lost time and cost, to avoid processing mistakesof software strategic applications,

(c) determine and analyze the relative deviation errors inmeasurement,

(d) minimize the size of uncertainty in measurement.

2. Experimental Work

The experiment in this paper is designed and performedwithin the research plan of the Engineering and SurfaceMetrology Department at NIS. The experimental investi-gations including measurement procedures are designed tostudy the effect of geometric fitting algorithm and probescanning speed on CMM measurement. This plan has beendone through two main stages. The first stage is verificationof CMM probing system. The main stage in the experimentsis made to verify the CMMmachine in measurement quality.This second stage based on the measurement of the diameterand roundness of the Flick standard at different strategies inmeasurement. The dynamic influences of geometric fittingalgorithms and probe canning speeds strategies on diameterand roundness deviations have been studied using CMM.

2.1. Verification of CMM Probing System. Dynamic verifica-tion of probing system is the very important recommendationtask for an acceptance before study CMM performanceaccuracy [7, 8]. The standard verification method of bothprobing error and scanning probing error using referencesphere is used. To determine the probing error, it must beprobed twenty five recommended points on the reference testsphere surface. To determine the scanning probing error, itmust scan four recommended scanning lines on the surfaceof test sphere and compute. The Gaussian center point of thesphere using all measured points of all four scan lines. Beforemaking measurements with the CMM in the cylindrical

Table 1: Output verification data of CMM probes and sphere.

CMM element Measured value, mm SD, mmMaster probe 𝑅 = 3.9999 0.0002Reference sphere 𝑅 = 14.9942 0.0002Used stylus probe 𝑅 = 1.4990, 𝐿 = 33.5 0.0002With Al. extension 𝐿 = 100

Step width 1.3861Number of probe points 104 points

feature of Flick standard, the CMM was verified usingmaster probe to evaluate standard sphere and using standardsphere to evaluate used probe [9]. The standard deviation(SD) of program output result in the CMM verification testconditions is presented in Table 1. The CMM has limitedspecific values as follows:

MPE = 𝐴 + 𝐿𝐾, 𝜇m,

MPE𝑃= 1.00 𝜇m;

MPET𝑖𝑗= 1.90 𝜇m,

(1)

where MPE is the maximum permissible measurement error,A is the constant machine uncertainty equal to 0.9𝜇m, K isthe length constant or slope of line equal to 350, L is the lengthmeasurement in mm, MPE

𝑃is the maximum permissible

probing error, andMPET𝑖𝑗is themaximum permissible error

when measuring a part by using scanning mode which iscalled maximum permissible scanning probing error.

2.2. Flick Standard. The best method of dynamic verificationis made through the use of a high precision Flick standardin roundness Talyrond machines. Existing outer surface ofFlick standard is only one applicable embodiment type ofsensitivity verification for dimensional and geometrical form.Flick standard is a cylindrical artifactwith a knownnominallyone flat face through the outer perimeter. Flick standard havesome major disadvantages like a small dynamical contactfor larger wave numbers [10]. The amplitude of signals for atypical Flick with a cylinder diameter of about 44mm anda deepness of the flattening of (294.9 : 283.3) 𝜇m has beenmeasured; see Figure 1. While the measured signal to noiseratio is actually low for reference standard Flicks, thereforethe author predicted that the use of Flick standard will maybe a new suitable tool for verification method of CMMmeasurement quality.

The CMM output significant contributions from largerwave numbers than 25 UPR (undulation per revolution)are demonstrated in Figure 2. The output result has beenperformed in an appropriate measurement range for thediameter and roundness form deviations. But the calibrationresult depends on the individual wave amplitudes.The outputsignal to noise ratio is quite low for Flick, because the fullmeasured signal is carried by a single wave number.

2.3. CMMMeasurement Procedures. Themeasurements havebeen carried out at the same transverse section on the outer

Page 3: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

Journal of Quality and Reliability Engineering 3

Figure 1: Flick standard reference.

surface of the Flick standard at location of 18mm fromthe top to detect any errors of the surface. The measuredsample errors were obtained for five probe scanning speeds1, 2, 3, 4, and 5mm/s during 360∘ angle range trace ofthe Flick standard, respectively, while CMM traveling speedwas constant at 70mm/s and the number of scan fittingpoints also was constant with about 104 ± 2 points duringmeasurement tests at temperature condition of 20 ± 0.5∘C.Each measurement point has 10 times repetitions for thesame transverse circle (𝑥, 𝑦, and 𝑧) positions. The feature ofroundness measurement has been determined at each probescanning speed, where the CMM PRISMO navigator hasbeen selected corresponding to evaluation fitting algorithms.Four main geometric fitting algorithms are recommendedin CMM measurement for cylindrical parts, LSQ, ME, MC,and MI, where LSQmeans least square fitting algorithm, MEmeans minimum element fitting algorithm, MC is minimumcircumscribed fitting algorithm, and MI means maximuminscribed fitting algorithm. Determination of measured devi-ation errors have included 200 experimental measuringtests to differentiate between evaluation qualities of differ-ent measurement strategies. The diameter and roundnessdeviation errors of corresponding evaluation algorithm atdifferent scanning speeds are obtained through comparison,judgment, and repeated arrangement. However, a questionarises, which algorithm is suitable to be chosen and whatcriterion should be taken at which probe scanning speed.

3. Measurement Results and Discussion

The distribution density of measured points of both diameterand roundness variation is presented in Figures 3, 5, 7, 9, 11,13, 15, and 17. The errors’ average of results obtained for bothdiameter and roundness variation are reduced and presentedin a more practical and explicit form in Figures 4, 6, 8, 10,12, 14, 16, and 18.The diameter and roundness error results asfunctions of the probe scanning speed and fitting algorithmare given as follows.

3.1. Least Square Fitting Technique. The results in Figure 3present the distribution density of measured points fordiameter error of ten reputed test results for Flick transversecircle using LSQ fitting algorithm at different probe scanningspeeds. Figure 4 shows the average variation of diametererror for different probe scanning speeds using LSQ fittingalgorithm. Analysis of the fifty given results indicates that:

(i) detection LSQ fitting algorithm at probe speed of1mm/s measurements has a diameter error rangeof 0.2 𝜇m from 43.9705mm to 43.9703mm, while

measurements at 2mm/s have diameter error limitsof 43.9704 and 43.9703mm with an error range of0.1 𝜇m, while the error ranges at 3, 4, and 5mm/smeasurements became stable within the same valueof 0.1 𝜇m;

(ii) according to the application of the LSQ fitting tech-nique to all measuring speeds, the evaluated dif-ference of average error between probe speeds asrepresenting values to the scanning speed quality has0.1 and 0.2 𝜇m for 1, 2, and 3mm/s (43.9703mm)and 4mm/s (43.9702mm) and 5mm/s (43.9701mm),respectively;

(iii) measurements at 4mm/s have low diameter error forthe LSQ fitting algorithm, while at 5mm/s have thelowest diameter error.

The presentation results in Figure 5 show the distributiondensity ofmeasured points for roundness error of ten reputedtest results for Flick transverse circle using LSQ fittingalgorithm at different probe scanning speeds. Figure 6 showsthe average variation of roundness error for different probescanning speeds using LSQ fitting algorithm. The resultanalysis of the fifty tests indicates that:

(i) roundnessmeasurements of LSQ algorithmat 2mm/shave a minimum error of 0.2 𝜇m andmaximum errorof 0.5 𝜇m at 5mm/s;

(ii) according to the application of the LSQ fitting tech-nique to all measuring speeds, the evaluated dif-ference of average error between probe speeds asrepresenting values to the scanning speed qualityhas 0.3 for 1 and 4mm/s (288.8𝜇m) and 2mm/s(288.5 𝜇m) and at 3mm/s (288.9𝜇m) and (288.7𝜇m)at 5mm/s;

(iii) the value of roundness measurement errors for2mm/s has the lowest significant variation comparedto all other probe speeds.

3.2. Minimum Element Fitting Technique. Figure 7 showsthe density of measured points for diameter error of tenreputed test results for Flick transverse circle usingME fittingalgorithm at different probe scanning speeds. Figure 8 showsthe average variation of diameter error for different probescanning speeds using ME fitting algorithm. Analysis of thefifty given results indicates that:

(i) detection ME fitting algorithm at probe speed of1mm/s measurements has a diameter error rangeof 0.5𝜇m from 43.9868mm to 43.9863mm, whilemeasurements at 2mm/s have diameter error limitsof 43.9865 and 43.9864mm with an error rangeof 0.1 𝜇m. While the error ranges at 3mm/s mea-surements have the highest diameter error range of0.6 𝜇m, measurements at 2mm/s have the lowestdiameter error for the ME fitting algorithm, whileat 1mm/s they have the highest diameter error of0.5 𝜇m;

Page 4: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

4 Journal of Quality and Reliability Engineering

200.0 𝜇m

X

Y

(a) Outer roundness form profile

1 3 10 32 1000

20.00

40.00

60.00

80.00

100.00

120.00

316

Am

plitu

de (𝜇

m)

[—]

(b) Spectrum amplitude at max frequency 51Hz

Figure 2: Typical result of a Flick standard reference at scanning speed of 1mm/sec at LSQ fitting algorithm and Gaussian filter using CMMwith magnification factor 100.

43.9700

43.9701

43.9702

43.9703

43.9704

43.9705

43.9706

1 2 3 4 5Probe scanning speed (mm/s)

Dia

met

er er

ror (

mm

)

Test1Test2Test3Test4Test5

Test6Test7Test8Test9Test10

Least square

Figure 3: Diameter errors variation of LSQ fitting algorithm atdifferent probe scanning speeds.

(ii) according to the application of the ME fitting algo-rithm to all measuring speeds, the evaluated dif-ference of average error between probe speeds asrepresenting values to the scanning speed quality has0.1 and 0.6 𝜇m for 1, 3, and 4mm/s (43.9865mm)andat 2 and 5mm/s (43.9864mm), respectively;

(iii) the measuring error range has the lowest significantvariation at 2mm/s compared to 1, 3, 4, and 5mm/stesting speed, which may be due to probe response atresonance traveling speed.

Least square

43.9700

43.9701

43.9702

43.9703

43.9704

1 2 3 4 5Probe scanning speed (mm/s)

Dia

met

er er

ror (

mm

)

Figure 4: Measuring errors average of diameter at different probescanning speeds with LSQ fitting algorithm.

The presented results in Figure 9 show the distributiondensity of measured points for roundness error of tenreputed test results for Flick transverse circle usingME fittingalgorithm at different probe scanning speeds. Figure 10 showsthe average variation of roundness error for different probescanning speeds using ME fitting algorithm. Analysis of thegiven result indicates that:

(i) roundness measurements of ME algorithm at 2mm/shave a minimum average error of 285.7 𝜇m relative toall other probe speeds;

(ii) according to the application of the ME fitting tech-nique to all measuring speeds, the evaluated dif-ference of average error between probe speeds asrepresenting values to the scanning speed quality has(286.0 𝜇m) for 1, 3, 4, and 5mm/s;

Page 5: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

Journal of Quality and Reliability Engineering 5

288.3288.4288.5288.6288.7288.8288.9289.0289.1

1 2 3 4 5Probe scanning speed (mm/s)

Least square

Roun

dnes

s err

or (𝜇

m)

Test1Test2Test3Test4Test5

Test6Test7Test8Test9Test10

Figure 5: Roundness errors variation of LSQ fitting algorithm atdifferent probe scanning speeds.

Least square

288.4

288.5

288.6

288.7

288.8

288.9

289.0

1 2 3 4 5

Roun

dnes

s err

or (𝜇

m)

Probe scanning speed (mm/s)

Figure 6: Measuring errors average of roundness at different probescanning speeds with LSQ fitting algorithm.

43.986

43.98615

43.9863

43.98645

43.9866

43.98675

43.9869

1 2 3 4 5

Minimum element

Probe scanning speed (mm/s)

Test1Test2Test3Test4Test5

Test6Test7Test8Test9Test10

Dia

met

er er

ror (

mm

)

Figure 7: Diameter errors variation of ME fitting algorithm atdifferent probe scanning speeds.

Minimum element

43.9862

43.9863

43.9864

43.9865

43.9866

43.9867

1 2 3 4 5Probe scanning speed (mm/s)

Dia

met

er er

ror (

mm

)

Figure 8: Measuring errors average of diameter at different probescanning speeds with ME fitting algorithm.

285.0285.2285.4285.6285.8286.0286.2286.4286.6

1 2 3 4 5

Minimum element

Roun

dnes

s err

or (𝜇

m)

Probe scanning speed (mm/s)

Test1Test2Test3Test4Test5

Test6Test7Test8Test9Test10

Figure 9: Roundness errors variation of ME fitting algorithm atdifferent probe scanning speeds.

(iii) the value of roundness measurement errors for2mm/s has the lowest significant variation of 0.3 𝜇mcompared to all other probe scanning speeds.

3.3. Minimum Circumscribed Fitting Technique. The resultsin Figure 11 indicate the density of measured points fordiameter error of ten reputed test results for Flick transversecircle using MC fitting algorithm at different probe scanningspeeds. Figure 12 shows the average variation of diametererror for different probe scanning speeds using MC fittingalgorithm. Analysis of the fifty given results indicates that:

(i) detection MC fitting technique at probe speed of1mm/s measurements has a diameter error rangeof 0.3 𝜇m from 43.9907mm to 43.9904mm, whilemeasurements at 2mm/s have diameter error limitsof 43.9907 and 43.9905mm with an error range of0.2 𝜇m. While the error ranges at 3 and 5mm/smeasurements have the lowest diameter variationerror range of 0.1 𝜇m, measurements at 5mm/s have

Page 6: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

6 Journal of Quality and Reliability Engineering

Minimum element

285.6

285.7

285.8

285.9

286.0

286.1

1 2 3 4 5

Roun

dnes

s err

or (𝜇

m)

Probe scanning speed (mm/s)

Figure 10: Measuring errors average of roundness at different probescanning speeds with ME fitting algorithm.

43.9901

43.9902

43.9903

43.9904

43.9905

43.9906

43.9907

43.9908

1 2 3 4 5

Minimum circumscribed

Probe scanning speed (mm/s)

Test1Test2Test3Test4Test5

Test6Test7Test8Test9Test10

Dia

met

er er

ror (

mm

)

Figure 11: Diameter errors variation of MC fitting algorithm atdifferent probe scanning speeds.

the lowest average diameter error of 43.9903mm forthe ME fitting algorithm;

(ii) according to the application of the MC fitting algo-rithm to all measuring speeds, the evaluated differ-ence of average error between probe speeds as rep-resenting values to the scanning speed quality has 0.1and 0.3 𝜇m for 1 and 2mm/s (43.9906mm), for 3 and4mm/s (43.9905mm), and 5mm/s (43.9903mm),respectively.

(iii) the measuring error range has the lowest significantvariation at 3mm/s compared to 1mm/s testingspeed, which may be due to probe response at reso-nance traveling speed at MC algorithm design.

The presentation results in Figure 13 show the densityof measured points for roundness error of ten reputed testresults for Flick transverse circle usingMCfitting algorithmatdifferent probe scanning speeds. Figure 14 shows the averagevariation of roundness error for different probe scanningspeeds using MC fitting technique. The result analysis of thefifty tests indicates that:

Minimum circumscribed

43.9902

43.9903

43.9904

43.9905

43.9906

43.9907

1 2 3 4 5Probe scanning speed (mm/s)

Dia

met

er er

ror (

mm

)

Figure 12: Measuring errors average of diameter at different probescanning speeds with MC fitting algorithm.

288.8

289.0

289.2

289.4

289.6

289.8

1 2 3 4 5

Minimum circumscribed

Probe scanning speed (mm/s)

Test1Test2Test3Test4Test5

Test6Test7Test8Test9Test10

Dia

met

er er

ror (

mm

)

Figure 13: Roundness errors variation of MC fitting algorithm atdifferent probe scanning speeds.

(i) roundness measurements of MC algorithm at 1, 3,and 4mm/s have a minimum error of 0.4 𝜇m andmaximum error of 0.5 𝜇m at 2 and 5mm/s;

(ii) according to the application of the MC fitting tech-nique to all measuring speeds, the evaluated dif-ference of average error between probe speeds asrepresenting values to the scanning speed qualityhas 0.4𝜇m for 1 and 4mm/s (289.4 𝜇m) and 2mm/s(289.1𝜇m) and at 3mm/s (289.5𝜇m) and (289.3𝜇m)at 5mm/s;

(iii) the value of roundness measurement errors for2mm/s has the lowest significant variation comparedto all other probe scanning speeds.

3.4. Maximum Inscribed Fitting Technique. As in Figure 15,the density of measured points for diameter error of tenreputed test results for Flick transverse circle using MI fittingalgorithm at different probe scanning speeds. Figure 16 showsthe average variation of diameter error for different probe

Page 7: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

Journal of Quality and Reliability Engineering 7

Minimum circumscribed

289.0

289.1

289.2

289.3

289.4

289.5

289.6

1 2 3 4 5

Roun

dnes

s err

or (𝜇

m)

Probe scanning speed (mm/s)

Figure 14: Measuring errors average of roundness at different probescanning speeds with MC fitting algorithm.

43.7006

43.7008

43.701

43.7012

43.7014

43.7016

1 2 3 4 5

Maximum inscribed

Probe scanning speed (mm/s)

Test1Test2Test3Test4Test5

Test6Test7Test8Test9Test10

Dia

met

er er

ror (

mm

)

Figure 15: Diameter errors variation of MI fitting algorithm atdifferent probe scanning speeds.

scanning speeds using MI fitting algorithm. Analysis of thefifty given results indicates that:

(i) detection MI fitting technique at probe speed of1mm/s measurements has a diameter error rangeof 0.5 𝜇m from 43.7013mm to 43.7008mm, whilemeasurements at 2mm/s have diameter error limits of43.7015 and 43.7011mmwith an error range of 0.4𝜇m.While the error ranges at 3 and 5mm/smeasurementshave the highest diameter error range of 0.5 𝜇m,measurements at 4mm/s have the lowest diametererror variation of 0.2 𝜇m for the MI fitting algorithm;

(ii) according to the application of the MI fitting algo-rithm to all measuring speeds, the evaluated dif-ference of average error between probe speeds asrepresenting values to the scanning speed qualityhas 0.5 and 0.2𝜇m for 1 and 4mm/s (43.7010mm),for 3 and 5mm/s (43.9709mm) and for 2mm/s(43.9713mm), respectively.

The presented results in Figure 17 present the distri-bution density of measured points for roundness error

Maximum inscribed

43.7008

43.7009

43.7010

43.7011

43.7012

43.7013

43.7014

1 2 3 4 5Probe scanning speed (mm/s)

Dim

. err

or (m

m)

Figure 16: Measuring errors average of diameter at different probescanning speeds with MI fitting algorithm.

286.5

286.7

286.9

287.1

287.3

287.5

287.7

287.9

1 2 3 4 5

Maximum inscribed

Probe scanning speed (mm/s)

Roun

dnes

s err

or (𝜇

m)

Test1Test2Test3Test4Test5

Test6Test7Test8Test9Test10

Figure 17: Roundness errors variation of MI fitting algorithm atdifferent probe scanning speeds.

of ten reputed test results for Flick transverse circleusing MI fitting algorithm at different probe scanningspeeds. Figure 18 shows the average variation of round-ness error for different probe scanning speeds using MIfitting algorithm. Analysis of the fifty given results indicatesthat:

(i) roundness measurements of MI algorithm at 4mm/shave a minimum error of 0.4 𝜇m andmaximum errorof 0.6 𝜇m at 1, 3 and 5mm/s;

(ii) according to the application of the MI fitting tech-nique to all measuring speeds, the evaluated dif-ference of average error between probe speeds asrepresenting values to the scanning speed quality,has287.3𝜇m at 3 and 4mm/s, while at 1mm/s (287.4𝜇m)and (287.2𝜇m) at 2 and 5mm/s;

(iii) the value of roundness measurement errors for 2and 5mm/s has lowest significant average variationcompared to other probe speeds.

Page 8: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

8 Journal of Quality and Reliability Engineering

Table 2: Average variation and standard deviation of Flick diameter (mm).

Scanning speed, mm/s LSQ ME MC MI Average of𝐷 SD1 43.9703 43.9865 43.9906 43.7010 43.9121 0.12212 43.9703 43.9864 43.9906 43.7013 43.9122 0.12203 43.9703 43.9865 43.9905 43.7009 43.9121 0.12214 43.9702 43.9865 43.9905 43.7010 43.9121 0.12215 43.9701 43.9864 43.9903 43.7009 43.9119 0.1221Average of𝐷 43.9702 43.9865 43.9905 43.7010 43.9120SD 0.0001 0.0000 0.0001 0.0001𝑢1= SD/√𝑛 0.0001 0.0000 0.0001 0.0001

Maximum inscribed

287.1

287.2

287.3

287.4

287.5

1 2 3 4 5Probe scanning speed (mm/s)

Roun

dnes

s err

or (𝜇

m)

Figure 18: Measuring errors average of roundness at different probescanning speeds with MI fitting algorithm.

4. Statistical Analysis

To give more reliable analysis of the experimental data of theinfluence of fitting algorithm and probe scanning speed onthe CMM measurement accuracy should be using statisticaltests. Statistical test analysis of diameter and roundness erroraverages, standard deviation and combined uncertainty (u

1)

due to repeatability are calculated. Expanded uncertainty forselected parameters in measurement has been evaluated.Thestatistical results obtained are reduced and presented in amore practical and explicit error form in Figures 19 and 20.The errors averages in diameter and roundness results asfunction of the five probe scanning speed using four fittingtechniques are given as follows.

4.1. The Error in Diameter Measurement. The error responseof diameter variation for Flick standard related to differentCMMmeasurements strategies of different fitting algorithmsof probe scanning speeds is indicated in Table 2 and presentedin Figure 19. The indicated results include the following:

(i) signal measured for 200 evaluated samples has aglobal average of Flick diameter of 43.9120mm. Thesamples at probe scanning speeds of 1, 2, 3, 4, and5mm/s have the roundness averaged values 43.9121,43.9122, 43.9121, 43.9121, and 43.9119mm, which cor-respond to represent 100.0002, 100.0005, 100.0002,100.0002, and 99.9998% of the global average, respec-tively. It ensures that measurements at probe speed

5mm/s are the suitable case for this work piece Flickto satisfy the high level of accuracy;

(ii) averaged percentage errors as a function of the fittingalgorithm response with respect to the global meanvalue are 100.1325, 100.1697, 100.1788, and 99.5195%,which are corresponding to LSQ, ME, MC, and MI,respectively. Quality of measurements indicates thatMC and ME algorithms have about 0.1788% and0.1697% inaccuracy, while the LSQ andMI algorithmshave accurate responses with the error range of0.1325% and −0.4805%, respectively;

(iii) signal measured for evaluated test samples at probescanning speeds from 1 to 5mm/s (with 1mm/sinterval value) has the standard deviation (SD) valuesof 0.1221, 0.1220, 0.1221, 0.1221, and 0.1221mm, respec-tively. It ensures that measurement at probe speedof 2mm/s is the suitable case for this work piece tosatisfy the high level of accuracy;

(iv) the diameter standard deviation has the best signifi-cant variation at 2mm/s for all fitting algorithms; thismay be due to probe response at resonance travelingspeed;

(v) the ME algorithm has the lowest standard deviationaverage response, while all other algorithms have thesame accurate response within the application range.

(vi) the uncertainty due to repeatability in diameter mea-surement for four fitting algorithms does not exceed0.0001;

(vii) from the data presented in Figure 19, the values havebeen treated statistically using fit linear regressiontype to get general formulae of the diameter error inmm as a function of probe scanning speed V for thedifferent four fitting algorithms as follows:

RON(MC) = −1𝐸 − 05𝑉 + 43.9860,

RON(ME) = −7𝐸 − 05𝑉 + 43.9910,

RON(LSQ) = −5𝐸 − 05𝑉 + 43.9700,

RON(MI) = −5𝐸 − 05𝑉 + 43.7010.

(2)

From the linear regression equations (2), the empiricalformulae illustrate that the ME algorithm has high error

Page 9: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

Journal of Quality and Reliability Engineering 9

Table 3: Average variation and standard deviation of Flick roundness (𝜇m).

Scanning speed, mm/s LSQ ME MC MI Average of RON SD1 288.8 286.0 289.4 287.4 287.9 1.31532 288.5 285.7 289.1 287.2 287.6 1.30653 288.9 286.0 289.5 287.3 287.9 1.37184 288.8 286.0 289.4 287.3 287.9 1.32555 288.7 286.0 289.3 287.2 287.8 1.2903Average of RON 288.7 285.9 289.3 287.3 287.8SD 0.1356 0.1200 0.1356 0.0748𝑢1= SD/√𝑛 0.0606 0.0537 0.0606 0.0335

Average

1 2 3 4 543.6

43.7

43.8

43.9

44.0

44.1

Flic

k di

amet

er er

ror (

mm

)

LSQME

MCMI

Probe scanning speed (mm/s)

Figure 19: Average variation of Flick diameter measurements atdifferent CMM strategies.

potential of 43.991mm at sensitivity coefficients of −7𝐸 − 05to the probe scanning speed (V), whereMI technique has thelowest error potential of 43.701mm at sensitivity coefficientsof −5𝐸 − 05 to the probe scanning speed. The LSQ algorithmhas an error potential of 43.970mm at sensitivity coefficientsof −5𝐸 − 05 to the probe scanning speed. The MC algorithmhas an error potential of 43.986mm at sensitivity coefficientsof −1𝐸 − 05 to the probe scanning speed.

4.2.TheError in RoundnessMeasurement. Theerror responseof roundness variation for Flick standard related to differentCMMmeasurements strategies of different fitting algorithmsof probe scanning speeds is presented in Table 3 and shownin Figure 20. The statistical results indicate that:

(i) signal measured for 200 evaluated samples has aglobal average of Flick roundness of 287.8 𝜇m. Thesamples at probe scanning speeds of 1, 2, 3, 4, and5mm/s have the roundness averaged values 287.9,287.6, 287.9, 287.9, and 287.8𝜇m, which correspond torepresent 100.04, 99.93, 100.04, 100.04, and 100.00%of the global average, respectively. It ensures thatmeasurement at probe speed 2mm/s is the suitablecase for this work piece Flick to satisfy the high levelof accuracy.

1 2 3 4 5285

286

287

288

289

290

Flic

k ro

undn

ess e

rror

(𝜇m

)

Probe scanning speed (mm/s)

AverageLSQME

MCMI

Figure 20: Average variation of Flick roundness measurements atdifferent CMM strategies.

(ii) averaged percentage errors as a function of the fittingalgorithm response with respect to the global meanvalue are 100.31, 99.48, 100.73, and 99.83% which arecorresponding to LSQ,ME,MC, andMI, respectively.Quality of measurements indicates that LSQ and MCalgorithms have about 0.31% and 0.73% inaccuracy,while the ME algorithm has accurate responses withthe error range of 0.52%.

(iii) signal measured for evaluated test samples at probescanning speeds from 1 to 5mm/s (with 1mm/sinterval value) has the standard deviation (SD) valuesof 1.3153, 1.3065, 1.3718, 1.3255 and 1.2903mm, respec-tively. It ensures that measurement at probe speedof 5mm/s is the suitable case for this work piece tosatisfy the high level of accuracy.

(iv) the roundness standard deviation has significant vari-ation at 3mm/s for all fitting algorithms; this may bedue to probe response at resonance traveling speed.

(v) the LSQandMCalgorithmshave the highest standarddeviation average response, while the MI algorithmhas accurate response within the application range.

(vi) the uncertainty due to repeatability in roundnessmeasurement for four fitting algorithms lies between0.0606 and 0.0335.

Page 10: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

10 Journal of Quality and Reliability Engineering

Table 4: Evaluation of uncertainty budget in roundness deviation measurements.

CMM fitting algorithms LSQ ME MC MICombined standards uncertainty, 𝑢

𝑐0.52650 0.52555 0.52650 0.523873

Expanded uncertainty 𝑈exp 1.0526 1.0511 1.0526 1.0478Average of 𝑈exp 1.0510

(vii) from the data presented in Figure 20, the valueshave been treated statistically using fourth orderpolynomial regression fit type to get general formulaeof the roundness error (RON) in 𝜇m as a function ofprobe scanning speed 𝑉 for the different four fittingalgorithms as follows:

RON(MC) = 0.0708𝑉

4− 0.9083𝑉

3+ 4.0292𝑉

2

− 7.0917𝑉 + 293.3000,

RON(LSQ) = 0.0708𝑉

4− 0.9083𝑉

3+ 4.0292𝑉

2

− 7.0917𝑉 + 292.7000,

RON(MI) = 0.0167𝑉

4− 0.2333𝑉

3+ 1.1333𝑉

2

− 2.2167𝑉 + 288.7000;

RON(ME) = 0.0500𝑉

4− 0.6500𝑉

3+ 2.9500𝑉

2

− 5.3500𝑉 + 289.0000.

(3)

From polynomial regression equations (3), the empiricalformulae illustrate that theMCandLSQalgorithms have higherror potentials of 293.3 and 292.7 𝜇m to the probe scanningspeed, where MI technique has the lowest error potential of288.7 𝜇m at high sensitivity coefficients of −2.2 to the probescanning speed (V). The MC and LSQ algorithm have thesame sensitivity coefficient of 4.0 to the probe scanning accel-eration.TheME algorithm has an error potential of 289.0 𝜇mat sensitivity coefficients of −5.4 to the probe scanning speedand sensitivity coefficients of 3.0𝜇m to the probe scanningacceleration.

4.3. Uncertainty Evaluation. In coordinate metrology, tomeasure the cylindrical artifact, as in our work, the mea-surement uncertainty mainly results from CMM machine,measurement environment, and sampling strategies [11].Whilst the uncertainty influence of sampling strategy hasnot been considered here due to its complexity, while theuncertainty significant contributions include the followingparameters.

4.3.1. Repeatability. The statistical study based on designof experiment was carried out in order to evaluate theexpanded uncertainty in measurement. Statistical study ofthe repeatability in measurement has been calculated andevaluated for four fitting algorithms LSQ, ME, MC, and MI,which have the type (A) uncertainty (u

1) values of 0.0606,

0.0537, 0.0606, and 0.0335 𝜇m, respectively, as shown inTable 3.

4.3.2. Resolution. The resolution 𝑟 of CMM in last digit of ameasured value, is causing an uncertainty component u

2:

𝑢2=𝑟

2√3=0.1

2√3= 0.0289 𝜇m (4)

4.3.3. Indication Error. The maximum permissible error ofindication is 0.9 𝜇m.When a normal distribution is assumed,the uncertainty component is

𝑢3=0.9

√3= 0.5196 𝜇m (5)

4.3.4. Temperature. The standard reference temperature formeasurement is 20∘C. During the implementation, the envi-ronmental temperature in the coordinate metrology labora-tory at NIS was controlled within 20 ± 0.5∘C, the uncertaintycomponent u

4from temperature and dirt is estimated at

0.05 𝜇m.The previous components are all uncorrelated, so the

uncertainties of measured points for roundness variation arecalculated as follows:

𝑢𝑐= √𝑢21+ 𝑢22+ 𝑢23+ 𝑢24, (6)

𝑈exp = 𝐾 (𝑢𝑐) . (7)

Eventually, the evaluation of uncertainties in this workreflects confidence in the high credibility of the proposedmeasurement method as shown obviously in Table 4. Thus,the author can say that the MI fitting technique is closerto reality and very much more accurate than the othertechniques. In addition, the evaluation results illustrate thatthe LSQ and MC algorithms have higher uncertainties of1.0526 𝜇m, where MI technique has the lowest uncertaintiesof 1.0478𝜇m.

5. Conclusions

This paper presents new experimental investigations toimprove themeasurement accuracy of dimension and round-ness form of a coordinate measuring machine for particularmeasurement tasks. The proposed method is very importantin the development issue of CMMs dynamic verification to bemore accurate and precise machines in measurement. Someerror formulae of two main data sets have been postulated tocorrelate the diameter and roundness measurements within

Page 11: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

Journal of Quality and Reliability Engineering 11

the application range.Themethod proposed requires just twoselections of geometric fitting algorithms and probe scanningspeeds in cylindrical measurements. From this research toimprove dimension and roundness measurements quality,some conclusions can be drawn as follows:

(i) the investigation of dimension and form deviationusing Flick standard may new suitable use as a highsensitive verification method for CMM performance[5];

(ii) suitable scanning speed of the machine touch probeshould be well selected in accordance with the fittingalgorithm to get high response quality with accuratedimension and roundness measurements;

(iii) for diameter and roundness measurements, MI algo-rithm shows lowest error potential as an indication tohigh quality response;

(iv) for diametermeasurement both treatment algorithmsensure stable measuring accuracy at all probe scanspeeds. This is probably due to the geometric per-fection resonance of the circular surface of the Flickstandard;

(v) for roundness measurement, both geometric fittingtreatment algorithms ensure high measuring accu-racy at 2mm/s probe scan speed may be due to probedesign resonance;

(vi) for similar diameter measurements, MI algorithmshow high quality response, while other algorithmsshow low quality response to the probe scanningspeed;

(vii) for similar roundness measurements, ME algorithmshow high quality response beside MC algorithmshow low quality response to the probe scanningspeed;

(viii) the expanded uncertainty result has been evaluatedwithin 1.051 𝜇m, which was less than the maximumlimit of the CMM permissible probing error (1.9 𝜇m),confirming the high degree of confidence in themeasurement results. Thus, the Flick standard mayuse as a simple reliable artifact for CMM evaluation;

(ix) from the investigations analysis, the author observedthat there are proportional relationship between eachof uncertainty and error potential of the empiri-cal error formulae in measurement, Thus, the LSQand MC algorithms have higher uncertainties whichagreed with the higher rate of there error potentials,where MI technique has a lowest uncertainty of1.0478 𝜇m that agreed with the lowest rate of theirerror potential;

(x) using a Flick standard should be practical highprecision incremental in dimension (diameter) androundness form indicator with high stability in mea-surement. This means a better transfer stability ofCMM quality could be significantly improved;

(xi) eventually, the most basic measuring applications,the metrology of supplemental Flick standard, will

increase the knowledge about the state of CMMmeasuring strategies.

Conflict of Interests

The authors declare that there is no conflict of interestsregarding the publication of this paper.

References

[1] J. Buajarern, T. Somthong, S. H. R. Ali, and A. Tonmeuanwai,“Effect of step number on roundness determination usingmulti-step method,” International Journal of Precision Engineer-ing and Manufacturing, vol. 14, no. 11, pp. 2047–2050, 2013.

[2] S. H. R. Ali, H. H. Mohamed, and M. K. Bedewy, “Identifyingcylinder liner wear using precise coordinate measurements,”International Journal of Precision Engineering and Manufactur-ing, vol. 10, no. 5, pp. 19–25, 2009.

[3] I. Vrba, R. Palencar, M. Hadzistevic, B. Strbac, and J. Hodolic,“The influence of the sampling strategy and the evaluationmethod on the cylindricity error on a coordinate measurementmachine,” Journal of Production Engineering, vol. 16, no. 2, 2013.

[4] S. H. R. Ali and J. Buajarerm, “New measurement methodand uncertainty estimation for plate dimensions and surfacequality,” Advances in Materials Science and Engineering, vol.2013, Article ID 918380, 10 pages, 2013.

[5] H. Nouira and P. Bourdet, “Evaluation of roundness errorusing a new method based on a small displacement screw,”Measurement Science and Technology, vol. 25, no. 4, 2014.

[6] International StandardOrganization, ISO/IECGuide 98-3:2008,Uncertainty of Measurement—Part 3: Guide to the Expression ofUncertainty in Measurement, GUM, 1995.

[7] International Standard Organization: Geometrical ProductSpecifications (GPS), “Acceptance and reverification tests forCoordinate Measuring Machines (CMM)—Part 2: CMMs usedfor Measuring Size,” ISO 10360-2, 2nd ed, 2001.

[8] International StandardOrganization,Geometrical Product Spec-ifications (GPS)—Acceptance and Reverification Tests for Coor-dinate Measuring Machines (CMM)—Part 4: CMMs Used inScanning Measuring Mode, ISO 10360-4, 2nd edition, 2000.

[9] S. H. R. Ali, “Probing system characteristics in coordinatemetrology,”Measurement Science Review, vol. 10, no. 4, pp. 120–129, 2010.

[10] H. S. Nielsen and M. C. Malburg, “Traceability and correlationin roundness measurement,” Precision Engineering, vol. 19, no.2-3, pp. 175–179, 1996.

[11] X. Wen, Y. Xu, H. Li, F. Wang, and D. Sheng, “Monte Carlomethod for the uncertainty evaluation of spatial straightnesserror based on new generation geometrical product specifica-tion,” Chinese Journal of Mechanical Engineering, vol. 25, no. 5,pp. 875–881, 2012.

Page 12: Research Article Performance Investigation of CMM …downloads.hindawi.com/archive/2014/960649.pdf · 2019-07-31 · Research Article Performance Investigation of CMM Measurement

International Journal of

AerospaceEngineeringHindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

RoboticsJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Active and Passive Electronic Components

Control Scienceand Engineering

Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

International Journal of

RotatingMachinery

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporation http://www.hindawi.com

Journal ofEngineeringVolume 2014

Submit your manuscripts athttp://www.hindawi.com

VLSI Design

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Shock and Vibration

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Civil EngineeringAdvances in

Acoustics and VibrationAdvances in

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Electrical and Computer Engineering

Journal of

Advances inOptoElectronics

Hindawi Publishing Corporation http://www.hindawi.com

Volume 2014

The Scientific World JournalHindawi Publishing Corporation http://www.hindawi.com Volume 2014

SensorsJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Modelling & Simulation in EngineeringHindawi Publishing Corporation http://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Chemical EngineeringInternational Journal of Antennas and

Propagation

International Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Navigation and Observation

International Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

DistributedSensor Networks

International Journal of