slide intra unikl
TRANSCRIPT
INDUSTRIAL TRAININGHealthtronics (M) Sdn
Bhd
To understand and experience the real-life
situation in industrial organization and their
related environment.
Exposure to practice and apply the acquired
knowledge “hands on” in the working
environment.
OBJECTIVE
HEALTHTRONICS (M) SDN BHD
Well-established Malaysian based Biomedical
Engineering company
A subsidiary of Faber Medi-Serve (M) Sdn Bhd (FMS),
the largest integrated hospital support service provider
in Malaysia
Provides comprehensive biomedical equipment
maintenance and management services to over 100
hospitals in the ASEAN region for the past 10 years
COMPANY PROFILE
COMPANY ORGANIZATION CHART
(PENANG REGION)
V. Sangaran - Technician
Mohd Azmi - TA
Regional Op Executive
Hospital Sg Bakap
Hospital Bukit MertajamHospital Seberang Jaya
Team Leader
Regional Technical Exec
Snr Engineer
Haslinda Che Hassan
Radzi Mohamed - TA
M.K. Suhaimi - TE
based in PGH
Lim Leong Choon - Snr. TE
Rajesh Kumar - Snr. TE
Hospital Pulau Pinang
Asst. Regional HeadLakmal C. Maitipe
Hospital Balik PulauDevananthan Gunasegaran - TEShukour Kamarudin - TE
Regional MMD
Rosnita Noordin - Admin Asst
PENANG ORGANIZATION CHART
Regional Admin/Finance/HR
Stephanie Quah
Regional SODA
Khoo Ooi Koon
Hospital Bukit Mertajam
Regional Head
Hospital Pulau Pinang
HEALTHTRONICS SDN BHD
Admin Exec
Yusoff Harun - Technician
Jetendrajee - TE
Mohd Idi - Technician
Mohd Azlan Zainol - Technician
Hospital Kepala Batas
Azhar Zainuddin - TE
Shaharuddin - Technician
Nailah - Admin Asst
V.R. Kailasabathy
Team Leader
Snr Assistant
Hospital Seberang JayaSyed Ahmad - TE
Team Leader
Marzuki Manaf - TE
Mohd. Sabarudin - TechnicianMohd. Faizul - TA
Nursina Abdul Sidik - Logistic Asst
Yaakob Bakar - TE
Faisal Che Pa - Technician
Wan Rafizi - TA
Zunita - Admin Asst
Saravanan Rajagopal - TE
Mahendran Pullaiyar - TE
Teoh Jiunn Eng - Technician
Muhamad Azam Mat Isa - TE
R. Rajarathinam - TE
Niramon Eliam - Admin Asst.
Noorzatul Iffa - TE
Edw in Gonzales - TE
Azam Fadli - TA
Hasnizar - Logistic Asst
1. Maintenance Management for BEMS◦ Planned Preventive Maintenance (PPM)◦ Scheduled Corrective Maintenance (SCM)◦ Breakdown Maintenance (BM)
2. Biomedical Equipment Planning and Consultation Service
3. User Training4. Training Centre
COMPANY SERVICES
Testing and commissioning on installation of new
biomedical equipment.
Maintenance management for biomedical
equipment.
Technical support and giving technical advice to
user and customer.
Supplying biomedical parts to user and customer.
BEMS - BIOMEDICAL ENGINEERING MAINTENANCE SERVICES
To ensure all the documents are being prepared by vendors.
To ensure that the vendor will provide a user training program for the equipment the hospital bought.
To run a test on any medical equipment before the hospital agreed to purchased.◦ Electrical Safety Test◦ Equipment Accuracy Test.
Testing and commissioning
Planned Preventive Maintenance (PPM)◦ To ensure that the medical equipment operates in
accordance with its design parameters and safe working conditions.
◦ To reduce the breakdown of the working medical equipment.
◦ Regular PPM inspection and equipment calibration will be performed in accordance with ISO accepted standards.
Maintenance management
Electrical safety test
Tester/analyzer
Defibrillator analyzer and tester.
Tester/analyzer
Oximeter analyzer and tester
Tester/analyzer
Infusion pump tester
Tester/analyzer
Breakdown Maintenance (BM) Corrective Maintenance (CM)
◦ To restore the equipment to its specified condition when it fails.
Maintenance management
CONDITION APPRAISAL OF CELL COUNTING ANALYZER BECKMAN COULTER
◦ Describe the usage or working principles of the
machine
◦ Details of the machine’s breakdown and
preventive maintenance lists
◦ Giving recommendation to improve the machine’s
services
INDUSTRIAL TRAINING PROJECT
Procedures
Attend breakdown and preventive
maintenance for the cell counting analyzer.
Study the principle of the machine.
Doing research on the machine
maintenance history.
Basic princple
BREAKDOWN LISTDate
(work order)
Problems
09/02/2002 Vacuum error
20/08/2002 Not functioning
24/02/2004 Probe stuck
22/05/2004 WBC high count, PLT high count. Change PPM kit.
23/07/2004 WBC test failed. Clean WBC chamber.
17/01/2005 Platelet count failed. Clean probe.
17/03/2005 Platelet result not accurate. Clean RBC chamber.
11/08/2006 WBC count not accurate. Clean the transducer chamber.
Causes of the breakdown
Parts having a wear and tear condition
Fluid, reagents or blood clotting inside the
tube, probe and chambers.
Recommendation to improve the service.
Do more user training courses.
Frequent cleaning.
Stock ready for the parts.
LEARN MORE ABOUT ELECTRONIC
TROUBLESHOOT.
NEW KNOWLEDGE CAN BE OBTAIN
ESPECIALLY IN BIOMEDICAL ENGINEERING.
CONCLUSION