spect: new technologies and applications
TRANSCRIPT
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
SPECT: New Technologies and Applications
Eric C. Frey, Ph.D., ProfessorDivision of Medical Imaging Physics
Russell H. Morgan Department of Radiology and Radiological Science
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Disclosures
Johns Hopkins University Licenses an iterative reconstructioncode for SPECT to GE Healthcare. Dr. Frey is entitled to a shareof royalty received on sales by GE Healthcare of thisreconstruction code. The terms of this arrangement are beingmanaged by the Johns Hopkins University in accordance with itsconflict of interest policies
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Learning Objectives
• Review current and future advances in instrumentation and reconstruction
• Describe new applications, especially those related to cancer therapy
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Outline
• New Technologies– Instrumentation– Reconstruction
• New Applications
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
New Technologies: Instrumentation
• Collimators• Detectors• Photodetectors• Electronics• Multimodality Systems• Commercial Systems
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Developments in Collimators
• Goal: improve resolution sensitivity tradeoff– Parallel-hole collimator Sensitivity ~ (FWHM Res)2
• Pinhole collimators• Slit/Slat collimators• Multi-focal collimators
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Pinhole Collimators
θ
g de2 sin316Z 2
Rph B Z de
B (in object plane)
Rt Rph2
BZ
Ri
2
Field-of-view=F D ZB
Magnification M BZ
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Pinhole vs. Parallel
• What is ratio of sensitivity for pinhole and parallel for equal resolution? gph
gp4k
1M 1
2
if dZL
2
MRi,ph 2
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Pinhole vs. Parallel
• Pinhole sensitivity is better than parallel ifM > 0.42
and
• For fixed size object F, DR to be small• D is limit by floor on M• Thus, need small intrinsic resolution for pinhole
camera
dZL
MRi,ph
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Multiple Pinholes
• Pinhole can use minification to reduce required size of detector (e.g., by factor of 2)
• Can thus use multiple pinholes (e.g., 4x) to image same size object
• Further improves sensitivity tradeoff
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Limitation of Pinholes for Tomography
• Pinhole geometry is similar to cone-beam• Rotating pinhole in plan around object only
provides data sufficient to reconstruct central slice of object
• Requires more complicated orbits• Multiple pinholes partly remove this restriction
because they are not in the same plane
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Slit-Slat Collimator
• Slit (like 1D pinhole) in transaxial direction
• Slats (like 1D parallel hole collimator) in axial direction
• Partial benefits of pinhole– Better resolution/sensitivity tradeoff– Magnification/minification in one
direction (multiple slits)
• Complete data for reconstruction (fan-beam)
Slit
Slat
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Multifocal Collimators
• Converging collimators provide better sensitivity for same resolution
• Problem: Truncation• For small volumes of interest (VOI)
in a large object, multifocal collimators may offer benefits– Improved sensitivity in VOI– No truncation artivacts
• Require motion that keeps VOI in region of high sensitivity
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Converging Collimators
• Converging collimators provide better sensitivity for same resolution
• Problem: Truncation
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Multifocal Collimators
• For small volumes of interest (VOI) in a large object, multifocal collimators may offer benefits– Improved sensitivity in VOI– No truncation artifacts
• Require motion that keeps VOI in region of high sensitivity
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
New Detectors: CZT
• CZT (CdZnTe)– Semiconductor– Direct Conversion– Better energy resolution– Tailing:
• Trapping• Lower hole mobility
– Pixel sized determined by contact spacing• Charge sharing a challenge• Requires sophisticated per-pixel electronics (ASICs)
– Cost and growing large crystals still a challenge
0�
0.2�
0.4�
0.6�
0.8�
1�
1.2�
0� 20� 40� 60� 80� 100� 120� 140� 160� 180�Intens
ity�
Energy�(keV)�
Measured�Model�
FWHM=4.1 keV
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Pixelated Scintillators
• Cut scintillators into discrete crystals• Opaque light isolation• Resolution determined by size of crystal• Sizes < 1 mm• Pixel identification instead of position estimation
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
New Scintillators:
• LaBr3(Ce)– 65% higher light output than NaI(Tl)– Denser and similar effective Z (47 vs 50)– Substantially more expensive– Available in much smaller crystal sizes– Possibly useful for small, modular camera better
intrinsic resolution
SpecificGravity
Max nm
Index ofRefraction
Decay Time(µs)
Light Yield(photons/MeV)
Rel Yieldw/PMT
NaI(Tl) 3.67 415 1.85 0.23 38,000 1.00LaBr3(Ce) 5.08 380 1.9 0.016 63,000 1.65
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Advances in Photodetectors
• Position Sensitive/Multi-Anode PMTs• P-I-N Photodiodes• Silicon Drift Detectors• Avalance Photodiodes• SiPMTs
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Position Sensitive PMTs (PSPMTs)
• PMT with multiple outputs (4x4 to 16x16)• Output signals preserve spatial information about
photoelectron production• Similar properties to PMTs• Up to 50x50 mm active area (3x3 mm pixel size)• Can be tiled (~6 mm gaps)• Larger and lower efficiency than solid-state
alternatives
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
P-I-N Photodiode
• Reverse-biased P-I-N diode• Production of photoelectrons produces charge• No amplification, require very low noise preamp• Low noise (dark current)• Simple, stable, compact• Different absorption peak than PMT photocathodes• Available in arrays
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Silicon Drift Detectors
• Series of electrodes in concentric rings creates drift region
• Photoelectrons drift toward anode in center of device
• Small anode gives low capacitance and good energy resolution
• Signal collected and amplified using integrated JFET provides low noise
• Arrays and large area devices have been produced
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Avalanche Photodiodes
• Special designs to allow large reverse bias (100-2000 volts)
• Avalanche multiplication by impact ionization yields amplification of signal (gain of 100-1000)
• Sensitive to temperature and bias voltage fluctuations
• Available in arrays and position-sensitive versions
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
SiPM
• Arrays of many small photodiodes operating in GM-Mode– Each diode acts as binary photodetector with large gain– Using array of many photodiodes provides analog
behavior to scintillation light
• Relatively new• Modest size arrays available
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Comparison of Photodetectors
PMT PIN PD SDD APD SiPMlog10(gain) 6‐7 0 0* Lo 6Noise Low Low Low Moderate LowPhotodetectionefficiency(420nm)
~20% 70% ~60% ~30%
FormFactor Bulky Compact Compact Compact CompactSensitivitytomagneticfields
High Low Possible Low Low
*Built‐inpreamp.
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Advances in Electronics
• Move from analog to digital• Front-End Electronics• Improved position and energy estimation
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Moving from Analog to Digital
γ Detector
hν Detector
ImageFormation
DigitalX,y,E Estimation
Analog Pulse Shaping
Digital Pulse ProcessingAnalog
X,y,E Estimation
ADC
ADC
ADC
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Front End Electronics
• Achieving high intrinsic resolution requires processing more analog outputs from detectors
• This requires many more chains of analog and digital electronics
• Application specific integrated circuits are important for miniaturizing and reducing costs
• Field-programmable gate arrays (FPGAs) provide front-end digital logic
• Digital single processors often used to perform digital pulse processing
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Improved Position and Energy Estimation• Traditional position and energy estimation
– Simple Anger equations– Apply energy, spatial, linearity corrections– Limits accuracy and precision of energy and position
estimates– ‘Dead area’ at adge of detector
• Modern approaches– Treat as estimation problem– Apply maximum likelihood techniques– Better energy and position estimates– Greater uniformity– Larger useful FOV
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Multimodality: SPECT/CT
• First generation (GE Hawkeye):– Low power CT– 1-4 slice detector
• Slice thickness: 2.5-10 mm
– Attached to Slip-Ring SPECT Gantry
• 2.5 rpm->up to 9 minutes to acquire data for entire SPECT FOV
• Breathing artifacts
X-rayTube
X-rayDetector
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
SPECT/CT
• Second Generation (Philips Precedence)– CT scanner joined to
SPECT system– Diagnostic-quality CT
and acquisition times– Large footprint
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
SPECT/CT
• Third Generation (Siemens Symbia T, GE NM/CT 670)– Integration of XCT and SPECT– Smaller footprint than 2nd
generation– Diagnostic quality CT
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
SPECT/CT
• Flat Panel CT and x-ray source (Philips BrightView XCT)– High resolution– 80 mA tube– Not diagnostic CT quality
• Other (Digirad Cardius X-ACT)– Pb x-ray fluorescence source and NM detectors– Fan-beam geometry– Low x-ray flux/low resolution– Suitable for attenuation map
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Multimodality: SPECT/MR
• Semiconductor detectors or solid state photodetectors
• Collimators a challenge (tungsten epoxy, tungsten beads, …)
• Investigational for small animals• Potential for true simultaneous imaging
– Monitor motion– Dynamic imaging
• At developmental stage
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Novel Commercial Systems
• Digirad Cardius X-ACT– Three cameras with solid-state photodetectors, fan-beam
collimators transmission CT using x-ray fluorescence source
• GE NM530/570– Pinhole, CZT
• D-SPECT– Scanning small FOV CZT detectors with parallel-hole
collimators
• Cardiarc HD– Multiple moving slit/slat collimators , curved NaI
detector
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Digirad Cardius X-ACT
Modular solid state detectorpixelated CsI(Tl) scintillator
photodiode array
Figures courtesy of C. Bai and R. Conwell, Digirad
3 Camera, fan-beam collimators, seated imaging position
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Digirad Cardius X-ACT
Figures courtesy of C. Bai and R. Conwell, Digirad
Sample TCTImages
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
GE Discovery NM 530c & NM/CT 570c
• CZT detectors (2.5 mm pixels)• Multiple (~16) pinholes• No rotation required• True 3D dynamic SPECT• Improved energy resolution• >5x reduction in scan time
99mTc: 140 keV
123I: 159 keV
50 100 150 keV 200
point sources in air Alcyone 6.2% Anger 9.5%
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Spectrum Dynamics DSPECT
• 10 CZT detectors (2.5 mm pixels)
• Parallel-hole collimation• Detectors moveindividually• Very high sensitivity (10x?)
Images courtesy of Spectrum Dynamics
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Spectrum Dynamics D-SPECT
64 elements ~160 mm
16 elements ~40 mm
2.46 mm
2.46 mm
5 mm thick CZT Detector Element
1024 elements
per column
Detector Column
Column Collimator
Detector Array
Direct conversion of photon energy to voltage pulse
Figures courtesy of Spectrum Dynamics
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Cardiarc HD-SPECT
• Conventional (curved) NaI/PMT detector
• Slit/slat collimator (fan beam geometry)
• Resolution determined by slit width and slat spacing
• Multiple slits (equivalent to multiple cameras)
• Only slats (aperture arc) move
• Seated imaging position
6 slitsHorizontal
Slats (one per slice)
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Outline
• New Technologies– Instrumentation– Reconstruction
• New Applications
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Developments in Reconstruction
• Quantitative SPECT• Simultaneous Dual Isotope SPECT• Quantitative Bremsstrahlung SPECT
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Quantitative SPECT
• Requires– Attenuation and scatter compensation– CDR compensation desirable for small objects and high
energy photons– Calibration
• Sensitivity measurement (source in air)• Calibration phantom
• Applicable to a range of radionuclides• Technologies available, but not typically
implemented by manufacturers
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Quantitative SPECT with In-111
NC A AS AGS ADS Atn Map
NC=No CompensationA=Attenuation Compensation
AD=Attenuation and CDR Comp
AS=Attenuation and Scatter CompensationADS=Attenuation, CDR and Scatter Comp
B. He, Y. Du, X. Song, W.P. Segars and E.C. Frey, “A Monte Carlo and physical phantom evaluation of quantitative In-111 SPECT,” Phys Med Biol, 50(2005): 4169-4185, 2005.
GE Millenium VG w/Hawkeye SPECT/CT system, MEGP Collimator
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Quantitative SPECT with In-111
Organ NoComp
AttenComp
Atn+Scat
Comp
Atn +CDR
+ ScatComp
Atn +CDR
+ Scat+ PVC
Heart -77.60% 24.63% -11.76% -3.72% -2.11%
Lungs -62.78% 31.39% -0.96% 4.23% 6.45%
Liver -74.38% 29.22% -7.47% 2.71% 4.14%
20.6 ccsphere -78.88% -14.85% -29.81% -3.36% -1.97%
5.6 ccsphere -88.24% -51.53% -56.75% -21.55% -11.95%
% Error in total activity estimation: (true-estimate)/true x 100%
PVC using pGTM method
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
128 projection views Acquisition time: 40s / view
Heart Chamber Myocardium Large
SphereSmall
Sphere Background
Volume (ml) 59.7 115.3 17.5(r =1.61 cm)
5.7(r =1.11 cm) 9580
Activity(mCi) 0.562 0.471 0.136 0.044 8.15
Activity concentration(mCi/μl)
9.38 4.08 7.77 7.72 0.851
I-131 Physical Phantom
• Philips Precedence SPECT/CT system with HEGP collimator
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
I-131 Physical Phantom
(%) Heart Large sphere(r = 1.61 cm)
Small sphere(r = 1.11 cm)
AGS -15.21 -26.12 -32.72
ADS 4.75 -17.63 -25.77
ADS+Dwn+ -5.20 -21.10 -31.17
ADS+Dwn+PVC* -2.88 -15.49 -19.28
Percent errors of activity estimates for Anthropomorphic torso phantom
50 iterations 24
subsets/iteration
AGS ADS ADS + Dwn ADS+Dwn+PVE+DWN=model-based downscatter compensation*PVC=reconstruction-based PVC compensation
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Non-specific background uptake
Left putamenRight putamen
Left caudateRight caudate• GE Millennium VG/Hawkeye
(5/8” thick crystal)• LEHR Collimator
• 128 views/360°, 128*128 projection w/ 0.24 cm pixels • CT attenuation maps
• Manually defined VOIs using registered MR Images
• Activity concentrations:• Bkg: 110 kBq/ml
• Left Caudate: 212 kBq/ml• Left Putamen: 154 kBq/ml
• Right Caudate: 1770 kBq/ml• Left Putamen: 222 kBq/ml
Quantitative SPECT with I-123RSD Striatal Phantom
Y. Du†, B.M.W. Tsui, E.C. Frey, “Model-based compensation for quantitative 123I brain SPECT imaging,” Phys Med Biol, 2006, Vol. 51(5): 1269-1282, 2006.
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
OS-EM w/AttenuationScatter &
CDRF Compensation Post-Reconstruction
pGTM PVC
-3
-2
-1
0
1
2
3
Bac
kgro
und
Left
Cau
date
Left
Put
amen
Rig
ht C
auda
te
Rig
ht P
utam
en
% E
rror
in A
ctiv
ity E
stim
ate
Quantitative SPECT with I-123
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Simultaneous Dual-Isotope SPECT
• Requires– Compensation for cross-talk
• Scatter in body• Scatter (and Pb x-rays) in collimator-detector
– Publications on various pairs of radionuclides• Tc-99m/Tl-201 (rest/stress perfusion, brain tumors)• Tc-99m/I-123 (perfusion/innervation,
perfusion/receptor, …)• Tc-99m/In-111 (blood pool, antibody)• Tc-99m/F-188 (perfusion/viability)• In-111/I-131 (therapy)
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
MIN0 30 60
EX
2 mCi Tl
0
1
2
3
4
5
6
40 80 120 160 200Energy (keV)
Det
ecte
d C
ount
s
Tl
10 mCi Tc
Tc
Dual IsotopeSPECT
TcTl
Simultaneous Dual Isotope Rest/Stress MPS
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Tc-99m/Tl-201 Dual Isotope SPECTValidation in Dogs
Sep Acq
DISA-No Comp
DISA-w/ Comp
Comparison of CircumferentialProfiles for One Dog
Correlations betweenIshemic/Non-Ischemic
Activity Ratios for 25 Dogsfor Imaging and Ex Vivo
Measurements
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Quantitative Bremsstrahlung SPECT
• Continuous and broad energy spectrum
Tc-99m90Y
bremsstrahlung
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Quantitative Bremsstrahlung SPECTPhantom Study
Large sphere Medium sphere Small sphere
Error -7.0% -9.7% -10.2%
Error = (EstimatedActivity – TrueActivity) / TrueActivity ×100%
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
True Activity Distribution
MERw/o added noise
OS-EM w/ attenuation compensation alone
MERw/ added noise
Total activity: 80 mCi (2.96 GBq)Imaging time: 30 min10 iterations with 16 subsets per iterationTwo rightmost images were filtered using a Butterworth filter.
Quantitative Bremsstrahlung SPECTSimulation Study
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Error = (EstimatedActivity – TrueActivity) / TrueActivity * 100%
Lung Spleen Kidneys Liver Heart
MER 100-500keV 11.9% -5.9% -3.2% -1.6% -2.4%
attenuation compensation 324.4% 121.5% 400.9% 254.1% 133.8%
Errors in organ activity estimates at 200th iteration (16subsets per iteration) for data w/o added noise.
Quantitative Bremsstrahlung SPECTSimulation Study
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Quantitative Bremsstrahlung SPECTPatient Study
BremsstrahlungSPECT
BremsstrahlungQSPECT
MAAQSPECT
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Outline
• New Technologies– Instrumentation– Reconstruction
• New Applications
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Common Imaging AgentsRadiopharmaceutical Applications99mTc sestamibi/tetrofosmin myocardial perfusion, breast cancer, parathyroid201TlCl Myocardial perfusion, glioma99mTc MDP Bone metasteses/stress fractures99mTc DMSA Renal function99mTc MAG-3 Citrate Renal function99mTcO4- Salivary glands, thyroid, parathyroid99mTc Red blood cells Blood: bleeding, cardiac blood pool/ejection fraction99mTc microaggregated albumin (MAA) pulmonary perfusion
99mTc sulfur colloid reticuloendothelial system, lymphatic system, gastric emptying
Xe-133 pulmonary ventilation111In oxime (label stem cells, white blood cells, eggs) stem cell tracking, infection, gastric emptying111In pentetreotide (Octreoscan) somatostatin analog: neuroendocrine tumors
Ga-67 cintrate lymphoma, inflammation
I-123/I-131 MIBG neuroendocrine tumors (neuroblastoma/pheochromocytomas), cardiac ennervation
I 123/I 131Cl Th id
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
New Applications
• Myocardial Innervation Imaging• Neuroreceptor Imaging• Targeted Radionuclide Therapy Treatment
Planning– Radioimmunotherapy– Peptide recepter radionuclude therapy– Radioembolization/Microsphere brachytherapy
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Myocardial Innervation Imaging
• I-123 MIBG (Iobenguane/AdreView)– Noradrenaline analog– Used to image cardiac innervation
• Increased washout rate in heart failure• Mismatch between innervation and perfusion post-
MI
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Neuroreceptor Imaging
• I-123 Ioflupane (DaTscan)– Targets presynaptic dopamine
terminals– Applications in diagnosis of
Parkinson’s disease– Reduction or asymmetry in uptake
in striatum
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Targeted Radionuclide Therapy
• Treatment planning– Goal: estimate administered activity (AA) to provide
therapeutic effect and avoid toxicities– Method:
• Estimate dose to organs and tumors from known AA of a planning dose
• Use this with RTD to calculate therapeutic AA
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Targeted Radionuclide Therapy
• Treatment planning– Complications
• Organ anatomy and biokinetics are patient dependent• Radiation transport is patient dependent• Dose rate is time varying and spatially nonuniform
– Approach• Use quantitative imaging to measure biodistribution
of planning dose as a functionof time for each patient• Use radiation transport calculation to calculate dose
to organs, tumors or voxels• Apply radiobiology-based concepts from external
beam therapy (RBE, EUD, NTCP, TCP)
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
TRT Treatment Planning Flow Chart
Administer Planning
Dose
Measure Distribution over Time
Calculate Dose (Dose Rate) Distribution
Calculate Therapeutic
Activity
Administer Therapeutic
Dose
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
MIRD Formalism (Organ Dosimetry)
DoseT As S(OT Os )s
All SourceOrgan
S-valueTime-Integrated Activity
A A(t)dtt0
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Time-Integrated Activity and Residence Time
A : Time-integrated activity (MBq sec)A0 : Injected activity (MBq) : Residence Time (sec)
A A t dtt A0
Activ
ity A
(t)(M
Bq)
A
Time t (sec)
where A / A0
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
SPECT Residence Time Estimation
ResidenceTime
CTSPECT
Proj.
Curve Fitting
SPECTActivity
Estimation
0 hr
4 hr
24 hr
72 hr
144 hr0
0.02
0.04
0.06
0.08
0.1
0.12
0 50 100 150Time (hours)
0
A tA
organ
0
0.02
0.04
0.06
0.08
0.1
0.12
0 50 100 150Time (hours)
0
A tA
organ
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Voxel Dosimetry
Measure Activity
Distribution at Set of Times
Calculate Dose Rate in Each Voxel at Each
Time
Register Dose-Rate Images
Integrate Dose Rate over Time
Integrate Dose Rate over
Organ
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Summary of Imaging Requirements for TRT Treatment Planning
• Organ Dosimetry: organ activity estimates at each time point
• Voxel Dosimetry: requires registered 3D activity distribution estimates at each time
• Activity can be estimated from PET or SPECT images acquired at eac time
• Radionuclides imaged often have non-ideal imaging properties
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Radioimmunotherapy
• Non-Hodgkins Lymphoma – I-131 tositumomab (Bexxar)
• Dosing based on whole body dose from planar imaging
– In-111/Y-90 ibritumomab tiuxetan (Zevalin)• Weight-based dosing• Imaging no longer required• Approved as first-line consolidation therapy
– Bone marrow is dose-limiting– Clinical trials of myeolablative therapy regimens
• Lungs (Bexxar) or liver (Zevalin) dose limiting• Imaging required for treatment planning
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Peptide Receptor Radionuclide Therapy
• Smaller moleculues– More rapid uptake and clearance– Kidneys dose-limiting
• Lu-177 or Y-90 Dotatate– Binds to somatostatin receptors– Targets neuroendocrine tumors (NETs)– Therapy planning using In-111 SPECT or Ga-68 PET
labeled agent– Currently in clinical trials (extensive experience in
Europe)
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Radioembolization of Liver Cancer
• Y-90 labeled microspheres– Delivered through cather to hepatic artery– Hepatic artery supplies provides larger blood flow to
tumors than normal tissue
• Tc-99m Macroaggregated Albumin (MAA) used to evaluate potential for extrahepatic radiation
• Dosing scheme depends on product• Recent AAPM Dosimetry Recommendations
– Dezarn et al, Recommendations of the American Association of Physicists in Medicine on dosimetry, imaging, and quality assurance procedures for 90Y microsphere brachytherapy in the treatment of hepatic malignancies , Med Phys, vol 38(8), 2011, pp 4824-45.
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Radioembolization of Liver Cancer
• Commercial Products– TheraSphere (glass microspheres containing Y-90)
• Activity based on average dose to treated volume• Approved for treatment of hepatocelluar carcinoma (HCC)
– SIR-Spheres (resin microspheres containing Y-90)• Several dosing schemes (some include dosimetry using
partition model)• Lower specific activity -> larger number of spheres-> greater
embolic effect• Approved for metastatic colorectal cancer (mCRC)
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Evaluation of Therapeutic Response• Various agents are available that target cancers
– I-123 MIBG • Noradrenaline analog• NETs such as neuroblastoma, phaeochromocytoma
– In-111 Octreotide• somatostatin receptor• Various NETs
– Tc-99m MDP• Bone metastases and cancers
• Quantitative imaging may be useful to follow treatment response
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Evaluation of Therapeutic Response
AAPM 2012 Summer School on Medical Imaging using Ionizing Radiation
Conclusions
• New instrumentation technologies are on the horizon that have the potential to improve the quality of SPECT images– Collimation– Detectors
• Technology is available to produce quantitative images for a variety of applications
• New agents have been introduced• New applications cancer therapy include targeted
therapy treatment planning and monitoring treatment response