www.anton-paar.com convection temperature device ctd 600

14
www.anton-paar.com Convection Temperature Device CTD 600

Upload: roxana-grammer

Post on 30-Mar-2015

241 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: Www.anton-paar.com Convection Temperature Device CTD 600

www.anton-paar.com

Convection Temperature DeviceCTD 600

Page 2: Www.anton-paar.com Convection Temperature Device CTD 600

2

Gas Convection heating and cooling

Typical working range: ambient – 600°C

With LN2: -150°C – 600°C

Advantages: - Large temperature range

- Fast heating and cooling

- Enough working space for special accessories like solid torsion bar fixture etc.

Disadvantages: - LN2 needed for below ambient temperatures

Convection Temperature Device - CTD 600

Page 3: Www.anton-paar.com Convection Temperature Device CTD 600

3

Gas convection and radiation heating and cooling

-150°C – 600°C

Symmetrical design

Extremely small temperature gradients

Controller temperature equals sample temperature

No temperature overshoots

Easy sample access

DMTA with solid torsion bar fixture etc.

Convection Temperature Device - CTD 600

Page 4: Www.anton-paar.com Convection Temperature Device CTD 600

4

Solid torsion bar fixture (SRF)

Convection Temperature Device- CTD 600

UXF SER 2

Page 5: Www.anton-paar.com Convection Temperature Device CTD 600

5

Temperature

Standard: ambient to 600°CLow temperature option: -150°C to 300°C (600°C)Temperature sensor: Pt100 Class B

Supply pressure

Gas: max. 5barLiquid nitrogen: up to 1.2 bar

(0.8 to 1.0 recommended)

Consumption rate

Gas heating: 1.5 Nm³/h up to 400°C0.9 Nm³/h from 400 to 500°C0.5 Nm³/h from 500 to 600°C

Air shaft cooling: 1.2 Nm³/hLiquid nitrogen: 2-5 L/h

Convection Temperature Device- CTD 600

Page 6: Www.anton-paar.com Convection Temperature Device CTD 600

6

Chamber coolingwith liquid or air

Gasor cooledNitrogen gas

Heater

Pt100 (Heater)Pt100 (Heater)Pt100 (Sample)

The Sample is heated byconvection and radiation

Convection Temperature Device - CTD 600

Page 7: Www.anton-paar.com Convection Temperature Device CTD 600

7

PowerReady

EVU10Set value

ActualTemp.

Heater

Heater

Set value

ControlledPowerSet value

Set value

ActualTemp.

ActualTemp.

Convection Temperature Device- CTD 600

Page 8: Www.anton-paar.com Convection Temperature Device CTD 600

8

TC30 Temperature Controll Unit

-150.0

P

25.0°C

1....GAS2....LSVE3....Precision pressure controller4....Dewar

1

1

5....EVU106....CTD6007....TC308....water bath or air cooling

2

3

4

5 6

7

8

P 3Schematic diagram

Convection Temperature Device - CTD 600

Page 9: Www.anton-paar.com Convection Temperature Device CTD 600

9

1....Holder for lower plates 2....lower measuring plate 8mm 3....lower measuring plate 25mm 4....lower measuring plate 50mm 5....lower measuring plate with groove

1

23

45

Lower measuring plates

Convection Temperature Device - CTD 600

Page 10: Www.anton-paar.com Convection Temperature Device CTD 600

10

2

1....Holder for lower plates 2....Fixture for disposable plates 3....Disposable lower measuring plate 50 mm 4....Disposable lower measuring plate25 mm 5....Fixture for disposable dishes6....Disposable dishes

2

3

4

5

6

1

Lower disposable systems

Convection Temperature Device - CTD 600

Page 11: Www.anton-paar.com Convection Temperature Device CTD 600

11

Based on the CTD 600 Temperatures up to 200°C Measuring plate up to

25 mm diameter recommended

UV-Option for Photo DMTA

Convection Temperature Device - CTD 600

Page 12: Www.anton-paar.com Convection Temperature Device CTD 600

12

Light guide diameter 3 mm

Glass plate diameter: 40 mmheight: 4 mm

Coupling ringwith O-ring

Light guide diameter 3 mm

UV accessory CTD 600 – lower part for Photo DMTA

Convection Temperature Device - CTD 600

Page 13: Www.anton-paar.com Convection Temperature Device CTD 600

13

The CTD 600 has mayor advantage:• Extremely small temperature gradients• No temperature overshoots• PT100 signal reflects true sample temperature• No icing at low temperatures• Compact, robust design• Automatic adjustment of LN2 consumption as evaporation unit

actively controls continuous flow of liquid nitrogen• Easy to open, good sample access• Sample observation during measurement• Temperature-isolated jacket (hand-warm for safe use)• Actively cooled jacket according to EN61010-1:2001• Modular configuration (DMTA torsion/tension, Photo DMTA/UV,

reaction kinetics, SER extensional viscosity, disposable plates, etc.

Features and Benefits

Convection Temperature Device - CTD 600

Page 14: Www.anton-paar.com Convection Temperature Device CTD 600

14

Working range:-40°C - 200°CCalibration Temperatures:10°C and 70°CMeasurement Uncertainty:100mK

Working range:-150°C – 650°CCalibration Temperatures:Eight temperatures between-40°C and 500°CMeasurement Uncertainty:< 200°C: 100mK; > 200°C 200mK

2 Sensors available for different temperature ranges

Temperature Calibration - Software supported Automatic Temperature Calibration with Certified Temperature Sensor

Convection Temperature Device - CTD 600