www.anton-paar.com convection temperature device ctd 600
TRANSCRIPT
www.anton-paar.com
Convection Temperature DeviceCTD 600
2
Gas Convection heating and cooling
Typical working range: ambient – 600°C
With LN2: -150°C – 600°C
Advantages: - Large temperature range
- Fast heating and cooling
- Enough working space for special accessories like solid torsion bar fixture etc.
Disadvantages: - LN2 needed for below ambient temperatures
Convection Temperature Device - CTD 600
3
Gas convection and radiation heating and cooling
-150°C – 600°C
Symmetrical design
Extremely small temperature gradients
Controller temperature equals sample temperature
No temperature overshoots
Easy sample access
DMTA with solid torsion bar fixture etc.
Convection Temperature Device - CTD 600
4
Solid torsion bar fixture (SRF)
Convection Temperature Device- CTD 600
UXF SER 2
5
Temperature
Standard: ambient to 600°CLow temperature option: -150°C to 300°C (600°C)Temperature sensor: Pt100 Class B
Supply pressure
Gas: max. 5barLiquid nitrogen: up to 1.2 bar
(0.8 to 1.0 recommended)
Consumption rate
Gas heating: 1.5 Nm³/h up to 400°C0.9 Nm³/h from 400 to 500°C0.5 Nm³/h from 500 to 600°C
Air shaft cooling: 1.2 Nm³/hLiquid nitrogen: 2-5 L/h
Convection Temperature Device- CTD 600
6
Chamber coolingwith liquid or air
Gasor cooledNitrogen gas
Heater
Pt100 (Heater)Pt100 (Heater)Pt100 (Sample)
The Sample is heated byconvection and radiation
Convection Temperature Device - CTD 600
7
PowerReady
EVU10Set value
ActualTemp.
Heater
Heater
Set value
ControlledPowerSet value
Set value
ActualTemp.
ActualTemp.
Convection Temperature Device- CTD 600
8
TC30 Temperature Controll Unit
-150.0
P
25.0°C
1....GAS2....LSVE3....Precision pressure controller4....Dewar
1
1
5....EVU106....CTD6007....TC308....water bath or air cooling
2
3
4
5 6
7
8
P 3Schematic diagram
Convection Temperature Device - CTD 600
9
1....Holder for lower plates 2....lower measuring plate 8mm 3....lower measuring plate 25mm 4....lower measuring plate 50mm 5....lower measuring plate with groove
1
23
45
Lower measuring plates
Convection Temperature Device - CTD 600
10
2
1....Holder for lower plates 2....Fixture for disposable plates 3....Disposable lower measuring plate 50 mm 4....Disposable lower measuring plate25 mm 5....Fixture for disposable dishes6....Disposable dishes
2
3
4
5
6
1
Lower disposable systems
Convection Temperature Device - CTD 600
11
Based on the CTD 600 Temperatures up to 200°C Measuring plate up to
25 mm diameter recommended
UV-Option for Photo DMTA
Convection Temperature Device - CTD 600
12
Light guide diameter 3 mm
Glass plate diameter: 40 mmheight: 4 mm
Coupling ringwith O-ring
Light guide diameter 3 mm
UV accessory CTD 600 – lower part for Photo DMTA
Convection Temperature Device - CTD 600
13
The CTD 600 has mayor advantage:• Extremely small temperature gradients• No temperature overshoots• PT100 signal reflects true sample temperature• No icing at low temperatures• Compact, robust design• Automatic adjustment of LN2 consumption as evaporation unit
actively controls continuous flow of liquid nitrogen• Easy to open, good sample access• Sample observation during measurement• Temperature-isolated jacket (hand-warm for safe use)• Actively cooled jacket according to EN61010-1:2001• Modular configuration (DMTA torsion/tension, Photo DMTA/UV,
reaction kinetics, SER extensional viscosity, disposable plates, etc.
Features and Benefits
Convection Temperature Device - CTD 600
14
Working range:-40°C - 200°CCalibration Temperatures:10°C and 70°CMeasurement Uncertainty:100mK
Working range:-150°C – 650°CCalibration Temperatures:Eight temperatures between-40°C and 500°CMeasurement Uncertainty:< 200°C: 100mK; > 200°C 200mK
2 Sensors available for different temperature ranges
Temperature Calibration - Software supported Automatic Temperature Calibration with Certified Temperature Sensor
Convection Temperature Device - CTD 600