TRANSCRIPT

ALIGNMENT VERIFICATION

NOTE: Verification possibilities: femur first, tibia first, femur only, tibia only.

• Position the cutting block in the required position using standard (intramedullary) instrumentation. Do not use more pins than required for preliminary fixation. • Place the cutting block adapter in the saw blade slot, with the reflective marker spheres facing the camera.

Perform femur registration by acquiring:• Femoral head center• Femoral mechanical axis• Rotational references

NOTE: Pressing the Clip-On Remote Control for a long time repeats entire registration.

• Check the displayed information and if required, adjust the femoral cutting block’s position and/or orientation.

• Repeat registration.

• Place the pointer on the distal condyles to measure the distance to the cutting slot as an indicator for the resection height.

FEMUR SETUP

FEMORAL REGISTRATION

ALIGNMENT VERIFICATION FEMUR

RESECTION HEIGHT MEASUREMENT FEMUR

• Position the cutting block in the required position using standard (extramedullary) instrumentation. Do not use more pins than required for preliminary fixation.

• Place the cutting block adapter in the saw blade slot, with the reflective marker spheres facing the camera.

Perform tibia registration by acquiring:• Malleolar points• Tibial mechanical axis• AP direction

• Check the displayed information and if required, adjust the tibial cutting block’s position and/or orientation.

• Repeat registration.

• Use the pointer to measure the distance from the tibial resection plane to the cutting slot as an indicator for the resection height.

MANUFACTURER INFORMATION:

Brainlab AGKapellenstr. 12, 85622 Feldkirchen, Germany

Europe, Africa, Asia, Australia: +49 89 99 15 68 44USA and Canada: +1 800 597 5911Japan: +81 3 3769 6900Latin America: +55 11 33 55 33 70France: +33-800-67-60-30

E-mail: support@brainlab.com

COPYRIGHT:

This guide contains proprietary information protected by copyright. No part of this guide may be reproduced or translated without the express written permission of Brainlab.

LIABILITY:

This guide is subject to change without notice and does not represent a commitment on the part of Brainlab.

For further information, please refer to the “Limitations of Liability” section in the Brainlab Standard Terms and Conditions of Sale.

OR_FL_EN_Flyer Knee Express_1011_Rev1

Document Revision: 1.0

ALIGNMENT VERIFICATION

copyright 2012, Brainlab AG Germany. All rights reserved.

TIBIA SETUP

TIBIA REGISTRATION

ALIGNMENT VERIFICATION TIBIA