gb suction system metasys excom z-eco, z-eco...
TRANSCRIPT
D� GBPage Practice personnel / Technician
Contents Key to symbols
The footnote found on each page defines the user group for which that particular information is aimed.
1. Contents:
Page
1. Contents 2
2. Keytosymbols 2
3. Generalinformation 3
4. Application 4
5. Construction 4
6. Keytotypeplates 4
7. Technicaldata 5
8. Functionaldescription 5
9. Installationofthesystem 6
10.Assemblyofthesystem 6
11. Guidelinesforinstallation 7
12.Hoseconnections 8
13.Electricalconnections 8
14.Maintenance,cleaninganddisinfection 9
2. Key to symbols:
Warning to ignore the following instructions could lead to personal injury, disruption of operation, or damage to the unit!
Therefore the servicing personnel or technical engineers are made aware of the following facts.
!
GB Page �Practice personnel / Technician
3. General information:
METASYScanonlyguaranteesafety,reliabilityandperformanceoftheunit,whenthefollowinginstructionsareadheredto:
qInstallation, changes or repairs are carried out byauthorized service personnel in compliance withStandardEN60601-1(InternationalStandardformedi-calelectricalapparatus,especiallypart1:Generalrulesforsafety).
qTheelectricalinstallationcomplieswiththeIEC(Inter-nationalCommissionforElectricalEngineering)regula-tions.
qTheapparatusisassembled,operatedandmaintainedaccordingtotheinstructionsprovided.
qOnly original parts are used for repairs or replace-ments.
qAllspecificationsofthemanufacturersofthehandlingunits,towhichthedry-suctionsystemZ-ECO,Z-ECOLightisattached,aretobeconsidered.
Following the commissioning of the amalgam separator, the installation report, contained with-in the apparatus documents must be filled in and sent to METASYS to confirm the time period of the guarantee.
q Everyinspection,serviceinterventionandexchangeofcollectioncontainershouldberecordedintheappara-tuslogbook.
qBylaw,everyamalgamwastedisposalcertificateshouldberetained.
qWhenrequestedbyanauthorizedengineer,METASYSagreestomakealldocumentsavailablefortheuseoftechnicallyqualifiedservicepersonnel.
qMETASYSacceptsnoresponsibilityfordamagescausedduetoexternalfactors,suchaswronginstallation,useoffalseinformation,improperuseoftheapparatusorunauthorizedtechnicalintervention.
qDuplicationanddistributionofthisdocumentmayonlybeundertakenafterapriorpermissionisobtainedfromMETASYS.
qWhenthecompletecentralsingleunitsuctionsystemEXCOMZ-ECO,Z-ECOLighthasbeendismantledattheendofitsworkinglife,itmustbereturnedtothemanu-facturerfordisposal.
!
General instructions
1
2
3
4.1
4.3
4.2
4.4
4.6
ZE - 12345 456
D� GBPage Practice personnel / Technician
Application, construction Key to type plates
4.5
4. Application:
TheMETASYSEXCOMZ-ECO,Z-ECOLightisanindividualcentralsingleunitsuctionsystemwithintegratedair-water-separationandplaceselectionvalveinconnectionwithadrysuctionmachine.
5. Construction:
Inordertoensureafastandefficientpartsexchangeinthecaseofserviceandmaintenance,theMETASYSEXCOMZ-ECO,Z-ECOLighthasamulti-modularconstruction.
1 Modul 1consistsofMETASYSEXCOMZ-ECO,Z-ECOLightair-water-separation,theplaceselectionvalveandthesuctionstreaminletconnection.
2 Modul �consistsofthedrysuctionmachine–sidechannelvacuumpump,condensateseparatorwithairby-passvalve,mountingpanelwiththeconnectionofthewateroutletandtheelectricalcontrolmodule(electricalconnec-tions,transformer,mainswitchandelectronicboard).
6. Explanation of the manufacturer’s plate.
3 Seediagram.
Themanufacturer’splateistobefoundonthemountingpanel.
4.1 Equipmentdesignation
4.2 Connectiondata
4.3 Serialnumber
4.4 Addressofthemanufacturer
4.5 Compliance,asperÜZVO
4.6 CEmark
4
5
470
mm
460mm
470
mm
385mm
5.2
5.1 5.3
5.6 5.5
5.4
GB Page �Practice personnel / Technician
Technical dataFunctional description
7. Technical data: Z-ECO Z-ECO Light
Powersupply: 230VAC 230VAC
Frequency: 50/60Hz 50/60Hz
Currentconsumption: 3,6/3,7A 3,3/3,6A
Powerconsumption: 0,37/0,44kW 0,55/0,63kW
Underpressurerange:180mbar 120mbar
Max.flowrate(air): 440l/min 400l/min
Max.flowrate(water):4,5l/min 4,5l/min
Permissibleambient operation5-40°C operation5-40°Ctemperature: storage0-55°C storage0-55°C
Averagenoiselevel: 65dB 60dB
Dryweight: 25kg 15kg
Dimensions(BxDxH): 460x385x470mm 460x385x470mm
EXCOM Z-ECO, Z-ECO Light must be stored in dry surroundings at normal humidity. It is recommended that the unit be stored in a closed wrapping with a drying agent if the relative humidity is more than 80%.
4 Seediagram
Totaldimensions(BxDxH):460x385x470mm
8. Description of the functioning:
5 Seepicture
Whenthesuctionhoseisliftedoffthemanifold,thedrysuction machine 5.1 starts and the place selectionvalve 5.4 opens.Thesuctionflowisintroduced 5.3 totheEXCOMZ-ECO,Z-ECOLightair-water-separation.Thesolidsandliquidareseparatedintheseparationcham-berthroughtheuseofacyclone.Theairgoesthroughthecondensateseparatorto 5.2 thedrysuctionmachine.Ifthesuction-hoseisreplaced,thedrysuctionmachinestops,theplaceselectionvalveclosesandthewaterflowsoutoftheEXCOMZ-ECO,Z-ECOLightair-water-separationtothewateroutlet 5.5 Ifmorethan3,5lwaterisintroducedtotheEXCOMZ-ECO,Z-ECOLightatatime,theinsidefloatergivesacontacttotheelectronicboard 5.6 andtheplaceselectionvalveclosesfor30seconds,thewaterwillrinseoutoftheEXCOMZ-ECO,Z-ECOLightair-water-separationatthewateroutletandthatmeansthesuctionstreamisintermittent!After30secondstheplaceselectionvalveisopenedandthesuctionstreamisopened!
3
1
2
D� GBPage Practice personnel / Technician
Installation and assembly
The installation of the METASYS EXCOM Z-ECO, Z-ECO Light must be carried out by authorised technicians only. All instructions and regulations made by the manufac-turer concerning the assembly of the system have to be followed.
9. Installation of the system:
Variations:
1 same level as dental unit
2 one level below dental unit
qThe movements of the dental staff and the patientshouldnotbehinderedbythemachineorbyitshoseandelectricconnections.Theconnectinghosefromthedentalunitshouldbeasshortaspossibleinordertopreventunnecessarypre-sedimentation.
The equipment may only be installed and operated in a dry, well ventilated and dust-free room. If there is insufficient air flow, a suction device or ventilator of the correct dimensions must be installed. The ideal ambient temperature lies between +10° C and +30° C. The motor fan ensures that the power unit is force cooled effec-tively. However, for this to happen, the air must be able to flow in and out unobstructed. If this is not the case, a separate ventilation system must be installed. The EXCOM Z-ECO, Z-ECO Light must be installed in such a way that it is easily accessi-ble for operation and maintenance and for access to the rating plate. The unit must stand on a level floor which is sufficiently robust.
The waste water must be able to flow freely, as with a back-wash the function of the central single unit suction system is not guaranteed. The waste water pipe must have a negative gradient of minimum 2%.
3 The connection to the waste water piping system must be under the waste water outlet of the EXCOM Z-ECO, Z-ECO Light.
10. Assembly of the system:
q Thesystemhastobeinstalledatgroundlevel(solidandeven)inordertopreventvibrations.Cablingandtubesshouldbeinstalledaccordingtoindividualinstallationsituations.
3
4
5
6
GB Page �Practice personnel / Technician
Guidelines for installation
11. Guidelines for installation:
3 Pre-filter:
Inthesuctionlineofadentalunittheremustbeapre-filterwithameshwidthofmax2mminstalled.Thecoarsefilterfittedintheoutletofthespittoonbowlshouldhaveamaximum3mmmeshwidthandmustnotberemovedwhilethecentralsuctionsystemisinoperation.
4 Hose connections:
qThewastewatermustbeabletoflowfreely,aswithaback-washthefunctionofthecentralsingleunitsuctionsystemisnotguaranteed.
qTherinsingprocessofthespittoonbowlshouldbelimitedbymeansofatimeroraspecialbuttontoamaximumof30secondsatamaximumwaterflowrateof3litres/minuteinordertofacilitatearinseoutofthewastewater.
qTheconnectionsfromthepipestothecentralsuctionsystemEXCOMZ-ECO,Z-ECOLightmustbewithflex-iblehosesandmustbeasshortaspossible.Toreducethelossofsuctionflowpowerit isrecommendedtousepipeswitha25to40mmdiameterandtoavoid90°turns(insteaduse2x45°).
5 Werecommendthatthewaterofthespittoonbowlbedirectlyguidedintothesuctionpipeusingthespittoonvalve.ThishastheadvantagethatonlyonehoseneedstobelaidfromthedentalunittoEXCOMZ-ECO,Z-ECOLight.Theinflowvalveisrequiredtocurbsuctionnoisescomingfromthespittoonbowl.
6 Secondary air valve:
Thebuiltinairby-passvalveprotectsthesuctionmachinefromoverheatingandmaintainsuniformsuction.Therecommendedvacuumsuitablefortheunitisapprox.300l/minor160mbarmeasuredatthebigsuctiontip(canula).Thissuctionflowvol-umemustbesetaccordinglyandmonitoredbyatechniciancalibratingtheairby-passvalve.
2
1
1.3
1.4
1.2
3
1.1
1M1 -1B1
-1Y1-1S2-AO
D� GBPage Practice personnel / Technician
Hose connections Electrical connections
12. Hose connections:
1 Seediagram
1.2 Connectionsuctionflowinlet,connectiontothedentalunitIDM25–32mm
1.3 ConnectionairoutletfromthedrysuctionmachineIDM32mm
1.4 ConnectionwastewateroutletwithconnectorIDM25mm(standardscopeofdelivery)
The distance between the EXCOM Z-ECO, Z-ECO Light and the dental unit must not exceed 25 metres.
All hose connections must be secured by using hose-clips!
1.1 Due to hygienic reasons and to avoid heat build-up it is recommended to guide the air exhaust out-doors and to fit a bio-filter in the exhaust pipe. (Item No.: 40060001)
2 Seediagram
Main switch:Please ensure that the amalgam separator is switched off at least once per working day! The connection to the electric current 230 V AC must be in circuit with the dental practice main switch.
13. Electrical connections:
Cable 3 x 1.5 mm2 for power supply 230 V.Cable 2 x 0.75 mm2 for starting signal of the suction system (suction contact from the dental unit).
3 Explanation:
-AO controlbox-1B1 reedrelay-1M1 motorsuctionsystem-1S2 mainswitch-1Y1 placeselectionvalve
4
5
W3
W2
F3 K1 W1
F1
F2
S1
V1
A1
V2
T1
W4W5
F2 / T
10 A
F1 / T
10 A
GB Page �Practice personnel / TechnicianTechnician
Electrical connections
4 electricalconnectiondiagram
5 electricalconnectioncontrolbox
Legend:
A1 electricalmainboard
F1 fuse1(T10A)
F2 fuse2(10A)
F3 fuse3(T2A)
K1 motorcontactor
S1 mainswitch
T1 transformer
V1 varistorK275
V2 varistorK35
W1 motorcable
W2 powersupply
W3 suckingcontact
W4 reedrelaywaterprotection
W5 valvewaterprotection
2
1
3
D10 GBPage Practice personnel / Technician
Care, cleaning and disinfection with Green&Clean M2
14. Care, cleaning and disinfection:
1 Seediagram
Twicedaily,afterthewaterhasbeendrainedoff,theMETASYSGreen&Clean M2,suctionlinedisfectantandcleaner.
Ideally,thedisinfectionshouldbedonepriortoarelativelylongperiodofnon-useofthetreatmentunit(afternoonbreak,off-dayorvacations).
2 Seediagram
In any case, the rinsing basin must be rinsed withGreen&Clean M2,suction linedisfectantandcleaner,twicedaily.
qCleaning the preliminary filter:
Thepreliminaryfiltermustbeemptiedatleastonceeveryweek,andcleaned.Dependingonyourworkingpatterns,youmayevencarryoutthiscleaningdaily.
3 Seediagram
Theresiduesemptiedfromthefilterbox,whichcontainamalgams,aretobecollectedinthesuitableMETASYScontainerEcoCenter,item:40150001,andbedisposedatDental ECO Service GmbH.
GB Page 11Practice personnel / Technician
Annual inspection
15. The annual inspection:
qAccordingtotheManufacturesGuidelines,theunithastobetestedandcleanedatleastonceeveryyear.
qForthisthecentralsingleunitsuctionsystemEXCOMZ-ECO,Z-ECOLighthastoundergoafunctionalcheckbyauthorisedtechnicians!
Ifthesuctionsystemcontinuouslystopswhileinoperation,anddoesnotturnonautomaticallyanymore,evenwhenthesuctionhoseisliftedoff,theprobablecauseisthatthefloaterandtheswitchhavebeenshort-circuitedduetocon-tamination!
Iftheunitsucksbackfromthewastewaterdrainduetoreflux,themembraneofthenon-returnvalveispollutedordamaged.Inthiscasecleanthemembraneorifnecessaryexchangethenon-returnvalve.
Data from the annual inspection must be entered into the equipment log-book!
AustriaMETASYS
Medizintechnik GmbHFlorianistraße 3
A-6063 Rum bei Innsbruck
1 +43/(0)512 /20 54 20-0
5 +43/(0)512/20 54 20-7
GermanyMETASYS
Medizintechnik GmbHAhornstraße 19
D-85614 Kirchseeon
1 +49/(0)2236 / 37 42 42
5 +49/(0)2236 / 37 42 41
Ihr METASYS Berater:/Your METASYS agent:
email: [email protected]
www.metasys.com
FranceMETASYS France S.a.r.l.
9, bd E. MicheletF-69008 Lyon
1 +33/(0)4 / 37 90 22 15
5 +33/(0)4 / 37 90 22 47e-mail: [email protected]://www.metasys.fr
Italy
Dental eCO ServiCe italia S.r.l.Florianistraße 3
A-6063 Rum bei Innsbruck
1 +39/045 / 981 4477
5 +39/045 / 981 4475e-mail: [email protected]
EBW
Sau
gsys
tem
Z-E
CO, Z
-ECO
Lig
ht, 0
1.04
.20
10
70
1003
16
G
B
ZK
-55.
218
/02
S
ubje
ct to
tech
nica
l cha
nges
, pri
ntin
g an
d se
ttin
g er
rors
!