mariana petris, cbm collaboration meeting, october 13-18, 2008, dubna mariana petris, nipne -...

22
Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass Resistive Plate Counter Prototypes

Upload: steven-stanley

Post on 17-Jan-2016

219 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Mariana Petris, NIPNE - Bucharest

CBM

In-Beam Test Results of the Pestov

Glass Resistive Plate Counter Prototypes

Page 2: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

• Motivation

• Standard Read-out Pestov Glass RPC Prototype

- Construction Details

- 60Co Source Test

- In-Beam tests

• Differential Read-out Pestov Glass RPC Prototype

- Construction Details

- In-Beam tests

• Conclusions and Outlook

Outline

Page 3: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

RPC Counting Rate Performance

• MSMGRPC based on commercial float glass (glass ~ 1012 cm) keeps

the performances up to ~1 kHz/cm2 ; this type of RPC could be a

solution for a major part of the CBM - TOF subdetector.

• CBM - TOF subdetector at small polar angles – high counting rate

environment (up to 20 kHz/cm2).

• Solutions:

– Electrodes with lower resistivity

– Smaller and many gaps

• Our prototypes were built using Pestov glass with ~1010 cm.

Page 4: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Volume Resistivity Pestov Glass

1.00E+09

1.00E+10

measuring point

ρ [

Ω c

m]

ρ [Ω cm]

Sample 1 Sample 2

1

2

3

glass sample

Volume Resistivity Measurements of Pestov Glass

Experimental SetupMeasurement Setup:

Device – Keithly Picoampermeter 487;

Measurement Duration – 100 ps;

Applied voltage – 100 V;

Condition – 22 0C, 42 % LF

Power supply

pA-Meter

Measurement electrode

Measurement

electrode

Guard-ring

Conductive rubber sample

Counter electrode

Point 1 Point 2 Point 3 Point 4

Page 5: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Electrode sequence:

- 2 mm aluminum – cathode

- 0.3 mm nylon fishing line

- 2 mm glass plate (supporting the spacers)

- 0.3 mm nylon fishing line

- 2 mm glass plate

- 0.5 mm double faced strip readout plate

symmetry plane

Readout electrode:

- support:0.5mm thick pcb plate

- 14 copper coated strips on each side

- pitch: 2.54 mm, 1.10 mm strip width, 1.44 mm gap width

Sizes:

glass electrodes: 40.6 x 300 mm2

readout plate: 0.5 x 40.6 x 360 mm3

housing: 40 x 80 x 330 mm3 Al box

Construction details of the standard – readout Pestov glass RPC prototype

Page 6: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

60Co Source Test85% C2F4H2 +10% SF6 + 5% izo-C4H10

≤ 60 ps

Page 7: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

In-Beam Tests@ ELBE (Electron Linac with high Brilliance and low Emittance)

Forschungszentrum Dresden-Rossendorf

Experimental setup:

- electron beam, 30 MeV, scattered @ 450 by a 18 m Al foil;

- plastic scintillators S5(XP2972), S12(XP2020), S34(XP2020),

(2 x 2 cm2) used for active collimation;

- signal amplification: FEE1 developed for FOPI at GSI.

- digital converters: CAEN ADC V965, CAEN TDC V1290N

- DAQ – MBS (Multi-Branch System – GSI – Darmstadt)

- information recorded for 4 central stripsV. Ammosov et al., Nucl. Instr. And Meth. A 576 (2007), 331

Page 8: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Time resolution

Page 9: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Time resolutions of the strips 6 and 7

Strip no. 6 Strip no. 7

Cond. QmaxCond. Qmax

Page 10: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Time resolution of the strips 8 and 9

Strip no. 8 Strip no. 9

Cond. QmaxCond. Qmax

Page 11: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Position dependence

Strip no. 8

No significant position dependence for strip # 6,7,9.

Page 12: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Time resolutions of the 4 measured strips

These results have been obtained

• No correction for the intrinsic electronic time resolution

• These results have been obtained with a uniform exposure of the counter active area

at a particle flux density of ~1kHz/cm2.

Page 13: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Rate dependence

No saturation effect of the number of events recorded by the Pestov glass RPC as a function of counting rate

Page 14: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Differential Strip – Readout Pestov Glass RPC Prototype

Page 15: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

In-Beam Tests@ ELBE (Electron Linac with high Brilliance and low Emittance)

Forschungszentrum Dresden-RossendorfExperimental set-up:

- electron beam, 28 MeV, scattered @ 450 by a 18 μm Al foil;

- plastic scintillators S5(XP2972), S12(XP2020), S34(XP2020),

(2 x 2 cm2) used for active collimation;

- signal amplification: differential readout based on NINO chip.

- digital converters: CAEN TDC V1290N

- DAQ – MBS (Multi-Branch System – GSI – Darmstadt)

- information recorded for 2 central strips

Page 16: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

100 nA ≡ 975 Hz/cm2 250 nA ≡ 1872 Hz/cm2

Time resolution as a function of counting rate

Page 17: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

500 nA ≡ 3100 Hz/cm2

Time resolution as a function of counting rate

750 nA ≡ 4430 Hz/cm2

Page 18: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Time resolution as a function of counting rate

1500 nA ≡ 8440 Hz/cm2 3000 nA ≈ 16400 Hz/cm2

Page 19: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Time resolution as a function of counting rate

No deterioration of the time resolution as a function of counting rate up

to ~ 16 kHz/cm2

Page 20: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Conclusions and Outlook•Three RPC prototypes based on PESTOV glass with a resistivity of ~1010 Ωcm were tested in-beam in the first part of this year @ the ELBE facility at Forschungszentrum Dresden-Rossendorf.• Standard readout Pestov glass RPC prototype.

• The 60Co source tests show that the prototype has a good time resolution;

•Its performances in terms of time resolution -52 ps, were confirmed by the in-beam tests, fulfilling the requirement for the CBM-TOF subdetector. This time resolution was obtained in condition of a uniform exposure of the counter active area at a particle flux density of ~1kHz/cm2.

• Differential strip readout RPC prototype was designed and built by us. •The in-beam test showed the good performance of the counter with differential readout based on NINO chip in terms of time resolution – 69 ps.• No significant deterioration of the time resolution as a function of counting rate was observed up to ~ 16 kHz/cm2.

• Differential pad readout RPC prototype was also tested in-beam @ ELBE facility. •The data are going to be analyzed .

Page 21: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Participants

NIPNE – BucharestD. BartosG. CaragheorgheopolM. PetrisM. PetroviciV. Simion

FZD – Dresden RossendorfD. StachF. DohrmannR. KotteL. NaumannJ. Wüstenfeld

GSI – Darmstadt

K. Hildenbrand

CERN - Geneva

C. J. Williams

Page 22: Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna Mariana Petris, NIPNE - Bucharest C B M In-Beam Test Results of the Pestov Glass

Mariana Petris, CBM Collaboration Meeting, October 13-18, 2008, Dubna

Diferential pad readout RPC prototype

Better granularity:

Cell size: 0.5 x 2 cm2