tri gas drawer type incubator - astec bio, usasteptoemd.com/products/incubators/in9.pdf · astec...

Post on 14-Nov-2020

9 Views

Category:

Documents

0 Downloads

Preview:

Click to see full reader

TRANSCRIPT

Drawer Type Incubator

%CO2

INJECT .%O2

INJECT .℃TEMP

HEAT .

OVER TEMP. CO2 ALARM O2 ALARM

TEMP. SET CO2 SET N2 / O2 SET

OTEMP. SET O2 SET

DOWN UPESCAPESILENCE SET

Drawer Type Incubator

%CO2

INJECT .%O2

INJECT .℃TEMP

HEAT .

OVER TEMP. CO2 ALARM O2 ALARM

TEMP. SET CO2 SET N2 / O2 SET

OTEMP. SET O2 SET

DOWN UPESCAPESILENCE SET

Drawer Type Incubator

%CO2

INJECT . %O2

INJECT .℃TEMP

HEAT .

OVER TEMP. CO2 ALARM O2 ALARM

TEMP. SET CO2 SET N2 / O2 SET

OTEMP. SETO2 SET

DOWN UPESCAPESILENCE SET

Drawer Type Incubator

%CO2

INJECT . %O2

INJECT .

℃TEMPHEAT .

OVER TEMP. CO2 ALARM O2 ALARM

TEMP. SET CO2 SET N2 / O2 SETOTEMP. SET

O2 SET

DOWN UP ESCAPESILENCE SET

Drawer Type Incubator

%CO2

INJECT .%O2

INJECT .℃TEMP

HEAT .

OVER TEMP. CO2 ALARM O2 ALARM

TEMP. SET CO2 SET N2 / O2 SET

OTEMP. SET O2 SET

DOWN UPESCAPESILENCE SET

URL http://www.astec-bio.com/global E-mail info@astec-bio.com

HEADQUARTERS/FACTORY4-6-15 Minamizato Shime Kasuya Fukuoka Japan 811-2207TEL +81-92-935-5585 FAX +81-92-936-6613

Accurate environmental control Exclusive and patented design

Flexible usage

Enhanced safety functions

Independently controlled chambers

Dry and wet cultureAdvanced cleanability

AD SERIES

Tri GasDrawer Type Incubator

ASTEC Co.,Ltd

Product SpecificationsMODEL Drawer Type Incubator (AD-3100)

Overall Dimensions W564 x D561 x H591* (mm) / W22.2 x D22.1 x H23.3* (in)

Drawer Chamber

Drawer Chamber Capacity4 x 4-Well dish / 6 x 60mm petri dish / 12 x 35mm petri dish775ml

Drawer Chamber Weight 2.8kg (6.2lbs) / per chamber

Number of Drawers 4 drawers

Temperature Control

Control System Digital PID control

Range 7℃ + ambient to 40℃

Accuracy ±0.2℃

Uniformity ±0.4℃

CO2 Control

Sensor Type Infra Red

Range 0% - 10.0%Uniformity ±0.3%

O2 Control

Sensor Type Galvanic Battery

Range 4.0% - 10.0%

Uniformity ±0.5%

Others

Humidity in Chamber 90%RH

Humidity method Natural Evaporation

Gas Used CO2 / N2

Gas Flow CO2: 40ml / min / N2: 120ml / min

Alarms User Programmable on Temperature, CO2, O2

Alarm Output DC12V 0.3A (Optional)

Safety Feature Temperature Cut-off (Software Controlled)Analog Output Temperature/CO2/O2 (Optional)

Weight 95kg / 210lbs

Power 110V, 117V, 220V, 230V 50/60Hz

*Without adjustorsNOTE: Specifications are verified at 28℃ ambient temperature.

Usable Dimensions W180 x D217 x H20 (mm) / W7.1 x D8.6 x H0.8 (in)