standard q rota/adeno ag - lab mark – klinická ... · standard q rota/adeno ag test is a rapid,...

2
EXPLANATION AND SUMMARY [Introduction] Rotavirus is the commonest virus cause acute diarrhea. It is most common in infants and young children. Children who get rotavirus infected may have severe watery diarrhea, often with vomiting, fever, and abdominal pain. However, older children and adults also can get sick from rotavirus. Adenovirus is the second most common cause of diarrhea following rotavirus. Adenoviruses cause various clinical diseases such as Common cold, Sore throat, Bronchitis, etc. Adenoviruses mainly implicated respiratory, ocular and the gastrointestinal systems of humans. Particularly, Enteric adenovirus types 40 and 41 cause gastroenteritis. Symptoms may include loss of appetite and dehydration (loss of body fluids), which can be especially dangerous for infants and young children. STANDARD Q Rota/Adeno Ag Test provides significantly fast, easy and accurate system to detect the specific antigens to rotavirus and adenovirus in human fecal specimen. It is essential for the reliable clinical diagnosis to identify Rotavirus/Adenovirus infection and enables supportive definite diagnosis of them. [Intended use] STANDARD Q Rota/Adeno Ag Test is a rapid, qualitative immunoassay to detect Rotavirus Ag and Adenovirus Ag in human fecal specimen. The test is for in vitro diagnostic use and is intended as an aid to early diagnosis of Rotavirus and Adenovirus infection. This is intended for professional use, only for an initial screening test. [Test principle] STANDARD Q Rota/Adeno Ag Test has two test lines (“R” for Rotavirus and “A” for Adenovirus) and a control line(“C”). Line “R” is coated with monoclonal anti-Rotavirus, “A” is coated with monoclonal anti-Adenovirus and “C” is coated with monoclonal anti-Rabbit IgG in the membrane of test device. For the test, the sample is added to the sample well and antigens in the sample interacts with monoclonal anti-Rotavirus-gold/monoclonal anti-Adenovirus-gold on the conjugation pad. Antigens in the sample and the gold conjugated antibodies make complexes and the complexes move along membrane to the test and control line chromatographically to react with the antibodies coated on the surface of membrane. If Rotavirus and Adenovirus are present in sample, visible lines in the test line would appear in the result window. The control line has to appear every time when the test has performed correctly. STANDARD Q Rota/Adeno Ag Test can also perform by using appropriate analyzer (STANDARD F Analyzers) manufactured by SD BIOSENSOR. The analyzers use the RGB optic sensor to measure reflected light from the membrane of the test device. The intensity of the reflected light is scanned and converted into an electric signal which is proportional to the intensity of reflected light produced on the membrane. STANDARD F analyzers can calculate the concentration of the analyte in the clinical specimen based on a pre-programmed calibration and display the test result on the screen. [Kit contents] Test device Sterile swab Extraction buffer tube Filter cap Instructions for use [Materials required but not provided] STANDARD F Analyzer Timer KIT STORAGE AND STABILITY Store the kit at 2-40°C /36-104°F, out of direct sunlight. Kit materials are stable until the expiration date printed on the outer box. Do not freeze. WARNINGS AND PRECAUTIONS 1. Do not re-use the test kit. 2. Do not use the test kit if the pouch is damaged or the seal is broken. 3. Do not use the extraction buffer of another lot. 4. Do not smoke, drink or eat while handling specimen. 5. Wear personal protective equipment, such as gloves and lab coats when handling kit reagents. Wash hands thoroughly when afterwards. 6. Clean up spills thoroughly using an appropriate disinfectant. 7. Handle all specimens as if they contain infectious agents. 8. Observe established precautions against microbiological hazards throughout testing procedures. 9. Dispose of all specimens and materials used to perform the test as bio-hazard waste. Laboratory chemical and bio-hazard wastes must be handled and discarded in accordance with all local, state, and national regulations. 10. Silica gel in foil pouch is to absorb moisture and keep humidity from affecting products. If the moisture indicating silica gel beads change from yellow to green, the pouch should be discarded. 11. Immediately use the test device after taking out of foil pouch. 12. Do not write on the bar code or damage the bar code of the test device. SPECIMEN COLLECTION AND PREPARATION [Feces] 1. Fecal specimen should be collected in a clean and dry container and collected at any time of the day may be used. 2. Urine should be excluded from the fecal specimen. 3. Fresh fecal specimen should be used in this test. Do not use any transport media to store and transport the fecal specimen. 4. Fecal specimen may be stored at room temperature (15-30°C/59-86°F) for up to 1 day or at 2-8°C/ 36-46°F for up to 2 days prior to testing. TEST PROCEDURE [Preparation] 1. Allow test device and collected sample to room temperature (15-30°C / 59-86°F ) prior to testing. 2. Carefully read instructions for using the STANDARD Q Rota/Adeno Ag Test. 3. Check the expiry date at the back of the foil pouch. Use another lot, if expiry date has passed. / LOT No. / MFG DATE / EXP DATE / REF No. / Item ML25RT7MLR1 Issue date : 2017.10 xxxxxxxxxxxx xxxx.xx.xx xxxx.xx.xx STANDARD Q Rota/Adeno Ag xxxxxxxxxxxx 4. Open the foil pouch, and check the test device and the silica gel pack in the foil pouch. C R A Rota/Adeno <Foil pouch> Yellow Yellow: Valid Green: Invalid Green <Silica gel> STANDARD Q Rapid Test Global standard test for rapid diagnostic test Result window Sample well <Test device> [Collection of sample] 1. Poke the 4 different areas of the fecal sample using the sterile swab to collect a portion of feces (about 40-70mg). In case of liquid sample, soak the sterile swab completely in the liquid fecal sample. Too little Good Too much NOTE • The amount of fecal swab may affect the results. It is required to follow the swab amount of feces as shown in the picture above. Excessive fecal amount may induce a false positive result and slow migration. 2. Take off the top of an Extraction buffer tube and insert the swab into the tube. 3. Swirl the swab at least ten times into the Extraction buffer tube to dissolve the sample. x10 4. Remove the swab and dispose of the used swab in accordance with your biohazard waste disposal protocol. 5. Tightly screw the filter cap onto the tube. CAUTION Do not use the extraction buffer tube without the filter cap. It affects the result value. [Analysis of sample] Visual inspection 1. Take the test device out of the foil pouch and place it on a flat and dry surface. 2. Apply 4 drops of mixed sample to the sample well in the test device holding the prepared extraction buffer tube upside down. C R A Rota/Adeno 4 drops 3. Read the test result after 20 minutes. 20 Incubate For 20 mins [Analysis of sample] Using a ‘STANDARD TEST’ mode - Applied STANDARD F100, F200 and F2400 analyzer 1. Prepare a STANDARD F Analyzer and select the ‘Standard Test’ mode according to the analyzer’s manual. In case of STANDARD F2400 analyzer, go to the ‘Workplace’ in the main screen. And select the ‘Run Test‘. 2. In case of STANDARD F200 and F2400 analyzer, input patient ID and/or operator ID on the analyzer. 3. Take the test device out of the foil pouch. 4. Insert the test device to the test slot of the analyzer. When inserting the test device to the analyzer, the analyzer will read the barcode data, and check the test device is valid. Quality approved by SD BIOSENSOR / For In vitro use only Rota/Adeno Ag STANDARD TM Q Rota/Adeno Ag Test STANDARD Q PLEASE READ BACK PAGE CAREFULLY BEFORE YOU PERFORM THE TEST QRAC01G

Upload: others

Post on 28-Jun-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: STANDARD Q Rota/Adeno Ag - LAB MARK – Klinická ... · STANDARD Q Rota/Adeno Ag Test is a rapid, qualitative immunoassay to detect Rotavirus Ag and Adenovirus Ag in human fecal

EXPLANATION AND SUMMARY[Introduction]Rotavirus is the commonest virus cause acute diarrhea. It is most common in infants and young children. Children who get rotavirus infected may have severe watery diarrhea, often with vomiting, fever, and abdominal pain. However, older children and adults also can get sick from rotavirus. Adenovirus is the second most common cause of diarrhea following rotavirus. Adenoviruses cause various clinical diseases such as Common cold, Sore throat, Bronchitis, etc. Adenoviruses mainly implicated respiratory, ocular and the gastrointestinal systems of humans. Particularly, Enteric adenovirus types 40 and 41 cause gastroenteritis. Symptoms may include loss of appetite and dehydration (loss of body fluids), which can be especially dangerous for infants and young children. STANDARD Q Rota/Adeno Ag Test provides significantly fast, easy and accurate system to detect the specific antigens to rotavirus and adenovirus in human fecal specimen. It is essential for the reliable clinical diagnosis to identify Rotavirus/Adenovirus infection and enables supportive definite diagnosis of them.

[Intended use]STANDARD Q Rota/Adeno Ag Test is a rapid, qualitative immunoassay to detect Rotavirus Ag and Adenovirus Ag in human fecal specimen. The test is for in vitro diagnostic use and is intended as an aid to early diagnosis of Rotavirus and Adenovirus infection. This is intended for professional use, only for an initial screening test.

[Test principle]STANDARD Q Rota/Adeno Ag Test has two test lines (“R” for Rotavirus and “A” for Adenovirus) and a control line(“C”). Line “R” is coated with monoclonal anti-Rotavirus, “A” is coated with monoclonal anti-Adenovirus and “C” is coated with monoclonal anti-Rabbit IgG in the membrane of test device. For the test, the sample is added to the sample well and antigens in the sample interacts with monoclonal anti-Rotavirus-gold/monoclonal anti-Adenovirus-gold on the conjugation pad. Antigens in the sample and the gold conjugated antibodies make complexes and the complexes move along membrane to the test and control line chromatographically to react with the antibodies coated on the surface of membrane. If Rotavirus and Adenovirus are present in sample, visible lines in the test line would appear in the result window. The control line has to appear every time when the test has performed correctly. STANDARD Q Rota/Adeno Ag Test can also perform by using appropriate analyzer (STANDARD F Analyzers) manufactured by SD BIOSENSOR. The analyzers use the RGB optic sensor to measure reflected light from the membrane of the test device. The intensity of the reflected light is scanned and converted into an electric signal which is proportional to the intensity of reflected light produced on the membrane. STANDARD F analyzers can calculate the concentration of the analyte in the clinical specimen based on a pre-programmed calibration and display the test result on the screen.

[Kit contents] ① Test device ② Sterile swab ③ Extraction buffer tube ④ Filter cap ⑤ Instructions for use

[Materials required but not provided]• STANDARD F Analyzer• Timer

KIT STORAGE AND STABILITYStore the kit at 2-40°C /36-104°F, out of direct sunlight. Kit materials are stable until the expiration date printed on the outer box. Do not freeze.

WARNINGS AND PRECAUTIONS1. Do not re-use the test kit.2. Do not use the test kit if the pouch is damaged or the seal is broken.3. Do not use the extraction buffer of another lot.4. Do not smoke, drink or eat while handling specimen.5. Wear personal protective equipment, such as gloves and lab coats when handling kit reagents. Wash hands thoroughly when

afterwards.6. Clean up spills thoroughly using an appropriate disinfectant.7. Handle all specimens as if they contain infectious agents.8. Observe established precautions against microbiological hazards throughout testing procedures.9. Dispose of all specimens and materials used to perform the test as bio-hazard waste. Laboratory chemical and bio-hazard

wastes must be handled and discarded in accordance with all local, state, and national regulations.10. Silica gel in foil pouch is to absorb moisture and keep humidity from affecting products. If the moisture indicating silica gel

beads change from yellow to green, the pouch should be discarded.11. Immediately use the test device after taking out of foil pouch.12. Do not write on the bar code or damage the bar code of the test device.

SPECIMEN COLLECTION AND PREPARATION [Feces] 1. Fecal specimen should be collected in a clean and dry container and collected at any time of the day may be used. 2. Urine should be excluded from the fecal specimen. 3. Fresh fecal specimen should be used in this test. Do not use any transport media to store and transport the fecal specimen. 4. Fecal specimen may be stored at room temperature (15-30°C/59-86°F) for up to 1 day or at 2-8°C/ 36-46°F for up to 2 days

prior to testing.

TEST PROCEDURE[Preparation]1. Allow test device and collected sample to room

temperature (15-30°C / 59-86°F ) prior to testing.2. Carefully read instructions for using the STANDARD Q

Rota/Adeno Ag Test.3. Check the expiry date at the back of the foil pouch. Use

another lot, if expiry date has passed./ Item

/ LOT No./ MFG DATE/ EXP DATE

/ REF No.

/ Item

ML25RT7MLR1Issue date : 2017.10

xxxxxxxxxxxxxxxx.xx.xxxxxx.xx.xx

STANDARD Q Rota/Adeno Agxxxxxxxxxxxx

4. Open the foil pouch, and check the test device and the silica gel pack in the foil pouch.

CR

A

Rota/Adeno

<Foil pouch>

Yellow

Yellow: ValidGreen: Invalid

Green

<Silica gel>

STAN

DA

RD Q

Rapid Test G

lobal standard test for rapid diagnostic test

40°C(104°F)

2°C(36°F)

Result window

Sample well

<Test device>

[Collection of sample]1. Poke the 4 different areas of the fecal sample using the

sterile swab to collect a portion of feces (about 40-70mg). In case of liquid sample, soak the sterile swab completely in the liquid fecal sample.

Too little

Good

Too much

NOTE

• The amount of fecal swab may affect the results. It is required to follow the swab amount of feces as shown in the picture above. Excessive fecal amount may induce a false positive result and slow migration.

2. Take off the top of an Extraction buffer tube and insert the swab into the tube.

3. Swirl the swab at least ten times into the Extraction buffer tube to dissolve the sample.

x10

4. Remove the swab and dispose of the used swab in accordance with your biohazard waste disposal protocol.

5. Tightly screw the filter cap onto the tube.

CAUTION

• Do not use the extraction buffer tube without the filter cap. It affects the result value.

[Analysis of sample]• Visual inspection

1. Take the test device out of the foil pouch and place it on a flat and dry surface.

2. Apply 4 drops of mixed sample to the sample well in the test device holding the prepared extraction buffer tube upside down.

CRA

Rota/Adeno

4 drops

3. Read the test result after 20 minutes.

20 IncubateFor 20 mins

[Analysis of sample]• Using a ‘STANDARD TEST’ mode - Applied STANDARD F100, F200 and F2400 analyzer

1. Prepare a STANDARD F Analyzer and select the ‘Standard Test’ mode according to the analyzer’s manual. In case of STANDARD F2400 analyzer, go to the ‘Workplace’ in the main screen. And select the ‘Run Test‘.

2. In case of STANDARD F200 and F2400 analyzer, input patient ID and/or operator ID on the analyzer.3. Take the test device out of the foil pouch.

4. Insert the test device to the test slot of the analyzer. When inserting the test device to the analyzer, the analyzer will read the barcode data, and check the test device is valid.

Quality approved by SD BIOSENSOR / For In vitro use only

Rota/Adeno AgSTANDARDTM Q Rota/Adeno Ag Test

STANDARD Q

PLEASE READ BACK PAGE CAREFULLY BEFORE YOU PERFORM THE TEST

QRAC01G

Page 2: STANDARD Q Rota/Adeno Ag - LAB MARK – Klinická ... · STANDARD Q Rota/Adeno Ag Test is a rapid, qualitative immunoassay to detect Rotavirus Ag and Adenovirus Ag in human fecal

STANDARDTM Q Rota/Adeno Ag Test

Reference number In vitro Diagnos�cs Do not re-use.

To indicate the temperature limita�ons in which the transport package has to be kept and handled.

Consult Instruc�ons for Use Contains Sufficient for <n> Tests

ManufacturerBatch codeUse byCau�on Date of manufacture Note

Fulfill the requirements ofDirec�ve 98/79/EC on in vitro diagnos�cmedical devices

5. Apply 4 drops of mixed sample to the sample well in the test device holding the prepared extraction buffer tube upside down.

Flu. A/B

4 drops

6. After applying the sample, immediately press the ‘TEST START’ button.

<F100> <F200> <F2400>

Test StartSTART

7. The analyzer will automatically display the test result within 20 minutes.

20Incubate20mins

• Using a ‘READ ONLY’ mode - Applied STANDARD F100 and F200 analyzer 1. Take the test device out of the foil pouch and place it on a flat and dry surface. Write a sample information on the

label of test device.

2. Apply 4 drops of mixed sample to the sample well in the test device holding the prepared extraction buffer tube upside down.

CMS G

Rota/Adeno4 drops

Assay diluent

3. Stay the test device for 20 minutes outside of the analyzer for incubation.

20Incubate20mins

4. To preparing the test with STANDARD F Analyzer, set the ‘READ ONLY’ mode following the instructions in the manual.

5. Insert the test device to the test slot of the analyzer. The analyzer automatically reads the barcode information of the the test device and outputs the resulting values based on the test information.

INTERPRETATION OF TEST RESULTS[Visual inspection]

Test result Example Description

NegativeC R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

Only band (“C” Control line) within the result window indicates a negative result.

Positive

Rotavirus Positive

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

Two colored bands ("C" Control line and "R" Test line) within the result window indicate Rotavirus positive.

Adenovirus Positive

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

Two colored bands ("C" Control line and "A" Test line) within the result window indicate Adenovirus positive

Rotavirus/Adenovirus

Positive

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

C R A

Rota/

Aden

o

Three colored bands ("C" Control line, "A" Test line and "R" Test line) within the result window indicate Rotavirus/Adenovirus positive.

Invalid

C R A

Rota/Ade

no

C R A

Rota/Ade

no

C R A

Rota/Ade

no

C R A

Rota/Ade

no

C R A

Rota/Ade

no

C R A

Rota/Ade

no

C R A

Rota/Ade

no

C R A

Rota/Ade

no

If the control band ("C" Control line) is not visible within the result window, the result is considered invalid. The directions may not have been followed correctly or the test may have deteriorated. Re-test with a new test device.

[STANDARD F100, F200, F2400 analyzers]

Result COI (Cutoff index) value Interpretation

Rota = Positive COI ≥ 1.0 Positive for Rotavirus

Adeno = Positive COI ≥ 1.0 Positive for Adenovirus

Rota = Negative COI < 1.0 Negative for Rotavirus

Adeno = Negative COI < 1.0 Negative for Adenovirus

Invalid COI value is not displayed Retest should be performed with a new test device and a new patient sample

NOTE

• The test result of a sample is given either as Positive(+)/Pos(+) or Negative(-)/Neg(-) with a COI(cutoff index) value. The COI is a numerical representation of the measured uorecence signal.

LIMITATION OF TEST1. The test should be used for the detection of Rotavirus and Adenovirus in human fecal specimens.2. Neither the quantitative value nor the rate of concentration for Rotavirus and Adenovirus can be determined by this

qualitative test.3. Failure to follow the test procedure and interpretation of test results may adversely affect test performance and/or produce

invalid results.4. A negative test result may occur if the level of extracted antibody in a sample is below the sensitivity of the test or if a poor-

quality specimen is obtained.5. For more accuracy of immune status, additional follow-up testing using other laboratory methods is recommended.6. The test result must always be evaluated with other data available to the physician.

BIBLIOGRAPHY1. Murphy TV, Gargiullo PM, Massoudi MS, Nelson DB, Jumaan AO, Okoro CA, et al. Intussusception among infants given an

oral Rotavirus vaccine. N Engl J Med 2001;344:564-72.2. CDC. Prevention of rotavirus gastroenteritis among infants and children. Recommendations of the Advisory Committee on

Immunization Practices (ACIP). MMWR 2006;55(No. RR-12).3. Parashar UD, Gibson CJ, Bresee JS, Glass RI. Rotavirus and severe childhood diarrhea. Emerg Infect Dis 2006;12:304–6.4. Parashar UD, Burton A, Lanata C, et al. Global mortality associated with rotavirus disease among children in 2004. J Infect

Dis 2009;200 Suppl 1:S9–15.5. Bon F, Fascia P, Dauvergne M, Tenenbaum D, Planson H, Petion AM, et al. Prevalence of group A rotavirus, human

calicivirus, astrovirus, and adenovirus type 40 and 41 infections among children with acute gastroenteritis in Dijon, France. J Clin Microbiol 1999;37: 3055-8.

Product DisclaimerWhilst every precaution has been taken to ensure the diagnostic ability and accuracy of this product, the product is used outside of the control of the SD BIOSENSOR and distributor and the result may accordingly be affected by environmental factors and/or user error. A person who is the subject of the diagnosis should consult a doctor for further confirmation of the result.

WarningThe SD BIOSENSOR and distributors of this product shall not be liable for any losses, liability, claims, costs or damages whether direct or indirect of consequential arising out of or related to an incorrect diagnosis, whether positive or negative, in the use of this product.

Manufactured by SD Biosensor, Inc.Head office : C-4th&5th, 16, Deogyeong-daero 1556beon-gil, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16690, REPUBLIC OF KOREA Manufacturing site : 74, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, 28161, REPUBLIC OF KOREA

Authorized Representative

Altenhofstrasse 80 D-66386 St. Ingbert GermanyPhone : +49 6894 581020, Fax : +49 6894 581021

Any inquiries regarding instructions provided should be addressed to: [email protected] you can also contact us through www.sdbiosensor.com

L23ROT2ENR2Issue date: 2018.09